{"id":{"repo_id":"fau","oai_identifier":"oai:digitalcommons.fau.edu:etd_general_6-3128"},"canonical_url":"https://search.ndltd.org/etd/fau/oai:digitalcommons.fau.edu:etd_general_6-3128","repository":{"repo_id":"fau","name":"title=\"Search in OATD\">Florida Atlantic University","base_url":"https://digitalcommons.fau.edu/do/oai/"},"display":{"title":"COMMISSIONING AND ACCEPTANCE TESTING OF A TRUEBEAM LINEAR ACCELERATOR","abstract":"Due to the difficulty of a complex commissioning technique for a multi energetic, multi-modality linear accelerator, I perform all the commissioning and acceptance testing for a TrueBeam linear accelerator with 4 megavoltage (MV) energies of which 2 are flattening filter-free (FFF) and 6 electron energies varying from 6 MeV to 20 MeV. A 2 dimensional (2D) water tank was used for scanning all the required field sizes for all the energies. The one dimensional (1D) water tank was used to collect all the output factors for all the photon fields sizes small to medium electron field sizes. For the large electron fields sizes, we had to use the 2D water tank. All the collected data was converted into a file type accepted by the planning system (Eclipse) and subsequently imported there. Treatment plans were generated using multiple forms of planning to verify the viability and quality of the beam data commissioned.","abstract_html":"Due to the difficulty of a complex commissioning technique for a multi energetic, multi-modality linear accelerator, I perform all the commissioning and acceptance testing for a TrueBeam linear accelerator with 4 megavoltage (MV) energies of which 2 are flattening filter-free (FFF) and 6 electron energies varying from 6 MeV to 20 MeV. A 2 dimensional (2D) water tank was used for scanning all the required field sizes for all the energies. The one dimensional (1D) water tank was used to collect all the output factors for all the photon fields sizes small to medium electron field sizes. For the large electron fields sizes, we had to use the 2D water tank. All the collected data was converted into a file type accepted by the planning system (Eclipse) and subsequently imported there. Treatment plans were generated using multiple forms of planning to verify the viability and quality of the beam data commissioned.","abstract_has_math":false,"creators":["Dumitru, Nicolae"],"institution":null,"degree_name":null,"degree_level":"Thesis","degree_discipline":null,"degree_department":null,"school":null,"contributors":[],"advisors":[],"committee_chairs":[],"committee_members":[],"year":null,"date_issued":"","date_published":null,"updated_at":"2026-11-26T13:53:19Z","subjects":["Linear accelerators","Cancer--Treatment","Radiation dosimetry"],"languages":[],"rights":[],"rights_urls":[],"identifier_entries":[],"person_identifier_entries":[]},"links":{"outbound_url":"https://digitalcommons.fau.edu/etd_general_6/2129","outbound_label":"Repository record","outbound_source":"dc:identifier","record_url":"https://digitalcommons.fau.edu/etd_general_6/2129","record_label":"Repository record","record_kind":"record","record_source":"dc:identifier","access_status":null,"embargo_until":null,"access_note":null},"oai_pmh_record":{"url":"https://union.ndltd.org/oai-pmh/?verb=GetRecord&metadataPrefix=oai_dc&identifier=oai%3Aunion.ndltd.org%3Arepo%3Afau%3Aoai%3Adigitalcommons.fau.edu%3Aetd_general_6-3128","prefix":"oai_dc"},"source_record":{"url":"https://digitalcommons.fau.edu/do/oai/?verb=GetRecord&metadataPrefix=oai_etdms&identifier=oai%3Adigitalcommons.fau.edu%3Aetd_general_6-3128","prefix":"oai_etdms"},"metadata_groups":[{"id":"people","label":"People","entries":[{"key":"dc:creator","label":"Author","values":["Dumitru, Nicolae"]}]},{"id":"academic_context","label":"Academic Context","entries":[{"key":"dc:date.created","label":"Dc Date Created","values":["2019-01-01T08:00:00Z"]},{"key":"thesis:degree_level","label":"Degree Level","values":["Thesis"]}]},{"id":"subjects_keywords","label":"Subjects and Keywords","entries":[{"key":"dc:subject","label":"Dc Subject","values":["Linear accelerators","Cancer--Treatment","Radiation dosimetry"]}]},{"id":"identifiers","label":"Identifiers","entries":[{"key":"dc:identifier","label":"Identifier","values":["https://digitalcommons.fau.edu/etd_general_6/2129"]}]},{"id":"additional","label":"Additional Metadata","entries":[{"key":"dc:description.abstract","label":"Abstract","values":["Due to the difficulty of a complex commissioning technique for a multi energetic, multi-modality linear accelerator, I perform all the commissioning and acceptance testing for a TrueBeam linear accelerator with 4 megavoltage (MV) energies of which 2 are flattening filter-free (FFF) and 6 electron energies varying from 6 MeV to 20 MeV. A 2 dimensional (2D) water tank was used for scanning all the required field sizes for all the energies. The one dimensional (1D) water tank was used to collect all the output factors for all the photon fields sizes small to medium electron field sizes. For the large electron fields sizes, we had to use the 2D water tank. All the collected data was converted into a file type accepted by the planning system (Eclipse) and subsequently imported there. Treatment plans were generated using multiple forms of planning to verify the viability and quality of the beam data commissioned."]},{"key":"dc:description.note","label":"Dc Description Note","values":["Includes bibliography.|Thesis (M.S.)--Florida Atlantic University, 2019.|FAU Electronic Theses and Dissertations Collection|Includes bibliography."]},{"key":"dc:title","label":"Title","values":["COMMISSIONING AND ACCEPTANCE TESTING OF A TRUEBEAM LINEAR ACCELERATOR"]}]}],"canonical_facts":{"dc:creator":["Dumitru, Nicolae"],"dc:date.created":["2019-01-01T08:00:00Z"],"dc:description.abstract":["Due to the difficulty of a complex commissioning technique for a multi energetic, multi-modality linear accelerator, I perform all the commissioning and acceptance testing for a TrueBeam linear accelerator with 4 megavoltage (MV) energies of which 2 are flattening filter-free (FFF) and 6 electron energies varying from 6 MeV to 20 MeV. A 2 dimensional (2D) water tank was used for scanning all the required field sizes for all the energies. The one dimensional (1D) water tank was used to collect all the output factors for all the photon fields sizes small to medium electron field sizes. For the large electron fields sizes, we had to use the 2D water tank. All the collected data was converted into a file type accepted by the planning system (Eclipse) and subsequently imported there. Treatment plans were generated using multiple forms of planning to verify the viability and quality of the beam data commissioned."],"dc:description.note":["Includes bibliography.|Thesis (M.S.)--Florida Atlantic University, 2019.|FAU Electronic Theses and Dissertations Collection|Includes bibliography."],"dc:identifier":["https://digitalcommons.fau.edu/etd_general_6/2129"],"dc:subject":["Linear accelerators","Cancer--Treatment","Radiation dosimetry"],"dc:title":["COMMISSIONING AND ACCEPTANCE TESTING OF A TRUEBEAM LINEAR ACCELERATOR"],"thesis:degree_level":["Thesis"]},"updated_at":"2026-11-26T13:53:19Z"}