{"id":{"repo_id":"guelph","oai_identifier":"oai:atrium.lib.uoguelph.ca:10214/28352"},"canonical_url":"https://search.ndltd.org/etd/guelph/oai:atrium.lib.uoguelph.ca:10214/28352","repository":{"repo_id":"guelph","name":"University of Guelph","base_url":"https://atrium.lib.uoguelph.ca/server/oai/request"},"display":{"title":"The Effect of Locking Head Inserts on the Biomechanical Properties of a 3.5 mm Broad Locking Compression Plate when used in an Open Fracture Gap Model","abstract":"Locking compression plates (LCP) have gained increasing popularity in veterinary and human medicine for their versatility, combining the functions of both locking and conventional plating systems. LCP are frequently used in minimally invasive plate osteosynthesis (MIPO). MIPO aims to preserve fracture biology and is the standard-of-care for fractures in which rigid anatomic reconstruction is not possible. MIPO increases the biomechanical demands on implants through reduced load-sharing between implant and bone. Empty screw holes act as a stress riser and can predispose LCP to premature failure. Locking head inserts (LHI) have been designed to fill empty screw holes in an attempt to improve plate longevity. The present study examined the effect of LHI in an open fracture gap model. LHI were applied to a 3.5 mm broad LCP in three different configurations, (1) no LHI, (2) three central LHI over the fracture gap, and (3) complete filling of empty holes with LHI. LHI were applied with 4 Nm of torque. LCP were attached to synthetic bone models and cyclically loaded with three-degrees of freedom. Surface strain was collected during testing using uniaxial strain gauges. Construct stiffness was calculated from load and deformation data collected by the biomechanical testing apparatus. No significant difference in implant peak-to-peak strain, or construct stiffness and peak-to-peak deformation were detected with the addition of LHI. Results from this study indicate that the addition of LHI did not significantly impact the biomechanical properties of a 3.5 mm broad LCP. In clinical high-load fracture fixation scenarios, alternative methods of reducing implant fatigue should be considered.","abstract_html":"Locking compression plates (LCP) have gained increasing popularity in veterinary and human medicine for their versatility, combining the functions of both locking and conventional plating systems. LCP are frequently used in minimally invasive plate osteosynthesis (MIPO). MIPO aims to preserve fracture biology and is the standard-of-care for fractures in which rigid anatomic reconstruction is not possible. MIPO increases the biomechanical demands on implants through reduced load-sharing between implant and bone. Empty screw holes act as a stress riser and can predispose LCP to premature failure. Locking head inserts (LHI) have been designed to fill empty screw holes in an attempt to improve plate longevity. The present study examined the effect of LHI in an open fracture gap model. LHI were applied to a 3.5 mm broad LCP in three different configurations, (1) no LHI, (2) three central LHI over the fracture gap, and (3) complete filling of empty holes with LHI. LHI were applied with 4 Nm of torque. LCP were attached to synthetic bone models and cyclically loaded with three-degrees of freedom. Surface strain was collected during testing using uniaxial strain gauges. Construct stiffness was calculated from load and deformation data collected by the biomechanical testing apparatus. No significant difference in implant peak-to-peak strain, or construct stiffness and peak-to-peak deformation were detected with the addition of LHI. Results from this study indicate that the addition of LHI did not significantly impact the biomechanical properties of a 3.5 mm broad LCP. In clinical high-load fracture fixation scenarios, alternative methods of reducing implant fatigue should be considered.","abstract_has_math":false,"creators":["Hawker, William"],"institution":"University of Guelph","degree_name":"Doctor of Veterinary Science","degree_level":null,"degree_discipline":null,"degree_department":"Department of Clinical Studies","school":null,"contributors":[],"advisors":["Moens, Noel"],"committee_chairs":[],"committee_members":[],"year":null,"date_issued":"","date_published":null,"updated_at":"2026-11-19T08:26:20Z","subjects":["biomechanics","orthopedic","locking","plate","inserts"],"languages":["en"],"rights":["Attribution-ShareAlike 4.0 International"],"rights_urls":["http://creativecommons.org/licenses/by-sa/4.0/"],"identifier_entries":[],"person_identifier_entries":[]},"links":{"outbound_url":"https://hdl.handle.net/10214/28352","outbound_label":"Handle","outbound_source":"dc:identifier.uri","record_url":"https://hdl.handle.net/10214/28352","record_label":"Repository record","record_kind":"handle","record_source":"dc:identifier.uri","download_url":null,"download_label":null,"download_source":null,"access_status":null,"embargo_until":null,"access_note":null},"source_record":{"url":"https://atrium.lib.uoguelph.ca/server/oai/request?verb=GetRecord&metadataPrefix=dim&identifier=oai%3Aatrium.lib.uoguelph.ca%3A10214%2F28352","prefix":"dim"},"metadata_groups":[{"id":"people","label":"People","entries":[{"key":"dc:contributor.advisor","label":"Advisor","values":["Moens, Noel"]},{"key":"dc:creator","label":"Author","values":["Hawker, William"]}]},{"id":"academic_context","label":"Academic Context","entries":[{"key":"dc:date.accessioned","label":"Dc Date Accessioned","values":["2024-05-17T13:11:49Z"]},{"key":"dc:date.available","label":"Dc Date Available","values":["2024-05-17T13:11:49Z"]},{"key":"dc:date.copyright","label":"Dc Date Copyright","values":["2024-05"]},{"key":"dc:date.created","label":"Dc Date Created","values":["2024-05-06"]},{"key":"dc:publisher","label":"Institution","values":["University of Guelph"]},{"key":"dc:type","label":"Dc