Return to search

Can handheld POC capillary lactate measurement be used with arterial and venous laboratory testing methods in the identification of sepsis?

No / The aim of this review was to examine whether the measurement of lactate in capillary blood samples using point-of-care handheld analysers corresponds sufficiently closely with arterial and venous whole-blood samples analysed by hospital central laboratory or blood gas analyser to be used interchangeably.

A systematic search, informed by focused inclusion/exclusion criteria, was performed using multiple databases up to October 2015. A total of 65 articles were considered to have potential relevance and were evaluated in full text, of which ultimately five articles met all inclusion/exclusion criteria, and a final four were selected after data extraction and quality appraisal.

All four studies found a predominantly upward bias in the measurement of lactate in capillary samples tested using a handheld point-of-care device over arterial or venous samples tested by laboratory methods or blood gas analyser. In terms of correlation, there was consensus between the studies that the strength of association between the two methods of measurement was statistically significant. Three studies directly examined the extent of agreement between point-of-care capillary lactate measurements and those of laboratory or blood gas analyser reference determined to ±2 standard deviations; 95% confidence intervals, and report contextually broad limits of agreement, identifying a potential for both over triage and, to a lesser extent, under triage. The findings of the review do not support interchangeable use of handheld fingertip point-of-care lactate measurement with laboratory or blood gas analyser methods in the identification of sepsis.

Identiferoai:union.ndltd.org:BRADFORD/oai:bradscholars.brad.ac.uk:10454/18605
Date22 September 2021
CreatorsLightowler, Bryan, Hoswell, Anthony
Source SetsBradford Scholars
LanguageEnglish
Detected LanguageEnglish
TypeArticle, No full-text in the repository

Page generated in 0.0021 seconds