Return to search

3D A-Mode Ultrasound Calibration and Registration of the Tibia and Femur for Computer-Assisted Robotic Surgery

Registration is a key component for computer-navigated robot-assisted surgery. Invasive approaches such as fiducial-based and surface matching with mechanical probes are common but ultrasound may provide a non-invasive alternative. If an A-mode ultrasound transducer can be used to percutaneously select data points on the bones, a registration can be determined without needing any incision. This study investigates selecting an A-mode ultrasound transducer, calibrating it, analyzing the ultrasound signal, and using it to register a phantom sawbone tibia and femur as well as cadaveric specimens. This study is performed in conjunction with MAKO Surgical Corp.'s Tactile Guidance System™ (TGS™) at their headquarters and at The South Florida Spine Clinic for cadaveric experiments. The results for phantom and cadaveric ultrasound registrations compared to a mechanical probe approach demonstrate that A-mode ultrasound registration is a viable option for registration of the bones of the knee.

Identiferoai:union.ndltd.org:UMIAMI/oai:scholarlyrepository.miami.edu:oa_dissertations-1113
Date11 June 2008
CreatorsMozes, Alon
PublisherScholarly Repository
Source SetsUniversity of Miami
Detected LanguageEnglish
Typetext
Formatapplication/pdf
SourceOpen Access Dissertations

Page generated in 0.0021 seconds