{"id":{"repo_id":"bu","oai_identifier":"oai:open.bu.edu:2144/53318"},"canonical_url":"https://search.ndltd.org/etd/bu/oai:open.bu.edu:2144/53318","repository":{"repo_id":"bu","name":"Boston University","base_url":"https://open.bu.edu/oai/request"},"display":{"title":"Assessment of optical coherence tomography image quality after repeated catheter insertions","abstract":"OBJECTIVES: To evaluate whether repeated insertion of an optical coherence tomography (OCT) catheter affects intracanal image quality, using signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR), and sharpness. MATERIALS AND METHODS: Three extracted single-rooted teeth with simulated intracanal defects were mounted in a fixed alignment block and imaged using an Abbott Dragonfly OPTIS catheter. Fifty OCT pullback scans were performed through the same three-tooth block without intermediate canal modification. Image quality was assessed at scans 1, 10, 20, 30, 40, and 50 using ImageJ with standardized regions of interest. RESULTS: SNR remained stable across all insertion cycles (mean range 2.55–2.99; p > 0.05). CNR fluctuated without a significant overall trend (mean range 0.120–0.310; p > 0.05). Sharpness deteriorated significantly with repeated use (p < 0.05), with mean edge-transition distance increasing from 2.55 ± 0.03 pixels at scan 1 to 10.20 ± 0.26 pixels at scan 40, with an abrupt rise across all three samples between scans 30 and 40. CONCLUSIONS: Repeated OCT catheter insertion preserved signal quality and contrast but produced significant sharpness deterioration after approximately 30–40 insertions. Repeated use may remain feasible for general visualization up to a threshold, but progressive loss of edge definition may reduce detection of fine anatomical features. Further studies with larger samples and wet canal conditions are needed","abstract_html":"OBJECTIVES: To evaluate whether repeated insertion of an optical coherence tomography (OCT) catheter affects intracanal image quality, using signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR), and sharpness. MATERIALS AND METHODS: Three extracted single-rooted teeth with simulated intracanal defects were mounted in a fixed alignment block and imaged using an Abbott Dragonfly OPTIS catheter. Fifty OCT pullback scans were performed through the same three-tooth block without intermediate canal modification. Image quality was assessed at scans 1, 10, 20, 30, 40, and 50 using ImageJ with standardized regions of interest. RESULTS: SNR remained stable across all insertion cycles (mean range 2.55–2.99; p &gt; 0.05). CNR fluctuated without a significant overall trend (mean range 0.120–0.310; p &gt; 0.05). Sharpness deteriorated significantly with repeated use (p &lt; 0.05), with mean edge-transition distance increasing from 2.55 ± 0.03 pixels at scan 1 to 10.20 ± 0.26 pixels at scan 40, with an abrupt rise across all three samples between scans 30 and 40. CONCLUSIONS: Repeated OCT catheter insertion preserved signal quality and contrast but produced significant sharpness deterioration after approximately 30–40 insertions. Repeated use may remain feasible for general visualization up to a threshold, but progressive loss of edge definition may reduce detection of fine anatomical features. Further studies with larger samples and wet canal conditions are needed","abstract_has_math":false,"creators":["Alzoobaee, Mohammed"],"institution":"Boston University","degree_name":"Master of Science in Design","degree_level":"masters","degree_discipline":"Endodontics","degree_department":null,"school":null,"contributors":[],"advisors":["Chogle, Sami"],"committee_chairs":[],"committee_members":[],"year":2026,"date_issued":"2026","date_published":"2026","updated_at":"2026-11-18T20:25:40Z","subjects":["Dentistry","Catheter durability","Endodontics","Image quality","Optical coherence tomography","Root canal imaging","Signal-to-noise ratio"],"languages":["en_US"],"rights":[],"rights_urls":[],"identifier_entries":[],"person_identifier_entries":[]},"links":{"outbound_url":"https://hdl.handle.net/2144/53318","outbound_label":"Handle","outbound_source":"dc:identifier.uri","record_url":"https://hdl.handle.net/2144/53318","record_label":"Repository record","record_kind":"handle","record_source":"dc:identifier.uri","download_url":null,"download_label":null,"download_source":null,"access_status":null,"embargo_until":null,"access_note":null},"source_record":{"url":"https://open.bu.edu/oai/request?verb=GetRecord&metadataPrefix=dim&identifier=oai%3Aopen.bu.edu%3A2144%2F53318","prefix":"dim"},"metadata_groups":[{"id":"people","label":"People","entries":[{"key":"dc:contributor.advisor","label":"Advisor","values":["Chogle, Sami"]},{"key":"dc:creator","label":"Author","values":["Alzoobaee, Mohammed"]}]},{"id":"academic_context","label":"Academic Context","entries":[{"key":"dc:date.accessioned","label":"Dc Date Accessioned","values":["2026-07-14T17:45:43Z"]},{"key":"dc:date.available","label":"Dc Date Available","values":["2026-07-14T17:45:43Z"]},{"key":"dc:date.issued","label":"Date","values":["2026"]},{"key":"dc:date.updated","label":"Dc Date Updated","values":["2026-07-14T16:15:40Z"]},{"key":"dc:type","label":"Dc