{"id":{"repo_id":"cornell","oai_identifier":"oai:ecommons.cornell.edu:1813/30675"},"canonical_url":"https://search.ndltd.org/etd/cornell/oai:ecommons.cornell.edu:1813/30675","repository":{"repo_id":"cornell","name":"Cornell University","base_url":"https://ecommons.cornell.edu/server/oai/request"},"display":{"title":"Bioanalytical Microsystems For Waterborne Pathogen Detection","abstract":"Waterborne pathogens are a global health concern, with 1 billion people having no access to uncontaminated water sources and 2.2 million annual deaths resulting from infection. Biosensors offer the possibility of rapid, sensitive and specific detection of pathogens and, thus, the ability to provide warning systems prior to consumption and a means to monitor the efficacy of disinfection systems. However, in order for biosensors to be useful, one must be able to attain clinically relevant detection limits in real environmental samples and do so on a reasonable timescale. Thus, methods to increase the analytical sensitivity via pre-concentration and microfluidic device design are described herein. Firstly, employing a series of pre-concentration and purification steps followed by an enzymatic amplification of an mRNA sequence, it is shown that a limit of detection of one Cryptosporidium parvum oocyst can be attained and that this assay does work in real water samples. Secondly, integrating a liposome-based immunoassay and a nanoporous membrane for electrokinetic pre-concentration into a single microfluidic device for the detection of a human norovirus surrogate yielded an order-of-magnitude decrease in the limit of detection when compared to an optimized device foregoing pre-concentration. Finally, the optimization of microfluidic channel dimensions for assays using liposomal signal amplification is described; illustrating that a 60% reduction in channel height produces an order-of-magnitude reduction in the limit of detection. Taken together, the described work provides tools to advance the field of biosensors by demonstrating means to improve analytical sensitivity and overcome the obstacles presented by environmental water samples.","abstract_html":"Waterborne pathogens are a global health concern, with 1 billion people having no access to uncontaminated water sources and 2.2 million annual deaths resulting from infection. Biosensors offer the possibility of rapid, sensitive and specific detection of pathogens and, thus, the ability to provide warning systems prior to consumption and a means to monitor the efficacy of disinfection systems. However, in order for biosensors to be useful, one must be able to attain clinically relevant detection limits in real environmental samples and do so on a reasonable timescale. Thus, methods to increase the analytical sensitivity via pre-concentration and microfluidic device design are described herein. Firstly, employing a series of pre-concentration and purification steps followed by an enzymatic amplification of an mRNA sequence, it is shown that a limit of detection of one Cryptosporidium parvum oocyst can be attained and that this assay does work in real water samples. Secondly, integrating a liposome-based immunoassay and a nanoporous membrane for electrokinetic pre-concentration into a single microfluidic device for the detection of a human norovirus surrogate yielded an order-of-magnitude decrease in the limit of detection when compared to an optimized device foregoing pre-concentration. Finally, the optimization of microfluidic channel dimensions for assays using liposomal signal amplification is described; illustrating that a 60% reduction in channel height produces an order-of-magnitude reduction in the limit of detection. Taken together, the described work provides tools to advance the field of biosensors by demonstrating means to improve analytical sensitivity and overcome the obstacles presented by environmental water samples.","abstract_has_math":false,"creators":["Connelly, John"],"institution":"Cornell University","degree_name":"Ph. D., Agricultural and Biological Engineering","degree_level":"Doctor of Philosophy","degree_discipline":"Agricultural and Biological Engineering","degree_department":null,"school":null,"contributors":[],"advisors":[],"committee_chairs":["Baeumner, Antje J"],"committee_members":["Weiland, Gregory Arthur","Durst, Richard