{"id":{"repo_id":"gsu","oai_identifier":"oai:digitalcommons.georgiasouthern.edu:etd-3131"},"canonical_url":"https://search.ndltd.org/etd/gsu/oai:digitalcommons.georgiasouthern.edu:etd-3131","repository":{"repo_id":"gsu","name":"Georgia Southern University","base_url":"https://digitalcommons.georgiasouthern.edu/do/oai/"},"display":{"title":"Development of an Impedance-Controlled Hot Snare Polypectomy Device to Minimize Tissue Damage","abstract":"This study explores the ability to measure the changing impedance, ex vivo, of a porcine colon sample while undergoing a high-frequency alternating current from an Olympus PSD-30 electrosurgical unit and stop the applied current before excessive tissue damage ensues. The causes of the thermal damage are first examined, followed by the construction and testing of the impedance-controlled feedback device. Perforation was observed to occur when the impedance of the tissue sample increased by 25% or more. Using this information, the device was tested for five power settings ranging from 10W-50W. In each trial, the feedback device stopped the applied current to the tissue samples when the measured impedance exceeded the cut-off threshold of a 25% increase from the starting impedance. The device was found to have an accuracy of ±5Ω. This successfully minimized undue tissue damage and proved able to prevent serious complications such as perforation from occurring.","abstract_html":"This study explores the ability to measure the changing impedance, ex vivo, of a porcine colon sample while undergoing a high-frequency alternating current from an Olympus PSD-30 electrosurgical unit and stop the applied current before excessive tissue damage ensues. The causes of the thermal damage are first examined, followed by the construction and testing of the impedance-controlled feedback device. Perforation was observed to occur when the impedance of the tissue sample increased by 25% or more. Using this information, the device was tested for five power settings ranging from 10W-50W. In each trial, the feedback device stopped the applied current to the tissue samples when the measured impedance exceeded the cut-off threshold of a 25% increase from the starting impedance. The device was found to have an accuracy of ±5Ω. This successfully minimized undue tissue damage and proved able to prevent serious complications such as perforation from occurring.","abstract_has_math":false,"creators":["Thornton, CurtisLee"],"institution":null,"degree_name":"Master of Science, Mechanical Engineering","degree_level":"Thesis (open access)","degree_discipline":"Department of Mechanical Engineering","degree_department":null,"school":null,"contributors":["Biswanath Samanta","Minchul Shin"],"advisors":[],"committee_chairs":[],"committee_members":[],"year":null,"date_issued":"","date_published":null,"updated_at":"2026-11-19T07:54:56Z","subjects":["Polypectomy","Snare","Impedance","Feedback control","Radiofrequency ablation","Biomedical Devices and Instrumentation"],"languages":[],"rights":[],"rights_urls":["http://creativecommons.org/licenses/by/4.0/"],"identifier_entries":[],"person_identifier_entries":[]},"links":{"outbound_url":"https://digitalcommons.georgiasouthern.edu/etd/1970","outbound_label":"Repository record","outbound_source":"dc:identifier","record_url":"https://digitalcommons.georgiasouthern.edu/etd/1970","record_label":"Repository record","record_kind":"record","record_source":"dc:identifier","download_url":null,"download_label":null,"download_source":null,"access_status":null,"embargo_until":null,"access_note":null},"source_record":{"url":"https://digitalcommons.georgiasouthern.edu/do/oai/?verb=GetRecord&metadataPrefix=oai_etdms&identifier=oai%3Adigitalcommons.georgiasouthern.edu%3Aetd-3131","prefix":"oai_etdms"},"metadata_groups":[{"id":"people","label":"People","entries":[{"key":"dc:contributor","label":"Contributor","values":["Biswanath Samanta","Minchul Shin"]},{"key":"dc:creator","label":"Author","values":["Thornton, CurtisLee"]}]},{"id":"academic_context","label":"Academic Context","entries":[{"key":"dc:date.available","label":"Dc