{"id":{"repo_id":"carleton","oai_identifier":"oai:carleton.scholaris.ca:20.500.14718/38518"},"canonical_url":"https://search.ndltd.org/etd/carleton/oai:carleton.scholaris.ca:20.500.14718/38518","repository":{"repo_id":"carleton","name":"Carleton University","base_url":"https://carleton.scholaris.ca/server/oai/request"},"display":{"title":"Automatic Localization of Trackable Features in Retinal Fundus Images","abstract":"Laser photocoagulation is used to reduce vision loss due to degenerative eye conditions by guiding a treatment laser over diseased tissue on the retina. However, due to natural motions of the eye healthy retinal tissue can be damaged by the laser. Compensating for this motion can reduce this risk. Motion analysis of scanning laser opthalmoscope (SLO) video provides information suitable for automatic laser aiming. Motion estimation from SLO video can be achieved using landmark-based motion tracking. In the ALET implementation, landmarks are manually selected from a reference image. This thesis work automates the landmark selection process in two steps: first, a set of candidate landmarks is found using a corner detection procedure on a vessel-enhanced retinal image; and second, the set of candidate landmarks are ranked in decreasing order of their estimated tracking integrity.","abstract_html":"Laser photocoagulation is used to reduce vision loss due to degenerative eye conditions by guiding a treatment laser over diseased tissue on the retina. However, due to natural motions of the eye healthy retinal tissue can be damaged by the laser. Compensating for this motion can reduce this risk. Motion analysis of scanning laser opthalmoscope (SLO) video provides information suitable for automatic laser aiming. Motion estimation from SLO video can be achieved using landmark-based motion tracking. In the ALET implementation, landmarks are manually selected from a reference image. This thesis work automates the landmark selection process in two steps: first, a set of candidate landmarks is found using a corner detection procedure on a vessel-enhanced retinal image; and second, the set of candidate landmarks are ranked in decreasing order of their estimated tracking integrity.","abstract_has_math":false,"creators":["Inder, Paul"],"institution":"Carleton University","degree_name":"Master of Applied Science (M.App.Sc.)","degree_level":"Master's","degree_discipline":"Engineering, Biomedical","degree_department":null,"school":null,"contributors":[],"advisors":[],"committee_chairs":[],"committee_members":[],"year":2014,"date_issued":"2014","date_published":"2014","updated_at":"2026-11-19T07:04:44Z","subjects":["Biomedical engineering"],"languages":["en"],"rights":["Copyright © 2014 the author(s). Theses may be used for non-commercial research, educational, or related academic purposes only. Such uses include personal study, research, scholarship, and teaching. Theses may only be shared by linking to Carleton University Institutional Repository and no part may be used without proper attribution to the author. No part may be used for commercial purposes directly or indirectly via a for-profit platform; no adaptation or derivative works are permitted without consent from the copyright owner."],"rights_urls":[],"identifier_entries":[{"key":"dc:identifier.uri","label":"Identifier URI","values":["https://hdl.handle.net/20.500.14718/38518"],"render_values":[{"text":"https://hdl.handle.net/20.500.14718/38518","href":"https://hdl.handle.net/20.500.14718/38518","code":true}]}],"person_identifier_entries":[]},"links":{"outbound_url":"https://doi.org/10.22215/etd/2014-10373","outbound_label":"DOI","outbound_source":"dc:identifier.doi","record_url":"https://doi.org/10.22215/etd/2014-10373","record_label":"Repository record","record_kind":"doi","record_source":"dc:identifier.doi","access_status":null,"embargo_until":null,"access_note":null},"source_record":{"url":"https://carleton.scholaris.ca/server/oai/request?verb=GetRecord&metadataPrefix=dim&identifier=oai%3Acarleton.scholaris.ca%3A20.500.14718%2F38518","prefix":"dim"},"metadata_groups":[{"id":"people","label":"People","entries":[{"key":"dc:creator","label":"Author","values":["Inder, Paul"]}]},{"id":"academic_context","label":"Academic Context","entries":[{"key":"dc:date.accessioned","label":"Dc Date Accessioned","values":["2025-04-08T19:24:56Z"]},{"key":"dc:date.available","label":"Dc Date Available","values":["2025-04-08T19:24:56Z"]},{"key":"dc:date.issued","label":"Date","values":["2014"]},{"key":"dc:publisher","label":"Institution","values":["Carleton University"]},{"key":"dc:type","label":"Dc Type","values":["thesis"]},{"key":"thesis:degree_discipline","label":"Discipline","values":["Engineering, Biomedical"]},{"key":"thesis:degree_level","label":"Degree Level","values":["Master's"]},{"key":"thesis:degree_name","label":"Degree Name","values":["Master of Applied Science (M.App.Sc.)"]