{"id":{"repo_id":"calgary","oai_identifier":"oai:ucalgary.scholaris.ca:11023/3484"},"canonical_url":"https://search.ndltd.org/etd/calgary/oai:ucalgary.scholaris.ca:11023/3484","repository":{"repo_id":"calgary","name":"University of Calgary","base_url":"https://ucalgary.scholaris.ca/server/oai/request"},"display":{"title":"Characterization of Planar Titanium Platforms for Flame Ionization Detection in Microfluidic Gas Chromatography","abstract":"Gas chromatography (GC) with a Flame Ionization Detector (FID) has been widely used for the analysis of volatile organic compounds in various fields. Instrument portability and miniaturization of the FID for micro GC (μGC) applications are of increasing interest. This thesis describes the design and characterization of a planar on-chip counter current micro FID (μ-FID). The μ-FID fabricated in both quartz and Titanium (Ti) substrates were characterized and the Ti μ-FID showed great improvements in detection limit and sensitivity over other prototypes. This Ti μ-FID design is further integrated with a Ti separation column to fabricate a planar μGC-FID Ti tile for the first time. This Ti μGC-FID tile is explored for its detector performance attributes and some routine GC applications are examined. This novel Ti μGC-FID tile, with an on-chip single monolithic design, is found to provide a very low detection limit (9 pgC/s) and a high sensitivity (60 mC/gC).","abstract_html":"Gas chromatography (GC) with a Flame Ionization Detector (FID) has been widely used for the analysis of volatile organic compounds in various fields. Instrument portability and miniaturization of the FID for micro GC (μGC) applications are of increasing interest. This thesis describes the design and characterization of a planar on-chip counter current micro FID (μ-FID). The μ-FID fabricated in both quartz and Titanium (Ti) substrates were characterized and the Ti μ-FID showed great improvements in detection limit and sensitivity over other prototypes. This Ti μ-FID design is further integrated with a Ti separation column to fabricate a planar μGC-FID Ti tile for the first time. This Ti μGC-FID tile is explored for its detector performance attributes and some routine GC applications are examined. This novel Ti μGC-FID tile, with an on-chip single monolithic design, is found to provide a very low detection limit (9 pgC/s) and a high sensitivity (60 mC/gC).","abstract_has_math":false,"creators":["Raut, Rahul Pandurang"],"institution":"Graduate Studies","degree_name":"Master of Science (MSc)","degree_level":null,"degree_discipline":"Chemistry","degree_department":null,"school":null,"contributors":[],"advisors":["Thurbide, Kevin"],"committee_chairs":[],"committee_members":["Hinman, Allen Scott","Gailer, Jürgen","Marriott, Robert"],"year":2016,"date_issued":"2016","date_published":"2016","updated_at":"2026-11-19T07:30:27Z","subjects":["Chemistry--Analytical"],"languages":["eng"],"rights":["University of Calgary graduate students retain copyright ownership and moral rights for their thesis. You may use this material in any way that is permitted by the Copyright Act or through licensing that has been assigned to the document. For uses that are not allowable under copyright legislation or licensing, you are required to seek permission."],"rights_urls":[],"identifier_entries":[{"key":"dc:identifier.uri","label":"Identifier URI","values":["http://hdl.handle.net/11023/3484"],"render_values":[{"text":"http://hdl.handle.net/11023/3484","href":"https://hdl.handle.net/11023/3484","code":true}]}],"person_identifier_entries":[]},"links":{"outbound_url":"https://doi.org/10.11575/PRISM/25286","outbound_label":"DOI","outbound_source":"dc:identifier.doi","record_url":"https://doi.org/10.11575/PRISM/25286","record_label":"Repository