{"id":{"repo_id":"calpoly","oai_identifier":"oai:digitalcommons.calpoly.edu:theses-4018"},"canonical_url":"https://search.ndltd.org/etd/calpoly/oai:digitalcommons.calpoly.edu:theses-4018","repository":{"repo_id":"calpoly","name":"Cal Poly","base_url":"https://digitalcommons.calpoly.edu/do/oai/"},"display":{"title":"Three-Dimensional Probabilistic Finite Element Model of a Cemented Hip Implant for Brittle Fracture","abstract":"Hip implant fractures are rare yet severe complications that follow total hip arthroplasty. These fractures often occur at the implant stem for cemented hip implants due to proximal debonding of the bone cement. While deterministic models offer some predictive power of implant durability, they often are based on average properties, failing to capture the variability in properties that can be drastically different from the mean. This study developed a probabilistic finite element model of a debonded total hip replacement to better account for this variability to determine the most likely failure mode in a particular case. The hypothesis was that fatigue was the more likely cause of failure than overloading. The model used Monte Carlo sampling to create 1000 possible simulations varying in material stiffness and loading magnitude. The analysis found virtually no possibility of failure due to overloading. In contrast, failure due to fatigue had a 99.36% chance of occurring by comparing the maximum stresses recorded at the stem for these simulations to distributions of failure criteria based on literature. Fracture mechanics predicted that the implant in this loading configuration would fail between 0.15-2.2 years with a mean life of 0.6 years. The model had good agreement with the case study outcomes, particularly at predicting the failure site and durability of the implant. Moreover, altering assumptions only moderately changed the probabilistic analysis results compared to a deterministic analysis. Future work could be spent adjusting the model to match the cement implant interface or applying the probabilistic approach to other stages of the implant’s degradation.","abstract_html":"Hip implant fractures are rare yet severe complications that follow total hip arthroplasty. These fractures often occur at the implant stem for cemented hip implants due to proximal debonding of the bone cement. While deterministic models offer some predictive power of implant durability, they often are based on average properties, failing to capture the variability in properties that can be drastically different from the mean. This study developed a probabilistic finite element model of a debonded total hip replacement to better account for this variability to determine the most likely failure mode in a particular case. The hypothesis was that fatigue was the more likely cause of failure than overloading. The model used Monte Carlo sampling to create 1000 possible simulations varying in material stiffness and loading magnitude. The analysis found virtually no possibility of failure due to overloading. In contrast, failure due to fatigue had a 99.36% chance of occurring by comparing the maximum stresses recorded at the stem for these simulations to distributions of failure criteria based on literature. Fracture mechanics predicted that the implant in this loading configuration would fail between 0.15-2.2 years with a mean life of 0.6 years. The model had good agreement with the case study outcomes, particularly at predicting the failure site and durability of the implant. Moreover, altering assumptions only moderately changed the probabilistic analysis results compared to a deterministic analysis. Future work could be spent adjusting the model to match the cement implant interface or applying the probabilistic approach to other stages of the implant’s degradation.","abstract_has_math":false,"creators":["Truong, Daniel J"],"institution":null,"degree_name":"MS in Biomedical Engineering","degree_level":null,"degree_discipline":"Biomedical Engineering","degree_department":null,"school":null,"contributors":["Lanny Griffin","Biomedical Engineering","College of Engineering"],"advisors":[],"committee_chairs":[],"committee_members":[],"year":null,"date_issued":"","date_published":null,"updated_at":"2026-11-19T07:11:08Z","subjects":["Biomaterials","Biomechanical Engineering","Biomechanics and Biotransport","Biomedical Devices and Instrumentation","Computer-Aided Engineering and Design"],"languages":[],"rights":[],"rights_urls":[],"identifier_entries":[{"key":"dc:identifier","label":"Identifier","values":["10.15368/theses.2022.19"],"render_values":[{"text":"10.15368/theses.2022.19","href":"https://doi.org/10.15368/theses.2022.19","code":true}]}],"person_identifier_entries":[]},"links":{"outbound_url":"https://digitalcommons.calpoly.edu/theses/2390","outbound_label":"Repository record","outbound_source":"dc:identifier","record_url":"https://digitalcommons.calpoly.edu/theses/2390","record_label":"Repository record","record_kind":"record","record_source":"dc:identifier","download_url":null,"download_label":null,"download_source":null,"access_status":null,"embargo_until":null,"access_note":null},"source_record":{"url":"https://digitalcommons.calpoly.edu/do/oai/?verb=GetRecord&metadataPrefix=oai_etdms&identifier=oai%3Adigitalcommons.calpoly.edu%3Atheses-4018","prefix":"oai_etdms"},"metadata_groups":[{"id":"people","label":"People","entries":[{"key":"dc:contributor","label":"Contributor","values":["Lanny