{"id":{"repo_id":"gatech","oai_identifier":"oai:repository.gatech.edu:1853/47675"},"canonical_url":"https://search.ndltd.org/etd/gatech/oai:repository.gatech.edu:1853/47675","repository":{"repo_id":"gatech","name":"Georgia Tech","base_url":"https://repository.gatech.edu/server/oai/request"},"display":{"title":"Small field dose measurements with Gafchromic film","abstract":"Purpose: To examine the dosimetric characteristics of Gafchromic EBT3 film when measuring small fields of radiation, and compare it against other common radiation detectors. Methods and Materials: EBT3 film was placed in a solid water phantom and irradiated with 6MV photons, field sizes from 10x10cm2 down to 6x6mm2. The films were scanned with a Vidar DosimetryPRO Advantage Red scanner, and analyzed with RIT113 software. The films were also scanned at different orientations and times to quantify the discrepancies associated with scanning orientation and post-exposure darkening. The same fields were measured with a PTW TN30013 farmer chamber, an Exradin T1 cylindrical ion chamber, a PTW parallel plate ion chamber, and a Sun Nuclear Edge Detector (diode). Output factors were calculated for each detector and compared for accuracy. The output factors were measured from a Varian Clinac iX, Clinac 21EX, Trilogy, and TrueBeam; as well as a Novalis Tx. The outputs from different machines at different clinics were compared. Results: The EBT3 film and Edge Detector were the only detectors that succeeded in accurately measuring the output from all field sizes; the ion chambers were too large and failed for field sizes below 4x4cm2 due to volume averaging. The dose measured with the film increased by an average of 8.8% after one week post-irradiation. The dose measured was also reduced by an average of 4.4% by scanning the film in landscape orientation, as opposed to portrait orientation. It was shown that the output factors for the smallest field of 6x6mm2--successfully measured with film and diode--varied between 0.54-0.74 for five different machines at three different clinics. Conclusions: The feasibility of using Gafchromic EBT3 film to measure very small fields of radiation is confirmed. Of the other 4 detectors used, only the diode was shown to be capable of accurately measuring small fields of radiation. The need to optimize the film dosimetry process--including the time films are scanned post-irradiation, the consistency of the scanning orientation of the calibration and subsequent films, and the measurement procedure on the computer software--is highlighted.","abstract_html":"Purpose: To examine the dosimetric characteristics of Gafchromic EBT3 film when measuring small fields of radiation, and compare it against other common radiation detectors. Methods and Materials: EBT3 film was placed in a solid water phantom and irradiated with 6MV photons, field sizes from 10x10cm2 down to 6x6mm2. The films were scanned with a Vidar DosimetryPRO Advantage Red scanner, and analyzed with RIT113 software. The films were also scanned at different orientations and times to quantify the discrepancies associated with scanning orientation and post-exposure darkening. The same fields were measured with a PTW TN30013 farmer chamber, an Exradin T1 cylindrical ion chamber, a PTW parallel plate ion chamber, and a Sun Nuclear Edge Detector (diode). Output factors were calculated for each detector and compared for accuracy. The output factors were measured from a Varian Clinac iX, Clinac 21EX, Trilogy, and TrueBeam; as well as a Novalis Tx. The outputs from different machines at different clinics were compared. Results: The EBT3 film and Edge Detector were the only detectors that succeeded in accurately measuring the output from all field sizes; the ion chambers were too large and failed for field sizes below 4x4cm2 due to volume averaging. The dose measured with the film increased by an average of 8.8% after one week post-irradiation. The dose measured was also reduced by an average of 4.4% by scanning the film in landscape orientation, as opposed to portrait orientation. It was shown that the output factors for the smallest field of 6x6mm2--successfully measured with film and diode--varied between 0.54-0.74 for five different machines at three different clinics. Conclusions: The feasibility of using Gafchromic EBT3 film to measure very small fields of radiation is confirmed. Of the other 4 detectors used, only the diode was shown to be capable of accurately measuring small fields of radiation. The need to optimize the film dosimetry process--including the time films are scanned post-irradiation, the consistency of the scanning orientation of the calibration and subsequent films, and the measurement procedure on the computer software--is highlighted.","abstract_has_math":false,"creators":["Underwood, Ryan John"],"institution":"Georgia Institute of Technology","degree_name":null,"degree_level":null,"degree_discipline":null,"degree_department":"Mechanical Engineering","school":null,"contributors":[],"advisors":["Wang, C.