{"id":{"repo_id":"gatech","oai_identifier":"oai:repository.gatech.edu:1853/60148"},"canonical_url":"https://search.ndltd.org/etd/gatech/oai:repository.gatech.edu:1853/60148","repository":{"repo_id":"gatech","name":"Georgia Tech","base_url":"https://repository.gatech.edu/server/oai/request"},"display":{"title":"Gait optimization with a real-time closed-loop artificial sensory feedback","abstract":"Individuals with unilateral lower limb musculoskeletal and neurological conditions, including prosthetic users, experience asymmetric walking. It may result in undesirable compensations by the body and secondary conditions (osteoarthritis, low back pain, etc.). My goal was to develop a real-time closed-loop control system for a sensing, bone-anchored transtibial prosthesis interfaced with residual peripheral nerves and muscles. The prosthesis and control system would allow users to sense ground contact during walking and control the prosthetic ankle using natural motor commands to automatically correct asymmetries and instability of walking. To inform the design of the prosthesis and control system, I investigated in walking cats the effects of manipulating tactile sensory feedback from paw pads and stretch-dependent feedback from thigh muscles on symmetry and stability of walking. I found that removal of tactile and muscle length-dependent feedback resulted in profound changes in symmetry and stability of walking. In addition, electrical stimulation of the distal tibial nerve, innervating paw pads, during the stance phase of walking substantially reduced and sometimes reversed effects of sensory feedback removal. I then demonstrated that a real-time closed-loop gait control system could control step length symmetry of walking cats by stimulation of the distal tibial nerve. This system could be used for correcting asymmetries during walking in people with sensorimotor deficits. I also developed a prototype of a sensing transtibial prosthesis with an ankle joint controlled by activity of residual muscles. The results of this study provide new insights into the role of sensory feedback in control of locomotion and offer new engineering solutions for bone-anchored, limb neuro-prostheses and for improving pathological gait.","abstract_html":"Individuals with unilateral lower limb musculoskeletal and neurological conditions, including prosthetic users, experience asymmetric walking. It may result in undesirable compensations by the body and secondary conditions (osteoarthritis, low back pain, etc.). My goal was to develop a real-time closed-loop control system for a sensing, bone-anchored transtibial prosthesis interfaced with residual peripheral nerves and muscles. The prosthesis and control system would allow users to sense ground contact during walking and control the prosthetic ankle using natural motor commands to automatically correct asymmetries and instability of walking. To inform the design of the prosthesis and control system, I investigated in walking cats the effects of manipulating tactile sensory feedback from paw pads and stretch-dependent feedback from thigh muscles on symmetry and stability of walking. I found that removal of tactile and muscle length-dependent feedback resulted in profound changes in symmetry and stability of walking. In addition, electrical stimulation of the distal tibial nerve, innervating paw pads, during the stance phase of walking substantially reduced and sometimes reversed effects of sensory feedback removal. I then demonstrated that a real-time closed-loop gait control system could control step length symmetry of walking cats by stimulation of the distal tibial nerve. This system could be used for correcting asymmetries during walking in people with sensorimotor deficits. I also developed a prototype of a sensing transtibial prosthesis with an ankle joint controlled by activity of residual muscles. The results of this study provide new insights into the role of sensory feedback in control of locomotion and offer new engineering solutions for bone-anchored, limb neuro-prostheses and for improving pathological gait.","abstract_has_math":false,"creators":["Park, Hangue"],"institution":"Georgia Institute of Technology","degree_name":null,"degree_level":"Doctoral","degree_discipline":null,"degree_department":"Electrical and Computer Engineering","school":null,"contributors":[],"advisors":["Prilutsky, Boris I.","DeWeerth, Stephen P."],"committee_chairs":[],"committee_members":["Butera, Robert J.","Chang, Young-Hui","Grover, Martha A.","Lemay, Michel