{"id":{"repo_id":"guelph","oai_identifier":"oai:atrium.lib.uoguelph.ca:10214/30035"},"canonical_url":"https://search.ndltd.org/etd/guelph/oai:atrium.lib.uoguelph.ca:10214/30035","repository":{"repo_id":"guelph","name":"University of Guelph","base_url":"https://atrium.lib.uoguelph.ca/server/oai/request"},"display":{"title":"Pop Goes the Needle: Measurement of Needle-Tissue Force Interactions During Canine Epidural Needle Insertions and Subsequent Design of a Teaching Simulator","abstract":"Canine epidural needle insertion is a technical procedure used to deliver drugs such as local anesthetics, analgesics, and anti-inflammatory agents to the epidural space. During insertion, the needle penetrates the skin, muscle, and ligamentum flavum, which “pops” upon penetration indicating entry into the epidural space. Achieving competency may require 60–90 insertions; however, training is often performed on live patients, limiting practice opportunities and raising ethical concerns. The objective of this thesis was to design and validate a novel, haptically accurate, canine epidural training simulator. A handheld device (EpiDyno) was developed to measure force–displacement data during needle insertion using a low-profile load cell and proximity sensor. This device was validated against a mechanical testing system (Instron) on a range of natural and synthetic materials at expected forces (0-11 N). The load cell was validated for four key points of interest including the maximum force (average error = 0.04 N) for each tissue, and the local minima (average error = 0.04 N) post-skin peak force. After validating the proximity sensor’s ability to measure insertion speed for flat matte (average error = 0.26 mm/s) and curved matte (average error = 0.13 mm/s) surfaces, the EpiDyno was used to collect data from six fresh, canine cadavers. Mean (±SD) peak forces were quantified for skin (4.1 ± 0.8 N), muscle (3.9 ± 1.8 N), and ligamentum flavum (6.2 ± 1.7 N), along with the post-skin puncture local minima (1.8 ± 0.8 N). The average insertion speed across all trials was 7.9 ± 3.4 mm/s (N=1 clinician). Artificial simulator materials were investigated to replicate tissue thickness and puncture forces, resulting in a composite structure of vinyl leather, foam, and 100% polyurethane that accurately (±0.6 N) reproduces the haptic profile of canine epidural needle insertion. Canine L6–S3 lumbar vertebrae were scanned, modified for articulation, 3D-printed, and integrated with the composite at the injection site within the canine simulator. A pull-up circuit using an LED was incorporated to confirm entry into the epidural space. This simulator demonstrates the potential to improve medical skills training by providing a repeatable, ethical learning environment for skill acquisition.","abstract_html":"Canine epidural needle insertion is a technical procedure used to deliver drugs such as local anesthetics, analgesics, and anti-inflammatory agents to the epidural space. During insertion, the needle penetrates the skin, muscle, and ligamentum flavum, which “pops” upon penetration indicating entry into the epidural space. Achieving competency may require 60–90 insertions; however, training is often performed on live patients, limiting practice opportunities and raising ethical concerns. The objective of this thesis was to design and validate a novel, haptically accurate, canine epidural training simulator. A handheld device (EpiDyno) was developed to measure force–displacement data during needle insertion using a low-profile load cell and proximity sensor. This device was validated against a mechanical testing system (Instron) on a range of natural and synthetic materials at expected forces (0-11 N). The load cell was validated for four key points of interest including the maximum force (average error = 0.04 N) for each tissue, and the local minima (average error = 0.04 N) post-skin peak force. After validating the proximity sensor’s ability to measure insertion speed for flat matte (average error = 0.26 mm/s) and curved matte (average error = 0.13 mm/s) surfaces, the EpiDyno was used to collect data from six fresh, canine cadavers. Mean (±SD) peak forces were quantified for skin (4.1 ± 0.8 N), muscle (3.9 ± 1.8 N), and ligamentum flavum (6.2 ± 1.7 N), along with the post-skin puncture local minima (1.8 ± 0.8 N). The average insertion speed across all trials was 7.9 ± 3.4 mm/s (N=1 clinician). Artificial simulator materials were investigated to replicate tissue thickness and puncture forces, resulting in a composite structure of vinyl leather, foam, and 100% polyurethane that accurately (±0.6 N) reproduces the haptic profile of canine epidural needle insertion. Canine L6–S3 lumbar vertebrae were scanned, modified for articulation, 3D-printed, and integrated with the composite at the injection site within