{"id":{"repo_id":"edinburgh","oai_identifier":"oai:era.ed.ac.uk:1842/4628"},"canonical_url":"https://search.ndltd.org/etd/edinburgh/oai:era.ed.ac.uk:1842/4628","repository":{"repo_id":"edinburgh","name":"University of Edinburgh","base_url":"https://era.ed.ac.uk/server/oai/request"},"display":{"title":"Tribometer with programmable motion and load to investigate the influence of molecular structure on wear of orthopaedic polyethylene","abstract":"Total hip arthroplasty commonly involves a hard metallic/ceramic femoral ball component articulating against an acetabular ultra-high molecular weight polyethylene (UHMWPE) counter-bearing. A novel six-station, wear tribometer, featuring programmable load and motion, was designed to further the investigation into wear, sub-surface plasticity and debris generation of UHMWPE. This thesis describes the pin-on-plate device, its validation and subsequent use to assess unirradiated (-PE) and gamma-irradiated highly crosslinked (+PE) UHMWPE wear behaviour. With the emphasis on dynamic loading and a closer gait matched open wear path, the tribometer improves on the clinical relevance of pin-on-plate testing. There is a requirement for this type of machine in order to investigate the directional dependence of wear and debris generation of UHMWPE more accurately, where “simplified” tribometers (adequate for constant load/constant velocity and constant load/sinusoidal velocity work) are not capable or suitable. For the first time in orthopaedic pin-on-plate studies, tests were conducted using an advanced dynamic load synchronised to a more physiologically accurate elliptical motion path. To validate the machine, three orthopaedic polymers of clinical relevance; Polytetrafluroethylene, Polyacetal, and UHMWPE were subjected to linear-reciprocating (LR) and novel elliptical motion paths under a Paul-type load profile. All three polymers showed higher wear factors under elliptical motion, by up to 2 orders of magnitude, agreeing well with explanted values. The UHMWPE elliptical wear factor was comparable to that reported for clinical, where kelliptical = 1.56 x 10-6 mm3/Nm. In the crosslinked study, the mean steady state wear of -PE and +PE groups under linear reciprocating motion was not significantly different. However, under elliptical motion, crosslinking reduced UHMWPE wear by up to 92% when compared to the unirradiated group. In –PE pins worn under LR motion and in +PE pins subjected to both motion paths a sub-surface damage zone with reduced crystallinity and increased strain was measured using Raman spectroscopy. This was attributed to large strain accumulation in the slower wearing surfaces providing a mechanism for de-crystallisation. The discovery of such a near-surface layer is in good agreement with critical strain wear models. In disagreement, however, we found the sliding induced layer to extend to greater depths than previous reported.","abstract_html":"Total hip arthroplasty commonly involves a hard metallic/ceramic femoral ball component articulating against an acetabular ultra-high molecular weight polyethylene (UHMWPE) counter-bearing. A novel six-station, wear tribometer, featuring programmable load and motion, was designed to further the investigation into wear, sub-surface plasticity and debris generation of UHMWPE. This thesis describes the pin-on-plate device, its validation and subsequent use to assess unirradiated (-PE) and gamma-irradiated highly crosslinked (+PE) UHMWPE wear behaviour. With the emphasis on dynamic loading and a closer gait matched open wear path, the tribometer improves on the clinical relevance of pin-on-plate testing. There is a requirement for this type of machine in order to investigate the directional dependence of wear and debris generation of UHMWPE more accurately, where “simplified” tribometers (adequate for constant load/constant velocity and constant load/sinusoidal velocity work) are not capable or suitable. For the first time in orthopaedic pin-on-plate studies, tests were conducted using an advanced dynamic load synchronised to a more physiologically accurate