{"id":{"repo_id":"carleton","oai_identifier":"oai:carleton.scholaris.ca:20.500.14718/38361"},"canonical_url":"https://search.ndltd.org/etd/carleton/oai:carleton.scholaris.ca:20.500.14718/38361","repository":{"repo_id":"carleton","name":"Carleton University","base_url":"https://carleton.scholaris.ca/server/oai/request"},"display":{"title":"Quantitative Multiplexed Multi-pinhole Small-Animal SPECT","abstract":"Small-animal microSPECT is an important research tool for the development of clinical diagnostic tests, evaluation of novel pharmaceuticals, and basic science research. Experimentally quantifying the accuracy of microSPECT — the reproducibility of measurements made using microSPECT, and how well the reconstructed microSPECT data correlates to the true tracer distribution — is therefore essential. Photon attenuation and scatter complicate the problem of acquiring quantitatively accurate information from SPECT.In this work, the reproducibility of cardiac measurements made in rats using the nanoSPECT/CT scanner (Bioscan, Washington, DC) is first evaluated without scatter or attenuation correction. For relative perfusion measurements of rat myocardia, normal databases for 99mTc-tetrofosmin and 201Tl imaging are created and it is shown that the relative myocardial perfusion measurements made using either tracer are reproducible over repeated scans and over multiple rats with standard deviations of 3% and 4% respectively.Attenuation and scatter compensation are implemented in the SPECT reconstruction algorithm to evaluate the accuracy of using nanoSPECT to recover absolute tracer concentrations in phantoms and in vivo. With attenuation and scatter compensation, absolute tracer concentration is measured using SPECT images with systematic and random uncertainties as low as 1% in phantoms, relative to dose calibrator measurements. The systematic and random uncertainties for measurements of total cardiac tracer uptake in vivo compared to well chamber measurements ex vivo were as small as –1% and 3% respectively.Reevaluation of the reproducibility of the cardiac measurements with attenuation and scatter correction demonstrates that neither correction strongly influences the relative perfusion measurements or their reproducibility. However, it is shown that attenuation correction does significantly modify the relationship between body weight and absolute tracer uptake.","abstract_html":"Small-animal microSPECT is an important research tool for the development of clinical diagnostic tests, evaluation of novel pharmaceuticals, and basic science research. Experimentally quantifying the accuracy of microSPECT — the reproducibility of measurements made using microSPECT, and how well the reconstructed microSPECT data correlates to the true tracer distribution — is therefore essential. Photon attenuation and scatter complicate the problem of acquiring quantitatively accurate information from SPECT.In this work, the reproducibility of cardiac measurements made in rats using the nanoSPECT/CT scanner (Bioscan, Washington, DC) is first evaluated without scatter or attenuation correction. For relative perfusion measurements of rat myocardia, normal databases for 99mTc-tetrofosmin and 201Tl imaging are created and it is shown that the relative myocardial perfusion measurements made using either tracer are reproducible over repeated scans and over multiple rats with standard deviations of 3% and 4% respectively.Attenuation and scatter compensation are implemented in the SPECT reconstruction algorithm to evaluate the accuracy of using nanoSPECT to recover absolute tracer concentrations in phantoms and in vivo. With attenuation and scatter compensation, absolute tracer concentration is measured using SPECT images with systematic and random uncertainties as low as 1% in phantoms, relative to dose calibrator measurements. The systematic and random uncertainties for measurements of total cardiac tracer uptake in vivo compared to well chamber measurements ex vivo were as small as –1% and 3% respectively.Reevaluation of the reproducibility of the cardiac measurements with attenuation and scatter correction demonstrates that neither correction strongly influences the relative perfusion measurements or their reproducibility. However, it is shown that attenuation correction does significantly modify the relationship between body weight and absolute tracer uptake.","abstract_has_math":false,"creators":["Strydhorst, Jared"],"institution":"Carleton University","degree_name":"Doctor of Philosophy (Ph.D.)","degree_level":"Doctoral","degree_discipline":"Physics - Medical","degree_department":null,"school":null,"contributors":[],"advisors":[],"committee_chairs":[],"committee_members":[],"year":2013,"date_issued":"2013","date_published":"2013","updated_at":"2026-11-19T07:06:12Z","subjects":["PHYSICAL SCIENCES Physics - Radiation"],"languages":["en"],"rights":["Copyright © 2013 the author(s). Theses may be used for non-commercial research, educational, or related academic purposes only. Such uses include personal study, research, scholarship, and teaching. Theses may only be shared by linking to Carleton University Institutional Repository and no part may be used without proper attribution to the author. No part may be used for commercial purposes directly or indirectly via a for-profit platform; no adaptation or derivative works are permitted without consent from the copyright owner."],"rights_urls":[],"identifier_entries":[{"key":"dc:identifier.uri","label":"Identifier URI","values":["https://hdl.handle.net/20.500.14718/38361"],"render_values":[{"text":"https://hdl.handle.net/20.500.14718/38361","href":"https://hdl.handle.net/20.500.14718/38361","code":true}]}],"person_identifier_entries":[]},"links":{"outbound_url":"https://doi.org/10.22215/etd/2013-09979","outbound_label":"DOI","outbound_source":"dc:identifier.doi","record_url":"https://doi.org/10.22215/etd/2013-09979","record_label":"Repository record","record_kind":"doi","record_source":"dc:identifier.doi","access_status":null,"embargo_until":null,"access_note":null},"source_record":{"url":"https://carleton.scholaris.ca/server/oai/request?verb=GetRecord&metadataPrefix=dim&identifier=oai%3Acarleton.scholaris.ca%3A20.500.14718%2F38361","prefix":"dim"},"metadata_groups":[{"id":"people","label":"People","entries":[{"key":"dc:creator","label":"Author","values":["Strydhorst, Jared"]}]},{"id":"academic_context","label":"Academic Context","entries":[{"key":"dc:date.accessioned","label":"Dc Date Accessioned","values":["2025-04-08T19:22:03Z"]},{"key":"dc:date.available","label":"Dc Date Available","values":["2025-04-08T19:22:03Z"]},{"key":"dc:date.issued","label":"Date","values":["2013"]},{"key":"dc:publisher","label":"Institution","values":["Carleton University"]},{"key":"dc:type","label":"Dc Type","values":["thesis"]},{"key":"thesis:degree_discipline","label":"Discipline","values":["Physics - Medical"]},{"key":"thesis:degree_level","label":"Degree Level","values":["Doctoral"]},{"key":"thesis:degree_name","label":"Degree Name","values":["Doctor of Philosophy (Ph.D.)"]}]},{"id":"subjects_keywords","label":"Subjects and Keywords","entries":[{"key":"dc:subject.lcsh","label":"Dc Subject Lcsh","values":["PHYSICAL SCIENCES Physics - Radiation"]}]},{"id":"language_rights","label":"Language and Rights","entries":[{"key":"dc:language.iso","label":"Language (ISO)","values":["en"]},{"key":"dc:rights","label":"Dc Rights","values":["Copyright © 2013 the author(s). Theses may be used for non-commercial research, educational, or related academic purposes only. Such uses include personal study, research, scholarship, and teaching. Theses may only be shared by linking to Carleton University Institutional Repository and no part may be used without proper attribution to the author. No part may be used for commercial purposes directly or indirectly via a for-profit platform; no adaptation or derivative works are permitted without consent from the copyright owner."]}]},{"id":"identifiers","label":"Identifiers","entries":[{"key":"dc:identifier.doi","label":"DOI","values":["10.22215/etd/2013-09979"]},{"key":"dc:identifier.uri","label":"Identifier URI","values":["https://hdl.handle.net/20.500.14718/38361"]}]},{"id":"additional","label":"Additional Metadata","entries":[{"key":"dc:description.abstract","label":"Abstract","values":["Small-animal microSPECT is an important research tool for the development of clinical diagnostic tests, evaluation of novel pharmaceuticals, and basic science research. Experimentally quantifying the accuracy of microSPECT — the reproducibility of measurements made using microSPECT, and how well the reconstructed microSPECT data correlates to the true tracer distribution — is therefore essential. Photon attenuation and scatter complicate the problem of acquiring quantitatively accurate information from SPECT.In this work, the reproducibility of cardiac measurements made in rats using the nanoSPECT/CT scanner (Bioscan, Washington, DC) is first evaluated without scatter or attenuation correction. For relative perfusion measurements of rat myocardia, normal databases for 99mTc-tetrofosmin and 201Tl imaging are created and it is shown that the relative myocardial perfusion measurements made using either tracer are reproducible