{"id":{"repo_id":"gsu","oai_identifier":"oai:digitalcommons.georgiasouthern.edu:etd-2672"},"canonical_url":"https://search.ndltd.org/etd/gsu/oai:digitalcommons.georgiasouthern.edu:etd-2672","repository":{"repo_id":"gsu","name":"Georgia Southern University","base_url":"https://digitalcommons.georgiasouthern.edu/do/oai/"},"display":{"title":"Validity and Reliability of the Balance Tracking System™ During Static Stance","abstract":"Introduction: The Balance Tracking System™ (BTrackS™) is a balance board designed to quickly analyze postural control through center of pressure (COP) analysis. Purpose: Examine concurrent validity and test-retest reliability of the BTrackS™ during static stance. Methods: A convenience sample of 51 healthy collegiate students between 18 - 25 years old (21.8 ± 3.1 years) completed four, 20 second (s) trials of feet together static stance during both eyes open and closed trials. Data was simultaneously collected on the BTrackS™ and a laboratory force plate to establish concurrent validity. A second testing session was administered 48 - 72 hours later to establish test-retest reliability. Independent variables were device (BTrackS™/force plate) and time (Time Point 1/Time Point 2). Dependent variables were anterior-posterior (AP) and medial-lateral (ML) COP excursion. Eight Pearson’s Product Correlations were used to compare the relationship of dependent variables between the BTrackS™ and laboratory force plate. Four Intra-Class Correlations (ICC) were used to compare the relationship of dependent variables between Time Point 1 and Time Point 2 measured by the BTrackS™. Four 2 x 2 (device x time) repeated measures ANOVA’s were used to determine the magnitude of differences within independent variables. Results: Pearson Product Correlations showed an excellent relationship (r = 0.867 – 0.968) between the BTrackS™ and laboratory force plate. However, the 2 x 2 repeated measures ANOVA’s showed a significant difference between devices during both eyes open and closed conditions for all dependent variables (p < 0.001). Intra-Class Correlations showed an excellent relationship (ICC = 0.859 – 0.984) between time points for the BTrackS™. The 2 x 2 repeated measures ANOVA’s showed no significant differences between time points for the BTrackS™ during eyes open and closed conditions for all dependent variables (p = 0.185 – 0.976). Conclusion: Findings suggest that the BTrackS™ is strongly correlated with a laboratory force plate, but is significantly different. The BTrackS™ is strongly correlated and not significantly different within 48 - 72 hours. Additional research regarding an acceptable difference between the BTrackS™ and a laboratory force plate is warranted before it can be used clinically.","abstract_html":"Introduction: The Balance Tracking System™ (BTrackS™) is a balance board designed to quickly analyze postural control through center of pressure (COP) analysis. Purpose: Examine concurrent validity and test-retest reliability of the BTrackS™ during static stance. Methods: A convenience sample of 51 healthy collegiate students between 18 - 25 years old (21.8 ± 3.1 years) completed four, 20 second (s) trials of feet together static stance during both eyes open and closed trials. Data was simultaneously collected on the BTrackS™ and a laboratory force plate to establish concurrent validity. A second testing session was administered 48 - 72 hours later to establish test-retest reliability. Independent variables were device (BTrackS™/force plate) and time (Time Point 1/Time Point 2). Dependent variables were anterior-posterior (AP) and medial-lateral (ML) COP excursion. Eight Pearson’s Product Correlations were used to compare the relationship of dependent variables between the BTrackS™ and laboratory force plate. Four Intra-Class Correlations (ICC) were used to compare the relationship of dependent variables between Time Point 1 and Time Point 2 measured by the BTrackS™. Four 2 x 2 (device x time) repeated measures ANOVA’s were used to determine the magnitude of differences within independent variables. Results: Pearson Product Correlations showed an excellent relationship (r = 0.867 – 0.968) between the BTrackS™ and laboratory force plate. However, the 2 x 2 repeated measures ANOVA’s showed a significant difference between devices during both eyes open and closed conditions for all dependent variables (p &lt; 0.001). Intra-Class Correlations showed an excellent relationship (ICC = 0.859 – 0.984) between time