{"id":{"repo_id":"botswana","oai_identifier":"oai:ubrisa.ub.bw:10311/2441"},"canonical_url":"https://search.ndltd.org/etd/botswana/oai:ubrisa.ub.bw:10311/2441","repository":{"repo_id":"botswana","name":"University Botswana","base_url":"https://ubrisa.ub.bw/oai/request"},"display":{"title":"A validated analytical method for the determination of lumefantrine in selected time modulated dried blood spot samples from malaria patients in Botswana","abstract":"A simple liquid chromatography- diode array detector (LC-DAD) method for the determination of lumefantrine in whole blood collected in dried blood spot (DBS) filters was developed and validated. The validation was done using the United States Food and Drug Administration (USFDA) guidelines. Sample preparation was done using solid liquid extraction (SLE) followed by separation using high performance liquid chromatography (HPLC) and diode array detection (DAD). Separation was done at 25ºC using an XTerra C18 column with dimensions of 50 mm x 4.6 mm x 5μm (length x internal diameter x particle size) and a binary solvent system of acetonitrile and water adjusted to pH of 2.3 with formic acid as the mobile phase. The mobile phase was pumped at a flow rate of 0.570 mL/min using a gradient elution program. The analysis time was 2 minutes and the calibration curve obtained was linear over the concentration range of 1-8 μg/mL with a correlation coefficient (R2) of 0.9980. The limit of detection (LOD) and the lower limit of quantification (LLOQ) were 0.4 and 0.8 μg/mL respectively. The extraction efficiency estimated as percent recovery was greater than 60 %. Both the intra and inter-day precision of this method were ˂ ±15% as prescribed by the USFDA guidelines. The method was successfully applied for the quantification of lumefantrine in time modulated dried blood spot samples, previously collected from patients on malarial treatment with the artemisinin/ lumefantrine combination therapy.","abstract_html":"A simple liquid chromatography- diode array detector (LC-DAD) method for the determination of lumefantrine in whole blood collected in dried blood spot (DBS) filters was developed and validated. The validation was done using the United States Food and Drug Administration (USFDA) guidelines. Sample preparation was done using solid liquid extraction (SLE) followed by separation using high performance liquid chromatography (HPLC) and diode array detection (DAD). Separation was done at 25ºC using an XTerra C18 column with dimensions of 50 mm x 4.6 mm x 5μm (length x internal diameter x particle size) and a binary solvent system of acetonitrile and water adjusted to pH of 2.3 with formic acid as the mobile phase. The mobile phase was pumped at a flow rate of 0.570 mL/min using a gradient elution program. The analysis time was 2 minutes and the calibration curve obtained was linear over the concentration range of 1-8 μg/mL with a correlation coefficient (R2) of 0.9980. The limit of detection (LOD) and the lower limit of quantification (LLOQ) were 0.4 and 0.8 μg/mL respectively. The extraction efficiency estimated as percent recovery was greater than 60 %. Both the intra and inter-day precision of this method were ˂ ±15% as prescribed by the USFDA guidelines. The method was successfully applied for the quantification of lumefantrine in time modulated dried blood spot samples, previously collected from patients on malarial treatment with the artemisinin/ lumefantrine combination therapy.","abstract_has_math":false,"creators":["Brooks, Blondie O."],"institution":"University of Botswana, www.ub.bw","degree_name":null,"degree_level":null,"degree_discipline":null,"degree_department":null,"school":null,"contributors":[],"advisors":[],"committee_chairs":[],"committee_members":[],"year":2020,"date_issued":"2020-03","date_published":"2020-03","updated_at":"2026-11-19T17:38:58Z","subjects":["Liquid chromatography- diode array detector","lumefantrine","blood","United States Food and Drug Administration (USFDA)","dried blood spot samples","malaria