{"id":{"repo_id":"calpoly","oai_identifier":"oai:digitalcommons.calpoly.edu:theses-4636"},"canonical_url":"https://search.ndltd.org/etd/calpoly/oai:digitalcommons.calpoly.edu:theses-4636","repository":{"repo_id":"calpoly","name":"Cal Poly","base_url":"https://digitalcommons.calpoly.edu/do/oai/"},"display":{"title":"Assessing Arterial Compliance: Utilizing a Surrogate Arm Bench Top Model for Oscillometric Blood Pressure Monitoring","abstract":"A study was conducted to evaluate the compliance of artificial artery systems utilizing an adaptation of the Yong-Geddes Surrogate Arm model, composed of a blood pressure cuff, artificial artery, pressure transducers, and flow to replicate physiological conditions to validate oscillometric non-invasive blood pressure measurement methods. The experiment phases included tube, blood pressure cuff, and the surrogate arm compliance tests both in static and dynamic conditions. Tube compliance tests showed that larger diameters resulted in increased compliance, with Sample 1 (smaller inner diameter, ID) resulting in an average compliance of 0.0912 mL/mmHg, and Sample 2 (bigger ID) resulting in 0.1597 mL/mmHg. Arterial compliance curves were fit to the data of Samples 1 and 2 with correlations of R2 = 96.56% and R2 = 98.55% respectively. Cuff compliance testing yielded an average compliance of 0.8991 cm2/mmHg, with a second-order inverse polynomial model achieving a strong fit (R2 = 85.02%). Static surrogate arm compliance tests revealed variability, with average compliance values of 0.1102 and 0.1862 for Samples 1 and 2, respectively with errors of 15.99–20.88% compared to tube compliance, highlighting the need for refinements in system design. Overall, the experiments demonstrated the model’s potential for compliance measurement, but identified variability and error, requiring further refinement. These findings provide a foundation for advancing surrogate arm systems and new techniques for validating noninvasive blood pressure systems models.","abstract_html":"A study was conducted to evaluate the compliance of artificial artery systems utilizing an adaptation of the Yong-Geddes Surrogate Arm model, composed of a blood pressure cuff, artificial artery, pressure transducers, and flow to replicate physiological conditions to validate oscillometric non-invasive blood pressure measurement methods. The experiment phases included tube, blood pressure cuff, and the surrogate arm compliance tests both in static and dynamic conditions. Tube compliance tests showed that larger diameters resulted in increased compliance, with Sample 1 (smaller inner diameter, ID) resulting in an average compliance of 0.0912 mL/mmHg, and Sample 2 (bigger ID) resulting in 0.1597 mL/mmHg. Arterial compliance curves were fit to the data of Samples 1 and 2 with correlations of R2 = 96.56% and R2 = 98.55% respectively. Cuff compliance testing yielded an average compliance of 0.8991 cm2/mmHg, with a second-order inverse polynomial model achieving a strong fit (R2 = 85.02%). Static surrogate arm compliance tests revealed variability, with average compliance values of 0.1102 and 0.1862 for Samples 1 and 2, respectively with errors of 15.99–20.88% compared to tube compliance, highlighting the need for refinements in system design. Overall, the experiments demonstrated the model’s potential for compliance measurement, but identified variability and error, requiring further refinement. These findings provide a foundation for advancing surrogate arm systems and new techniques for validating noninvasive blood pressure systems models.","abstract_has_math":false,"creators":["Cardenas, Krystal"],"institution":null,"degree_name":"MS in Biomedical Engineering","degree_level":null,"degree_discipline":"Biomedical Engineering","degree_department":null,"school":null,"contributors":["Michael Whitt","Biomedical Engineering","College of Engineering"],"advisors":[],"committee_chairs":[],"committee_members":[],"year":null,"date_issued":"","date_published":null,"updated_at":"2026-11-19T07:12:29Z","subjects":["Bench Top Models","Oscillometry","Blood Pressure Measurement","Endothelial Dysfunction","Atherosclerosis","Biomedical Engineering and Bioengineering"],"languages":[],"rights":[],"rights_urls":[],"identifier_entries":[],"person_identifier_entries":[]},"links":{"outbound_url":"https://digitalcommons.calpoly.edu/theses/2978","outbound_label":"Repository record","outbound_source":"dc:identifier","record_url":"https://digitalcommons.calpoly.edu/theses/2978","record_label":"Repository record","record_kind":"record","record_source":"dc:identifier","download_url":null,"download_label":null,"download_source":null,"access_status":null,"embargo_until":null,"access_note":null},"source_record":{"url":"https://digitalcommons.calpoly.edu/do/oai/?verb=GetRecord&metadataPrefix=oai_etdms&identifier=oai%3Adigitalcommons.calpoly.edu%3Atheses-4636","prefix":"oai_etdms"},"metadata_groups":[{"id":"people","label":"People","entries":[{"key":"dc:contributor","label":"Contributor","values":["Michael Whitt","Biomedical Engineering","College of