{"id":{"repo_id":"fau","oai_identifier":"oai:digitalcommons.fau.edu:etd_general_3-1560"},"canonical_url":"https://search.ndltd.org/etd/fau/oai:digitalcommons.fau.edu:etd_general_3-1560","repository":{"repo_id":"fau","name":"title=\"Search in OATD\">Florida Atlantic University","base_url":"https://digitalcommons.fau.edu/do/oai/"},"display":{"title":"MICROPROCESSOR BASED DRUG INFUSION CONTROL SYSTEM, EMPLOYING A MODEL REFERENCE ADAPTIVE CONTROL ALGORITHM, TO REGULATE BLOOD PRESSURE IN INTENSIVE CARE UNIT PATIENTS","abstract":"This microprocessor based drug infusion control system is intended to aid doctors and nurses in the care of critically ill cardiac patients. The patient's arterial blood pressure is monitored and the infusion rate of the vasodilator sodium nitroprusside is regulated based on a model reference adaptive control algorithm. The algorithm employs a reference model to approximate the patient drug response. The reference model output is compared with the patient blood pressure change and the adaptive controller parameters are changed bringing the patient drug response in closer agreement with the reference model. Drug infusion is digitally controlled by a microprocessor based system and employs a stepper motor driven peristaltic pump. Simulation studies have validated the system. Animal experiments and clinical studies will be conducted later.","abstract_html":"This microprocessor based drug infusion control system is intended to aid doctors and nurses in the care of critically ill cardiac patients. The patient&#x27;s arterial blood pressure is monitored and the infusion rate of the vasodilator sodium nitroprusside is regulated based on a model reference adaptive control algorithm. The algorithm employs a reference model to approximate the patient drug response. The reference model output is compared with the patient blood pressure change and the adaptive controller parameters are changed bringing the patient drug response in closer agreement with the reference model. Drug infusion is digitally controlled by a microprocessor based system and employs a stepper motor driven peristaltic pump. Simulation studies have validated the system. Animal experiments and clinical studies will be conducted later.","abstract_has_math":false,"creators":["HERNANDEZ, LEO."],"institution":null,"degree_name":null,"degree_level":"Thesis","degree_discipline":null,"degree_department":null,"school":null,"contributors":[],"advisors":[],"committee_chairs":[],"committee_members":[],"year":null,"date_issued":"","date_published":null,"updated_at":"2026-11-26T13:51:09Z","subjects":["Blood pressure--Regulation","Drug infusion pumps"],"languages":[],"rights":[],"rights_urls":[],"identifier_entries":[],"person_identifier_entries":[]},"links":{"outbound_url":"https://digitalcommons.fau.edu/etd_general_3/561","outbound_label":"Repository record","outbound_source":"dc:identifier","record_url":"https://digitalcommons.fau.edu/etd_general_3/561","record_label":"Repository record","record_kind":"record","record_source":"dc:identifier","access_status":null,"embargo_until":null,"access_note":null},"oai_pmh_record":{"url":"https://union.ndltd.org/oai-pmh/?verb=GetRecord&metadataPrefix=oai_dc&identifier=oai%3Aunion.ndltd.org%3Arepo%3Afau%3Aoai%3Adigitalcommons.fau.edu%3Aetd_general_3-1560","prefix":"oai_dc"},"source_record":{"url":"https://digitalcommons.fau.edu/do/oai/?verb=GetRecord&metadataPrefix=oai_etdms&identifier=oai%3Adigitalcommons.fau.edu%3Aetd_general_3-1560","prefix":"oai_etdms"},"metadata_groups":[{"id":"people","label":"People","entries":[{"key":"dc:creator","label":"Author","values":["HERNANDEZ, LEO."]}]},{"id":"academic_context","label":"Academic Context","entries":[{"key":"dc:date.created","label":"Dc Date Created","values":["1987-01-01T08:00:00Z"]},{"key":"thesis:degree_level","label":"Degree Level","values":["Thesis"]}]},{"id":"subjects_keywords","label":"Subjects and Keywords","entries":[{"key":"dc:subject","label":"Dc Subject","values":["Blood pressure--Regulation","Drug infusion pumps"]}]},{"id":"identifiers","label":"Identifiers","entries":[{"key":"dc:identifier","label":"Identifier","values":["https://digitalcommons.fau.edu/etd_general_3/561"]}]},{"id":"additional","label":"Additional Metadata","entries":[{"key":"dc:description.abstract","label":"Abstract","values":["This microprocessor based drug infusion control system is intended to aid doctors and nurses in the care of critically ill cardiac patients. The patient's arterial blood pressure is monitored and the infusion rate of the vasodilator sodium nitroprusside is regulated based on a model reference adaptive control algorithm. The algorithm employs a reference model to approximate the patient drug response. The reference model output is compared with the patient blood pressure change and the adaptive controller parameters are changed bringing the patient drug response in closer agreement with the reference model. Drug infusion is digitally controlled by a microprocessor based system and employs a stepper motor driven peristaltic pump. Simulation studies have validated the system. Animal experiments and clinical studies will be conducted later."]},{"key":"dc:description.note","label":"Dc Description Note","values":["College of Engineering and Computer Science|Thesis (M.S.E.)--Florida Atlantic University, 1987.|FAU Electronic Theses and Dissertations Collection|College of Engineering and Computer Science"]},{"key":"dc:title","label":"Title","values":["MICROPROCESSOR BASED DRUG INFUSION CONTROL SYSTEM, EMPLOYING A MODEL REFERENCE ADAPTIVE CONTROL ALGORITHM, TO REGULATE BLOOD PRESSURE IN INTENSIVE CARE UNIT PATIENTS"]}]}],"canonical_facts":{"dc:creator":["HERNANDEZ, LEO."],"dc:date.created":["1987-01-01T08:00:00Z"],"dc:description.abstract":["This microprocessor based drug infusion control system is intended to aid doctors and nurses in the care of critically ill cardiac patients. The patient's arterial blood pressure is monitored and the infusion rate of the vasodilator sodium nitroprusside is regulated based on a model reference adaptive control algorithm. The algorithm employs a reference model to approximate the patient drug response. The reference model output is compared with the patient blood pressure change and the adaptive controller parameters are changed bringing the patient drug response in closer agreement with the reference model. Drug infusion is digitally controlled by a microprocessor based system and employs a stepper motor driven peristaltic pump. Simulation studies have validated the system. Animal experiments and clinical studies will be conducted later."],"dc:description.note":["College of Engineering and Computer Science|Thesis (M.S.E.)--Florida Atlantic University, 1987.|FAU Electronic Theses and Dissertations Collection|College of Engineering and Computer Science"],"dc:identifier":["https://digitalcommons.fau.edu/etd_general_3/561"],"dc:subject":["Blood pressure--Regulation","Drug infusion pumps"],"dc:title":["MICROPROCESSOR BASED DRUG INFUSION CONTROL SYSTEM, EMPLOYING A MODEL REFERENCE ADAPTIVE CONTROL ALGORITHM, TO REGULATE BLOOD PRESSURE IN INTENSIVE CARE UNIT PATIENTS"],"thesis:degree_level":["Thesis"]},"updated_at":"2026-11-26T13:51:09Z"}