Type","values":["Thesis"]},{"key":"thesis:degree_department","label":"Department","values":["Department of Clinical Studies"]},{"key":"thesis:degree_name","label":"Degree Name","values":["Doctor of Veterinary Science"]},{"key":"thesis:degree_program","label":"Thesis Degree Program","values":["Clinical Studies"]},{"key":"thesis:institution_name","label":"Thesis Institution Name","values":["University of Guelph"]}]},{"id":"subjects_keywords","label":"Subjects and Keywords","entries":[{"key":"dc:subject","label":"Dc Subject","values":["biomechanics","orthopedic","locking","plate","inserts"]}]},{"id":"language_rights","label":"Language and Rights","entries":[{"key":"dc:language.iso","label":"Language (ISO)","values":["en"]},{"key":"dc:rights","label":"Dc Rights","values":["Attribution-ShareAlike 4.0 International"]},{"key":"dc:rights.uri","label":"Rights URI","values":["http://creativecommons.org/licenses/by-sa/4.0/"]}]},{"id":"identifiers","label":"Identifiers","entries":[{"key":"dc:identifier.uri","label":"Identifier URI","values":["https://hdl.handle.net/10214/28352"]}]},{"id":"additional","label":"Additional Metadata","entries":[{"key":"dc:description.abstract","label":"Abstract","values":["Locking compression plates (LCP) have gained increasing popularity in veterinary and human medicine for their versatility, combining the functions of both locking and conventional plating systems. LCP are frequently used in minimally invasive plate osteosynthesis (MIPO). MIPO aims to preserve fracture biology and is the standard-of-care for fractures in which rigid anatomic reconstruction is not possible. MIPO increases the biomechanical demands on implants through reduced load-sharing between implant and bone. Empty screw holes act as a stress riser and can predispose LCP to premature failure. Locking head inserts (LHI) have been designed to fill empty screw holes in an attempt to improve plate longevity. The present study examined the effect of LHI in an open fracture gap model. LHI were applied to a 3.5 mm broad LCP in three different configurations, (1) no LHI, (2) three central LHI over the fracture gap, and (3) complete filling of empty holes with LHI. LHI were applied with 4 Nm of torque. LCP were attached to synthetic bone models and cyclically loaded with three-degrees of freedom. Surface strain was collected during testing using uniaxial strain gauges. Construct stiffness was calculated from load and deformation data collected by the biomechanical testing apparatus. No significant difference in implant peak-to-peak strain, or construct stiffness and peak-to-peak deformation were detected with the addition of LHI. Results from this study indicate that the addition of LHI did not significantly impact the biomechanical properties of a 3.5 mm broad LCP. In clinical high-load fracture fixation scenarios, alternative methods of reducing implant fatigue should be considered."]},{"key":"dc:title","label":"Title","values":["The Effect of Locking Head Inserts on the Biomechanical Properties of a 3.5 mm Broad Locking Compression Plate when used in an Open Fracture Gap Model"]},{"key":"others:access_status","label":"Others Access Status","values":["open.access"]}]}],"canonical_facts":{"dc:contributor.advisor":["Moens, Noel"],"dc:creator":["Hawker, William"],"dc:date.accessioned":["2024-05-17T13:11:49Z"],"dc:date.available":["2024-05-17T13:11:49Z"],"dc:date.copyright":["2024-05"],"dc:date.created":["2024-05-06"],"dc:description.abstract":["Locking compression plates (LCP) have gained increasing popularity in veterinary and human medicine for their versatility, combining the functions of both locking and conventional plating systems. LCP are frequently used in minimally invasive plate osteosynthesis (MIPO). MIPO aims to preserve fracture biology and is the standard-of-care for fractures in which rigid anatomic reconstruction is not possible. MIPO increases the biomechanical demands on implants through reduced load-sharing between implant and bone. Empty screw holes act as a stress riser and can predispose LCP to premature failure. Locking head inserts (LHI) have been designed to fill empty screw holes in an attempt to improve plate longevity. The present study examined the effect of LHI in an open fracture gap model. LHI were applied to a 3.5 mm broad LCP in three different configurations, (1) no LHI, (2) three central LHI over the fracture gap, and (3) complete filling of empty holes with LHI. LHI were applied with 4 Nm of torque. LCP were attached to synthetic bone models and cyclically loaded with three-degrees of freedom. Surface strain was collected during testing using uniaxial strain gauges. Construct stiffness was calculated from load and deformation data collected by the biomechanical testing apparatus. No significant difference in implant peak-to-peak strain, or construct stiffness and peak-to-peak deformation were detected with the addition of LHI. Results from this study indicate that the addition of LHI did not significantly impact the biomechanical properties of a 3.5 mm broad LCP. In clinical high-load fracture fixation scenarios, alternative methods of reducing implant fatigue should be considered."],"dc:identifier.uri":["https://hdl.handle.net/10214/28352"],"dc:language.iso":["en"],"dc:publisher":["University of Guelph"],"dc:rights":["Attribution-ShareAlike 4.0 International"],"dc:rights.uri":["http://creativecommons.org/licenses/by-sa/4.0/"],"dc:subject":["biomechanics","orthopedic","locking","plate","inserts"],"dc:title":["The Effect of Locking Head Inserts on the Biomechanical Properties of a 3.5 mm Broad Locking Compression Plate when used in an Open Fracture Gap Model"],"dc:type":["Thesis"],"others:access_status":["open.access"],"thesis:degree_department":["Department of Clinical Studies"],"thesis:degree_name":["Doctor of Veterinary Science"],"thesis:degree_program":["Clinical Studies"],"thesis:institution_name":["University of Guelph"]},"updated_at":"2026-11-19T08:26:20Z"}