{"id":{"repo_id":"buffalo","oai_identifier":"oai:ubir.buffalo.edu:10477/89964"},"canonical_url":"https://search.ndltd.org/etd/buffalo/oai:ubir.buffalo.edu:10477/89964","repository":{"repo_id":"buffalo","name":"Buffalo","base_url":"https://ubir.buffalo.edu/oai/request"},"display":{"title":"Trace Mixture Analysis with Standoff Photothermal Cantilever Deflection Spectroscopy","abstract":"Most chemical sensors work by collecting and transporting targeted chemicals for interaction with their sensing elements. When analyte molecules appear in very low concentrations, pre-concentrators are used for enriching the sample. This thesis describes a non-contact method which eliminates the need for time consuming pre-concentrator techniques for trace level chemical sensors. This technique combines the high selectivity of infrared spectroscopy with the high sensitivity of microcantilevers. The target molecules physisorbed on the surface are irradiated with tunable infrared radiation and the diffused reflected light is collected and focused on a thermally sensitive cantilever which bends sensitively as a function of the impinging IR beam intensity. The bending of the cantilever as a function of illumination wavelength therefore shows the IR absorption spectrum of the target molecules. Unlike in direct photothermal cantilever deflection spectroscopy, where physisorbed molecules on a bi-material cantilever are resonantly excited using tunable IR source for generating heat to bend the cantilever, the sensitivity of this method detailed in the present thesis is not limited by the area of the cantilever. To demonstrate the advantages, the non-contact IR spectra obtained with hazardous environmental chemicals and volatile organic compounds (VOC) are presented. It is shown that a limit of detection (LOD) of 4.5x1011 molecules/cm2 for deposited analytes can be achieved using the present method.","abstract_html":"Most chemical sensors work by collecting and transporting targeted chemicals for interaction with their sensing elements. When analyte molecules appear in very low concentrations, pre-concentrators are used for enriching the sample. This thesis describes a non-contact method which eliminates the need for time consuming pre-concentrator techniques for trace level chemical sensors. This technique combines the high selectivity of infrared spectroscopy with the high sensitivity of microcantilevers. The target molecules physisorbed on the surface are irradiated with tunable infrared radiation and the diffused reflected light is collected and focused on a thermally sensitive cantilever which bends sensitively as a function of the impinging IR beam intensity. The bending of the cantilever as a function of illumination wavelength therefore shows the IR absorption spectrum of the target molecules. Unlike in direct photothermal cantilever deflection spectroscopy, where physisorbed molecules on a bi-material cantilever are resonantly excited using tunable IR source for generating heat to bend the cantilever, the sensitivity of this method detailed in the present thesis is not limited by the area of the cantilever. To demonstrate the advantages, the non-contact IR spectra obtained with hazardous environmental chemicals and volatile organic compounds (VOC) are presented. It is shown that a limit of detection (LOD) of 4.5x1011 molecules/cm2 for deposited analytes can be achieved using the present method.","abstract_has_math":false,"creators":["Leatt, Kyle; 0009-0009-1511-2340"],"institution":"State University of New York at Buffalo","degree_name":null,"degree_level":null,"degree_discipline":null,"degree_department":null,"school":null,"contributors":["Thundat, Thomas","Chemical and Biological Engineering"],"advisors":[],"committee_chairs":[],"committee_members":[],"year":2024,"date_issued":"2024","date_published":"2024","updated_at":"2026-11-19T07:10:53Z","subjects":["nanotechnology","analytical chemistry","chemical engineering"],"languages":["eng"],"rights":["Copyright retained by author. Users of works found in University at Buffalo Institutional Repository (UBIR) are responsible for identifying and contacting the copyright owner for permission to reuse. University at Buffalo Libraries do not manage rights for copyright-protected works and cannot assist with permissions."],"rights_urls":[],"identifier_entries":[],"person_identifier_entries":[]},"links":{"outbound_url":"https://hdl.handle.net/10477/89964","outbound_label":"Handle","outbound_source":"dc:identifier","record_url":"https://hdl.handle.net/10477/89964","record_label":"Repository record","record_kind":"handle","record_source":"dc:identifier","access_status":null,"embargo_until":null,"access_note":null},"oai_pmh_record":{"url":"https://union.ndltd.org/oai-pmh/?verb=GetRecord&metadataPrefix=oai_dc&identifier=oai%3Aunion.ndltd.org%3Arepo%3Abuffalo%3Aoai%3Aubir.buffalo.edu%3A10477%2F89964","prefix":"oai_dc"},"source_record":{"url":"https://ubir.buffalo.edu/oai/request?verb=GetRecord&metadataPrefix=qdc&identifier=oai%3Aubir.buffalo.edu%3A10477%2F89964","prefix":"qdc"},"metadata_groups":[{"id":"people","label":"People","entries":[{"key":"dc:contributor","label":"Contributor","values":["Thundat, Thomas","Chemical and Biological Engineering"]},{"key":"dc:creator","label":"Author","values":["Leatt, Kyle; 0009-0009-1511-2340"]}]},{"id":"academic_context","label":"Academic Context","entries":[{"key":"dc:date","label":"Dc Date","values":["2026-01-13T17:58:01Z"]},{"key":"dc:date.available","label":"Dc Date Available","values":["2026-01-13T17:58:01Z"]},{"key":"dc:date.created","label":"Dc Date Created","values":["2026-01-13T17:58:01Z"]},{"key":"dc:date.issued","label":"Date","values":["2024"]},{"key":"dc:publisher","label":"Institution","values":["State University of New York at Buffalo"]},{"key":"dc:type","label":"Dc Type","values":["Text","Thesis"]}]},{"id":"subjects_keywords","label":"Subjects and Keywords","entries":[{"key":"dc:subject","label":"Dc Subject","values":["nanotechnology","analytical chemistry","chemical engineering"]}]},{"id":"language_rights","label":"Language and Rights","entries":[{"key":"dc:language","label":"Dc Language","values":["eng"]},{"key":"dc:rights","label":"Dc Rights","values":["Copyright retained by author. Users of works found in University at Buffalo Institutional Repository (UBIR) are responsible for identifying and contacting the copyright owner for permission to reuse. University at Buffalo Libraries do not manage rights for copyright-protected works and cannot assist with permissions."]