points for the BTrackS™. The 2 x 2 repeated measures ANOVA’s showed no significant differences between time points for the BTrackS™ during eyes open and closed conditions for all dependent variables (p = 0.185 – 0.976). Conclusion: Findings suggest that the BTrackS™ is strongly correlated with a laboratory force plate, but is significantly different. The BTrackS™ is strongly correlated and not significantly different within 48 - 72 hours. Additional research regarding an acceptable difference between the BTrackS™ and a laboratory force plate is warranted before it can be used clinically.","abstract_has_math":false,"creators":["Grotewold, Cody L"],"institution":null,"degree_name":"Master of Science in Kinesiology (M.S.)","degree_level":"Thesis (open access)","degree_discipline":"Department of Health and Kinesiology","degree_department":null,"school":null,"contributors":["Nicholas Murray","Tamerah Hunt"],"advisors":[],"committee_chairs":[],"committee_members":[],"year":null,"date_issued":"","date_published":null,"updated_at":"2026-11-19T07:54:40Z","subjects":["Postural Control","Balance Tracking System","Validity","Reliability","Equipment and Supplies","Sports Medicine"],"languages":[],"rights":[],"rights_urls":["http://creativecommons.org/licenses/by/4.0/"],"identifier_entries":[],"person_identifier_entries":[]},"links":{"outbound_url":"https://digitalcommons.georgiasouthern.edu/etd/1564","outbound_label":"Repository record","outbound_source":"dc:identifier","record_url":"https://digitalcommons.georgiasouthern.edu/etd/1564","record_label":"Repository record","record_kind":"record","record_source":"dc:identifier","download_url":null,"download_label":null,"download_source":null,"access_status":null,"embargo_until":null,"access_note":null},"source_record":{"url":"https://digitalcommons.georgiasouthern.edu/do/oai/?verb=GetRecord&metadataPrefix=oai_etdms&identifier=oai%3Adigitalcommons.georgiasouthern.edu%3Aetd-2672","prefix":"oai_etdms"},"metadata_groups":[{"id":"people","label":"People","entries":[{"key":"dc:contributor","label":"Contributor","values":["Nicholas Murray","Tamerah Hunt"]},{"key":"dc:creator","label":"Author","values":["Grotewold, Cody L"]}]},{"id":"academic_context","label":"Academic Context","entries":[{"key":"dc:date.available","label":"Dc Date Available","values":["2017-04-12T07:00:00Z"]},{"key":"dc:date.created","label":"Dc Date Created","values":["2017-01-01T08:00:00Z"]},{"key":"thesis:degree_discipline","label":"Discipline","values":["Department of Health and Kinesiology"]},{"key":"thesis:degree_level","label":"Degree Level","values":["Thesis (open access)"]},{"key":"thesis:degree_name","label":"Degree Name","values":["Master of Science in Kinesiology (M.S.)"]}]},{"id":"subjects_keywords","label":"Subjects and Keywords","entries":[{"key":"dc:subject","label":"Dc Subject","values":["Postural Control","Balance Tracking System","Validity","Reliability","Equipment and Supplies","Sports Medicine"]}]},{"id":"language_rights","label":"Language and Rights","entries":[{"key":"dc:rights.license","label":"Dc Rights License","values":["http://creativecommons.org/licenses/by/4.0/"]}]},{"id":"identifiers","label":"Identifiers","entries":[{"key":"dc:identifier","label":"Identifier","values":["https://digitalcommons.georgiasouthern.edu/etd/1564"]}]},{"id":"additional","label":"Additional Metadata","entries":[{"key":"dc:description.abstract","label":"Abstract","values":["Introduction: The Balance Tracking System™ (BTrackS™) is a balance board designed to quickly analyze postural control through center of pressure (COP) analysis. Purpose: Examine concurrent validity and test-retest reliability of the BTrackS™ during static stance. Methods: A convenience sample of 51 healthy collegiate students between 18 - 25 years old (21.8 ± 3.1 years) completed four, 20 second (s) trials of feet together static stance during both eyes open and closed trials. Data was simultaneously collected on the BTrackS™ and a laboratory force plate to establish concurrent validity. A second testing session was administered 48 - 72 hours later to establish test-retest reliability. Independent variables were device (BTrackS™/force plate) and time (Time Point 1/Time Point 2). Dependent variables were anterior-posterior (AP) and medial-lateral (ML) COP excursion. Eight Pearson’s Product Correlations were used to compare the relationship of dependent variables between the BTrackS™ and laboratory force plate. Four Intra-Class Correlations (ICC) were used to