patients","Botswana"],"languages":["en"],"rights":[],"rights_urls":[],"identifier_entries":[],"person_identifier_entries":[]},"links":{"outbound_url":"https://hdl.handle.net/10311/2441","outbound_label":"Handle","outbound_source":"dc:identifier.uri","record_url":"https://hdl.handle.net/10311/2441","record_label":"Repository record","record_kind":"handle","record_source":"dc:identifier.uri","access_status":null,"embargo_until":null,"access_note":null},"oai_pmh_record":{"url":"https://union.ndltd.org/oai-pmh/?verb=GetRecord&metadataPrefix=oai_dc&identifier=oai%3Aunion.ndltd.org%3Arepo%3Abotswana%3Aoai%3Aubrisa.ub.bw%3A10311%2F2441","prefix":"oai_dc"},"source_record":{"url":"https://ubrisa.ub.bw/oai/request?verb=GetRecord&metadataPrefix=dim&identifier=oai%3Aubrisa.ub.bw%3A10311%2F2441","prefix":"dim"},"metadata_groups":[{"id":"people","label":"People","entries":[{"key":"dc:creator","label":"Author","values":["Brooks, Blondie O."]}]},{"id":"academic_context","label":"Academic Context","entries":[{"key":"dc:date.accessioned","label":"Dc Date Accessioned","values":["2022-06-13T14:15:06Z"]},{"key":"dc:date.available","label":"Dc Date Available","values":["2022-06-13T14:15:06Z"]},{"key":"dc:date.issued","label":"Date","values":["2020-03"]},{"key":"dc:publisher","label":"Institution","values":["University of Botswana, www.ub.bw"]},{"key":"dc:type","label":"Dc Type","values":["Masters Thesis/Dissertation"]}]},{"id":"subjects_keywords","label":"Subjects and Keywords","entries":[{"key":"dc:subject","label":"Dc Subject","values":["Liquid chromatography- diode array detector","lumefantrine","blood","United States Food and Drug Administration (USFDA)","dried blood spot samples","malaria patients","Botswana"]}]},{"id":"language_rights","label":"Language and Rights","entries":[{"key":"dc:language.iso","label":"Language (ISO)","values":["en"]}]},{"id":"identifiers","label":"Identifiers","entries":[{"key":"dc:identifier.uri","label":"Identifier URI","values":["http://hdl.handle.net/10311/2441"]}]},{"id":"additional","label":"Additional Metadata","entries":[{"key":"dc:description","label":"Description","values":["A dissertation submitted to the Dept. of Chemistry, Faculty of Science, University of Botswana in partial fulfillment of the requirement of the degree of Masters in Analytical Chemistry. Citation: Brooks, B.O. (2020) A validated analytical method for the determination of lumefantrine in selected time modulated dried blood spot samples from malaria patients in Botswana, University of Botswana."]},{"key":"dc:description.abstract","label":"Abstract","values":["A simple liquid chromatography- diode array detector (LC-DAD) method for the determination of lumefantrine in whole blood collected in dried blood spot (DBS) filters was developed and validated. The validation was done using the United States Food and Drug Administration (USFDA) guidelines. Sample preparation was done using solid liquid extraction (SLE) followed by separation using high performance liquid chromatography (HPLC) and diode array detection (DAD). Separation was done at 25ºC using an XTerra C18 column with dimensions of 50 mm x 4.6 mm x 5μm (length x internal diameter x particle size) and a binary solvent system of acetonitrile and water adjusted to pH of 2.3 with formic acid as the mobile phase. The mobile phase was pumped at a flow rate of 0.570 mL/min using a gradient elution program. The analysis time was 2 minutes and the calibration curve obtained was linear over the concentration range of 1-8 μg/mL with a correlation coefficient (R2) of 0.9980. The limit of detection (LOD) and the lower limit of quantification (LLOQ) were 0.4 and 0.8 μg/mL respectively. The extraction efficiency estimated as percent recovery was greater than 60 %. Both the intra and inter-day precision of this method were ˂ ±15% as prescribed by the USFDA guidelines. The method was successfully applied for the quantification of lumefantrine in time modulated dried blood spot samples, previously collected from patients on malarial treatment with the artemisinin/ lumefantrine combination therapy."]