{"id":{"repo_id":"dcu","oai_identifier":"oai:doras.dcu.ie:18008"},"canonical_url":"https://search.ndltd.org/etd/dcu/oai:doras.dcu.ie:18008","repository":{"repo_id":"dcu","name":"Dublin City University","base_url":"http://doras.dcu.ie/cgi/oai2"},"display":{"title":"Generic embedded sensor development: high-speed analogue preconditioning and data acquisition systems","abstract":"The design of a high-speed, low-cost, generic analogue preconditioning system, which is capable of interfacing with a wide variety of transducers, is described. A multi-purpose, configurable data acquisition module, which is also capable of data generation, is presented. Software support is provided by advanced signal analysis and innovative presentation algorithms that are implemented on a powerful embedded digital signal processor. Precision voltage-feedback and high-speed current-feedback operational amplifiers are combined to form adaptable, front-end multi-stage composite amplifiers. These distinctive composite amplifiers retain the individual qualities of their incorporated parts. The amplifier’s output voltage range is designed to match a standard ADC input voltage range. Easy to implement data converter circuitry and front-end signal amplification components were implemented on a detachable printed circuit daughterboard. A resourceful method of interfacing allows a series of Texas Instruments DSP mother-boards to be accessible. The PCB features single and dual input channel operation, adjustable input voltage ranges and sampling rates under software control. Bi-directional capability adds diagnostic functionality to the already versatile highperformance data acquisition system. The high-speed preconditioning and data converter element’s vulnerability to inconsistent and non-ideal real world effects such as noise, signal interference, parasitics, proximity effects and other layout problems is discussed. Real-time digital filtering, spectrum and phase analysis sub-routines form codeintensive multi-functional programmes that simultaneously display visually and record numerically the input signal properties in a comprehensive fashion. A user directed, automated, calibration software programme was developed to correct for precision-based errors caused by the signal conditioning instrumentation.","abstract_html":"The design of a high-speed, low-cost, generic analogue preconditioning system, which is capable of interfacing with a wide variety of transducers, is described. A multi-purpose, configurable data acquisition module, which is also capable of data generation, is presented. Software support is provided by advanced signal analysis and innovative presentation algorithms that are implemented on a powerful embedded digital signal processor. Precision voltage-feedback and high-speed current-feedback operational amplifiers are combined to form adaptable, front-end multi-stage composite amplifiers. These distinctive composite amplifiers retain the individual qualities of their incorporated parts. The amplifier’s output voltage range is designed to match a standard ADC input voltage range. Easy to implement data converter circuitry and front-end signal amplification components were implemented on a detachable printed circuit daughterboard. A resourceful method of interfacing allows a series of Texas Instruments DSP mother-boards to be accessible. The PCB features single and dual input channel operation, adjustable input voltage ranges and sampling rates under software control. Bi-directional capability adds diagnostic functionality to the already versatile highperformance data acquisition system. The high-speed preconditioning and data converter element’s vulnerability to inconsistent and non-ideal real world effects such as noise, signal interference, parasitics, proximity effects and other layout problems is discussed. Real-time digital filtering, spectrum and phase analysis sub-routines form codeintensive multi-functional programmes that simultaneously display visually and record numerically the input signal properties in a comprehensive fashion. A user directed, automated, calibration software programme was developed to correct for precision-based errors caused by the signal conditioning instrumentation.","abstract_has_math":false,"creators":["Leonard, Francis"],"institution":"Dublin City University","degree_name":"meng","degree_level":"masters","degree_discipline":null,"degree_department":null,"school":null,"contributors":[],"advisors":[],"committee_chairs":[],"committee_members":[],"year":2006,"date_issued":"2006","date_published":"2006","updated_at":"2026-11-25T19:58:32Z","subjects":[],"languages":["en"],"rights":[],"rights_urls":[],"identifier_entries":[],"person_identifier_entries":[]},"links":{"outbound_url":"https://doras.dcu.ie/18008/","outbound_label":"Repository record","outbound_source":"dc:identifier.uri","record_url":"https://doras.dcu.ie/18008/","record_label":"Repository record","record_kind":"record","record_source":"dc:identifier.uri","access_status":null,"embargo_until":null,"access_note":null},"oai_pmh_record":{"url":"https://union.ndltd.org/oai-pmh/?verb=GetRecord&metadataPrefix=oai_dc&identifier=oai%3Aunion.ndltd.org%3Arepo%3Adcu%3Aoai%3Adoras.dcu.ie%3A18008","prefix":"oai_dc"},"source_record":{"url":"http://doras.dcu.ie/cgi/oai2?verb=GetRecord&metadataPrefix=uketd_dc&identifier=oai%3Adoras.dcu.ie%3A18008","prefix":"uketd_dc"},"metadata_groups":[{"id":"people","label":"People","entries":[{"key":"dc:creator","label":"Author","values":["Leonard, Francis"]}]},{"id":"academic_context","label":"Academic