the canine simulator. A pull-up circuit using an LED was incorporated to confirm entry into the epidural space. This simulator demonstrates the potential to improve medical skills training by providing a repeatable, ethical learning environment for skill acquisition.","abstract_has_math":false,"creators":["Smith, Claudia do Ceu Vieira Sousadias de"],"institution":"University of Guelph","degree_name":"Doctor of Philosophy","degree_level":null,"degree_discipline":null,"degree_department":"Department of Interdisciplinary Engineering","school":null,"contributors":[],"advisors":["Brandon, Scott"],"committee_chairs":[],"committee_members":[],"year":null,"date_issued":"","date_published":null,"updated_at":"2026-11-19T08:25:52Z","subjects":["canine","epidural","simulator","force","haptic feedback","tissue mechanics","needle-tissue interactions","force-displacement"],"languages":["en"],"rights":["Attribution-NonCommercial-ShareAlike 4.0 International"],"rights_urls":["http://creativecommons.org/licenses/by-nc-sa/4.0/"],"identifier_entries":[],"person_identifier_entries":[]},"links":{"outbound_url":"https://hdl.handle.net/10214/30035","outbound_label":"Handle","outbound_source":"dc:identifier.uri","record_url":"https://hdl.handle.net/10214/30035","record_label":"Repository record","record_kind":"handle","record_source":"dc:identifier.uri","access_status":null,"embargo_until":null,"access_note":null},"oai_pmh_record":{"url":"https://union.ndltd.org/oai-pmh/?verb=GetRecord&metadataPrefix=oai_dc&identifier=oai%3Aunion.ndltd.org%3Arepo%3Aguelph%3Aoai%3Aatrium.lib.uoguelph.ca%3A10214%2F30035","prefix":"oai_dc"},"source_record":{"url":"https://atrium.lib.uoguelph.ca/server/oai/request?verb=GetRecord&metadataPrefix=dim&identifier=oai%3Aatrium.lib.uoguelph.ca%3A10214%2F30035","prefix":"dim"},"metadata_groups":[{"id":"people","label":"People","entries":[{"key":"dc:contributor.advisor","label":"Advisor","values":["Brandon, Scott"]},{"key":"dc:creator","label":"Author","values":["Smith, Claudia do Ceu Vieira Sousadias de"]}]},{"id":"academic_context","label":"Academic Context","entries":[{"key":"dc:date.accessioned","label":"Dc Date Accessioned","values":["2026-09-10T13:47:32Z"]},{"key":"dc:date.copyright","label":"Dc Date Copyright","values":["2026-09"]},{"key":"dc:publisher","label":"Institution","values":["University of Guelph"]},{"key":"dc:type","label":"Dc Type","values":["Thesis"]},{"key":"thesis:degree_department","label":"Department","values":["Department of Interdisciplinary Engineering"]},{"key":"thesis:degree_name","label":"Degree Name","values":["Doctor of Philosophy"]},{"key":"thesis:degree_program","label":"Thesis Degree Program","values":["Engineering"]},{"key":"thesis:institution_name","label":"Thesis Institution Name","values":["University of Guelph"]}]},{"id":"subjects_keywords","label":"Subjects and Keywords","entries":[{"key":"dc:subject","label":"Dc Subject","values":["canine","epidural","simulator","force","haptic feedback","tissue mechanics","needle-tissue interactions","force-displacement"]}]},{"id":"language_rights","label":"Language and Rights","entries":[{"key":"dc:language.iso","label":"Language (ISO)","values":["en"]},{"key":"dc:rights","label":"Dc Rights","values":["Attribution-NonCommercial-ShareAlike 4.0 International"]},{"key":"dc:rights.uri","label":"Rights URI","values":["http://creativecommons.org/licenses/by-nc-sa/4.0/"]}]},{"id":"identifiers","label":"Identifiers","entries":[{"key":"dc:identifier.uri","label":"Identifier URI","values":["https://hdl.handle.net/10214/30035"]}]},{"id":"additional","label":"Additional Metadata","entries":[{"key":"dc:description.abstract","label":"Abstract","values":["Canine epidural needle insertion is a technical procedure used to deliver drugs such as local anesthetics, analgesics, and anti-inflammatory agents to the epidural space. During insertion, the needle penetrates the skin, muscle, and ligamentum flavum, which “pops” upon penetration indicating entry into the epidural space. Achieving competency may require 60–90 insertions; however, training is often performed on live patients, limiting practice opportunities and raising ethical concerns. The objective of this thesis was to design and validate a novel, haptically accurate, canine epidural training simulator. A handheld device (EpiDyno) was developed to measure force–displacement data during needle insertion using a low-profile load cell and proximity sensor. This device was validated against a mechanical testing system (Instron) on a range of natural and synthetic materials at expected forces (0-11 N). The load cell was validated for four key points of interest including the maximum force (average error = 0.04 N) for each tissue, and the local minima (average error = 0.04 N) post-skin peak force. After validating the proximity sensor’s ability to measure