over repeated scans and over multiple rats with standard deviations of 3% and 4% respectively.Attenuation and scatter compensation are implemented in the SPECT reconstruction algorithm to evaluate the accuracy of using nanoSPECT to recover absolute tracer concentrations in phantoms and in vivo. With attenuation and scatter compensation, absolute tracer concentration is measured using SPECT images with systematic and random uncertainties as low as 1% in phantoms, relative to dose calibrator measurements. The systematic and random uncertainties for measurements of total cardiac tracer uptake in vivo compared to well chamber measurements ex vivo were as small as –1% and 3% respectively.Reevaluation of the reproducibility of the cardiac measurements with attenuation and scatter correction demonstrates that neither correction strongly influences the relative perfusion measurements or their reproducibility. However, it is shown that attenuation correction does significantly modify the relationship between body weight and absolute tracer uptake."]},{"key":"dc:title","label":"Title","values":["Quantitative Multiplexed Multi-pinhole Small-Animal SPECT"]},{"key":"local:lac.licence","label":"Local Lac Licence","values":["true"]},{"key":"others:access_status","label":"Others Access Status","values":["open.access"]}]}],"canonical_facts":{"dc:creator":["Strydhorst, Jared"],"dc:date.accessioned":["2025-04-08T19:22:03Z"],"dc:date.available":["2025-04-08T19:22:03Z"],"dc:date.issued":["2013"],"dc:description.abstract":["Small-animal microSPECT is an important research tool for the development of clinical diagnostic tests, evaluation of novel pharmaceuticals, and basic science research. Experimentally quantifying the accuracy of microSPECT — the reproducibility of measurements made using microSPECT, and how well the reconstructed microSPECT data correlates to the true tracer distribution — is therefore essential. Photon attenuation and scatter complicate the problem of acquiring quantitatively accurate information from SPECT.In this work, the reproducibility of cardiac measurements made in rats using the nanoSPECT/CT scanner (Bioscan, Washington, DC) is first evaluated without scatter or attenuation correction. For relative perfusion measurements of rat myocardia, normal databases for 99mTc-tetrofosmin and 201Tl imaging are created and it is shown that the relative myocardial perfusion measurements made using either tracer are reproducible over repeated scans and over multiple rats with standard deviations of 3% and 4% respectively.Attenuation and scatter compensation are implemented in the SPECT reconstruction algorithm to evaluate the accuracy of using nanoSPECT to recover absolute tracer concentrations in phantoms and in vivo. With attenuation and scatter compensation, absolute tracer concentration is measured using SPECT images with systematic and random uncertainties as low as 1% in phantoms, relative to dose calibrator measurements. The systematic and random uncertainties for measurements of total cardiac tracer uptake in vivo compared to well chamber measurements ex vivo were as small as –1% and 3% respectively.Reevaluation of the reproducibility of the cardiac measurements with attenuation and scatter correction demonstrates that neither correction strongly influences the relative perfusion measurements or their reproducibility. However, it is shown that attenuation correction does significantly modify the relationship between body weight and absolute tracer uptake."],"dc:identifier.doi":["10.22215/etd/2013-09979"],"dc:identifier.uri":["https://hdl.handle.net/20.500.14718/38361"],"dc:language.iso":["en"],"dc:publisher":["Carleton University"],"dc:rights":["Copyright © 2013 the author(s). Theses may be used for non-commercial research, educational, or related academic purposes only. Such uses include personal study, research, scholarship, and teaching. Theses may only be shared by linking to Carleton University Institutional Repository and no part may be used without proper attribution to the author. No part may be used for commercial purposes directly or indirectly via a for-profit platform; no adaptation or derivative works are permitted without consent from the copyright owner."],"dc:subject.lcsh":["PHYSICAL SCIENCES Physics - Radiation"],"dc:title":["Quantitative Multiplexed Multi-pinhole Small-Animal SPECT"],"dc:type":["thesis"],"local:lac.licence":["true"],"others:access_status":["open.access"],"thesis:degree_discipline":["Physics - Medical"],"thesis:degree_level":["Doctoral"],"thesis:degree_name":["Doctor of Philosophy (Ph.D.)"]},"updated_at":"2026-11-19T07:06:12Z"}