Engineering"]},{"key":"dc:creator","label":"Author","values":["Cardenas, Krystal"]}]},{"id":"academic_context","label":"Academic Context","entries":[{"key":"dc:date.available","label":"Dc Date Available","values":["2025-03-21T07:00:00Z"]},{"key":"dc:date.created","label":"Dc Date Created","values":["2025-03-01T08:00:00Z"]},{"key":"thesis:degree_discipline","label":"Discipline","values":["Biomedical Engineering"]},{"key":"thesis:degree_name","label":"Degree Name","values":["MS in Biomedical Engineering"]}]},{"id":"subjects_keywords","label":"Subjects and Keywords","entries":[{"key":"dc:subject","label":"Dc Subject","values":["Bench Top Models","Oscillometry","Blood Pressure Measurement","Endothelial Dysfunction","Atherosclerosis","Biomedical Engineering and Bioengineering"]}]},{"id":"identifiers","label":"Identifiers","entries":[{"key":"dc:identifier","label":"Identifier","values":["https://digitalcommons.calpoly.edu/theses/2978"]}]},{"id":"additional","label":"Additional Metadata","entries":[{"key":"dc:description.abstract","label":"Abstract","values":["A study was conducted to evaluate the compliance of artificial artery systems utilizing an adaptation of the Yong-Geddes Surrogate Arm model, composed of a blood pressure cuff, artificial artery, pressure transducers, and flow to replicate physiological conditions to validate oscillometric non-invasive blood pressure measurement methods. The experiment phases included tube, blood pressure cuff, and the surrogate arm compliance tests both in static and dynamic conditions. Tube compliance tests showed that larger diameters resulted in increased compliance, with Sample 1 (smaller inner diameter, ID) resulting in an average compliance of 0.0912 mL/mmHg, and Sample 2 (bigger ID) resulting in 0.1597 mL/mmHg. Arterial compliance curves were fit to the data of Samples 1 and 2 with correlations of R2 = 96.56% and R2 = 98.55% respectively. Cuff compliance testing yielded an average compliance of 0.8991 cm2/mmHg, with a second-order inverse polynomial model achieving a strong fit (R2 = 85.02%). Static surrogate arm compliance tests revealed variability, with average compliance values of 0.1102 and 0.1862 for Samples 1 and 2, respectively with errors of 15.99–20.88% compared to tube compliance, highlighting the need for refinements in system design. Overall, the experiments demonstrated the model’s potential for compliance measurement, but identified variability and error, requiring further refinement. These findings provide a foundation for advancing surrogate arm systems and new techniques for validating noninvasive blood pressure systems models."]},{"key":"dc:title","label":"Title","values":["Assessing Arterial Compliance: Utilizing a Surrogate Arm Bench Top Model for Oscillometric Blood Pressure Monitoring"]}]}],"canonical_facts":{"dc:contributor":["Michael Whitt","Biomedical Engineering","College of Engineering"],"dc:creator":["Cardenas, Krystal"],"dc:date.available":["2025-03-21T07:00:00Z"],"dc:date.created":["2025-03-01T08:00:00Z"],"dc:description.abstract":["A study was conducted to evaluate the compliance of artificial artery systems utilizing an adaptation of the Yong-Geddes Surrogate Arm model, composed of a blood pressure cuff, artificial artery, pressure transducers, and flow to replicate physiological conditions to validate oscillometric non-invasive blood pressure measurement methods. The experiment phases included tube, blood pressure cuff, and the surrogate arm compliance tests both in static and dynamic conditions. Tube compliance tests showed that larger diameters resulted in increased compliance, with Sample 1 (smaller inner diameter, ID) resulting in an average compliance of 0.0912 mL/mmHg, and Sample 2 (bigger ID) resulting in 0.1597 mL/mmHg. Arterial compliance curves were fit to the data of Samples 1 and 2 with correlations of R2 = 96.56% and R2 = 98.55% respectively. Cuff compliance testing yielded an average compliance of 0.8991 cm2/mmHg, with a second-order inverse polynomial model achieving a strong fit (R2 = 85.02%). Static surrogate arm compliance tests revealed variability, with average compliance values of 0.1102 and 0.1862 for Samples 1 and 2, respectively with errors of 15.99–20.88% compared to tube compliance, highlighting the need for refinements in system design. Overall, the experiments demonstrated the model’s potential for compliance measurement, but identified variability and error, requiring further refinement. These findings provide a foundation for advancing surrogate arm systems and new techniques for validating noninvasive blood pressure systems models."],"dc:identifier":["https://digitalcommons.calpoly.edu/theses/2978"],"dc:subject":["Bench Top Models","Oscillometry","Blood Pressure Measurement","Endothelial Dysfunction","Atherosclerosis","Biomedical Engineering and Bioengineering"],"dc:title":["Assessing Arterial Compliance: Utilizing a Surrogate Arm Bench Top Model for Oscillometric Blood Pressure Monitoring"],"thesis:degree_discipline":["Biomedical Engineering"],"thesis:degree_name":["MS in Biomedical Engineering"]},"updated_at":"2026-11-19T07:12:29Z"}