}]},{"id":"subjects_keywords","label":"Subjects and Keywords","entries":[{"key":"dc:subject.lcsh","label":"Dc Subject Lcsh","values":["Biomedical engineering"]}]},{"id":"language_rights","label":"Language and Rights","entries":[{"key":"dc:language.iso","label":"Language (ISO)","values":["en"]},{"key":"dc:rights","label":"Dc Rights","values":["Copyright © 2014 the author(s). Theses may be used for non-commercial research, educational, or related academic purposes only. Such uses include personal study, research, scholarship, and teaching. Theses may only be shared by linking to Carleton University Institutional Repository and no part may be used without proper attribution to the author. No part may be used for commercial purposes directly or indirectly via a for-profit platform; no adaptation or derivative works are permitted without consent from the copyright owner."]}]},{"id":"identifiers","label":"Identifiers","entries":[{"key":"dc:identifier.doi","label":"DOI","values":["10.22215/etd/2014-10373"]},{"key":"dc:identifier.uri","label":"Identifier URI","values":["https://hdl.handle.net/20.500.14718/38518"]}]},{"id":"additional","label":"Additional Metadata","entries":[{"key":"dc:description.abstract","label":"Abstract","values":["Laser photocoagulation is used to reduce vision loss due to degenerative eye conditions by guiding a treatment laser over diseased tissue on the retina. However, due to natural motions of the eye healthy retinal tissue can be damaged by the laser. Compensating for this motion can reduce this risk. Motion analysis of scanning laser opthalmoscope (SLO) video provides information suitable for automatic laser aiming. Motion estimation from SLO video can be achieved using landmark-based motion tracking. In the ALET implementation, landmarks are manually selected from a reference image. This thesis work automates the landmark selection process in two steps: first, a set of candidate landmarks is found using a corner detection procedure on a vessel-enhanced retinal image; and second, the set of candidate landmarks are ranked in decreasing order of their estimated tracking integrity."]},{"key":"dc:title","label":"Title","values":["Automatic Localization of Trackable Features in Retinal Fundus Images"]},{"key":"local:lac.licence","label":"Local Lac Licence","values":["true"]},{"key":"others:access_status","label":"Others Access Status","values":["open.access"]}]}],"canonical_facts":{"dc:creator":["Inder, Paul"],"dc:date.accessioned":["2025-04-08T19:24:56Z"],"dc:date.available":["2025-04-08T19:24:56Z"],"dc:date.issued":["2014"],"dc:description.abstract":["Laser photocoagulation is used to reduce vision loss due to degenerative eye conditions by guiding a treatment laser over diseased tissue on the retina. However, due to natural motions of the eye healthy retinal tissue can be damaged by the laser. Compensating for this motion can reduce this risk. Motion analysis of scanning laser opthalmoscope (SLO) video provides information suitable for automatic laser aiming. Motion estimation from SLO video can be achieved using landmark-based motion tracking. In the ALET implementation, landmarks are manually selected from a reference image. This thesis work automates the landmark selection process in two steps: first, a set of candidate landmarks is found using a corner detection procedure on a vessel-enhanced retinal image; and second, the set of candidate landmarks are ranked in decreasing order of their estimated tracking integrity."],"dc:identifier.doi":["10.22215/etd/2014-10373"],"dc:identifier.uri":["https://hdl.handle.net/20.500.14718/38518"],"dc:language.iso":["en"],"dc:publisher":["Carleton University"],"dc:rights":["Copyright © 2014 the author(s). Theses may be used for non-commercial research, educational, or related academic purposes only. Such uses include personal study, research, scholarship, and teaching. Theses may only be shared by linking to Carleton University Institutional Repository and no part may be used without proper attribution to the author. No part may be used for commercial purposes directly or indirectly via a for-profit platform; no adaptation or derivative works are permitted without consent from the copyright owner."],"dc:subject.lcsh":["Biomedical engineering"],"dc:title":["Automatic Localization of Trackable Features in Retinal Fundus Images"],"dc:type":["thesis"],"local:lac.licence":["true"],"others:access_status":["open.access"],"thesis:degree_discipline":["Engineering, Biomedical"],"thesis:degree_level":["Master's"],"thesis:degree_name":["Master of Applied Science (M.App.Sc.)"]},"updated_at":"2026-11-19T07:04:44Z"}