{"id":{"repo_id":"brno-tech","oai_identifier":"oai:dspace.vut.cz:11012/259222"},"canonical_url":"https://search.ndltd.org/etd/brno-tech/oai:dspace.vut.cz:11012/259222","repository":{"repo_id":"brno-tech","name":"Brno University of Technology","base_url":"https://dspace.vut.cz/oai/request"},"display":{"title":"Metody testování inhalátorů","abstract":"Inhalation therapy is an effective route for drug administration to the respiratory system, and the reliable testing of inhalation devices is essential to ensure therapeutic efficacy. The first part of this thesis provides a theoretical overview of inhalation systems and their working principles, aerosol and particle transport fundamentals and the methods used for testing inhalers and nebulisers. The second part presents experimental measurements performed using the Aerogen Solo vibrating mesh nebuliser with fluorescently labelled particles. The aerodynamic particle size distribution was characterised using both the Next Generation Impactor and Andersen cascade Impactor, the regional deposition was assessed using a 3D printed airway replica of the human respiratory tract. The obtained results are discussed in terms of measurement methodology and compared with computational predictions from the MPPD model, highlighting both the potential and the current limitations of integrating airway replicas into standard inhaler testing.","abstract_html":"Inhalation therapy is an effective route for drug administration to the respiratory system, and the reliable testing of inhalation devices is essential to ensure therapeutic efficacy. The first part of this thesis provides a theoretical overview of inhalation systems and their working principles, aerosol and particle transport fundamentals and the methods used for testing inhalers and nebulisers. The second part presents experimental measurements performed using the Aerogen Solo vibrating mesh nebuliser with fluorescently labelled particles. The aerodynamic particle size distribution was characterised using both the Next Generation Impactor and Andersen cascade Impactor, the regional deposition was assessed using a 3D printed airway replica of the human respiratory tract. The obtained results are discussed in terms of measurement methodology and compared with computational predictions from the MPPD model, highlighting both the potential and the current limitations of integrating airway replicas into standard inhaler testing.","abstract_has_math":false,"creators":["Petrakis, Nikolaos"],"institution":"Vysoké učení technické v Brně. Fakulta strojního inženýrství","degree_name":null,"degree_level":"Bakalářský","degree_discipline":"bez specializace","degree_department":null,"school":null,"contributors":[],"advisors":["Lízal, František"],"committee_chairs":[],"committee_members":["Bělka, Miloslav","doc. Ing. František Lízal, Ph.D. (člen) doc. Ing. Vít Jan, Ph.D. (člen) doc. Ing. Jiří Šremr, Ph.D. (člen) prof. Ing. Pavel Novotný, Ph.D. (předseda) doc. Ing. Pavel Charvát, Ph.D. (místopředseda)"],"year":null,"date_issued":"","date_published":null,"updated_at":"2026-11-19T05:04:58Z","subjects":["Inhalation therapy","inhaler","nebuliser","aerosol","aerodynamic particle size distribution","fine particle fraction","cascade impactor","airway replica","particle deposition","fluorescein"],"languages":["en","angličtina (English)"],"rights":["Standardní licenční smlouva - přístup k plnému textu bez omezení"],"rights_urls":[],"identifier_entries":[{"key":"dc:identifier.citation","label":"Dc Identifier Citation","values":["PETRAKIS, N. Metody testování inhalátorů [online]. Brno: Vysoké učení technické v Brně. Fakulta strojního inženýrství. 2026."],"render_values":[{"text":"PETRAKIS, N. Metody testování inhalátorů [online]. Brno: Vysoké učení technické v Brně. Fakulta strojního inženýrství. 2026.","href":null,"code":true}]},{"key":"dc:identifier.other","label":"Dc Identifier Other","values":["170647"],"render_values":[{"text":"170647","href":null,"code":true}]}],"person_identifier_entries":[]},"links":{"outbound_url":"https://hdl.handle.net/11012/259222","outbound_label":"Handle","outbound_source":"dc:identifier.uri","record_url":"https://hdl.handle.net/11012/259222","record_label":"Repository record","record_kind":"handle","record_source":"dc:identifier.uri","access_status":null,"embargo_until":null,"access_note":null},"oai_pmh_record":{"url":"https://union.ndltd.org/oai-pmh/?verb=GetRecord&metadataPrefix=oai_dc&identifier=oai%3Aunion.ndltd.org%3Arepo%3Abrno-tech%3Aoai%3Adspace.vut.cz%3A11012%2F259222","prefix":"oai_dc"},"source_record":{"url":"https://dspace.vut.cz/oai/request?verb=GetRecord&metadataPrefix=dim&identifier=oai%3Adspace.vut.cz%3A11012%2F259222","prefix":"dim"},"metadata_groups":[{"id":"people","label":"People","entries":[{"key":"dc:contributor.advisor","label":"Advisor","values":["Lízal, František"]},{"key":"dc:contributor.committee","label":"Committee