A"],"year":2011,"date_issued":"2011-08-31","date_published":"2011-08-31","updated_at":"2026-11-19T03:15:03Z","subjects":[],"languages":["en_US"],"rights":[],"rights_urls":[],"identifier_entries":[{"key":"dc:identifier.bibid","label":"Dc Identifier Bibid","values":["7955484"],"render_values":[{"text":"7955484","href":null,"code":true}]}],"person_identifier_entries":[]},"links":{"outbound_url":"https://hdl.handle.net/1813/30675","outbound_label":"Handle","outbound_source":"dc:identifier.uri","record_url":"https://hdl.handle.net/1813/30675","record_label":"Repository record","record_kind":"handle","record_source":"dc:identifier.uri","download_url":null,"download_label":null,"download_source":null,"access_status":null,"embargo_until":null,"access_note":null},"source_record":{"url":"https://ecommons.cornell.edu/server/oai/request?verb=GetRecord&metadataPrefix=dim&identifier=oai%3Aecommons.cornell.edu%3A1813%2F30675","prefix":"dim"},"metadata_groups":[{"id":"people","label":"People","entries":[{"key":"dc:contributor.chair","label":"Chair","values":["Baeumner, Antje J"]},{"key":"dc:contributor.committeemember","label":"Committee Member","values":["Weiland, Gregory Arthur","Durst, Richard A"]},{"key":"dc:creator","label":"Author","values":["Connelly, John"]}]},{"id":"academic_context","label":"Academic Context","entries":[{"key":"dc:date.accessioned","label":"Dc Date Accessioned","values":["2012-12-17T13:50:54Z"]},{"key":"dc:date.available","label":"Dc Date Available","values":["2016-12-30T06:46:57Z"]},{"key":"dc:date.issued","label":"Date","values":["2011-08-31"]},{"key":"dc:type","label":"Dc Type","values":["dissertation or thesis"]},{"key":"thesis:degree_discipline","label":"Discipline","values":["Agricultural and Biological Engineering"]},{"key":"thesis:degree_level","label":"Degree Level","values":["Doctor of Philosophy"]},{"key":"thesis:degree_name","label":"Degree Name","values":["Ph. D., Agricultural and Biological Engineering"]},{"key":"thesis:institution_name","label":"Thesis Institution Name","values":["Cornell University"]}]},{"id":"language_rights","label":"Language and Rights","entries":[{"key":"dc:language.iso","label":"Language (ISO)","values":["en_US"]}]},{"id":"identifiers","label":"Identifiers","entries":[{"key":"dc:identifier.bibid","label":"Dc Identifier Bibid","values":["7955484"]},{"key":"dc:identifier.uri","label":"Identifier URI","values":["https://hdl.handle.net/1813/30675"]}]},{"id":"additional","label":"Additional Metadata","entries":[{"key":"cerif:openaire","label":"Cerif Openaire","values":["<Publication xmlns=\"https://www.openaire.eu/cerif-profile/1.1/\" id=\"2155fd7f-5c73-4bdf-9869-574d13f6a153\"> <Type xmlns=\"https://www.openaire.eu/cerif-profile/vocab/COAR_Publication_Types\">http://purl.org/coar/resource_type/c_1843</Type> <Language>en_US</Language> <Title>Bioanalytical Microsystems For Waterborne Pathogen Detection</Title> <PublishedIn> <Publication> </Publication> </PublishedIn> <PublicationDate>2011-08-31</PublicationDate> <Authors> <Author> <DisplayName>Connelly, John</DisplayName> <Affiliation> <OrgUnit> </OrgUnit> </Affiliation> </Author> </Authors> <Editors> </Editors> <Publishers> <Publisher> <OrgUnit /> </Publisher> </Publishers> <Abstract>Waterborne pathogens are a global health concern, with 1 billion people having no access to uncontaminated water sources and 2.2 million annual deaths resulting from infection. Biosensors offer the possibility of rapid, sensitive and specific detection of pathogens and, thus, the ability to provide warning systems prior to consumption and a means to monitor the efficacy of disinfection systems. However, in order for biosensors to be useful, one must be able to attain clinically relevant detection limits in real environmental samples and do so on a reasonable timescale. Thus, methods to increase the analytical sensitivity via pre-concentration and microfluidic device design are described herein. Firstly, employing a series of pre-concentration and purification steps followed by an enzymatic amplification of an mRNA sequence, it is shown that a limit of detection of one Cryptosporidium parvum oocyst can be attained and that this assay does work in real water samples. Secondly, integrating a liposome-based immunoassay and a nanoporous membrane for electrokinetic pre-concentration into a single microfluidic device for the detection of a human norovirus surrogate yielded an order-of-magnitude decrease in the limit of detection when compared to an optimized device foregoing pre-concentration. Finally, the optimization of microfluidic channel dimensions for assays using liposomal signal amplification is described; illustrating that a 60% reduction in channel height produces an order-of-magnitude reduction in the limit of detection. Taken together, the described work provides tools to advance the field of biosensors by demonstrating means to improve analytical sensitivity and overcome the obstacles presented by environmental water samples.