{"id":{"repo_id":"njit","oai_identifier":"oai:digitalcommons.njit.edu:theses-2172"},"canonical_url":"https://search.ndltd.org/etd/njit/oai:digitalcommons.njit.edu:theses-2172","repository":{"repo_id":"njit","name":"NJIT","base_url":"https://digitalcommons.njit.edu/do/oai/"},"display":{"title":"Wear of polyethylene and hylamer on cobalt-chromium : a knee simulator study","abstract":"Two tests were conducted to examine the wear characteristics of tibial bearings used in total knee replacement systems. Each test consisted of six A/P Glide Tibial Bearings each having a conical control arm. The plastic portion of the conical bearings were all made of UHMWPe. Each of these bearing systems was mounted onto a Co-Cr alloy tibial platform and Co-Cr alloy LCS (low contact stress) femoral component . These test samples were mounted onto the New Jersey Mark III Knee Simulator System. The simulator was configured to produce flexion and axial rotation to simulates normal gait. Each test ran at 2 Hz with saline being sprayed between articulating surfaces. Simulation I tested six UHMWPe bearings with an off-center load applied to the bearing by the femoral component 25 from the articulating surface segment tangent. Simulation II tested three Hylamer and three UHMWPe bearings without an off-center load. Hylamer 's volumetric loss and wear rate were found to be higher then UHMWPe. Hylamer had a maximum volumetric loss of 12.86 mm^3 and a maximum wear rate of 6.19 mm^3/million cycles while UHMWPe had a maximum volumetric loss of 3.57 mm^3 and a maximum wear rate of 1.67 mm^3/million cycles. Hylamer increase in crystallinity slightly increases its yield strength and ultimate tensile strength. However, by increasing the crystallinity, stiffness is also increased. This increase in stiffness increases the contact stress which in turn increases the wear. Although a slight increase in strength is gained when using Hylamer , wear resistance, an important characteristic for total knee replacement systems, is reduced.","abstract_html":"Two tests were conducted to examine the wear characteristics of tibial bearings used in total knee replacement systems. Each test consisted of six A/P Glide Tibial Bearings each having a conical control arm. The plastic portion of the conical bearings were all made of UHMWPe. Each of these bearing systems was mounted onto a Co-Cr alloy tibial platform and Co-Cr alloy LCS (low contact stress) femoral component . These test samples were mounted onto the New Jersey Mark III Knee Simulator System. The simulator was configured to produce flexion and axial rotation to simulates normal gait. Each test ran at 2 Hz with saline being sprayed between articulating surfaces. Simulation I tested six UHMWPe bearings with an off-center load applied to the bearing by the femoral component 25 from the articulating surface segment tangent. Simulation II tested three Hylamer and three UHMWPe bearings without an off-center load. Hylamer &#x27;s volumetric loss and wear rate were found to be higher then UHMWPe. Hylamer had a maximum volumetric loss of 12.86 mm^3 and a maximum wear rate of 6.19 mm^3/million cycles while UHMWPe had a maximum volumetric loss of 3.57 mm^3 and a maximum wear rate of 1.67 mm^3/million cycles. Hylamer increase in crystallinity slightly increases its yield strength and ultimate tensile strength. However, by increasing the crystallinity, stiffness is also increased. This increase in stiffness increases the contact stress which in turn increases the wear. Although a slight increase in strength is gained when using Hylamer , wear resistance, an important characteristic for total knee replacement systems, is reduced.","abstract_has_math":false,"creators":["Canonaco, Alessandro F."],"institution":null,"degree_name":"Master of Science in Biomedical Engineering - (M.S.)","degree_level":null,"degree_discipline":"Biomedical Engineering Committee","degree_department":null,"school":null,"contributors":["Clarence W. Mayott","Michael Pappas","Georghios