elliptical motion path. To validate the machine, three orthopaedic polymers of clinical relevance; Polytetrafluroethylene, Polyacetal, and UHMWPE were subjected to linear-reciprocating (LR) and novel elliptical motion paths under a Paul-type load profile. All three polymers showed higher wear factors under elliptical motion, by up to 2 orders of magnitude, agreeing well with explanted values. The UHMWPE elliptical wear factor was comparable to that reported for clinical, where kelliptical = 1.56 x 10-6 mm3/Nm. In the crosslinked study, the mean steady state wear of -PE and +PE groups under linear reciprocating motion was not significantly different. However, under elliptical motion, crosslinking reduced UHMWPE wear by up to 92% when compared to the unirradiated group. In –PE pins worn under LR motion and in +PE pins subjected to both motion paths a sub-surface damage zone with reduced crystallinity and increased strain was measured using Raman spectroscopy. This was attributed to large strain accumulation in the slower wearing surfaces providing a mechanism for de-crystallisation. The discovery of such a near-surface layer is in good agreement with critical strain wear models. In disagreement, however, we found the sliding induced layer to extend to greater depths than previous reported.","abstract_has_math":false,"creators":["Kilgour, Alastair Scott"],"institution":"The University of Edinburgh","degree_name":null,"degree_level":null,"degree_discipline":null,"degree_department":null,"school":null,"contributors":[],"advisors":["Elfick, Alistair"],"committee_chairs":[],"committee_members":[],"year":2010,"date_issued":"2010","date_published":"2010","updated_at":"2026-11-19T03:35:23Z","subjects":["total hip arthroplasty","Polymer","Scanning electron","wear"],"languages":["en"],"rights":[],"rights_urls":[],"identifier_entries":[],"person_identifier_entries":[]},"links":{"outbound_url":"http://hdl.handle.net/1842/4628","outbound_label":"Handle","outbound_source":"dc:identifier.uri","record_url":"http://hdl.handle.net/1842/4628","record_label":"Repository record","record_kind":"handle","record_source":"dc:identifier.uri","download_url":null,"download_label":null,"download_source":null,"access_status":null,"embargo_until":null,"access_note":null},"source_record":{"url":"https://era.ed.ac.uk/server/oai/request?verb=GetRecord&metadataPrefix=dim&identifier=oai%3Aera.ed.ac.uk%3A1842%2F4628","prefix":"dim"},"metadata_groups":[{"id":"people","label":"People","entries":[{"key":"dc:contributor.advisor","label":"Advisor","values":["Elfick, Alistair"]},{"key":"dc:contributor.sponsor","label":"Sponsor","values":["Royal Society of Edinburgh"]},{"key":"dc:creator","label":"Author","values":["Kilgour, Alastair Scott"]}]},{"id":"academic_context","label":"Academic Context","entries":[{"key":"dc:date.accessioned","label":"Dc Date Accessioned","values":["2011-01-18T14:08:18Z"]},{"key":"dc:date.available","label":"Dc Date Available","values":["2011-01-18T14:08:18Z"]},{"key":"dc:date.issued","label":"Date","values":["2010"]},{"key":"dc:publisher","label":"Institution","values":["The University of Edinburgh"]},{"key":"dc:relation.hasversion","label":"Dc Relation Hasversion","values":["Atwood, S.A., Kilgour, A., Hoang, M., Furmanski, J., Pruitt, L.A., Lamellar orientation of plastically strained ultra-high molecular weight polyethylene, (Poster) presented at: Second International Conference on Mechanics of Biomaterials and Tissues, Lihue, Kaua'l Hawaii, USA, Dec. 2007.","Kilgour, A. and Elfick, A.P., Design and test of novel tribometer: investigation into wear performance of materials used in total hip replacements. Tribology - Materials, Surfaces & Interfaces, 2008. 2: p. 161-168.","Kilgour, A. and Elfick, A.P., Investigation of Polyethylene structure and its bearing on wear of joint replacements. (Abstract). Oral presentation In: World Biomaterials Congress. Amsterdam, 2008.","Kilgour, A. and Elfick, A.P., Influence of Crosslinked Polyethylene Structure on Wear of Joint Replacements. (Abstract). Oral presentation In: 35th Leeds-Lyon Symposium on Tribology, Leeds, 2008.","Kilgour, A. and Elfick, A.P., Influence of crosslinked polyethylene structure on wear of joint replacements. Tribology International, 2009. 