{"id":{"repo_id":"manitoba","oai_identifier":"oai:mspace.lib.umanitoba.ca:1993/31980"},"canonical_url":"https://search.ndltd.org/etd/manitoba/oai:mspace.lib.umanitoba.ca:1993/31980","repository":{"repo_id":"manitoba","name":"University of Manitoba","base_url":"https://mspace.lib.umanitoba.ca/oai/request"},"display":{"title":"The characterization of 3D printed plastics sterilized by hydrogen peroxide vapour","abstract":"3D printers that precisely fuse plastic filament are enabling medical manufacturers to produce high-quality plastic medical devices and implants. However, the low-temperature fusing process implies that post-production sterilization must also occur at a low temperature or destroy the precision of the product. This study characterizes the effects of hydrogen peroxide (H2O2) vapour sterilization on ASTM-compliant tensile samples of polylactic acid, polycaprolactone, and polycarbonate. The sterilization process caused physical deformations in polycaprolactone. Additionally, increases were observed in polycaprolactone and polycarbonate sample thickness, and in polycarbonate sample width. Decreases in E were found in all three materials, while UTS decreased in polycarbonate, and strain at UTS increased in polycaprolactone. The findings demonstrate that the materials can be compatible with H2O2 vapour sterilization, but products must be designed to accommodate for changes that occur due to sterilization.","abstract_html":"3D printers that precisely fuse plastic filament are enabling medical manufacturers to produce high-quality plastic medical devices and implants. However, the low-temperature fusing process implies that post-production sterilization must also occur at a low temperature or destroy the precision of the product. This study characterizes the effects of hydrogen peroxide (H2O2) vapour sterilization on ASTM-compliant tensile samples of polylactic acid, polycaprolactone, and polycarbonate. The sterilization process caused physical deformations in polycaprolactone. Additionally, increases were observed in polycaprolactone and polycarbonate sample thickness, and in polycarbonate sample width. Decreases in E were found in all three materials, while UTS decreased in polycarbonate, and strain at UTS increased in polycaprolactone. The findings demonstrate that the materials can be compatible with H2O2 vapour sterilization, but products must be designed to accommodate for changes that occur due to sterilization.","abstract_has_math":false,"creators":["Sosnowski, Emil-Peter"],"institution":null,"degree_name":null,"degree_level":"Master of Science (M.Sc.)","degree_discipline":"Biosystems Engineering","degree_department":null,"school":null,"contributors":[],"advisors":["Morrison, Jason (Biosystems Engineering)"],"committee_chairs":[],"committee_members":["Hewko, Mark (Biosystems Engineering)","Herbert, David (Chemistry)"],"year":2016,"date_issued":"2016","date_published":"2016","updated_at":"2026-11-19T09:08:41Z","subjects":["3D printing","Hydrogen Peroxide Sterilization","Polylactic Acid","Polycaprolactone","Polycarbonate","Mechanical properties"],"languages":["eng"],"rights":[],"rights_urls":[],"identifier_entries":[],"person_identifier_entries":[]},"links":{"outbound_url":"https://hdl.handle.net/1993/31980","outbound_label":"Handle","outbound_source":"dc:identifier.uri","record_url":"https://hdl.handle.net/1993/31980","record_label":"Repository record","record_kind":"handle","record_source":"dc:identifier.uri","access_status":null,"embargo_until":null,"access_note":null},"oai_pmh_record":{"url":"https://union.ndltd.org/oai-pmh/?verb=GetRecord&metadataPrefix=oai_dc&identifier=oai%3Aunion.ndltd.org%3Arepo%3Amanitoba%3Aoai%3Amspace.lib.umanitoba.ca%3A1993%2F31980","prefix":"oai_dc"},"source_record":{"url":"https://mspace.lib.umanitoba.ca/oai/request?verb=GetRecord&metadataPrefix=dim&identifier=oai%3Amspace.lib.umanitoba.ca%3A1993%2F31980","prefix":"dim"},"metadata_groups":[{"id":"people","label":"People","entries":[{"key":"dc:contributor.committeemember","label":"Committee Member","values":["Hewko, Mark (Biosystems Engineering)","Herbert, David (Chemistry)"]},{"key":"dc:contributor.supervisor","label":"Supervisor","values":["Morrison, Jason (Biosystems Engineering)"]},{"key":"dc:creator","label":"Author","values":["Sosnowski, Emil-Peter"]}]},{"id":"academic_context","label":"Academic Context","entries":[{"key":"dc:date.accessioned","label":"Dc Date Accessioned","values":["2017-01-05T16:08:18Z"]},{"key":"dc:date.available","label":"Dc Date Available","values":["2017-01-05T16:08:18Z"]},{"key":"dc:date.issued","label":"Date","values":["2016"]},{"key":"dc:type","label":"Dc Type","values":["master thesis"]},{"key":"thesis:degree_discipline","label":"Discipline","values":["Biosystems Engineering"]},{"key":"thesis:degree_level","label":"Degree Level","values":["Master of Science (M.Sc.)"]