Griffin","Biomedical Engineering","College of Engineering"]},{"key":"dc:creator","label":"Author","values":["Truong, Daniel J"]}]},{"id":"academic_context","label":"Academic Context","entries":[{"key":"dc:date.available","label":"Dc Date Available","values":["2022-03-23T07:00:00Z"]},{"key":"dc:date.created","label":"Dc Date Created","values":["2022-03-01T08:00:00Z"]},{"key":"thesis:degree_discipline","label":"Discipline","values":["Biomedical Engineering"]},{"key":"thesis:degree_name","label":"Degree Name","values":["MS in Biomedical Engineering"]}]},{"id":"subjects_keywords","label":"Subjects and Keywords","entries":[{"key":"dc:subject","label":"Dc Subject","values":["Biomaterials","Biomechanical Engineering","Biomechanics and Biotransport","Biomedical Devices and Instrumentation","Computer-Aided Engineering and Design"]}]},{"id":"identifiers","label":"Identifiers","entries":[{"key":"dc:identifier","label":"Identifier","values":["https://digitalcommons.calpoly.edu/theses/2390","10.15368/theses.2022.19"]}]},{"id":"additional","label":"Additional Metadata","entries":[{"key":"dc:description.abstract","label":"Abstract","values":["Hip implant fractures are rare yet severe complications that follow total hip arthroplasty. These fractures often occur at the implant stem for cemented hip implants due to proximal debonding of the bone cement. While deterministic models offer some predictive power of implant durability, they often are based on average properties, failing to capture the variability in properties that can be drastically different from the mean. This study developed a probabilistic finite element model of a debonded total hip replacement to better account for this variability to determine the most likely failure mode in a particular case. The hypothesis was that fatigue was the more likely cause of failure than overloading. The model used Monte Carlo sampling to create 1000 possible simulations varying in material stiffness and loading magnitude. The analysis found virtually no possibility of failure due to overloading. In contrast, failure due to fatigue had a 99.36% chance of occurring by comparing the maximum stresses recorded at the stem for these simulations to distributions of failure criteria based on literature. Fracture mechanics predicted that the implant in this loading configuration would fail between 0.15-2.2 years with a mean life of 0.6 years. The model had good agreement with the case study outcomes, particularly at predicting the failure site and durability of the implant. Moreover, altering assumptions only moderately changed the probabilistic analysis results compared to a deterministic analysis. Future work could be spent adjusting the model to match the cement implant interface or applying the probabilistic approach to other stages of the implant’s degradation."]},{"key":"dc:title","label":"Title","values":["Three-Dimensional Probabilistic Finite Element Model of a Cemented Hip Implant for Brittle Fracture"]}]}],"canonical_facts":{"dc:contributor":["Lanny Griffin","Biomedical Engineering","College of Engineering"],"dc:creator":["Truong, Daniel J"],"dc:date.available":["2022-03-23T07:00:00Z"],"dc:date.created":["2022-03-01T08:00:00Z"],"dc:description.abstract":["Hip implant fractures are rare yet severe complications that follow total hip arthroplasty. These fractures often occur at the implant stem for cemented hip implants due to proximal debonding of the bone cement. While deterministic models offer some predictive power of implant durability, they often are based on average properties, failing to capture the variability in properties that can be drastically different from the mean. This study developed a probabilistic finite element model of a debonded total hip replacement to better account for this variability to determine the most likely failure mode in a particular case. The hypothesis was that fatigue was the more likely cause of failure than overloading. The model used Monte Carlo sampling to create 1000 possible simulations varying in material stiffness and loading magnitude. The analysis found virtually no possibility of failure due to overloading. In contrast, failure due to fatigue had a 99.36% chance of occurring by comparing the maximum stresses recorded at the stem for these simulations to distributions of failure criteria based on literature. Fracture mechanics predicted that the implant in this loading configuration would fail between 0.15-2.2 years with a mean life of 0.6 years. The model had good agreement with the case study outcomes, particularly at predicting the failure site and durability of the implant. Moreover, altering assumptions only moderately changed the probabilistic analysis results compared to a deterministic analysis. Future work could be spent adjusting the model to match the cement implant interface or applying the probabilistic approach to other stages of the implant’s degradation."],"dc:identifier":["https://digitalcommons.calpoly.edu/theses/2390","10.15368/theses.2022.19"],"dc:subject":["Biomaterials","Biomechanical Engineering","Biomechanics and Biotransport","Biomedical Devices and Instrumentation","Computer-Aided Engineering and Design"],"dc:title":["Three-Dimensional Probabilistic Finite Element Model of a Cemented Hip Implant for Brittle Fracture"],"thesis:degree_discipline":["Biomedical Engineering"],"thesis:degree_name":["MS in Biomedical Engineering"]},"updated_at":"2026-11-19T07:11:08Z"}