{"id":{"repo_id":"mit","oai_identifier":"oai:dspace.mit.edu:1721.1/40444"},"canonical_url":"https://search.ndltd.org/etd/mit/oai:dspace.mit.edu:1721.1/40444","repository":{"repo_id":"mit","name":"MIT","base_url":"https://dspace.mit.edu/oai/request"},"display":{"title":"Design and modeling of an exhaust gas waste heat autoclave","abstract":"In order to provide proper sterilization and cleaning of medical equipment for field hospitals and third-world countries while also decreasing the reliance on electricity of traditional sterilization methods, a new steam sterilizer/autoclave system was designed and modeled. This system uses waste engine heat from the exhaust system of a diesel generator set to boil water and produce the pressurized steam conditions necessary for effective medical sterilization. Currently, the design utilizes a 0.59 meter, concentric tube cross-flow heat exchanger and high-temperature heat transfer fluid to draw thermal energy from the exhaust pipe and deposit it into the autoclave pressure vessel to create steam. The system is designed to run a 35-minute sterilization cycle, requiring 15 minutes to produce saturated steam at 2 atmospheres within a 50- liter autoclave, and 20 minutes to sterilize medical instruments in the steam environment. Furthermore, the system uses basic, off-the-shelf fluid transfer materials to provide a robust, effective system that can be easily maintained in the field without need for specialized parts or technicians.","abstract_html":"In order to provide proper sterilization and cleaning of medical equipment for field hospitals and third-world countries while also decreasing the reliance on electricity of traditional sterilization methods, a new steam sterilizer/autoclave system was designed and modeled. This system uses waste engine heat from the exhaust system of a diesel generator set to boil water and produce the pressurized steam conditions necessary for effective medical sterilization. Currently, the design utilizes a 0.59 meter, concentric tube cross-flow heat exchanger and high-temperature heat transfer fluid to draw thermal energy from the exhaust pipe and deposit it into the autoclave pressure vessel to create steam. The system is designed to run a 35-minute sterilization cycle, requiring 15 minutes to produce saturated steam at 2 atmospheres within a 50- liter autoclave, and 20 minutes to sterilize medical instruments in the steam environment. Furthermore, the system uses basic, off-the-shelf fluid transfer materials to provide a robust, effective system that can be easily maintained in the field without need for specialized parts or technicians.","abstract_has_math":false,"creators":["Jiricek, Joshua A"],"institution":"Massachusetts Institute of Technology","degree_name":null,"degree_level":null,"degree_discipline":null,"degree_department":null,"school":null,"contributors":[],"advisors":["Todd Thorsen."],"committee_chairs":[],"committee_members":[],"year":2007,"date_issued":"2007","date_published":"2007","updated_at":"2026-11-27T05:21:22Z","subjects":["Mechanical Engineering."],"languages":["en"],"rights":["M.I.T. theses are protected by copyright. They may be viewed from this source for any purpose, but reproduction or distribution in any format is prohibited without written permission. See provided URL for inquiries about permission."],"rights_urls":["http://dspace.mit.edu/handle/1721.1/7582"],"identifier_entries":[{"key":"dc:identifier.oclc","label":"OCLC","values":["191730700"],"render_values":[{"text":"191730700","href":null,"code":true}]}],"person_identifier_entries":[]},"links":{"outbound_url":"https://hdl.handle.net/1721.1/40444","outbound_label":"Handle","outbound_source":"dc:identifier.uri","record_url":"https://hdl.handle.net/1721.1/40444","record_label":"Repository record","record_kind":"handle","record_source":"dc:identifier.uri","access_status":null,"embargo_until":null,"access_note":null},"oai_pmh_record":{"url":"https://union.ndltd.org/oai-pmh/?verb=GetRecord&metadataPrefix=oai_dc&identifier=oai%3Aunion.ndltd.org%3Arepo%3Amit%3Aoai%3Adspace.mit.edu%3A1721.1%2F40444","prefix":"oai_dc"},"source_record":{"url":"https://dspace.mit.edu/oai/request?verb=GetRecord&metadataPrefix=dim&identifier=oai%3Adspace.mit.edu%3A1721.1%2F40444","prefix":"dim"},"metadata_groups":[{"id":"people","label":"People","entries":[{"key":"dc:contributor.advisor","label":"Advisor","values":["Todd Thorsen."]},{"key":"dc:contributor.other","label":"Dc Contributor Other","values":["Massachusetts Institute of Technology. Dept. of Mechanical Engineering."]},{"key":"dc:creator","label":"Author","values":["Jiricek, Joshua A"]}]},{"id":"academic_context","label":"Academic Context","entries":[{"key":"dc:date.accessioned","label":"Dc Date Accessioned","values":["2008-02-27T22:26:23Z"]},{"key":"dc:date.available","label":"Dc Date Available","values":["2008-02-27T22:26:23Z"]},{"key":"dc:date.copyright","label":"Dc Date Copyright","values":["2007"]},{"key":"dc:date.issued","label":"Date","values":["2007"]},{"key":"dc:publisher","label":"Institution","values":["Massachusetts Institute of Technology"]},{"key":"dc:type","label":"Dc Type","values":["Thesis"]}]},{"id":"subjects_keywords","label":"Subjects and Keywords","entries":[{"key":"dc:subject","label":"Dc Subject","values":["Mechanical Engineering."]