{"id":{"repo_id":"greece","oai_identifier":"oai:10442/35911"},"canonical_url":"https://search.ndltd.org/etd/greece/oai:10442/35911","repository":{"repo_id":"greece","name":"Greek National Archive of PhD Theses","base_url":"https://phdtheses.ekt.gr/eadd_oai/request"},"display":{"title":"In vivo δοσιμετρία ακτινοθεραπευτικής δέσμης φωτονίων με χρήση a-Si EPID.","abstract":"Purpose: The aim of this study is the validation of a new easily applicable EPID dosimetry method based on the back projection approach and the comparison of this method with the EBT2 film dosimetry method.Methods: The dosimetric characteristics of the EPID were determined. The methodology to calculate the dose on the central axis of the photon beam was described. The EPID dosimetry method was validated with phantom measurements using an ionization chamber. In vivo measurements of ten prostate patients treated with 3D conformal plans were also performed both with the EPID and the EBT2 films. Results: The phantom measurements revealed a mean dispersion of 1.7% between EPID and ionization chamber doses and 2.2% between EPID and EBT2 film doses. While, the in vivo measurements in prostate patients revealed a mean dispersion of 0.8% between EPID doses and treatment planning calculated doses and 1% between EPID and EBT2 film doses.Conclusions: The presented EPID dosimetry method is accurate, easily applicable to all types of flat panels without the use of any sophisticated software and is not time consuming.","abstract_html":"Purpose: The aim of this study is the validation of a new easily applicable EPID dosimetry method based on the back projection approach and the comparison of this method with the EBT2 film dosimetry method.Methods: The dosimetric characteristics of the EPID were determined. The methodology to calculate the dose on the central axis of the photon beam was described. The EPID dosimetry method was validated with phantom measurements using an ionization chamber. In vivo measurements of ten prostate patients treated with 3D conformal plans were also performed both with the EPID and the EBT2 films. Results: The phantom measurements revealed a mean dispersion of 1.7% between EPID and ionization chamber doses and 2.2% between EPID and EBT2 film doses. While, the in vivo measurements in prostate patients revealed a mean dispersion of 0.8% between EPID doses and treatment planning calculated doses and 1% between EPID and EBT2 film doses.Conclusions: The presented EPID dosimetry method is accurate, easily applicable to all types of flat panels without the use of any sophisticated software and is not time consuming.","abstract_has_math":false,"creators":["Kalliopi, Mandourari","Μανδουράρη, Καλλιόπη"],"institution":"National and Kapodistrian University of Athens","degree_name":null,"degree_level":null,"degree_discipline":null,"degree_department":null,"school":null,"contributors":[],"advisors":[],"committee_chairs":[],"committee_members":[],"year":2015,"date_issued":"2015","date_published":"2015","updated_at":"2026-11-18T21:23:57Z","subjects":["EBT2 φιλμ","EPID","In vivo δοσιμετρία","διερχόμενη δόση","Βαθμονόμηση","EBT2 films","in vivo dosimetry","transit dose","Calibration","Φυσικές Επιστήμες","Φυσική","Ιατρική και Επιστήμες Υγείας","Άλλες Ιατρικές Επιστήμες","Natural Sciences","Physical Sciences","Medical and Health Sciences","Other Medical Sciences"],"languages":["gre"],"rights":[],"rights_urls":[],"identifier_entries":[{"key":"dc:identifier","label":"Identifier","values":["10.12681/eadd/35911"],"render_values":[{"text":"10.12681/eadd/35911","href":"https://doi.org/10.12681/eadd/35911","code":true}]}],"person_identifier_entries":[]},"links":{"outbound_url":"http://hdl.handle.net/10442/hedi/35911","outbound_label":"Handle","outbound_source":"dc:identifier","record_url":"http://hdl.handle.net/10442/hedi/35911","record_label":"Repository