},{"key":"dc:description.provenance","label":"Dc Description Provenance","values":["Submitted by Isaac Thothe (thotheim@ub.ac.bw) on 2022-05-16T11:05:06Z No. of bitstreams: 1 Brooks_Unpublished (MSc)_2020.pdf: 1568216 bytes, checksum: fa635e8a54ffaaa50b8a9819cb3e96b4 (MD5)","Approved for entry into archive by Oarabile Rakgamanyane (rakgamanyane@ub.ac.bw) on 2022-06-13T14:15:06Z (GMT) No. of bitstreams: 1 Brooks_Unpublished (MSc)_2020.pdf: 1568216 bytes, checksum: fa635e8a54ffaaa50b8a9819cb3e96b4 (MD5)","Made available in DSpace on 2022-06-13T14:15:06Z (GMT). No. of bitstreams: 1 Brooks_Unpublished (MSc)_2020.pdf: 1568216 bytes, checksum: fa635e8a54ffaaa50b8a9819cb3e96b4 (MD5) Previous issue date: 2020-03"]},{"key":"dc:title","label":"Title","values":["A validated analytical method for the determination of lumefantrine in selected time modulated dried blood spot samples from malaria patients in Botswana"]}]}],"canonical_facts":{"dc:creator":["Brooks, Blondie O."],"dc:date.accessioned":["2022-06-13T14:15:06Z"],"dc:date.available":["2022-06-13T14:15:06Z"],"dc:date.issued":["2020-03"],"dc:description":["A dissertation submitted to the Dept. of Chemistry, Faculty of Science, University of Botswana in partial fulfillment of the requirement of the degree of Masters in Analytical Chemistry. Citation: Brooks, B.O. (2020) A validated analytical method for the determination of lumefantrine in selected time modulated dried blood spot samples from malaria patients in Botswana, University of Botswana."],"dc:description.abstract":["A simple liquid chromatography- diode array detector (LC-DAD) method for the determination of lumefantrine in whole blood collected in dried blood spot (DBS) filters was developed and validated. The validation was done using the United States Food and Drug Administration (USFDA) guidelines. Sample preparation was done using solid liquid extraction (SLE) followed by separation using high performance liquid chromatography (HPLC) and diode array detection (DAD). Separation was done at 25ºC using an XTerra C18 column with dimensions of 50 mm x 4.6 mm x 5μm (length x internal diameter x particle size) and a binary solvent system of acetonitrile and water adjusted to pH of 2.3 with formic acid as the mobile phase. The mobile phase was pumped at a flow rate of 0.570 mL/min using a gradient elution program. The analysis time was 2 minutes and the calibration curve obtained was linear over the concentration range of 1-8 μg/mL with a correlation coefficient (R2) of 0.9980. The limit of detection (LOD) and the lower limit of quantification (LLOQ) were 0.4 and 0.8 μg/mL respectively. The extraction efficiency estimated as percent recovery was greater than 60 %. Both the intra and inter-day precision of this method were ˂ ±15% as prescribed by the USFDA guidelines. The method was successfully applied for the quantification of lumefantrine in time modulated dried blood spot samples, previously collected from patients on malarial treatment with the artemisinin/ lumefantrine combination therapy."],"dc:description.provenance":["Submitted by Isaac Thothe (thotheim@ub.ac.bw) on 2022-05-16T11:05:06Z No. of bitstreams: 1 Brooks_Unpublished (MSc)_2020.pdf: 1568216 bytes, checksum: fa635e8a54ffaaa50b8a9819cb3e96b4 (MD5)","Approved for entry into archive by Oarabile Rakgamanyane (rakgamanyane@ub.ac.bw) on 2022-06-13T14:15:06Z (GMT) No. of bitstreams: 1 Brooks_Unpublished (MSc)_2020.pdf: 1568216 bytes, checksum: fa635e8a54ffaaa50b8a9819cb3e96b4 (MD5)","Made available in DSpace on 2022-06-13T14:15:06Z (GMT). No. of bitstreams: 1 Brooks_Unpublished (MSc)_2020.pdf: 1568216 bytes, checksum: fa635e8a54ffaaa50b8a9819cb3e96b4 (MD5) Previous issue date: 2020-03"],"dc:identifier.uri":["http://hdl.handle.net/10311/2441"],"dc:language.iso":["en"],"dc:publisher":["University of Botswana, www.ub.bw"],"dc:subject":["Liquid chromatography- diode array detector","lumefantrine","blood","United States Food and Drug Administration (USFDA)","dried blood spot samples","malaria patients","Botswana"],"dc:title":["A validated analytical method for the determination of lumefantrine in selected time modulated dried blood spot samples from malaria patients in Botswana"],"dc:type":["Masters Thesis/Dissertation"]},"updated_at":"2026-11-19T17:38:58Z"}