Context","entries":[{"key":"dc:date","label":"Dc Date","values":["2006"]},{"key":"dc:date.issued","label":"Date","values":["2006"]},{"key":"dc:publisher.institution","label":"Dc Publisher Institution","values":["Dublin City University"]},{"key":"dc:type","label":"Dc Type","values":["Thesis"]},{"key":"dc:type.qualificationlevel","label":"Dc Type Qualificationlevel","values":["masters"]},{"key":"dc:type.qualificationname","label":"Dc Type Qualificationname","values":["meng"]}]},{"id":"language_rights","label":"Language and Rights","entries":[{"key":"dc:language","label":"Dc Language","values":["en"]}]},{"id":"identifiers","label":"Identifiers","entries":[{"key":"dc:identifier.uri","label":"Identifier URI","values":["https://doras.dcu.ie/18008/"]}]},{"id":"additional","label":"Additional Metadata","entries":[{"key":"dc:description.abstract","label":"Abstract","values":["The design of a high-speed, low-cost, generic analogue preconditioning system, which is capable of interfacing with a wide variety of transducers, is described. A multi-purpose, configurable data acquisition module, which is also capable of data generation, is presented. Software support is provided by advanced signal analysis and innovative presentation algorithms that are implemented on a powerful embedded digital signal processor. Precision voltage-feedback and high-speed current-feedback operational amplifiers are combined to form adaptable, front-end multi-stage composite amplifiers. These distinctive composite amplifiers retain the individual qualities of their incorporated parts. The amplifier’s output voltage range is designed to match a standard ADC input voltage range. Easy to implement data converter circuitry and front-end signal amplification components were implemented on a detachable printed circuit daughterboard. A resourceful method of interfacing allows a series of Texas Instruments DSP mother-boards to be accessible. The PCB features single and dual input channel operation, adjustable input voltage ranges and sampling rates under software control. Bi-directional capability adds diagnostic functionality to the already versatile highperformance data acquisition system. The high-speed preconditioning and data converter element’s vulnerability to inconsistent and non-ideal real world effects such as noise, signal interference, parasitics, proximity effects and other layout problems is discussed. Real-time digital filtering, spectrum and phase analysis sub-routines form codeintensive multi-functional programmes that simultaneously display visually and record numerically the input signal properties in a comprehensive fashion. A user directed, automated, calibration software programme was developed to correct for precision-based errors caused by the signal conditioning instrumentation."]},{"key":"dc:format","label":"Dc Format","values":["application/pdf"]},{"key":"dcterms:accessrights","label":"Dcterms Accessrights","values":["public"]},{"key":"dc:title","label":"Title","values":["Generic embedded sensor development: high-speed analogue preconditioning and data acquisition systems"]}]}],"canonical_facts":{"dc:creator":["Leonard, Francis"],"dc:date":["2006"],"dc:date.issued":["2006"],"dc:description.abstract":["The design of a high-speed, low-cost, generic analogue preconditioning system, which is capable of interfacing with a wide variety of transducers, is described. A multi-purpose, configurable data acquisition module, which is also capable of data generation, is presented. Software support is provided by advanced signal analysis and innovative presentation algorithms that are implemented on a powerful embedded digital signal processor. Precision voltage-feedback and high-speed current-feedback operational amplifiers are combined to form adaptable, front-end multi-stage composite amplifiers. These distinctive composite amplifiers retain the individual qualities of their incorporated parts. The amplifier’s output voltage range is designed to match a standard ADC input voltage range. Easy to implement data converter circuitry and front-end signal amplification components were implemented on a detachable printed circuit daughterboard. A resourceful method of interfacing allows a series of Texas Instruments DSP mother-boards to be accessible. The PCB features single and dual input channel operation, adjustable input voltage ranges and sampling rates under software control. Bi-directional capability adds diagnostic functionality to the already versatile highperformance data acquisition system. The high-speed preconditioning and data converter element’s vulnerability to inconsistent and non-ideal real world effects such as noise, signal interference, parasitics, proximity effects and other layout problems is discussed. Real-time digital filtering, spectrum and phase analysis sub-routines form codeintensive multi-functional programmes that simultaneously display visually and record numerically the input signal properties in a comprehensive fashion. A user directed, automated, calibration software programme was developed to correct for precision-based errors caused by the signal conditioning instrumentation."],"dc:format":["application/pdf"],"dc:identifier.uri":["https://doras.dcu.ie/18008/"],"dc:language":["en"],"dc:publisher.institution":["Dublin City University"],"dcterms:accessrights":["public"],"dc:title":["Generic embedded sensor development: high-speed analogue preconditioning and data acquisition systems"],"dc:type":["Thesis"],"dc:type.qualificationlevel":["masters"],"dc:type.qualificationname":["meng"]},"updated_at":"2026-11-25T19:58:32Z"}