A dosimetric study of a heterogeneous phantom for lung stereotactic body radiation therapy comparing Monte Carlo and pencil beam calculations to dose distributions measured with a 2-d diode array

Abstract

dc:description.abstract

Monte Carlo (MC) and Pencil Beam (PB) calculations are compared to their measured planar dose distributions using a 2-D diode array for lung Stereotactic Body Radiation Therapy (SBRT). The planar dose distributions were studied for two different phantom types: an in-house heterogeneous phantom and a homogeneous phantom. The motivation is to mimic the human anatomy during a lung SBRT treatment and incorporate heterogeneities into the pre-treatment Quality Assurance process, where measured and calculated planar dose distributions are compared before the radiation treatment. Individual and combined field dosimetry has been performed for both fixed gantry angle (anterior to posterior) and planned gantry angle delivery. A gamma analysis has been performed for all beam arrangements. The measurements were obtained using the 2-D diode array MapCHECK 2™.

Degree

thesis:*
Level thesis:degree_level
Thesis

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Curley, Casey Michael

Subjects

dc:subject × 8

Identifiers

dc:identifier.*
Repository record dc:identifier
https://digitalcommons.fau.edu/etd_general_6/1154

Chain of custody

source
Harvested from
title="Search in OATD">Florida Atlantic University
Base URL
digitalcommons.fau.edu/do/oai/
Last updated
2026-11-26
Source record
OAI-PMH GetRecord
citation

Curley, Casey Michael. A dosimetric study of a heterogeneous phantom for lung stereotactic body radiation therapy comparing Monte Carlo and pencil beam calculations to dose distributions measured with a 2-d diode array. Thesis thesis, https://digitalcommons.fau.edu/etd_general_6/1154