{"id":{"repo_id":"calgary","oai_identifier":"oai:ucalgary.scholaris.ca:11023/2582"},"canonical_url":"https://search.ndltd.org/etd/calgary/oai:ucalgary.scholaris.ca:11023/2582","repository":{"repo_id":"calgary","name":"University of Calgary","base_url":"https://ucalgary.scholaris.ca/server/oai/request"},"display":{"title":"Ultra-Wideband Microwave Cardiac Output Monitoring: Approaches to Modeling the Heart at Microwave Frequencies and Validation of a Proof of Concept System","abstract":"Cardiac output is an important hemodynamic parameter in critical care and has proven difficult to measure accurately, non-invasively, and without a highly skilled operator. Microwave technologies have been shown to be a possible alternative to current methods for cardiac output determination. However, previously reported microwave systems had limitations including low signal change over the cardiac cycle. In this thesis, we specifically address improving signal change over the cardiac cycle by introducing an ultra-wideband transmission system. We characterize the system using heart models and show that wave propagation through the chest is complex, making the system sensitive to antenna placement. Additionally, a semi-dynamic heart model for characterization of the system in hardware and in simulation is developed. This model shows comparable performance to a complex MRI based heart model. On a human test subject, ultra-wideband readings showed up to 6 dB of magnitude variation over the cardiac cycle.","abstract_html":"Cardiac output is an important hemodynamic parameter in critical care and has proven difficult to measure accurately, non-invasively, and without a highly skilled operator. Microwave technologies have been shown to be a possible alternative to current methods for cardiac output determination. However, previously reported microwave systems had limitations including low signal change over the cardiac cycle. In this thesis, we specifically address improving signal change over the cardiac cycle by introducing an ultra-wideband transmission system. We characterize the system using heart models and show that wave propagation through the chest is complex, making the system sensitive to antenna placement. Additionally, a semi-dynamic heart model for characterization of the system in hardware and in simulation is developed. This model shows comparable performance to a complex MRI based heart model. On a human test subject, ultra-wideband readings showed up to 6 dB of magnitude variation over the cardiac cycle.","abstract_has_math":false,"creators":["Seguin, Marcel"],"institution":"Graduate Studies","degree_name":"Master of Science (MSc)","degree_level":null,"degree_discipline":"Electrical and Computer Engineering","degree_department":null,"school":null,"contributors":[],"advisors":["Okoniewski, Michal"],"committee_chairs":[],"committee_members":[],"year":2015,"date_issued":"2015-10-01","date_published":"2015-10-01","updated_at":"2026-11-19T07:23:41Z","subjects":["Engineering--Biomedical","Engineering--Electronics and Electrical"],"languages":["eng"],"rights":["University of Calgary graduate students retain copyright ownership and moral rights for their thesis. You may use this material in any way that is permitted by the Copyright Act or through licensing that has been assigned to the document. For uses that are not allowable under copyright legislation or licensing, you are required to seek permission."],"rights_urls":[],"identifier_entries":[{"key":"dc:identifier.uri","label":"Identifier URI","values":["http://hdl.handle.net/11023/2582"],"render_values":[{"text":"http://hdl.handle.net/11023/2582","href":"https://hdl.handle.net/11023/2582","code":true}]}],"person_identifier_entries":[]},"links":{"outbound_url":"https://doi.org/10.11575/PRISM/28549","outbound_label":"DOI","outbound_source":"dc:identifier.doi","record_url":"https://doi.org/10.11575/PRISM/28549","record_label":"Repository record","record_kind":"doi","record_source":"dc:identifier.doi","access_status":null,"embargo_until":null,"access_note":null},"oai_pmh_record":{"url":"https://union.ndltd.org/oai-pmh/?verb=GetRecord&metadataPrefix=oai_dc&identifier=oai%3Aunion.ndltd.org%3Arepo%3Acalgary%3Aoai%3Aucalgary.scholaris.ca%3A11023%2F2582","prefix":"oai_dc"},"source_record":{"url":"https://ucalgary.scholaris.ca/server/oai/request?verb=GetRecord&metadataPrefix=dim&identifier=oai%3Aucalgary.scholaris.ca%3A11023%2F2582","prefix":"dim"},"metadata_groups":[{"id":"people","label":"People","entries":[{"key":"dc:contributor.advisor","label":"Advisor","values":["Okoniewski, Michal"]},{"key":"dc:creator","label":"Author","values":["Seguin, Marcel"]}]},{"id":"academic_context","label":"Academic Context","entries":[{"key":"dc:date.accessioned","label":"Dc Date Accessioned","values":["2015-10-01T17:18:36Z"]},{"key":"dc:date.issued","label":"Date","values":["2015-10-01"]},{"key":"dc:date.liftdate","label":"Dc Date Liftdate","values":["2016-02-22T07:00:00Z"]},{"key":"dc:date.submitted","label":"Dc Date Submitted","values":["2015"]},{"key":"dc:publisher.faculty","label":"Dc Publisher Faculty","values":["Graduate Studies"]},{"key":"dc:publisher.institution","label":"Dc Publisher Institution","values":["University of Calgary"]},{"key":"dc:publisher.place","label":"Dc Publisher Place","values":["Calgary"]},{"key":"dc:type","label":"Dc Type","values":["master thesis"]},{"key":"thesis:degree_discipline","label":"Discipline","values":["Electrical and Computer Engineering"]},{"key":"thesis:degree_name","label":"Degree Name","values":["Master of Science (MSc)"]},{"key":"thesis:institution_name","label":"Thesis Institution Name","values":["University of Calgary"]}]},{"id":"subjects_keywords","label":"Subjects and Keywords","entries":[{"key":"dc:subject","label":"Dc Subject","values":["Engineering--Biomedical","Engineering--Electronics and Electrical"]},{"key":"dc:subject.classification","label":"Dc Subject Classification","values":["Cardiac Output","Microwave Monitoring","Ultra-Wideband"]}]},{"id":"language_rights","label":"Language and Rights","entries":[{"key":"dc:language.iso","label":"Language (ISO)","values":["eng"]},{"key":"dc:rights","label":"Dc Rights","values":["University of Calgary graduate students retain copyright ownership and moral rights for their thesis. You may use this material in any way that is permitted by the Copyright Act or through licensing that has been assigned to the document. For uses that are not allowable under copyright legislation or licensing, you are required to seek permission."]