Global ETD Search

Search theses and dissertations gathered from participating repositories worldwide. Every result links back to the library that holds it. No account is needed.

Results

Showing 1 to 3 of 3 for “"Hot snare"”.

  1. Controlled Voltage of Hot Snare Polypectomy Device in Electrosurgical Device

    The study is used to understand the working procedure of the Olympus PSD-30 Electrosurgical Unit which a high-frequency alternating current is to measure the voltage and power output from the unit when used for a surgical operation to determine the extent of tissue damage. The output power and …

    gsu Repository record for Controlled Voltage of Hot Snare Polypectomy Device in Electrosurgical Device (opens in a new tab)

  2. Development of an Impedance-Controlled Hot Snare Polypectomy Device to Minimize Tissue Damage

    This study explores the ability to measure the changing impedance, ex vivo, of a porcine colon sample while undergoing a high-frequency alternating current from an Olympus PSD-30 electrosurgical unit and stop the applied current before excessive tissue damage ensues. The causes of the thermal …

    gsu Repository record for Development of an Impedance-Controlled Hot Snare Polypectomy Device to Minimize Tissue Damage (opens in a new tab)

  3. Ψυχρός βρόχος πολυποδεκτομής έναντι θερμού βρόχου πολυποδεκτομής έναντι APC για την πλήρη αφαίρεση των μικρών πολύποδων από το αριστερό κόλον

    … to one of three treatment modalities: (1) cold snare polypectomy (CSP), (2) hot snare polypectomy (HSP) and (3) argon plasma coagulation (APC) ablation (50–60 W, flow: 2 l/min). The polyp site was marked with an endoscopic tattoo, and a follow-up colonoscopy with scar biopsies was performed >6 …

    greece Repository record for Ψυχρός βρόχος πολυποδεκτομής έναντι θερμού βρόχου πολυποδεκτομής έναντι APC για την πλήρη αφαίρεση των μικρών πολύποδων από το αριστερό κόλον (opens in a new tab)