Member","values":["doc. Ing. František Lízal, Ph.D. (člen) doc. Ing. Vít Jan, Ph.D. (člen) doc. Ing. Jiří Šremr, Ph.D. (člen) prof. Ing. Pavel Novotný, Ph.D. (předseda) doc. Ing. Pavel Charvát, Ph.D. (místopředseda)"]},{"key":"dc:contributor.committeemember","label":"Committee Member","values":["Bělka, Miloslav"]},{"key":"dc:creator","label":"Author","values":["Petrakis, Nikolaos"]}]},{"id":"academic_context","label":"Academic Context","entries":[{"key":"dc:date.acceptance","label":"Dc Date Acceptance","values":["2026-06-17"]},{"key":"dc:date.created","label":"Dc Date Created","values":["2026"]},{"key":"dc:date.updated","label":"Dc Date Updated","values":["2026-06-18-08:46:54"]},{"key":"dc:publisher","label":"Institution","values":["Vysoké učení technické v Brně. Fakulta strojního inženýrství"]},{"key":"dc:publisher.faculty","label":"Dc Publisher Faculty","values":["Fakulta strojního inženýrství"]},{"key":"dc:type","label":"Dc Type","values":["Text","bachelorThesis","bakalářská práce"]},{"key":"thesis:defence_status","label":"Thesis Defence Status","values":["práce byla úspěšně obhájena"]},{"key":"thesis:degree_discipline","label":"Discipline","values":["bez specializace"]},{"key":"thesis:degree_level","label":"Degree Level","values":["Bakalářský"]},{"key":"thesis:degree_program","label":"Thesis Degree Program","values":["Fundamentals of Mechanical Engineering"]},{"key":"thesis:examination_report","label":"Thesis Examination Report","values":["The student presented the committee with the progress of the work, the results and conclusions of the thesis, and answered the reviewer's questions. Doc. Jan asked what causes two droplets to merge, once they are close together, what makes them coalesce into one. The student was uncertain, and Doc. Jan guided him towards surface energy, the tendency to minimise it as far as possible. Doc. Charvát asked about the filter, why it had been included when it caused such difficulties. The student explained that the reason was purely practical, to prevent particles from contaminating the downstream flow meter. Doc. Jan then asked about the fluorescence method, and the student explained how the measurement was carried out."]},{"key":"thesis:institution_name","label":"Thesis Institution Name","values":["Bc.","Vysoké učení technické v Brně. Fakulta strojního inženýrství. Energetický ústav"]}]},{"id":"subjects_keywords","label":"Subjects and Keywords","entries":[{"key":"dc:subject","label":"Dc Subject","values":["Inhalation therapy","inhaler","nebuliser","aerosol","aerodynamic particle size distribution","fine particle fraction","cascade impactor","airway replica","particle deposition","fluorescein"]}]},{"id":"language_rights","label":"Language and Rights","entries":[{"key":"dc:language","label":"Dc Language","values":["angličtina (English)"]},{"key":"dc:language.iso","label":"Language (ISO)","values":["en"]},{"key":"dc:rights","label":"Dc Rights","values":["Standardní licenční smlouva - přístup k plnému textu bez omezení"]}]},{"id":"identifiers","label":"Identifiers","entries":[{"key":"dc:identifier.citation","label":"Dc Identifier Citation","values":["PETRAKIS, N. Metody testování inhalátorů [online]. Brno: Vysoké učení technické v Brně. Fakulta strojního inženýrství. 2026."]},{"key":"dc:identifier.other","label":"Dc Identifier Other","values":["170647"]},{"key":"dc:identifier.uri","label":"Identifier URI","values":["http://hdl.handle.net/11012/259222"]}]},{"id":"additional","label":"Additional Metadata","entries":[{"key":"dc:description.abstract","label":"Abstract","values":["Inhalation therapy is an effective route for drug administration to the respiratory system, and the reliable testing of inhalation devices is essential to ensure therapeutic efficacy. The first part of this thesis provides a theoretical overview of inhalation systems and their working principles, aerosol and particle transport fundamentals and the methods used for testing inhalers and nebulisers. The second part presents experimental measurements performed using the Aerogen Solo vibrating mesh nebuliser with fluorescently labelled particles. The aerodynamic particle size distribution was characterised using both the Next Generation Impactor and Andersen cascade Impactor, the regional deposition was assessed using a 3D printed airway replica of the human respiratory tract. The obtained results are discussed in terms of measurement methodology and compared with computational predictions from the MPPD model, highlighting both the potential and the current limitations of integrating airway replicas into standard inhaler testing."]