42(11-12): p. 1582-1594."]},{"key":"dc:type","label":"Dc Type","values":["Thesis or Dissertation"]},{"key":"dc:type.qualificationlevel","label":"Dc Type Qualificationlevel","values":["Doctoral"]},{"key":"dc:type.qualificationname","label":"Dc Type Qualificationname","values":["PhD Doctor of Philosophy"]}]},{"id":"subjects_keywords","label":"Subjects and Keywords","entries":[{"key":"dc:subject","label":"Dc Subject","values":["total hip arthroplasty","Polymer","Scanning electron","wear"]}]},{"id":"language_rights","label":"Language and Rights","entries":[{"key":"dc:language.iso","label":"Language (ISO)","values":["en"]}]},{"id":"identifiers","label":"Identifiers","entries":[{"key":"dc:identifier.uri","label":"Identifier URI","values":["http://hdl.handle.net/1842/4628"]}]},{"id":"additional","label":"Additional Metadata","entries":[{"key":"dc:description.abstract","label":"Abstract","values":["Total hip arthroplasty commonly involves a hard metallic/ceramic femoral ball component articulating against an acetabular ultra-high molecular weight polyethylene (UHMWPE) counter-bearing. A novel six-station, wear tribometer, featuring programmable load and motion, was designed to further the investigation into wear, sub-surface plasticity and debris generation of UHMWPE. This thesis describes the pin-on-plate device, its validation and subsequent use to assess unirradiated (-PE) and gamma-irradiated highly crosslinked (+PE) UHMWPE wear behaviour. With the emphasis on dynamic loading and a closer gait matched open wear path, the tribometer improves on the clinical relevance of pin-on-plate testing. There is a requirement for this type of machine in order to investigate the directional dependence of wear and debris generation of UHMWPE more accurately, where “simplified” tribometers (adequate for constant load/constant velocity and constant load/sinusoidal velocity work) are not capable or suitable. For the first time in orthopaedic pin-on-plate studies, tests were conducted using an advanced dynamic load synchronised to a more physiologically accurate elliptical motion path. To validate the machine, three orthopaedic polymers of clinical relevance; Polytetrafluroethylene, Polyacetal, and UHMWPE were subjected to linear-reciprocating (LR) and novel elliptical motion paths under a Paul-type load profile. All three polymers showed higher wear factors under elliptical motion, by up to 2 orders of magnitude, agreeing well with explanted values. The UHMWPE elliptical wear factor was comparable to that reported for clinical, where kelliptical = 1.56 x 10-6 mm3/Nm. In the crosslinked study, the mean steady state wear of -PE and +PE groups under linear reciprocating motion was not significantly different. However, under elliptical motion, crosslinking reduced UHMWPE wear by up to 92% when compared to the unirradiated group. In –PE pins worn under LR motion and in +PE pins subjected to both motion paths a sub-surface damage zone with reduced crystallinity and increased strain was measured using Raman spectroscopy. This was attributed to large strain accumulation in the slower wearing surfaces providing a mechanism for de-crystallisation. The discovery of such a near-surface layer is in good agreement with critical strain wear models. In disagreement, however, we found the sliding induced layer to extend to greater depths than previous reported."]