</Abstract> <Access xmlns=\"http://purl.org/coar/access_right\" > </Access> </Publication>"]},{"key":"cris:virtual.author","label":"Cris Virtual Author","values":["Connelly, John"]},{"key":"cris:virtualsource.collection","label":"Cris Virtualsource Collection","values":["5893a6ea-7af3-41d7-abc6-04bcd26ab5df"]},{"key":"dc:description.abstract","label":"Abstract","values":["Waterborne pathogens are a global health concern, with 1 billion people having no access to uncontaminated water sources and 2.2 million annual deaths resulting from infection. Biosensors offer the possibility of rapid, sensitive and specific detection of pathogens and, thus, the ability to provide warning systems prior to consumption and a means to monitor the efficacy of disinfection systems. However, in order for biosensors to be useful, one must be able to attain clinically relevant detection limits in real environmental samples and do so on a reasonable timescale. Thus, methods to increase the analytical sensitivity via pre-concentration and microfluidic device design are described herein. Firstly, employing a series of pre-concentration and purification steps followed by an enzymatic amplification of an mRNA sequence, it is shown that a limit of detection of one Cryptosporidium parvum oocyst can be attained and that this assay does work in real water samples. Secondly, integrating a liposome-based immunoassay and a nanoporous membrane for electrokinetic pre-concentration into a single microfluidic device for the detection of a human norovirus surrogate yielded an order-of-magnitude decrease in the limit of detection when compared to an optimized device foregoing pre-concentration. Finally, the optimization of microfluidic channel dimensions for assays using liposomal signal amplification is described; illustrating that a 60% reduction in channel height produces an order-of-magnitude reduction in the limit of detection. Taken together, the described work provides tools to advance the field of biosensors by demonstrating means to improve analytical sensitivity and overcome the obstacles presented by environmental water samples."]},{"key":"dc:description.collection","label":"Dc Description Collection","values":["Cornell Theses and Dissertations"]},{"key":"dc:format.mimetype","label":"Dc Format Mimetype","values":["application/pdf"]},{"key":"dc:title","label":"Title","values":["Bioanalytical Microsystems For Waterborne Pathogen Detection"]},{"key":"dspace:entity.type","label":"Dspace Entity Type","values":["Publication"]},{"key":"others:access_status","label":"Others Access Status","values":["open.access"]}]}],"canonical_facts":{"cerif:openaire":["<Publication xmlns=\"https://www.openaire.eu/cerif-profile/1.1/\" id=\"2155fd7f-5c73-4bdf-9869-574d13f6a153\"> <Type xmlns=\"https://www.openaire.eu/cerif-profile/vocab/COAR_Publication_Types\">http://purl.org/coar/resource_type/c_1843</Type> <Language>en_US</Language> <Title>Bioanalytical Microsystems For Waterborne Pathogen Detection</Title> <PublishedIn> <Publication> </Publication> </PublishedIn> <PublicationDate>2011-08-31</PublicationDate> <Authors> <Author> <DisplayName>Connelly, John</DisplayName> <Affiliation> <OrgUnit> </OrgUnit> </Affiliation> </Author> </Authors> <Editors> </Editors> <Publishers> <Publisher> <OrgUnit /> </Publisher> </Publishers> <Abstract>Waterborne pathogens are a global health concern, with 1 billion people having no access to uncontaminated water sources and 2.2 million annual deaths resulting from infection. Biosensors offer the possibility of rapid, sensitive and specific detection of pathogens and, thus, the ability to provide warning systems prior to consumption and a means to monitor the efficacy