record","record_kind":"doi","record_source":"dc:identifier.doi","access_status":null,"embargo_until":null,"access_note":null},"oai_pmh_record":{"url":"https://union.ndltd.org/oai-pmh/?verb=GetRecord&metadataPrefix=oai_dc&identifier=oai%3Aunion.ndltd.org%3Arepo%3Acalgary%3Aoai%3Aucalgary.scholaris.ca%3A11023%2F3484","prefix":"oai_dc"},"source_record":{"url":"https://ucalgary.scholaris.ca/server/oai/request?verb=GetRecord&metadataPrefix=dim&identifier=oai%3Aucalgary.scholaris.ca%3A11023%2F3484","prefix":"dim"},"metadata_groups":[{"id":"people","label":"People","entries":[{"key":"dc:contributor.advisor","label":"Advisor","values":["Thurbide, Kevin"]},{"key":"dc:contributor.committeemember","label":"Committee Member","values":["Hinman, Allen Scott","Gailer, Jürgen","Marriott, Robert"]},{"key":"dc:creator","label":"Author","values":["Raut, Rahul Pandurang"]}]},{"id":"academic_context","label":"Academic Context","entries":[{"key":"dc:date.accessioned","label":"Dc Date Accessioned","values":["2016-12-13T16:06:40Z"]},{"key":"dc:date.available","label":"Dc Date Available","values":["2016-12-13T16:06:40Z"]},{"key":"dc:date.issued","label":"Date","values":["2016"]},{"key":"dc:date.submitted","label":"Dc Date Submitted","values":["2016"]},{"key":"dc:publisher.faculty","label":"Dc Publisher Faculty","values":["Graduate Studies"]},{"key":"dc:publisher.institution","label":"Dc Publisher Institution","values":["University of Calgary"]},{"key":"dc:publisher.place","label":"Dc Publisher Place","values":["Calgary"]},{"key":"dc:type","label":"Dc Type","values":["master thesis"]},{"key":"thesis:degree_discipline","label":"Discipline","values":["Chemistry"]},{"key":"thesis:degree_name","label":"Degree Name","values":["Master of Science (MSc)"]},{"key":"thesis:institution_name","label":"Thesis Institution Name","values":["University of Calgary"]}]},{"id":"subjects_keywords","label":"Subjects and Keywords","entries":[{"key":"dc:subject","label":"Dc Subject","values":["Chemistry--Analytical"]},{"key":"dc:subject.classification","label":"Dc Subject Classification","values":["Miniature","Gas Chromatography","Micro","Flame Ionization Detection","Titanium"]}]},{"id":"language_rights","label":"Language and Rights","entries":[{"key":"dc:language.iso","label":"Language (ISO)","values":["eng"]},{"key":"dc:rights","label":"Dc Rights","values":["University of Calgary graduate students retain copyright ownership and moral rights for their thesis. You may use this material in any way that is permitted by the Copyright Act or through licensing that has been assigned to the document. For uses that are not allowable under copyright legislation or licensing, you are required to seek permission."]}]},{"id":"identifiers","label":"Identifiers","entries":[{"key":"dc:identifier.citation","label":"Dc Identifier Citation","values":["Raut, R. P. (2016). Characterization of Planar Titanium Platforms for Flame Ionization Detection in Microfluidic Gas Chromatography (Master's thesis, University of Calgary, Calgary, Canada). Retrieved from https://prism.ucalgary.ca. doi:10.11575/PRISM/25286"]},{"key":"dc:identifier.doi","label":"DOI","values":["http://dx.doi.org/10.11575/PRISM/25286"]},{"key":"dc:identifier.uri","label":"Identifier URI","values":["http://hdl.handle.net/11023/3484"]}]},{"id":"additional","label":"Additional Metadata","entries":[{"key":"atmire:migration.oldid","label":"Atmire Migration Oldid","values":["5148"]},{"key":"dc:description.abstract","label":"Abstract","values":["Gas chromatography (GC) with a Flame Ionization Detector (FID) has been widely used for the analysis of volatile organic compounds in various fields. Instrument portability and miniaturization of the FID for micro GC (μGC) applications are of increasing interest. This thesis describes the design and characterization of a planar on-chip counter current micro FID (μ-FID). The μ-FID fabricated in both quartz and Titanium (Ti) substrates were characterized and the Ti μ-FID showed great improvements in detection limit and sensitivity over other prototypes. This Ti μ-FID design is further integrated with a Ti separation column to fabricate a planar μGC-FID Ti tile for the first time. This Ti μGC-FID tile is explored for its detector performance attributes and some routine GC applications are examined. This novel Ti μGC-FID tile, with an on-chip single monolithic design, is found to provide a very low detection limit (9 pgC/s) and a high sensitivity (60 mC/gC)."]