insertion speed for flat matte (average error = 0.26 mm/s) and curved matte (average error = 0.13 mm/s) surfaces, the EpiDyno was used to collect data from six fresh, canine cadavers. Mean (±SD) peak forces were quantified for skin (4.1 ± 0.8 N), muscle (3.9 ± 1.8 N), and ligamentum flavum (6.2 ± 1.7 N), along with the post-skin puncture local minima (1.8 ± 0.8 N). The average insertion speed across all trials was 7.9 ± 3.4 mm/s (N=1 clinician). Artificial simulator materials were investigated to replicate tissue thickness and puncture forces, resulting in a composite structure of vinyl leather, foam, and 100% polyurethane that accurately (±0.6 N) reproduces the haptic profile of canine epidural needle insertion. Canine L6–S3 lumbar vertebrae were scanned, modified for articulation, 3D-printed, and integrated with the composite at the injection site within the canine simulator. A pull-up circuit using an LED was incorporated to confirm entry into the epidural space. This simulator demonstrates the potential to improve medical skills training by providing a repeatable, ethical learning environment for skill acquisition."]},{"key":"dc:title","label":"Title","values":["Pop Goes the Needle: Measurement of Needle-Tissue Force Interactions During Canine Epidural Needle Insertions and Subsequent Design of a Teaching Simulator"]},{"key":"others:access_status","label":"Others Access Status","values":["open.access"]}]}],"canonical_facts":{"dc:contributor.advisor":["Brandon, Scott"],"dc:creator":["Smith, Claudia do Ceu Vieira Sousadias de"],"dc:date.accessioned":["2026-09-10T13:47:32Z"],"dc:date.copyright":["2026-09"],"dc:description.abstract":["Canine epidural needle insertion is a technical procedure used to deliver drugs such as local anesthetics, analgesics, and anti-inflammatory agents to the epidural space. During insertion, the needle penetrates the skin, muscle, and ligamentum flavum, which “pops” upon penetration indicating entry into the epidural space. Achieving competency may require 60–90 insertions; however, training is often performed on live patients, limiting practice opportunities and raising ethical concerns. The objective of this thesis was to design and validate a novel, haptically accurate, canine epidural training simulator. A handheld device (EpiDyno) was developed to measure force–displacement data during needle insertion using a low-profile load cell and proximity sensor. This device was validated against a mechanical testing system (Instron) on a range of natural and synthetic materials at expected forces (0-11 N). The load cell was validated for four key points of interest including the maximum force (average error = 0.04 N) for each tissue, and the local minima (average error = 0.04 N) post-skin peak force. After validating the proximity sensor’s ability to measure insertion speed for flat matte (average error = 0.26 mm/s) and curved matte (average error = 0.13 mm/s) surfaces, the EpiDyno was used to collect data from six fresh, canine cadavers. Mean (±SD) peak forces were quantified for skin (4.1 ± 0.8 N), muscle (3.9 ± 1.8 N), and ligamentum flavum (6.2 ± 1.7 N), along with the post-skin puncture local minima (1.8 ± 0.8 N). The average insertion speed across all trials was 7.9 ± 3.4 mm/s (N=1 clinician). Artificial simulator materials were investigated to replicate tissue thickness and puncture forces, resulting in a composite structure of vinyl leather, foam, and 100% polyurethane that accurately (±0.6 N) reproduces the haptic profile of canine epidural needle insertion. Canine L6–S3 lumbar vertebrae were scanned, modified for articulation, 3D-printed, and integrated with the composite at the injection site within the canine simulator. A pull-up circuit using an LED was incorporated to confirm entry into the epidural space. This simulator demonstrates the potential to improve medical skills training by providing a repeatable, ethical learning environment for skill acquisition."],"dc:identifier.uri":["https://hdl.handle.net/10214/30035"],"dc:language.iso":["en"],"dc:publisher":["University of Guelph"],"dc:rights":["Attribution-NonCommercial-ShareAlike 4.0 International"],"dc:rights.uri":["http://creativecommons.org/licenses/by-nc-sa/4.0/"],"dc:subject":["canine","epidural","simulator","force","haptic feedback","tissue mechanics","needle-tissue interactions","force-displacement"],"dc:title":["Pop Goes the Needle: Measurement of Needle-Tissue Force Interactions During Canine Epidural Needle Insertions and Subsequent Design of a Teaching Simulator"],"dc:type":["Thesis"],"others:access_status":["open.access"],"thesis:degree_department":["Department of Interdisciplinary Engineering"],"thesis:degree_name":["Doctor of Philosophy"],"thesis:degree_program":["Engineering"],"thesis:institution_name":["University of Guelph"]},"updated_at":"2026-11-19T08:25:52Z"}