Return to search

PNEUMATIC ARTIFICIAL HEART DRIVER PARAMETER EFFECTS ON THE RATE OF PRESSURE CHANGE ((+) DP/DT MAX)

The aim of the research was to investigate the effects of three parameters of the artificial heart on the (+) dP/dt max. The study was conducted using a mock circulation which was connected to an artificial heart. The data were collected using the COMDU software developed for the computer which monitors the artificial heart. Stepwise regression analysis was utilized to test the three hypotheses. Two of the null hypotheses for the study could not be rejected, as the independent variable did not significantly affect (+) dP/dt max. Although the third hypothesis was accepted, the results were not clinically significant. Limitations in the study were multicollinearity among the independent variables, small sample size, and the inability of the mock circulation to represent human responses.

Identiferoai:union.ndltd.org:arizona.edu/oai:arizona.openrepository.com:10150/276466
Date January 1987
CreatorsHenker, Richard
PublisherThe University of Arizona.
Source SetsUniversity of Arizona
Languageen_US
Detected LanguageEnglish
Typetext, Thesis-Reproduction (electronic)
RightsCopyright © is held by the author. Digital access to this material is made possible by the University Libraries, University of Arizona. Further transmission, reproduction or presentation (such as public display or performance) of protected items is prohibited except with permission of the author.

Page generated in 0.0017 seconds