},{"key":"dc:description.grade","label":"Dc Description Grade","values":["B"]},{"key":"dc:title","label":"Title","values":["Metody testování inhalátorů"]},{"key":"dc:title.alternative","label":"Alternative Title","values":["Methods of testing inhalers"]},{"key":"others:access_status","label":"Others Access Status","values":["open.access"]},{"key":"sync:item.dbid","label":"Sync Item Dbid","values":["170647"]},{"key":"sync:item.dbtype","label":"Sync Item Dbtype","values":["ZP"]},{"key":"sync:item.insts","label":"Sync Item Insts","values":["2026.07.07 19:00:50"]},{"key":"sync:item.modts","label":"Sync Item Modts","values":["2026.07.07 12:47:44"]}]}],"canonical_facts":{"dc:contributor.advisor":["Lízal, František"],"dc:contributor.committee":["doc. Ing. František Lízal, Ph.D. (člen) doc. Ing. Vít Jan, Ph.D. (člen) doc. Ing. Jiří Šremr, Ph.D. (člen) prof. Ing. Pavel Novotný, Ph.D. (předseda) doc. Ing. Pavel Charvát, Ph.D. (místopředseda)"],"dc:contributor.committeemember":["Bělka, Miloslav"],"dc:creator":["Petrakis, Nikolaos"],"dc:date.acceptance":["2026-06-17"],"dc:date.created":["2026"],"dc:date.updated":["2026-06-18-08:46:54"],"dc:description.abstract":["Inhalation therapy is an effective route for drug administration to the respiratory system, and the reliable testing of inhalation devices is essential to ensure therapeutic efficacy. The first part of this thesis provides a theoretical overview of inhalation systems and their working principles, aerosol and particle transport fundamentals and the methods used for testing inhalers and nebulisers. The second part presents experimental measurements performed using the Aerogen Solo vibrating mesh nebuliser with fluorescently labelled particles. The aerodynamic particle size distribution was characterised using both the Next Generation Impactor and Andersen cascade Impactor, the regional deposition was assessed using a 3D printed airway replica of the human respiratory tract. The obtained results are discussed in terms of measurement methodology and compared with computational predictions from the MPPD model, highlighting both the potential and the current limitations of integrating airway replicas into standard inhaler testing."],"dc:description.grade":["B"],"dc:identifier.citation":["PETRAKIS, N. Metody testování inhalátorů [online]. Brno: Vysoké učení technické v Brně. Fakulta strojního inženýrství. 2026."],"dc:identifier.other":["170647"],"dc:identifier.uri":["http://hdl.handle.net/11012/259222"],"dc:language":["angličtina (English)"],"dc:language.iso":["en"],"dc:publisher":["Vysoké učení technické v Brně. Fakulta strojního inženýrství"],"dc:publisher.faculty":["Fakulta strojního inženýrství"],"dc:rights":["Standardní licenční smlouva - přístup k plnému textu bez omezení"],"dc:subject":["Inhalation therapy","inhaler","nebuliser","aerosol","aerodynamic particle size distribution","fine particle fraction","cascade impactor","airway replica","particle deposition","fluorescein"],"dc:title":["Metody testování inhalátorů"],"dc:title.alternative":["Methods of testing inhalers"],"dc:type":["Text","bachelorThesis","bakalářská práce"],"others:access_status":["open.access"],"sync:item.dbid":["170647"],"sync:item.dbtype":["ZP"],"sync:item.insts":["2026.07.07 19:00:50"],"sync:item.modts":["2026.07.07 12:47:44"],"thesis:defence_status":["práce byla úspěšně obhájena"],"thesis:degree_discipline":["bez specializace"],"thesis:degree_level":["Bakalářský"],"thesis:degree_program":["Fundamentals of Mechanical Engineering"],"thesis:examination_report":["The student presented the committee with the progress of the work, the results and conclusions of the thesis, and answered the reviewer's questions. Doc. Jan asked what causes two droplets to merge, once they are close together, what makes them coalesce into one. The student was uncertain, and Doc. Jan guided him towards surface energy, the tendency to minimise it as far as possible. Doc. Charvát asked about the filter, why it had been included when it caused such difficulties. The student explained that the reason was purely practical, to prevent particles from contaminating the downstream flow meter. Doc. Jan then asked about the fluorescence method, and the student explained how the measurement was carried out."],"thesis:institution_name":["Bc.","Vysoké učení technické v Brně. Fakulta strojního inženýrství. Energetický ústav"]},"updated_at":"2026-11-19T05:04:58Z"}