compare the relationship of dependent variables between Time Point 1 and Time Point 2 measured by the BTrackS™. Four 2 x 2 (device x time) repeated measures ANOVA’s were used to determine the magnitude of differences within independent variables. Results: Pearson Product Correlations showed an excellent relationship (r = 0.867 – 0.968) between the BTrackS™ and laboratory force plate. However, the 2 x 2 repeated measures ANOVA’s showed a significant difference between devices during both eyes open and closed conditions for all dependent variables (p < 0.001). Intra-Class Correlations showed an excellent relationship (ICC = 0.859 – 0.984) between time points for the BTrackS™. The 2 x 2 repeated measures ANOVA’s showed no significant differences between time points for the BTrackS™ during eyes open and closed conditions for all dependent variables (p = 0.185 – 0.976). Conclusion: Findings suggest that the BTrackS™ is strongly correlated with a laboratory force plate, but is significantly different. The BTrackS™ is strongly correlated and not significantly different within 48 - 72 hours. Additional research regarding an acceptable difference between the BTrackS™ and a laboratory force plate is warranted before it can be used clinically."]},{"key":"dc:title","label":"Title","values":["Validity and Reliability of the Balance Tracking System™ During Static Stance"]}]}],"canonical_facts":{"dc:contributor":["Nicholas Murray","Tamerah Hunt"],"dc:creator":["Grotewold, Cody L"],"dc:date.available":["2017-04-12T07:00:00Z"],"dc:date.created":["2017-01-01T08:00:00Z"],"dc:description.abstract":["Introduction: The Balance Tracking System™ (BTrackS™) is a balance board designed to quickly analyze postural control through center of pressure (COP) analysis. Purpose: Examine concurrent validity and test-retest reliability of the BTrackS™ during static stance. Methods: A convenience sample of 51 healthy collegiate students between 18 - 25 years old (21.8 ± 3.1 years) completed four, 20 second (s) trials of feet together static stance during both eyes open and closed trials. Data was simultaneously collected on the BTrackS™ and a laboratory force plate to establish concurrent validity. A second testing session was administered 48 - 72 hours later to establish test-retest reliability. Independent variables were device (BTrackS™/force plate) and time (Time Point 1/Time Point 2). Dependent variables were anterior-posterior (AP) and medial-lateral (ML) COP excursion. Eight Pearson’s Product Correlations were used to compare the relationship of dependent variables between the BTrackS™ and laboratory force plate. Four Intra-Class Correlations (ICC) were used to compare the relationship of dependent variables between Time Point 1 and Time Point 2 measured by the BTrackS™. Four 2 x 2 (device x time) repeated measures ANOVA’s were used to determine the magnitude of differences within independent variables. Results: Pearson Product Correlations showed an excellent relationship (r = 0.867 – 0.968) between the BTrackS™ and laboratory force plate. However, the 2 x 2 repeated measures ANOVA’s showed a significant difference between devices during both eyes open and closed conditions for all dependent variables (p < 0.001). Intra-Class Correlations showed an excellent relationship (ICC = 0.859 – 0.984) between time points for the BTrackS™. The 2 x 2 repeated measures ANOVA’s showed no significant differences between time points for the BTrackS™ during eyes open and closed conditions for all dependent variables (p = 0.185 – 0.976). Conclusion: Findings suggest that the BTrackS™ is strongly correlated with a laboratory force plate, but is significantly different. The BTrackS™ is strongly correlated and not significantly different within 48 - 72 hours. Additional research regarding an acceptable difference between the BTrackS™ and a laboratory force plate is warranted before it can be used clinically."],"dc:identifier":["https://digitalcommons.georgiasouthern.edu/etd/1564"],"dc:rights.license":["http://creativecommons.org/licenses/by/4.0/"],"dc:subject":["Postural Control","Balance Tracking System","Validity","Reliability","Equipment and Supplies","Sports Medicine"],"dc:title":["Validity and Reliability of the Balance Tracking System™ During Static Stance"],"thesis:degree_discipline":["Department of Health and Kinesiology"],"thesis:degree_level":["Thesis (open access)"],"thesis:degree_name":["Master of Science in Kinesiology (M.S.)"]},"updated_at":"2026-11-19T07:54:40Z"}