I. An isothermal titration microcalorimeter having a colume of 4 ml and capable of temerature control to ±2 x 10-5°C is described. Major components include a constant temperature water bath controlled to ±3 x 10-4°C, a platinum reation vessel, and an isothermal control circuit consisting of constant Peltier thermoelectric cooling and variable Joule heating controlled by a thermistor in an AC Wheatstone bridge circuit. The calorimeter was tested by measuring the heat of ionization of water and was found to produce data accurate to ± 0.1% where small samples are used such as in the investigation of many biological systems.
Identifer | oai:union.ndltd.org:BGMYU2/oai:scholarsarchive.byu.edu:etd-9210 |
Date | 01 August 1973 |
Creators | Gardner, John Willard |
Publisher | BYU ScholarsArchive |
Source Sets | Brigham Young University |
Detected Language | English |
Type | text |
Format | application/pdf |
Source | Theses and Dissertations |
Rights | http://lib.byu.edu/about/copyright/ |