of disinfection systems. However, in order for biosensors to be useful, one must be able to attain clinically relevant detection limits in real environmental samples and do so on a reasonable timescale. Thus, methods to increase the analytical sensitivity via pre-concentration and microfluidic device design are described herein. Firstly, employing a series of pre-concentration and purification steps followed by an enzymatic amplification of an mRNA sequence, it is shown that a limit of detection of one Cryptosporidium parvum oocyst can be attained and that this assay does work in real water samples. Secondly, integrating a liposome-based immunoassay and a nanoporous membrane for electrokinetic pre-concentration into a single microfluidic device for the detection of a human norovirus surrogate yielded an order-of-magnitude decrease in the limit of detection when compared to an optimized device foregoing pre-concentration. Finally, the optimization of microfluidic channel dimensions for assays using liposomal signal amplification is described; illustrating that a 60% reduction in channel height produces an order-of-magnitude reduction in the limit of detection. Taken together, the described work provides tools to advance the field of biosensors by demonstrating means to improve analytical sensitivity and overcome the obstacles presented by environmental water samples.</Abstract> <Access xmlns=\"http://purl.org/coar/access_right\" > </Access> </Publication>"],"cris:virtual.author":["Connelly, John"],"cris:virtualsource.collection":["5893a6ea-7af3-41d7-abc6-04bcd26ab5df"],"dc:contributor.chair":["Baeumner, Antje J"],"dc:contributor.committeemember":["Weiland, Gregory Arthur","Durst, Richard A"],"dc:creator":["Connelly, John"],"dc:date.accessioned":["2012-12-17T13:50:54Z"],"dc:date.available":["2016-12-30T06:46:57Z"],"dc:date.issued":["2011-08-31"],"dc:description.abstract":["Waterborne pathogens are a global health concern, with 1 billion people having no access to uncontaminated water sources and 2.2 million annual deaths resulting from infection. Biosensors offer the possibility of rapid, sensitive and specific detection of pathogens and, thus, the ability to provide warning systems prior to consumption and a means to monitor the efficacy of disinfection systems. However, in order for biosensors to be useful, one must be able to attain clinically relevant detection limits in real environmental samples and do so on a reasonable timescale. Thus, methods to increase the analytical sensitivity via pre-concentration and microfluidic device design are described herein. Firstly, employing a series of pre-concentration and purification steps followed by an enzymatic amplification of an mRNA sequence, it is shown that a limit of detection of one Cryptosporidium parvum oocyst can be attained and that this assay does work in real water samples. Secondly, integrating a liposome-based immunoassay and a nanoporous membrane for electrokinetic pre-concentration into a single microfluidic device for the detection of a human norovirus surrogate yielded an order-of-magnitude decrease in the limit of detection when compared to an optimized device foregoing pre-concentration. Finally, the optimization of microfluidic channel dimensions for assays using liposomal signal amplification is described; illustrating that a 60% reduction in channel height produces an order-of-magnitude reduction in the limit of detection. Taken together, the described work provides tools to advance the field of biosensors by demonstrating means to improve analytical sensitivity and overcome the obstacles presented by environmental water samples."],"dc:description.collection":["Cornell Theses and Dissertations"],"dc:format.mimetype":["application/pdf"],"dc:identifier.bibid":["7955484"],"dc:identifier.uri":["https://hdl.handle.net/1813/30675"],"dc:language.iso":["en_US"],"dc:title":["Bioanalytical Microsystems For Waterborne Pathogen Detection"],"dc:type":["dissertation or thesis"],"dspace:entity.type":["Publication"],"others:access_status":["open.access"],"thesis:degree_discipline":["Agricultural and Biological Engineering"],"thesis:degree_level":["Doctor of Philosophy"],"thesis:degree_name":["Ph. D., Agricultural and Biological Engineering"],"thesis:institution_name":["Cornell University"]},"updated_at":"2026-11-19T03:15:03Z"}