Date Available","values":["2019-06-28T07:00:00Z"]},{"key":"dc:date.created","label":"Dc Date Created","values":["2019-01-01T08:00:00Z"]},{"key":"thesis:degree_discipline","label":"Discipline","values":["Department of Mechanical Engineering"]},{"key":"thesis:degree_level","label":"Degree Level","values":["Thesis (open access)"]},{"key":"thesis:degree_name","label":"Degree Name","values":["Master of Science, Mechanical Engineering"]}]},{"id":"subjects_keywords","label":"Subjects and Keywords","entries":[{"key":"dc:subject","label":"Dc Subject","values":["Polypectomy","Snare","Impedance","Feedback control","Radiofrequency ablation","Biomedical Devices and Instrumentation"]}]},{"id":"language_rights","label":"Language and Rights","entries":[{"key":"dc:rights.license","label":"Dc Rights License","values":["http://creativecommons.org/licenses/by/4.0/"]}]},{"id":"identifiers","label":"Identifiers","entries":[{"key":"dc:identifier","label":"Identifier","values":["https://digitalcommons.georgiasouthern.edu/etd/1970"]}]},{"id":"additional","label":"Additional Metadata","entries":[{"key":"dc:description.abstract","label":"Abstract","values":["This study explores the ability to measure the changing impedance, ex vivo, of a porcine colon sample while undergoing a high-frequency alternating current from an Olympus PSD-30 electrosurgical unit and stop the applied current before excessive tissue damage ensues. The causes of the thermal damage are first examined, followed by the construction and testing of the impedance-controlled feedback device. Perforation was observed to occur when the impedance of the tissue sample increased by 25% or more. Using this information, the device was tested for five power settings ranging from 10W-50W. In each trial, the feedback device stopped the applied current to the tissue samples when the measured impedance exceeded the cut-off threshold of a 25% increase from the starting impedance. The device was found to have an accuracy of ±5Ω. This successfully minimized undue tissue damage and proved able to prevent serious complications such as perforation from occurring."]},{"key":"dc:title","label":"Title","values":["Development of an Impedance-Controlled Hot Snare Polypectomy Device to Minimize Tissue Damage"]}]}],"canonical_facts":{"dc:contributor":["Biswanath Samanta","Minchul Shin"],"dc:creator":["Thornton, CurtisLee"],"dc:date.available":["2019-06-28T07:00:00Z"],"dc:date.created":["2019-01-01T08:00:00Z"],"dc:description.abstract":["This study explores the ability to measure the changing impedance, ex vivo, of a porcine colon sample while undergoing a high-frequency alternating current from an Olympus PSD-30 electrosurgical unit and stop the applied current before excessive tissue damage ensues. The causes of the thermal damage are first examined, followed by the construction and testing of the impedance-controlled feedback device. Perforation was observed to occur when the impedance of the tissue sample increased by 25% or more. Using this information, the device was tested for five power settings ranging from 10W-50W. In each trial, the feedback device stopped the applied current to the tissue samples when the measured impedance exceeded the cut-off threshold of a 25% increase from the starting impedance. The device was found to have an accuracy of ±5Ω. This successfully minimized undue tissue damage and proved able to prevent serious complications such as perforation from occurring."],"dc:identifier":["https://digitalcommons.georgiasouthern.edu/etd/1970"],"dc:rights.license":["http://creativecommons.org/licenses/by/4.0/"],"dc:subject":["Polypectomy","Snare","Impedance","Feedback control","Radiofrequency ablation","Biomedical Devices and Instrumentation"],"dc:title":["Development of an Impedance-Controlled Hot Snare Polypectomy Device to Minimize Tissue Damage"],"thesis:degree_discipline":["Department of Mechanical Engineering"],"thesis:degree_level":["Thesis (open access)"],"thesis:degree_name":["Master of Science, Mechanical Engineering"]},"updated_at":"2026-11-19T07:54:56Z"}