-K. Chris"],"committee_chairs":[],"committee_members":["Dhabbaan, Anees","Elder, Eric"],"year":2013,"date_issued":"2013-04-09","date_published":"2013-04-09","updated_at":"2026-11-19T03:57:06Z","subjects":["Radiochromic film","Small field dosimetry","Radiation therapy","Output factors","Radiotherapy","Medical physics","Radioisotope brachytherapy","Radiation dosimetry"],"languages":[],"rights":[],"rights_urls":[],"identifier_entries":[],"person_identifier_entries":[]},"links":{"outbound_url":"http://hdl.handle.net/1853/47675","outbound_label":"Handle","outbound_source":"dc:identifier.uri","record_url":"http://hdl.handle.net/1853/47675","record_label":"Repository record","record_kind":"handle","record_source":"dc:identifier.uri","download_url":null,"download_label":null,"download_source":null,"access_status":null,"embargo_until":null,"access_note":null},"source_record":{"url":"https://repository.gatech.edu/server/oai/request?verb=GetRecord&metadataPrefix=dim&identifier=oai%3Arepository.gatech.edu%3A1853%2F47675","prefix":"dim"},"metadata_groups":[{"id":"people","label":"People","entries":[{"key":"dc:contributor.advisor","label":"Advisor","values":["Wang, C.-K. Chris"]},{"key":"dc:contributor.committeemember","label":"Committee Member","values":["Dhabbaan, Anees","Elder, Eric"]},{"key":"dc:contributor.department","label":"Department","values":["Mechanical Engineering"]},{"key":"dc:contributor.other","label":"Dc Contributor Other","values":["George W. Woodruff School of Mechanical Engineering","College of Engineering"]},{"key":"dc:creator","label":"Author","values":["Underwood, Ryan John"]}]},{"id":"academic_context","label":"Academic Context","entries":[{"key":"dc:date.accessioned","label":"Dc Date Accessioned","values":["2013-06-15T02:55:24Z"]},{"key":"dc:date.available","label":"Dc Date Available","values":["2013-06-15T02:55:24Z"]},{"key":"dc:date.issued","label":"Date","values":["2013-04-09"]},{"key":"dc:publisher","label":"Institution","values":["Georgia Institute of Technology"]},{"key":"dc:relation.advisorentity","label":"Dc Relation Advisorentity","values":["71d23899-d6d4-4d7f-b9bc-d30323a6f442"]},{"key":"dc:relation.ispartofseries","label":"Dc Relation Ispartofseries","values":["Master of Science in Medical Physics"]},{"key":"dc:relation.orgunit","label":"Dc Relation Orgunit","values":["c01ff908-c25f-439b-bf10-a074ed886bb7","7c022d60-21d5-497c-b552-95e489a06569"]},{"key":"dc:relation.seriesentity","label":"Dc Relation Seriesentity","values":["3173665c-cfb8-400b-bcf7-8fbfc614e0fb"]},{"key":"dc:type","label":"Dc Type","values":["Text","Thesis"]}]},{"id":"subjects_keywords","label":"Subjects and Keywords","entries":[{"key":"dc:subject","label":"Dc Subject","values":["Radiochromic film","Small field dosimetry","Radiation therapy","Output factors"]},{"key":"dc:subject.lcsh","label":"Dc Subject Lcsh","values":["Radiotherapy","Medical physics","Radioisotope brachytherapy","Radiation dosimetry"]}]},{"id":"identifiers","label":"Identifiers","entries":[{"key":"dc:identifier.uri","label":"Identifier URI","values":["http://hdl.handle.net/1853/47675"]}]},{"id":"additional","label":"Additional Metadata","entries":[{"key":"dc:description.abstract","label":"Abstract","values":["Purpose: To examine the dosimetric characteristics of Gafchromic EBT3 film when measuring small fields of radiation, and compare it against other common radiation detectors. Methods and Materials: EBT3 film was placed in a solid water phantom and irradiated with 6MV photons, field sizes from 10x10cm2 down to 6x6mm2. The films were scanned with a Vidar DosimetryPRO Advantage Red scanner, and analyzed with RIT113 software. The films were also scanned at different orientations and times to quantify the discrepancies associated with scanning orientation and post-exposure darkening. The same fields were measured with a PTW TN30013 farmer chamber, an Exradin T1 cylindrical ion chamber, a PTW parallel plate ion chamber, and a Sun Nuclear Edge Detector (diode). Output factors were calculated for each detector and compared for accuracy. The output factors were measured from a Varian Clinac iX, Clinac 21EX, Trilogy, and TrueBeam; as well as a Novalis Tx. The outputs from different machines at different clinics were compared. Results: The EBT3 film and Edge Detector were the only detectors that succeeded in accurately measuring the output from all field sizes; the ion chambers were too large and failed