A."],"year":2017,"date_issued":"2017-07-12","date_published":"2017-07-12","updated_at":"2026-11-19T04:54:18Z","subjects":["Neuroprosthesis","Sensory feedback","Gait optimization","Electrical nerve stimulation"],"languages":["en_US"],"rights":[],"rights_urls":[],"identifier_entries":[],"person_identifier_entries":[]},"links":{"outbound_url":"http://hdl.handle.net/1853/60148","outbound_label":"Handle","outbound_source":"dc:identifier.uri","record_url":"http://hdl.handle.net/1853/60148","record_label":"Repository record","record_kind":"handle","record_source":"dc:identifier.uri","download_url":null,"download_label":null,"download_source":null,"access_status":null,"embargo_until":null,"access_note":null},"source_record":{"url":"https://repository.gatech.edu/server/oai/request?verb=GetRecord&metadataPrefix=dim&identifier=oai%3Arepository.gatech.edu%3A1853%2F60148","prefix":"dim"},"metadata_groups":[{"id":"people","label":"People","entries":[{"key":"dc:contributor.advisor","label":"Advisor","values":["Prilutsky, Boris I.","DeWeerth, Stephen P."]},{"key":"dc:contributor.committeemember","label":"Committee Member","values":["Butera, Robert J.","Chang, Young-Hui","Grover, Martha A.","Lemay, Michel A."]},{"key":"dc:contributor.department","label":"Department","values":["Electrical and Computer Engineering"]},{"key":"dc:contributor.other","label":"Dc Contributor Other","values":["School of Electrical and Computer Engineering","College of Engineering"]},{"key":"dc:creator","label":"Author","values":["Park, Hangue"]}]},{"id":"academic_context","label":"Academic Context","entries":[{"key":"dc:date.accessioned","label":"Dc Date Accessioned","values":["2018-08-20T15:29:15Z"]},{"key":"dc:date.available","label":"Dc Date Available","values":["2018-08-20T15:29:15Z"]},{"key":"dc:date.created","label":"Dc Date Created","values":["2017-08"]},{"key":"dc:date.issued","label":"Date","values":["2017-07-12"]},{"key":"dc:date.submitted","label":"Dc Date Submitted","values":["August 2017"]},{"key":"dc:date.updated","label":"Dc Date Updated","values":["2018-08-20T15:29:16Z"]},{"key":"dc:publisher","label":"Institution","values":["Georgia Institute of Technology"]},{"key":"dc:relation.advisorentity","label":"Dc Relation Advisorentity","values":["0913f0f8-648a-4335-bcf2-915667e7bdd2","6b8c24a1-7328-4161-8715-b26e0231ae78"]},{"key":"dc:relation.ispartofseries","label":"Dc Relation Ispartofseries","values":["Doctor of Philosophy with a Major in Electrical and Computer Engineering"]},{"key":"dc:relation.orgunit","label":"Dc Relation Orgunit","values":["5b7adef2-447c-4270-b9fc-846bd76f80f2","7c022d60-21d5-497c-b552-95e489a06569"]},{"key":"dc:relation.seriesentity","label":"Dc Relation Seriesentity","values":["3890cd56-8ecf-424b-8074-8f5ebd8dbe15"]},{"key":"dc:type","label":"Dc Type","values":["Text","Dissertation"]},{"key":"thesis:degree_level","label":"Degree Level","values":["Doctoral"]}]},{"id":"subjects_keywords","label":"Subjects and Keywords","entries":[{"key":"dc:subject","label":"Dc Subject","values":["Neuroprosthesis","Sensory feedback","Gait optimization","Electrical nerve stimulation"]}]},{"id":"language_rights","label":"Language and Rights","entries":[{"key":"dc:language.iso","label":"Language (ISO)","values":["en_US"]}]},{"id":"identifiers","label":"Identifiers","entries":[{"key":"dc:identifier.uri","label":"Identifier URI","values":["http://hdl.handle.net/1853/60148"]}]},{"id":"additional","label":"Additional Metadata","entries":[{"key":"dc:description.abstract","label":"Abstract","values":["Individuals with unilateral lower limb musculoskeletal and neurological conditions, including prosthetic users, experience asymmetric walking. It may result in undesirable compensations by the body and secondary conditions (osteoarthritis, low back pain, etc.). My goal was to develop a real-time closed-loop control system for a sensing, bone-anchored transtibial prosthesis interfaced with residual peripheral nerves and muscles. The prosthesis and control system would allow users to sense ground contact during walking and control the prosthetic ankle using natural motor commands to automatically correct asymmetries and instability of walking. To inform the design of the prosthesis and control system, I investigated in walking cats the effects of manipulating tactile sensory feedback from paw pads and stretch-dependent feedback from thigh muscles on symmetry and stability of walking. I found that removal of tactile and muscle length-dependent feedback resulted in profound changes in symmetry and stability of walking. In addition, electrical stimulation of the distal tibial