},{"key":"dc:title","label":"Title","values":["Tribometer with programmable motion and load to investigate the influence of molecular structure on wear of orthopaedic polyethylene"]},{"key":"others:access_status","label":"Others Access Status","values":["open.access"]}]}],"canonical_facts":{"dc:contributor.advisor":["Elfick, Alistair"],"dc:contributor.sponsor":["Royal Society of Edinburgh"],"dc:creator":["Kilgour, Alastair Scott"],"dc:date.accessioned":["2011-01-18T14:08:18Z"],"dc:date.available":["2011-01-18T14:08:18Z"],"dc:date.issued":["2010"],"dc:description.abstract":["Total hip arthroplasty commonly involves a hard metallic/ceramic femoral ball component articulating against an acetabular ultra-high molecular weight polyethylene (UHMWPE) counter-bearing. A novel six-station, wear tribometer, featuring programmable load and motion, was designed to further the investigation into wear, sub-surface plasticity and debris generation of UHMWPE. This thesis describes the pin-on-plate device, its validation and subsequent use to assess unirradiated (-PE) and gamma-irradiated highly crosslinked (+PE) UHMWPE wear behaviour. With the emphasis on dynamic loading and a closer gait matched open wear path, the tribometer improves on the clinical relevance of pin-on-plate testing. There is a requirement for this type of machine in order to investigate the directional dependence of wear and debris generation of UHMWPE more accurately, where “simplified” tribometers (adequate for constant load/constant velocity and constant load/sinusoidal velocity work) are not capable or suitable. For the first time in orthopaedic pin-on-plate studies, tests were conducted using an advanced dynamic load synchronised to a more physiologically accurate elliptical motion path. To validate the machine, three orthopaedic polymers of clinical relevance; Polytetrafluroethylene, Polyacetal, and UHMWPE were subjected to linear-reciprocating (LR) and novel elliptical motion paths under a Paul-type load profile. All three polymers showed higher wear factors under elliptical motion, by up to 2 orders of magnitude, agreeing well with explanted values. The UHMWPE elliptical wear factor was comparable to that reported for clinical, where kelliptical = 1.56 x 10-6 mm3/Nm. In the crosslinked study, the mean steady state wear of -PE and +PE groups under linear reciprocating motion was not significantly different. However, under elliptical motion, crosslinking reduced UHMWPE wear by up to 92% when compared to the unirradiated group. In –PE pins worn under LR motion and in +PE pins subjected to both motion paths a sub-surface damage zone with reduced crystallinity and increased strain was measured using Raman spectroscopy. This was attributed to large strain accumulation in the slower wearing surfaces providing a mechanism for de-crystallisation. The discovery of such a near-surface layer is in good agreement with critical strain wear models. In disagreement, however, we found the sliding induced layer to extend to greater depths than previous reported."],"dc:identifier.uri":["http://hdl.handle.net/1842/4628"],"dc:language.iso":["en"],"dc:publisher":["The University of Edinburgh"],"dc:relation.hasversion":["Atwood, S.A., Kilgour, A., Hoang, M., Furmanski, J., Pruitt, L.A., Lamellar orientation of plastically strained ultra-high molecular weight polyethylene, (Poster) presented at: Second International Conference on Mechanics of Biomaterials and Tissues, Lihue, Kaua'l Hawaii, USA, Dec. 2007.","Kilgour, A. and Elfick, A.P., Design and test of novel tribometer: investigation into wear performance of materials used in total hip replacements. Tribology - Materials, Surfaces & Interfaces, 2008. 2: p. 161-168.","Kilgour, A. and Elfick, A.P., Investigation of Polyethylene structure and its bearing on wear of joint replacements. (Abstract). Oral presentation In: World Biomaterials Congress. Amsterdam, 2008.","Kilgour, A. and Elfick, A.P., Influence of Crosslinked Polyethylene Structure on Wear of Joint Replacements. (Abstract). Oral presentation In: 35th Leeds-Lyon Symposium on Tribology, Leeds, 2008.","Kilgour, A. and Elfick, A.P., Influence of crosslinked polyethylene structure on wear of joint replacements. Tribology International, 2009. 42(11-12): p. 1582-1594."],"dc:subject":["total hip arthroplasty","Polymer","Scanning electron","wear"],"dc:title":["Tribometer with programmable motion and load to investigate the influence of molecular structure on wear of orthopaedic polyethylene"],"dc:type":["Thesis or Dissertation"],"dc:type.qualificationlevel":["Doctoral"],"dc:type.qualificationname":["PhD Doctor of Philosophy"],"others:access_status":["open.access"]},"updated_at":"2026-11-19T03:35:23Z"}