record","record_kind":"handle","record_source":"dc:identifier","download_url":null,"download_label":null,"download_source":null,"access_status":null,"embargo_until":null,"access_note":null},"source_record":{"url":"https://phdtheses.ekt.gr/eadd_oai/request?verb=GetRecord&metadataPrefix=oai_dc&identifier=oai%3A10442%2F35911","prefix":"oai_dc"},"metadata_groups":[{"id":"people","label":"People","entries":[{"key":"dc:creator","label":"Author","values":["Kalliopi, Mandourari","Μανδουράρη, Καλλιόπη"]},{"key":"dc:creator.fathername","label":"Dc Creator Fathername","values":["Stylianos"]}]},{"id":"academic_context","label":"Academic Context","entries":[{"key":"dc:date","label":"Dc Date","values":["2015"]},{"key":"dc:publisher","label":"Institution","values":["National and Kapodistrian University of Athens","Εθνικό και Καποδιστριακό Πανεπιστήμιο Αθηνών (ΕΚΠΑ)"]},{"key":"dc:type","label":"Dc Type","values":["PhD Thesis"]}]},{"id":"subjects_keywords","label":"Subjects and Keywords","entries":[{"key":"dc:subject","label":"Dc Subject","values":["EBT2 φιλμ","EPID","In vivo δοσιμετρία","διερχόμενη δόση","Βαθμονόμηση","EBT2 films","in vivo dosimetry","transit dose","Calibration","Φυσικές Επιστήμες","Φυσική","Ιατρική και Επιστήμες Υγείας","Άλλες Ιατρικές Επιστήμες","Natural Sciences","Physical Sciences","Medical and Health Sciences","Other Medical Sciences"]}]},{"id":"language_rights","label":"Language and Rights","entries":[{"key":"dc:language","label":"Dc Language","values":["gre"]}]},{"id":"identifiers","label":"Identifiers","entries":[{"key":"dc:identifier","label":"Identifier","values":["10.12681/eadd/35911","http://hdl.handle.net/10442/hedi/35911"]}]},{"id":"additional","label":"Additional Metadata","entries":[{"key":"dc:description","label":"Description","values":["Purpose: The aim of this study is the validation of a new easily applicable EPID dosimetry method based on the back projection approach and the comparison of this method with the EBT2 film dosimetry method.Methods: The dosimetric characteristics of the EPID were determined. The methodology to calculate the dose on the central axis of the photon beam was described. The EPID dosimetry method was validated with phantom measurements using an ionization chamber. In vivo measurements of ten prostate patients treated with 3D conformal plans were also performed both with the EPID and the EBT2 films. Results: The phantom measurements revealed a mean dispersion of 1.7% between EPID and ionization chamber doses and 2.2% between EPID and EBT2 film doses. While, the in vivo measurements in prostate patients revealed a mean dispersion of 0.8% between EPID doses and treatment planning calculated doses and 1% between EPID and EBT2 film doses.Conclusions: The presented EPID dosimetry method is accurate, easily applicable to all types of flat panels without the use of any sophisticated software and is not time consuming.","Σκοπός: Η αξιολόγηση μιας εύκολα εφαρμόσιμης μεθόδου δοσιμετρίας EPID, η οποία βασίζεται στη προσέγγιση της οπισθοπροβολής, και η σύγκρισή της με τη μέθοδο δοσιμετρίας EBT2 φιλμ. Μέθοδος: Προσδιορίστηκαν τα δοσιμετρικά χαρακτηριστικά του EPID. Έγινε περιγραφή της μεθοδολογίας για τον προσδιορισμό της δόσης επί του κεντρικού άξονα της δέσμης φωτονίων. Η μέθοδος επικυρώθηκε με μετρήσεις σε ομοιογενές ομοίωμα με τη χρήση θαλάμου ιονισμού. Πραγματοποιήθηκαν in vivo μετρήσεις σε δέκα ασθενείς με καρκίνο του προστάτη (3D-CRT) και με τις δυο μεθόδους.Αποτελέσματα: Σύμφωνα με τα αποτελέσματα των μετρήσεων στο ομοιογενές ομοίωμα προκύπτει μέση απόκλιση 1.7% ανάμεσα στις δόσεις που μετρήθηκαν με το EPID και τον θάλαμο ιονισμού, και μέση απόκλιση 2.2% ανάμεσα στις δόσεις που μετρήθηκαν με το EPID και τα φιλμ EBT2 στο μέσο επίπεδο. Ενώ, σύμφωνα με τα αποτελέσματα των in vivo μετρήσεων προκύπτει μέση απόκλιση 0.8% ανάμεσα στις δόσεις που μετρήθηκαν με το EPID και υπολογίστηκαν από το TPS, και μέση απόκλιση 1% ανάμεσα στις δόσεις που μετρήθηκαν με το EPID και τα φιλμ EBT2 στο ισόκεντρο.