}]},{"id":"subjects_keywords","label":"Subjects and Keywords","entries":[{"key":"dc:subject","label":"Dc Subject","values":["3D printing","Hydrogen Peroxide Sterilization","Polylactic Acid","Polycaprolactone","Polycarbonate","Mechanical properties"]}]},{"id":"language_rights","label":"Language and Rights","entries":[{"key":"dc:language.iso","label":"Language (ISO)","values":["eng"]}]},{"id":"identifiers","label":"Identifiers","entries":[{"key":"dc:identifier.uri","label":"Identifier URI","values":["http://hdl.handle.net/1993/31980"]}]},{"id":"additional","label":"Additional Metadata","entries":[{"key":"dc:description.abstract","label":"Abstract","values":["3D printers that precisely fuse plastic filament are enabling medical manufacturers to produce high-quality plastic medical devices and implants. However, the low-temperature fusing process implies that post-production sterilization must also occur at a low temperature or destroy the precision of the product. This study characterizes the effects of hydrogen peroxide (H2O2) vapour sterilization on ASTM-compliant tensile samples of polylactic acid, polycaprolactone, and polycarbonate. The sterilization process caused physical deformations in polycaprolactone. Additionally, increases were observed in polycaprolactone and polycarbonate sample thickness, and in polycarbonate sample width. Decreases in E were found in all three materials, while UTS decreased in polycarbonate, and strain at UTS increased in polycaprolactone. The findings demonstrate that the materials can be compatible with H2O2 vapour sterilization, but products must be designed to accommodate for changes that occur due to sterilization."]},{"key":"dc:description.note","label":"Dc Description Note","values":["February 2017"]},{"key":"dc:title","label":"Title","values":["The characterization of 3D printed plastics sterilized by hydrogen peroxide vapour"]},{"key":"others:access_status","label":"Others Access Status","values":["open.access"]}]}],"canonical_facts":{"dc:contributor.committeemember":["Hewko, Mark (Biosystems Engineering)","Herbert, David (Chemistry)"],"dc:contributor.supervisor":["Morrison, Jason (Biosystems Engineering)"],"dc:creator":["Sosnowski, Emil-Peter"],"dc:date.accessioned":["2017-01-05T16:08:18Z"],"dc:date.available":["2017-01-05T16:08:18Z"],"dc:date.issued":["2016"],"dc:description.abstract":["3D printers that precisely fuse plastic filament are enabling medical manufacturers to produce high-quality plastic medical devices and implants. However, the low-temperature fusing process implies that post-production sterilization must also occur at a low temperature or destroy the precision of the product. This study characterizes the effects of hydrogen peroxide (H2O2) vapour sterilization on ASTM-compliant tensile samples of polylactic acid, polycaprolactone, and polycarbonate. The sterilization process caused physical deformations in polycaprolactone. Additionally, increases were observed in polycaprolactone and polycarbonate sample thickness, and in polycarbonate sample width. Decreases in E were found in all three materials, while UTS decreased in polycarbonate, and strain at UTS increased in polycaprolactone. The findings demonstrate that the materials can be compatible with H2O2 vapour sterilization, but products must be designed to accommodate for changes that occur due to sterilization."],"dc:description.note":["February 2017"],"dc:identifier.uri":["http://hdl.handle.net/1993/31980"],"dc:language.iso":["eng"],"dc:subject":["3D printing","Hydrogen Peroxide Sterilization","Polylactic Acid","Polycaprolactone","Polycarbonate","Mechanical properties"],"dc:title":["The characterization of 3D printed plastics sterilized by hydrogen peroxide vapour"],"dc:type":["master thesis"],"others:access_status":["open.access"],"thesis:degree_discipline":["Biosystems Engineering"],"thesis:degree_level":["Master of Science (M.Sc.)"]},"updated_at":"2026-11-19T09:08:41Z"}