A BF3 fast neutron dosimeter or long counter is constructed and calibrated for use with the Cockcroft-Walton accelerator and UTR-IO reactor at Virginia Polytechnic Institute. The counter consists of a paraffin cylinder 8 inches in diameter, having a BF3 proportional detector on the axis and a set of holes around the detector to improve the low energy response. External shields of paraffin, 3 inches thick, and boron-trioxide, 1/2 inch thick, enclose the paraffin cylinder. These shields serve to reduce the background effect and improve the high energy response.
Calibration of the counter is accomplished using as neutron sources: plutonium-berylliwn and the induced reactions H2(d,n)He3 and H3(d,n)He4. Long counter efficiencies for both a 1 inch and 1/2 inch BF) tube yield a flat response over the energy range, 2.5 to 4 Mev whereas an approximately 60% decrease in efficiency is observed for a 14.1 Mev neutron.
Count rate of long counter is expressed as a function of the incident neutron flux. Neutron fluxes, ranging from 0 to 1100 neutrons/cm2-second are monitored with the counter. / Master of Science
Identifer | oai:union.ndltd.org:VTETD/oai:vtechworks.lib.vt.edu:10919/42283 |
Date | 26 April 2010 |
Creators | Leyland, Walter A. |
Contributors | Physics |
Publisher | Virginia Tech |
Source Sets | Virginia Tech Theses and Dissertation |
Language | English |
Detected Language | English |
Type | Thesis, Text |
Format | 51 leaves, BTD, application/pdf, application/pdf |
Rights | In Copyright, http://rightsstatements.org/vocab/InC/1.0/ |
Relation | OCLC# 27251505, LD5655.V855_1960.L495.pdf |
Page generated in 0.0014 seconds