Συμπεράσματα: Η προτεινόμενη μέθοδος είναι ακριβής, εύκολα εφαρμόσιμη σε όλους τους τύπους των flat panel και ο υπολογισμός της δόσης μπορεί να πραγματοποιηθεί χωρίς τη χρήση κάποιου εξεζητημένου λογισμικού."]},{"key":"dc:title","label":"Title","values":["In vivo δοσιμετρία ακτινοθεραπευτικής δέσμης φωτονίων με χρήση a-Si EPID.","In vivo dosimetry of a photon radiotherapy beam with the use of an a-Si EPID"]}]}],"canonical_facts":{"dc:creator":["Kalliopi, Mandourari","Μανδουράρη, Καλλιόπη"],"dc:creator.fathername":["Stylianos"],"dc:date":["2015"],"dc:description":["Purpose: The aim of this study is the validation of a new easily applicable EPID dosimetry method based on the back projection approach and the comparison of this method with the EBT2 film dosimetry method.Methods: The dosimetric characteristics of the EPID were determined. The methodology to calculate the dose on the central axis of the photon beam was described. The EPID dosimetry method was validated with phantom measurements using an ionization chamber. In vivo measurements of ten prostate patients treated with 3D conformal plans were also performed both with the EPID and the EBT2 films. Results: The phantom measurements revealed a mean dispersion of 1.7% between EPID and ionization chamber doses and 2.2% between EPID and EBT2 film doses. While, the in vivo measurements in prostate patients revealed a mean dispersion of 0.8% between EPID doses and treatment planning calculated doses and 1% between EPID and EBT2 film doses.Conclusions: The presented EPID dosimetry method is accurate, easily applicable to all types of flat panels without the use of any sophisticated software and is not time consuming.","Σκοπός: Η αξιολόγηση μιας εύκολα εφαρμόσιμης μεθόδου δοσιμετρίας EPID, η οποία βασίζεται στη προσέγγιση της οπισθοπροβολής, και η σύγκρισή της με τη μέθοδο δοσιμετρίας EBT2 φιλμ. Μέθοδος: Προσδιορίστηκαν τα δοσιμετρικά χαρακτηριστικά του EPID. Έγινε περιγραφή της μεθοδολογίας για τον προσδιορισμό της δόσης επί του κεντρικού άξονα της δέσμης φωτονίων. Η μέθοδος επικυρώθηκε με μετρήσεις σε ομοιογενές ομοίωμα με τη χρήση θαλάμου ιονισμού. Πραγματοποιήθηκαν in vivo μετρήσεις σε δέκα ασθενείς με καρκίνο του προστάτη (3D-CRT) και με τις δυο μεθόδους.Αποτελέσματα: Σύμφωνα με τα αποτελέσματα των μετρήσεων στο ομοιογενές ομοίωμα προκύπτει μέση απόκλιση 1.7% ανάμεσα στις δόσεις που μετρήθηκαν με το EPID και τον θάλαμο ιονισμού, και μέση απόκλιση 2.2% ανάμεσα στις δόσεις που μετρήθηκαν με το EPID και τα φιλμ EBT2 στο μέσο επίπεδο. Ενώ, σύμφωνα με τα αποτελέσματα των in vivo μετρήσεων προκύπτει μέση απόκλιση 0.8% ανάμεσα στις δόσεις που μετρήθηκαν με το EPID και υπολογίστηκαν από το TPS, και μέση απόκλιση 1% ανάμεσα στις δόσεις που μετρήθηκαν με το EPID και τα φιλμ EBT2 στο ισόκεντρο.Συμπεράσματα: Η προτεινόμενη μέθοδος είναι ακριβής, εύκολα εφαρμόσιμη σε όλους τους τύπους των flat panel και ο υπολογισμός της δόσης μπορεί να πραγματοποιηθεί χωρίς τη χρήση κάποιου εξεζητημένου λογισμικού."],"dc:identifier":["10.12681/eadd/35911","http://hdl.handle.net/10442/hedi/35911"],"dc:language":["gre"],"dc:publisher":["National and Kapodistrian University of Athens","Εθνικό και Καποδιστριακό Πανεπιστήμιο Αθηνών (ΕΚΠΑ)"],"dc:subject":["EBT2 φιλμ","EPID","In vivo δοσιμετρία","διερχόμενη δόση","Βαθμονόμηση","EBT2 films","in vivo dosimetry","transit dose","Calibration","Φυσικές Επιστήμες","Φυσική","Ιατρική και Επιστήμες Υγείας","Άλλες Ιατρικές Επιστήμες","Natural Sciences","Physical Sciences","Medical and Health Sciences","Other Medical Sciences"],"dc:title":["In vivo δοσιμετρία ακτινοθεραπευτικής δέσμης φωτονίων με χρήση a-Si EPID.","In vivo dosimetry of a photon radiotherapy beam with the use of an a-Si EPID"],"dc:type":["PhD Thesis"]},"updated_at":"2026-11-18T21:23:57Z"}