}]},{"id":"identifiers","label":"Identifiers","entries":[{"key":"dc:identifier","label":"Identifier","values":["http://hdl.handle.net/10477/89964"]}]},{"id":"additional","label":"Additional Metadata","entries":[{"key":"dc:description.abstract","label":"Abstract","values":["Most chemical sensors work by collecting and transporting targeted chemicals for interaction with their sensing elements. When analyte molecules appear in very low concentrations, pre-concentrators are used for enriching the sample. This thesis describes a non-contact method which eliminates the need for time consuming pre-concentrator techniques for trace level chemical sensors. This technique combines the high selectivity of infrared spectroscopy with the high sensitivity of microcantilevers. The target molecules physisorbed on the surface are irradiated with tunable infrared radiation and the diffused reflected light is collected and focused on a thermally sensitive cantilever which bends sensitively as a function of the impinging IR beam intensity. The bending of the cantilever as a function of illumination wavelength therefore shows the IR absorption spectrum of the target molecules. Unlike in direct photothermal cantilever deflection spectroscopy, where physisorbed molecules on a bi-material cantilever are resonantly excited using tunable IR source for generating heat to bend the cantilever, the sensitivity of this method detailed in the present thesis is not limited by the area of the cantilever. To demonstrate the advantages, the non-contact IR spectra obtained with hazardous environmental chemicals and volatile organic compounds (VOC) are presented. It is shown that a limit of detection (LOD) of 4.5x1011 molecules/cm2 for deposited analytes can be achieved using the present method.","**To request an accessible version of the file(s) associated with this item, contact library@buffalo.edu. Please include the item's persistent URL [http://hdl.handle.net/. . .] in your request.**"]},{"key":"dc:title","label":"Title","values":["Trace Mixture Analysis with Standoff Photothermal Cantilever Deflection Spectroscopy"]}]}],"canonical_facts":{"dc:contributor":["Thundat, Thomas","Chemical and Biological Engineering"],"dc:creator":["Leatt, Kyle; 0009-0009-1511-2340"],"dc:date":["2026-01-13T17:58:01Z"],"dc:date.available":["2026-01-13T17:58:01Z"],"dc:date.created":["2026-01-13T17:58:01Z"],"dc:date.issued":["2024"],"dc:description.abstract":["Most chemical sensors work by collecting and transporting targeted chemicals for interaction with their sensing elements. When analyte molecules appear in very low concentrations, pre-concentrators are used for enriching the sample. This thesis describes a non-contact method which eliminates the need for time consuming pre-concentrator techniques for trace level chemical sensors. This technique combines the high selectivity of infrared spectroscopy with the high sensitivity of microcantilevers. The target molecules physisorbed on the surface are irradiated with tunable infrared radiation and the diffused reflected light is collected and focused on a thermally sensitive cantilever which bends sensitively as a function of the impinging IR beam intensity. The bending of the cantilever as a function of illumination wavelength therefore shows the IR absorption spectrum of the target molecules. Unlike in direct photothermal cantilever deflection spectroscopy, where physisorbed molecules on a bi-material cantilever are resonantly excited using tunable IR source for generating heat to bend the cantilever, the sensitivity of this method detailed in the present thesis is not limited by the area of the cantilever. To demonstrate the advantages, the non-contact IR spectra obtained with hazardous environmental chemicals and volatile organic compounds (VOC) are presented. It is shown that a limit of detection (LOD) of 4.5x1011 molecules/cm2 for deposited analytes can be achieved using the present method.","**To request an accessible version of the file(s) associated with this item, contact library@buffalo.edu. Please include the item's persistent URL [http://hdl.handle.net/. . .] in your request.**"],"dc:identifier":["http://hdl.handle.net/10477/89964"],"dc:language":["eng"],"dc:publisher":["State University of New York at Buffalo"],"dc:rights":["Copyright retained by author. Users of works found in University at Buffalo Institutional Repository (UBIR) are responsible for identifying and contacting the copyright owner for permission to reuse. University at Buffalo Libraries do not manage rights for copyright-protected works and cannot assist with permissions."],"dc:subject":["nanotechnology","analytical chemistry","chemical engineering"],"dc:title":["Trace Mixture Analysis with Standoff Photothermal Cantilever Deflection Spectroscopy"],"dc:type":["Text","Thesis"]},"updated_at":"2026-11-19T07:10:53Z"}