Return to search

Trapping and Removal of Bubbles in a Microfluidic Format

Unwanted gas bubbles are a challenge for microfluidic-based systems, as adherence to channel networks can disrupt fluid delivery. This is especially true for devices with biological applications, as the presence of a single bubble creates thin fluid films with extremely high shear stresses, which can damage biological samples. Current strategies to remove bubbles require complicated fabrication or off-chip components. This thesis describes an on-chip microfluidic strategy utilizing permeation for in-plane trapping and removal of occasional gas bubbles. The trap was demonstrated with nitrogen bubbles, which were consistently removed at a rate of 0.14 µL/min for a single trap, and shown to have long-term operation capability by removing approximately 4,000 bubbles during one day without failure. The trap was integrated with a microfluidic system for the study of small blood vessels. Experiments were complemented with analytical and numerical models to characterize the bubble removal process.

Identiferoai:union.ndltd.org:LACETR/oai:collectionscanada.gc.ca:OTU.1807/32258
Date21 March 2012
CreatorsLochovsky, Conrad
ContributorsGuenther, Axel
Source SetsLibrary and Archives Canada ETDs Repository / Centre d'archives des thèses électroniques de Bibliothèque et Archives Canada
Languageen_ca
Detected LanguageEnglish
TypeThesis