},{"key":"dc:title","label":"Title","values":["Characterization of Planar Titanium Platforms for Flame Ionization Detection in Microfluidic Gas Chromatography"]},{"key":"dspace:entity.type","label":"Dspace Entity Type","values":["Publication"]},{"key":"local:request.copy","label":"Local Request Copy","values":["true"]},{"key":"others:access_status","label":"Others Access Status","values":["open.access"]}]}],"canonical_facts":{"atmire:migration.oldid":["5148"],"dc:contributor.advisor":["Thurbide, Kevin"],"dc:contributor.committeemember":["Hinman, Allen Scott","Gailer, Jürgen","Marriott, Robert"],"dc:creator":["Raut, Rahul Pandurang"],"dc:date.accessioned":["2016-12-13T16:06:40Z"],"dc:date.available":["2016-12-13T16:06:40Z"],"dc:date.issued":["2016"],"dc:date.submitted":["2016"],"dc:description.abstract":["Gas chromatography (GC) with a Flame Ionization Detector (FID) has been widely used for the analysis of volatile organic compounds in various fields. Instrument portability and miniaturization of the FID for micro GC (μGC) applications are of increasing interest. This thesis describes the design and characterization of a planar on-chip counter current micro FID (μ-FID). The μ-FID fabricated in both quartz and Titanium (Ti) substrates were characterized and the Ti μ-FID showed great improvements in detection limit and sensitivity over other prototypes. This Ti μ-FID design is further integrated with a Ti separation column to fabricate a planar μGC-FID Ti tile for the first time. This Ti μGC-FID tile is explored for its detector performance attributes and some routine GC applications are examined. This novel Ti μGC-FID tile, with an on-chip single monolithic design, is found to provide a very low detection limit (9 pgC/s) and a high sensitivity (60 mC/gC)."],"dc:identifier.citation":["Raut, R. P. (2016). Characterization of Planar Titanium Platforms for Flame Ionization Detection in Microfluidic Gas Chromatography (Master's thesis, University of Calgary, Calgary, Canada). Retrieved from https://prism.ucalgary.ca. doi:10.11575/PRISM/25286"],"dc:identifier.doi":["http://dx.doi.org/10.11575/PRISM/25286"],"dc:identifier.uri":["http://hdl.handle.net/11023/3484"],"dc:language.iso":["eng"],"dc:publisher.faculty":["Graduate Studies"],"dc:publisher.institution":["University of Calgary"],"dc:publisher.place":["Calgary"],"dc:rights":["University of Calgary graduate students retain copyright ownership and moral rights for their thesis. You may use this material in any way that is permitted by the Copyright Act or through licensing that has been assigned to the document. For uses that are not allowable under copyright legislation or licensing, you are required to seek permission."],"dc:subject":["Chemistry--Analytical"],"dc:subject.classification":["Miniature","Gas Chromatography","Micro","Flame Ionization Detection","Titanium"],"dc:title":["Characterization of Planar Titanium Platforms for Flame Ionization Detection in Microfluidic Gas Chromatography"],"dc:type":["master thesis"],"dspace:entity.type":["Publication"],"local:request.copy":["true"],"others:access_status":["open.access"],"thesis:degree_discipline":["Chemistry"],"thesis:degree_name":["Master of Science (MSc)"],"thesis:institution_name":["University of Calgary"]},"updated_at":"2026-11-19T07:30:27Z"}