}]},{"id":"language_rights","label":"Language and Rights","entries":[{"key":"dc:language.iso","label":"Language (ISO)","values":["en"]},{"key":"dc:rights","label":"Dc Rights","values":["M.I.T. theses are protected by copyright. They may be viewed from this source for any purpose, but reproduction or distribution in any format is prohibited without written permission. See provided URL for inquiries about permission."]},{"key":"dc:rights.uri","label":"Rights URI","values":["http://dspace.mit.edu/handle/1721.1/7582"]}]},{"id":"identifiers","label":"Identifiers","entries":[{"key":"dc:identifier.oclc","label":"OCLC","values":["191730700"]},{"key":"dc:identifier.uri","label":"Identifier URI","values":["http://hdl.handle.net/1721.1/40444"]}]},{"id":"additional","label":"Additional Metadata","entries":[{"key":"cerif:openaire","label":"Cerif Openaire","values":["<Publication xmlns=\"https://www.openaire.eu/cerif-profile/1.1/\" id=\"28f79d65-ef5f-454f-8e9b-35f8dc2bf7da\"> <Type xmlns=\"https://www.openaire.eu/cerif-profile/vocab/COAR_Publication_Types\">http://purl.org/coar/resource_type/c_1843</Type> <Language>en</Language> <Title>Design and modeling of an exhaust gas waste heat autoclave</Title> <PublishedIn> <Publication> </Publication> </PublishedIn> <PublicationDate>2007</PublicationDate> <Authors> <Author> <DisplayName>Jiricek, Joshua A</DisplayName> <Affiliation> <OrgUnit> </OrgUnit> </Affiliation> </Author> </Authors> <Editors> </Editors> <Publishers> <Publisher> <DisplayName>Massachusetts Institute of Technology</DisplayName> <OrgUnit /> </Publisher> </Publishers> <License>http://dspace.mit.edu/handle/1721.1/7582</License> <Keyword>Mechanical Engineering.</Keyword> <Abstract>In order to provide proper sterilization and cleaning of medical equipment for field hospitals and third-world countries while also decreasing the reliance on electricity of traditional sterilization methods, a new steam sterilizer/autoclave system was designed and modeled. This system uses waste engine heat from the exhaust system of a diesel generator set to boil water and produce the pressurized steam conditions necessary for effective medical sterilization. Currently, the design utilizes a 0.59 meter, concentric tube cross-flow heat exchanger and high-temperature heat transfer fluid to draw thermal energy from the exhaust pipe and deposit it into the autoclave pressure vessel to create steam. The system is designed to run a 35-minute sterilization cycle, requiring 15 minutes to produce saturated steam at 2 atmospheres within a 50- liter autoclave, and 20 minutes to sterilize medical instruments in the steam environment. Furthermore, the system uses basic, off-the-shelf fluid transfer materials to provide a robust, effective system that can be easily maintained in the field without need for specialized parts or technicians.</Abstract> <Access xmlns=\"http://purl.org/coar/access_right\" > </Access> </Publication>"]},{"key":"dc:description","label":"Description","values":["Thesis (S.B.)--Massachusetts Institute of Technology, Dept. of Mechanical Engineering, 2007.","Includes bibliographical references (leaf 29)."]},{"key":"dc:description.abstract","label":"Abstract","values":["In order to provide proper sterilization and cleaning of medical equipment for field hospitals and third-world countries while also decreasing the reliance on electricity of traditional sterilization methods, a new steam sterilizer/autoclave system was designed and modeled. This system uses waste engine heat from the exhaust system of a diesel generator set to boil water and produce the pressurized steam conditions necessary for effective medical sterilization. Currently, the design utilizes a 0.59 meter, concentric tube cross-flow heat exchanger and high-temperature heat transfer fluid to draw thermal energy from the exhaust pipe and deposit it into the autoclave pressure vessel to create steam. The system is designed to run a 35-minute sterilization cycle, requiring 15 minutes to produce saturated steam at 2 atmospheres within a 50- liter autoclave, and 20 minutes to sterilize medical instruments in the steam environment. Furthermore, the system uses basic, off-the-shelf fluid transfer materials to provide a robust, effective system that can be easily maintained in the field without need for specialized parts or technicians."]},{"key":"dc:description.degree","label":"Dc Description Degree","values":["S.B."]},{"key":"dc:description.statementofresponsibility","label":"Dc Description Statementofresponsibility","values":["by Joshua A. Jiricek."]