Type","values":["Thesis/Dissertation"]},{"key":"thesis:degree_discipline","label":"Discipline","values":["Endodontics"]},{"key":"thesis:degree_level","label":"Degree Level","values":["masters"]},{"key":"thesis:degree_name","label":"Degree Name","values":["Master of Science in Design"]},{"key":"thesis:institution_name","label":"Thesis Institution Name","values":["Boston University"]}]},{"id":"subjects_keywords","label":"Subjects and Keywords","entries":[{"key":"dc:subject","label":"Dc Subject","values":["Dentistry","Catheter durability","Endodontics","Image quality","Optical coherence tomography","Root canal imaging","Signal-to-noise ratio"]}]},{"id":"language_rights","label":"Language and Rights","entries":[{"key":"dc:language.iso","label":"Language (ISO)","values":["en_US"]}]},{"id":"identifiers","label":"Identifiers","entries":[{"key":"dc:identifier.uri","label":"Identifier URI","values":["https://hdl.handle.net/2144/53318"]}]},{"id":"additional","label":"Additional Metadata","entries":[{"key":"dc:description","label":"Description","values":["2026"]},{"key":"dc:description.abstract","label":"Abstract","values":["OBJECTIVES: To evaluate whether repeated insertion of an optical coherence tomography (OCT) catheter affects intracanal image quality, using signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR), and sharpness. MATERIALS AND METHODS: Three extracted single-rooted teeth with simulated intracanal defects were mounted in a fixed alignment block and imaged using an Abbott Dragonfly OPTIS catheter. Fifty OCT pullback scans were performed through the same three-tooth block without intermediate canal modification. Image quality was assessed at scans 1, 10, 20, 30, 40, and 50 using ImageJ with standardized regions of interest. RESULTS: SNR remained stable across all insertion cycles (mean range 2.55–2.99; p > 0.05). CNR fluctuated without a significant overall trend (mean range 0.120–0.310; p > 0.05). Sharpness deteriorated significantly with repeated use (p < 0.05), with mean edge-transition distance increasing from 2.55 ± 0.03 pixels at scan 1 to 10.20 ± 0.26 pixels at scan 40, with an abrupt rise across all three samples between scans 30 and 40. CONCLUSIONS: Repeated OCT catheter insertion preserved signal quality and contrast but produced significant sharpness deterioration after approximately 30–40 insertions. Repeated use may remain feasible for general visualization up to a threshold, but progressive loss of edge definition may reduce detection of fine anatomical features. Further studies with larger samples and wet canal conditions are needed"]},{"key":"dc:title","label":"Title","values":["Assessment of optical coherence tomography image quality after repeated catheter insertions"]}]}],"canonical_facts":{"dc:contributor.advisor":["Chogle, Sami"],"dc:creator":["Alzoobaee, Mohammed"],"dc:date.accessioned":["2026-07-14T17:45:43Z"],"dc:date.available":["2026-07-14T17:45:43Z"],"dc:date.issued":["2026"],"dc:date.updated":["2026-07-14T16:15:40Z"],"dc:description":["2026"],"dc:description.abstract":["OBJECTIVES: To evaluate whether repeated insertion of an optical coherence tomography (OCT) catheter affects intracanal image quality, using signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR), and sharpness. MATERIALS AND METHODS: Three extracted single-rooted teeth with simulated intracanal defects were mounted in a fixed alignment block and imaged using an Abbott Dragonfly OPTIS catheter. Fifty OCT pullback scans were performed through the same three-tooth block without intermediate canal modification. Image quality was assessed at scans 1, 10, 20, 30, 40, and 50 using ImageJ with standardized regions of interest. RESULTS: SNR remained stable across all insertion cycles (mean range 2.55–2.99; p > 0.05). CNR fluctuated without a significant overall trend (mean range 0.120–0.310; p > 0.05). Sharpness deteriorated significantly with repeated use (p < 0.05), with mean edge-transition distance increasing from 2.55 ± 0.03 pixels at scan 1 to 10.20 ± 0.26 pixels at scan 40, with an abrupt rise across all three samples between scans 30 and 40. CONCLUSIONS: Repeated OCT catheter insertion preserved signal quality and contrast but produced significant sharpness deterioration after approximately 30–40 insertions. Repeated use may remain feasible for general visualization up to a threshold, but progressive loss of edge definition may reduce detection of fine anatomical features. Further studies with larger samples and wet canal conditions are needed"],"dc:identifier.uri":["https://hdl.handle.net/2144/53318"],"dc:language.iso":["en_US"],"dc:subject":["Dentistry","Catheter durability","Endodontics","Image quality","Optical coherence tomography","Root canal imaging","Signal-to-noise ratio"],"dc:title":["Assessment of optical coherence tomography image quality after repeated catheter insertions"],"dc:type":["Thesis/Dissertation"],"thesis:degree_discipline":["Endodontics"],"thesis:degree_level":["masters"],"thesis:degree_name":["Master of Science in Design"],"thesis:institution_name":["Boston University"]},"updated_at":"2026-11-18T20:25:40Z"}