}]},{"id":"identifiers","label":"Identifiers","entries":[{"key":"dc:identifier.citation","label":"Dc Identifier Citation","values":["Seguin, M. (2015). Ultra-Wideband Microwave Cardiac Output Monitoring: Approaches to Modeling the Heart at Microwave Frequencies and Validation of a Proof of Concept System (Master's thesis, University of Calgary, Calgary, Canada). Retrieved from https://prism.ucalgary.ca. doi:10.11575/PRISM/28549"]},{"key":"dc:identifier.doi","label":"DOI","values":["http://dx.doi.org/10.11575/PRISM/28549"]},{"key":"dc:identifier.uri","label":"Identifier URI","values":["http://hdl.handle.net/11023/2582"]}]},{"id":"additional","label":"Additional Metadata","entries":[{"key":"atmire:migration.oldid","label":"Atmire Migration Oldid","values":["3695"]},{"key":"dc:description.abstract","label":"Abstract","values":["Cardiac output is an important hemodynamic parameter in critical care and has proven difficult to measure accurately, non-invasively, and without a highly skilled operator. Microwave technologies have been shown to be a possible alternative to current methods for cardiac output determination. However, previously reported microwave systems had limitations including low signal change over the cardiac cycle. In this thesis, we specifically address improving signal change over the cardiac cycle by introducing an ultra-wideband transmission system. We characterize the system using heart models and show that wave propagation through the chest is complex, making the system sensitive to antenna placement. Additionally, a semi-dynamic heart model for characterization of the system in hardware and in simulation is developed. This model shows comparable performance to a complex MRI based heart model. On a human test subject, ultra-wideband readings showed up to 6 dB of magnitude variation over the cardiac cycle."]},{"key":"dc:title","label":"Title","values":["Ultra-Wideband Microwave Cardiac Output Monitoring: Approaches to Modeling the Heart at Microwave Frequencies and Validation of a Proof of Concept System"]},{"key":"dspace:entity.type","label":"Dspace Entity Type","values":["Publication"]},{"key":"local:request.copy","label":"Local Request Copy","values":["true"]},{"key":"others:access_status","label":"Others Access Status","values":["open.access"]}]}],"canonical_facts":{"atmire:migration.oldid":["3695"],"dc:contributor.advisor":["Okoniewski, Michal"],"dc:creator":["Seguin, Marcel"],"dc:date.accessioned":["2015-10-01T17:18:36Z"],"dc:date.issued":["2015-10-01"],"dc:date.liftdate":["2016-02-22T07:00:00Z"],"dc:date.submitted":["2015"],"dc:description.abstract":["Cardiac output is an important hemodynamic parameter in critical care and has proven difficult to measure accurately, non-invasively, and without a highly skilled operator. Microwave technologies have been shown to be a possible alternative to current methods for cardiac output determination. However, previously reported microwave systems had limitations including low signal change over the cardiac cycle. In this thesis, we specifically address improving signal change over the cardiac cycle by introducing an ultra-wideband transmission system. We characterize the system using heart models and show that wave propagation through the chest is complex, making the system sensitive to antenna placement. Additionally, a semi-dynamic heart model for characterization of the system in hardware and in simulation is developed. This model shows comparable performance to a complex MRI based heart model. On a human test subject, ultra-wideband readings showed up to 6 dB of magnitude variation over the cardiac cycle."],"dc:identifier.citation":["Seguin, M. (2015). Ultra-Wideband Microwave Cardiac Output Monitoring: Approaches to Modeling the Heart at Microwave Frequencies and Validation of a Proof of Concept System (Master's thesis, University of Calgary, Calgary, Canada). Retrieved from https://prism.ucalgary.ca. doi:10.11575/PRISM/28549"],"dc:identifier.doi":["http://dx.doi.org/10.11575/PRISM/28549"],"dc:identifier.uri":["http://hdl.handle.net/11023/2582"],"dc:language.iso":["eng"],"dc:publisher.faculty":["Graduate Studies"],"dc:publisher.institution":["University of Calgary"],"dc:publisher.place":["Calgary"],"dc:rights":["University of Calgary graduate students retain copyright ownership and moral rights for their thesis. You may use this material in any way that is permitted by the Copyright Act or through licensing that has been assigned to the document. For uses that are not allowable under copyright legislation or licensing, you are required to seek permission."],"dc:subject":["Engineering--Biomedical","Engineering--Electronics and Electrical"],"dc:subject.classification":["Cardiac Output","Microwave Monitoring","Ultra-Wideband"],"dc:title":["Ultra-Wideband Microwave Cardiac Output Monitoring: Approaches to Modeling the Heart at Microwave Frequencies and Validation of a Proof of Concept System"],"dc:type":["master thesis"],"dspace:entity.type":["Publication"],"local:request.copy":["true"],"others:access_status":["open.access"],"thesis:degree_discipline":["Electrical and Computer Engineering"],"thesis:degree_name":["Master of Science (MSc)"],"thesis:institution_name":["University of Calgary"]},"updated_at":"2026-11-19T07:23:41Z"}