},{"key":"dc:format.extent","label":"Dc Format Extent","values":["32 leaves"]},{"key":"dc:format.mimetype","label":"Dc Format Mimetype","values":["application/pdf"]},{"key":"dc:title","label":"Title","values":["Design and modeling of an exhaust gas waste heat autoclave"]},{"key":"dspace:entity.type","label":"Dspace Entity Type","values":["Publication"]},{"key":"others:access_status","label":"Others Access Status","values":["unknown"]}]}],"canonical_facts":{"cerif:openaire":["<Publication xmlns=\"https://www.openaire.eu/cerif-profile/1.1/\" id=\"28f79d65-ef5f-454f-8e9b-35f8dc2bf7da\"> <Type xmlns=\"https://www.openaire.eu/cerif-profile/vocab/COAR_Publication_Types\">http://purl.org/coar/resource_type/c_1843</Type> <Language>en</Language> <Title>Design and modeling of an exhaust gas waste heat autoclave</Title> <PublishedIn> <Publication> </Publication> </PublishedIn> <PublicationDate>2007</PublicationDate> <Authors> <Author> <DisplayName>Jiricek, Joshua A</DisplayName> <Affiliation> <OrgUnit> </OrgUnit> </Affiliation> </Author> </Authors> <Editors> </Editors> <Publishers> <Publisher> <DisplayName>Massachusetts Institute of Technology</DisplayName> <OrgUnit /> </Publisher> </Publishers> <License>http://dspace.mit.edu/handle/1721.1/7582</License> <Keyword>Mechanical Engineering.</Keyword> <Abstract>In order to provide proper sterilization and cleaning of medical equipment for field hospitals and third-world countries while also decreasing the reliance on electricity of traditional sterilization methods, a new steam sterilizer/autoclave system was designed and modeled. This system uses waste engine heat from the exhaust system of a diesel generator set to boil water and produce the pressurized steam conditions necessary for effective medical sterilization. Currently, the design utilizes a 0.59 meter, concentric tube cross-flow heat exchanger and high-temperature heat transfer fluid to draw thermal energy from the exhaust pipe and deposit it into the autoclave pressure vessel to create steam. The system is designed to run a 35-minute sterilization cycle, requiring 15 minutes to produce saturated steam at 2 atmospheres within a 50- liter autoclave, and 20 minutes to sterilize medical instruments in the steam environment. Furthermore, the system uses basic, off-the-shelf fluid transfer materials to provide a robust, effective system that can be easily maintained in the field without need for specialized parts or technicians.</Abstract> <Access xmlns=\"http://purl.org/coar/access_right\" > </Access> </Publication>"],"dc:contributor.advisor":["Todd Thorsen."],"dc:contributor.other":["Massachusetts Institute of Technology. Dept. of Mechanical Engineering."],"dc:creator":["Jiricek, Joshua A"],"dc:date.accessioned":["2008-02-27T22:26:23Z"],"dc:date.available":["2008-02-27T22:26:23Z"],"dc:date.copyright":["2007"],"dc:date.issued":["2007"],"dc:description":["Thesis (S.B.)--Massachusetts Institute of Technology, Dept. of Mechanical Engineering, 2007.","Includes bibliographical references (leaf 29)."],"dc:description.abstract":["In order to provide proper sterilization and cleaning of medical equipment for field hospitals and third-world countries while also decreasing the reliance on electricity of traditional sterilization methods, a new steam sterilizer/autoclave system was designed and modeled. This system uses waste engine heat from the exhaust system of a diesel generator set to boil water and produce the pressurized steam conditions necessary for effective medical sterilization. Currently, the design utilizes a 0.59 meter, concentric tube cross-flow heat exchanger and high-temperature heat transfer fluid to draw thermal energy from the exhaust pipe and deposit it into the autoclave pressure vessel to create steam. The system is designed to run a 35-minute sterilization cycle, requiring 15 minutes to produce saturated steam at 2 atmospheres within a 50- liter autoclave, and 20 minutes to sterilize medical instruments in the steam environment. Furthermore, the system uses basic, off-the-shelf fluid transfer materials to provide a robust, effective system that can be easily maintained in the field without need for specialized parts or technicians."],"dc:description.degree":["S.B."],"dc:description.statementofresponsibility":["by Joshua A. Jiricek."],"dc:format.extent":["32 leaves"],"dc:format.mimetype":["application/pdf"],"dc:identifier.oclc":["191730700"],"dc:identifier.uri":["http://hdl.handle.net/1721.1/40444"],"dc:language.iso":["en"],"dc:publisher":["Massachusetts Institute of Technology"],"dc:rights":["M.I.T. theses are protected by copyright. They may be viewed from this source for any purpose, but reproduction or distribution in any format is prohibited without written permission. See provided URL for inquiries about permission."],"dc:rights.uri":["http://dspace.mit.edu/handle/1721.1/7582"],"dc:subject":["Mechanical Engineering."],"dc:title":["Design and modeling of an exhaust gas waste heat autoclave"],"dc:type":["Thesis"],"dspace:entity.type":["Publication"],"others:access_status":["unknown"]},"updated_at":"2026-11-27T05:21:22Z"}