nerve, innervating paw pads, during the stance phase of walking substantially reduced and sometimes reversed effects of sensory feedback removal. I then demonstrated that a real-time closed-loop gait control system could control step length symmetry of walking cats by stimulation of the distal tibial nerve. This system could be used for correcting asymmetries during walking in people with sensorimotor deficits. I also developed a prototype of a sensing transtibial prosthesis with an ankle joint controlled by activity of residual muscles. The results of this study provide new insights into the role of sensory feedback in control of locomotion and offer new engineering solutions for bone-anchored, limb neuro-prostheses and for improving pathological gait."]},{"key":"dc:description.degree","label":"Dc Description Degree","values":["Ph.D."]},{"key":"dc:format.mimetype","label":"Dc Format Mimetype","values":["application/pdf"]},{"key":"dc:title","label":"Title","values":["Gait optimization with a real-time closed-loop artificial sensory feedback"]},{"key":"dspace:entity.type","label":"Dspace Entity Type","values":["Publication"]}]}],"canonical_facts":{"dc:contributor.advisor":["Prilutsky, Boris I.","DeWeerth, Stephen P."],"dc:contributor.committeemember":["Butera, Robert J.","Chang, Young-Hui","Grover, Martha A.","Lemay, Michel A."],"dc:contributor.department":["Electrical and Computer Engineering"],"dc:contributor.other":["School of Electrical and Computer Engineering","College of Engineering"],"dc:creator":["Park, Hangue"],"dc:date.accessioned":["2018-08-20T15:29:15Z"],"dc:date.available":["2018-08-20T15:29:15Z"],"dc:date.created":["2017-08"],"dc:date.issued":["2017-07-12"],"dc:date.submitted":["August 2017"],"dc:date.updated":["2018-08-20T15:29:16Z"],"dc:description.abstract":["Individuals with unilateral lower limb musculoskeletal and neurological conditions, including prosthetic users, experience asymmetric walking. It may result in undesirable compensations by the body and secondary conditions (osteoarthritis, low back pain, etc.). My goal was to develop a real-time closed-loop control system for a sensing, bone-anchored transtibial prosthesis interfaced with residual peripheral nerves and muscles. The prosthesis and control system would allow users to sense ground contact during walking and control the prosthetic ankle using natural motor commands to automatically correct asymmetries and instability of walking. To inform the design of the prosthesis and control system, I investigated in walking cats the effects of manipulating tactile sensory feedback from paw pads and stretch-dependent feedback from thigh muscles on symmetry and stability of walking. I found that removal of tactile and muscle length-dependent feedback resulted in profound changes in symmetry and stability of walking. In addition, electrical stimulation of the distal tibial nerve, innervating paw pads, during the stance phase of walking substantially reduced and sometimes reversed effects of sensory feedback removal. I then demonstrated that a real-time closed-loop gait control system could control step length symmetry of walking cats by stimulation of the distal tibial nerve. This system could be used for correcting asymmetries during walking in people with sensorimotor deficits. I also developed a prototype of a sensing transtibial prosthesis with an ankle joint controlled by activity of residual muscles. The results of this study provide new insights into the role of sensory feedback in control of locomotion and offer new engineering solutions for bone-anchored, limb neuro-prostheses and for improving pathological gait."],"dc:description.degree":["Ph.D."],"dc:format.mimetype":["application/pdf"],"dc:identifier.uri":["http://hdl.handle.net/1853/60148"],"dc:language.iso":["en_US"],"dc:publisher":["Georgia Institute of Technology"],"dc:relation.advisorentity":["0913f0f8-648a-4335-bcf2-915667e7bdd2","6b8c24a1-7328-4161-8715-b26e0231ae78"],"dc:relation.ispartofseries":["Doctor of Philosophy with a Major in Electrical and Computer Engineering"],"dc:relation.orgunit":["5b7adef2-447c-4270-b9fc-846bd76f80f2","7c022d60-21d5-497c-b552-95e489a06569"],"dc:relation.seriesentity":["3890cd56-8ecf-424b-8074-8f5ebd8dbe15"],"dc:subject":["Neuroprosthesis","Sensory feedback","Gait optimization","Electrical nerve stimulation"],"dc:title":["Gait optimization with a real-time closed-loop artificial sensory feedback"],"dc:type":["Text","Dissertation"],"dspace:entity.type":["Publication"],"thesis:degree_level":["Doctoral"]},"updated_at":"2026-11-19T04:54:18Z"}