Study of The Analytical Method for Chromium(VI) in Seawater / 海水中六價鉻分析方法之探討

碩士 / 國立中山大學 / 海洋生物科技暨資源學系研究所 / 94 / In this research, in order to set up the ultimate approaches to determine the content of chromium(Ⅵ) in seawater, different pH values, the amount of chelating agent, chelating time and extraction time were investigated in two different solvent extraction systems, APDC-MIBK and Aliquat 336-MIBK.
In APDC-MIBK system, the ultimate extracting condition is that 400 mL sample is adjusted to pH 3.0, added 4 % APDC 4 mL to chelate chromium(Ⅵ) for 60 minutes, and added 20 mL MIBK to partiton into two fractions for 5 minutes. The recovery in APDC-MIBK system is 68 %~ 123 %. The method detection limit is 0.2 ppb.
In Aliquat 336-MIBK system, the ultimate extracting condition is that 400 mL sample is adjusted to pH 2.0, added 5 % Aliquat 336 4 mL to chelate chromium(Ⅵ), added 20 mL MIBK to be shaken for 2 minutes, and partitoned into two fractions for 5 minutes. The recovery in Aliquat 336-MIBK system is 80 %~ 120 %. The method detection limit is 0.1ppb.
Although in APDC-MIBK system assorted metals can be extracted at same time, it is steadier in Aliquat 336-MIBK system than APDC-MIBK system. Therefore, it is a feasible extraction method to adopt Aliquat 336-MIBK system to determine the content of chromium(Ⅵ) in seawater.

Identiferoai:union.ndltd.org:TW/094NSYS5270030
Date January 2006
CreatorsLi-Li Huang, 黃麗莉
ContributorsWei-Hsien Wang, 王維賢
Source SetsNational Digital Library of Theses and Dissertations in Taiwan
Languagezh-TW
Detected LanguageEnglish
Type學位論文 ; thesis
Format79

Page generated in 0.0123 seconds