Return to search

Slurry nebulization ICP-OES for the determination of Cu, Fe, Mg, Mn and Zn in bovine liver

A method that involves analysis of bovine liver by slurry nebulization and ICP-OES has been
developed. This method permits rapid and accurate determination of Cu, Fe, Mg, Mn and Zn in bovine liver.
Aliquots of freeze-dried and powdered bovine liver sample were dispersed in 2.0 M HNO3 and sonicated to
homogenize the resulting slurries. Bovine liver samples were also microwave digested or subjected to aqueous
extraction for comparison of analytical results. Concentrations of Cu, Fe, Mg, Mn and Zn in aqueous slurries, the
digests, and aqueous extracts were determined by the ICP-OES using external calibration curves. A student’s ttest
showed that the results obtained using the slurry method were in good agreement at 95 % confidence level
(CL) with those of microwave digestion or aqueous extraction techniques, except for Fe. To check the accuracy
and precision of the slurry method, a bovine liver CRM was analyzed and good agreement was achieved with the
certified values at 95 % CL. The results demonstrate inefficiency of aqueous extraction technique for complete
removal of Fe in bovine liver sample.

Identiferoai:union.ndltd.org:netd.ac.za/oai:union.ndltd.org:tut/oai:encore.tut.ac.za:d1000695
Date17 November 2007
CreatorsMokgalaka, NS, Wondimu, T, McCrindle, RI
PublisherChemical Society of Ethiopia
Source SetsSouth African National ETD Portal
LanguageEnglish
Detected LanguageEnglish
TypeText
FormatPDF
RightsChemical Society of Ethiopia
RelationBulletin of the Chemical Society of Ethiopia

Page generated in 0.0023 seconds