for field sizes below 4x4cm2 due to volume averaging. The dose measured with the film increased by an average of 8.8% after one week post-irradiation. The dose measured was also reduced by an average of 4.4% by scanning the film in landscape orientation, as opposed to portrait orientation. It was shown that the output factors for the smallest field of 6x6mm2--successfully measured with film and diode--varied between 0.54-0.74 for five different machines at three different clinics. Conclusions: The feasibility of using Gafchromic EBT3 film to measure very small fields of radiation is confirmed. Of the other 4 detectors used, only the diode was shown to be capable of accurately measuring small fields of radiation. The need to optimize the film dosimetry process--including the time films are scanned post-irradiation, the consistency of the scanning orientation of the calibration and subsequent films, and the measurement procedure on the computer software--is highlighted."]},{"key":"dc:description.degree","label":"Dc Description Degree","values":["MS"]},{"key":"dc:title","label":"Title","values":["Small field dose measurements with Gafchromic film"]},{"key":"dspace:entity.type","label":"Dspace Entity Type","values":["Publication"]}]}],"canonical_facts":{"dc:contributor.advisor":["Wang, C.-K. Chris"],"dc:contributor.committeemember":["Dhabbaan, Anees","Elder, Eric"],"dc:contributor.department":["Mechanical Engineering"],"dc:contributor.other":["George W. Woodruff School of Mechanical Engineering","College of Engineering"],"dc:creator":["Underwood, Ryan John"],"dc:date.accessioned":["2013-06-15T02:55:24Z"],"dc:date.available":["2013-06-15T02:55:24Z"],"dc:date.issued":["2013-04-09"],"dc:description.abstract":["Purpose: To examine the dosimetric characteristics of Gafchromic EBT3 film when measuring small fields of radiation, and compare it against other common radiation detectors. Methods and Materials: EBT3 film was placed in a solid water phantom and irradiated with 6MV photons, field sizes from 10x10cm2 down to 6x6mm2. The films were scanned with a Vidar DosimetryPRO Advantage Red scanner, and analyzed with RIT113 software. The films were also scanned at different orientations and times to quantify the discrepancies associated with scanning orientation and post-exposure darkening. The same fields were measured with a PTW TN30013 farmer chamber, an Exradin T1 cylindrical ion chamber, a PTW parallel plate ion chamber, and a Sun Nuclear Edge Detector (diode). Output factors were calculated for each detector and compared for accuracy. The output factors were measured from a Varian Clinac iX, Clinac 21EX, Trilogy, and TrueBeam; as well as a Novalis Tx. The outputs from different machines at different clinics were compared. Results: The EBT3 film and Edge Detector were the only detectors that succeeded in accurately measuring the output from all field sizes; the ion chambers were too large and failed for field sizes below 4x4cm2 due to volume averaging. The dose measured with the film increased by an average of 8.8% after one week post-irradiation. The dose measured was also reduced by an average of 4.4% by scanning the film in landscape orientation, as opposed to portrait orientation. It was shown that the output factors for the smallest field of 6x6mm2--successfully measured with film and diode--varied between 0.54-0.74 for five different machines at three different clinics. Conclusions: The feasibility of using Gafchromic EBT3 film to measure very small fields of radiation is confirmed. Of the other 4 detectors used, only the diode was shown to be capable of accurately measuring small fields of radiation. The need to optimize the film dosimetry process--including the time films are scanned post-irradiation, the consistency of the scanning orientation of the calibration and subsequent films, and the measurement procedure on the computer software--is highlighted."],"dc:description.degree":["MS"],"dc:identifier.uri":["http://hdl.handle.net/1853/47675"],"dc:publisher":["Georgia Institute of Technology"],"dc:relation.advisorentity":["71d23899-d6d4-4d7f-b9bc-d30323a6f442"],"dc:relation.ispartofseries":["Master of Science in Medical Physics"],"dc:relation.orgunit":["c01ff908-c25f-439b-bf10-a074ed886bb7","7c022d60-21d5-497c-b552-95e489a06569"],"dc:relation.seriesentity":["3173665c-cfb8-400b-bcf7-8fbfc614e0fb"],"dc:subject":["Radiochromic film","Small field dosimetry","Radiation therapy","Output factors"],"dc:subject.lcsh":["Radiotherapy","Medical physics","Radioisotope brachytherapy","Radiation dosimetry"],"dc:title":["Small field dose measurements with Gafchromic film"],"dc:type":["Text","Thesis"],"dspace:entity.type":["Publication"]},"updated_at":"2026-11-19T03:57:06Z"}