Makris"],"advisors":[],"committee_chairs":[],"committee_members":[],"year":null,"date_issued":"","date_published":null,"updated_at":"2026-11-19T08:40:35Z","subjects":["Artificial knee.","Implants","Artificial--Materials.","Polyethylene.","Mechanical wear--Testing.","Cobalt alloys.","Chromium alloys.","Biomedical Engineering and Bioengineering"],"languages":[],"rights":[],"rights_urls":[],"identifier_entries":[],"person_identifier_entries":[]},"links":{"outbound_url":"https://digitalcommons.njit.edu/theses/1173","outbound_label":"Repository record","outbound_source":"dc:identifier","record_url":"https://digitalcommons.njit.edu/theses/1173","record_label":"Repository record","record_kind":"record","record_source":"dc:identifier","download_url":null,"download_label":null,"download_source":null,"access_status":null,"embargo_until":null,"access_note":null},"source_record":{"url":"https://digitalcommons.njit.edu/do/oai/?verb=GetRecord&metadataPrefix=oai_etdms&identifier=oai%3Adigitalcommons.njit.edu%3Atheses-2172","prefix":"oai_etdms"},"metadata_groups":[{"id":"people","label":"People","entries":[{"key":"dc:contributor","label":"Contributor","values":["Clarence W. Mayott","Michael Pappas","Georghios Makris"]},{"key":"dc:creator","label":"Author","values":["Canonaco, Alessandro F."]}]},{"id":"academic_context","label":"Academic Context","entries":[{"key":"dc:date.created","label":"Dc Date Created","values":["1995-01-31T08:00:00Z"]},{"key":"dc:type","label":"Dc Type","values":["Thesis"]},{"key":"thesis:degree_discipline","label":"Discipline","values":["Biomedical Engineering Committee"]},{"key":"thesis:degree_name","label":"Degree Name","values":["Master of Science in Biomedical Engineering - (M.S.)"]}]},{"id":"subjects_keywords","label":"Subjects and Keywords","entries":[{"key":"dc:subject","label":"Dc Subject","values":["Artificial knee.","Implants","Artificial--Materials.","Polyethylene.","Mechanical wear--Testing.","Cobalt alloys.","Chromium alloys.","Biomedical Engineering and Bioengineering"]}]},{"id":"identifiers","label":"Identifiers","entries":[{"key":"dc:identifier","label":"Identifier","values":["https://digitalcommons.njit.edu/theses/1173"]}]},{"id":"additional","label":"Additional Metadata","entries":[{"key":"dc:description.abstract","label":"Abstract","values":["Two tests were conducted to examine the wear characteristics of tibial bearings used in total knee replacement systems. Each test consisted of six A/P Glide Tibial Bearings each having a conical control arm. The plastic portion of the conical bearings were all made of UHMWPe. Each of these bearing systems was mounted onto a Co-Cr alloy tibial platform and Co-Cr alloy LCS (low contact stress) femoral component . These test samples were mounted onto the New Jersey Mark III Knee Simulator System. The simulator was configured to produce flexion and axial rotation to simulates normal gait. Each test ran at 2 Hz with saline being sprayed between articulating surfaces. Simulation I tested six UHMWPe bearings with an off-center load applied to the bearing by the femoral component 25 from the articulating surface segment tangent. Simulation II tested three Hylamer and three UHMWPe bearings without an off-center load. Hylamer 's volumetric loss and wear rate were found to be higher then UHMWPe. Hylamer had a maximum volumetric loss of 12.86 mm^3 and a maximum wear rate of 6.19 mm^3/million cycles while UHMWPe had a maximum volumetric loss of 3.57 mm^3 and a maximum wear rate of 1.67 mm^3/million cycles. Hylamer increase in crystallinity slightly increases its yield strength and ultimate tensile strength. However, by increasing the crystallinity, stiffness is also increased. This increase in stiffness increases the contact stress which in turn increases the wear. Although a slight increase in strength is gained when using Hylamer , wear resistance, an important characteristic for total knee replacement systems, is reduced."]},{"key":"dc:title","label":"Title","values":["Wear of polyethylene and hylamer on cobalt-chromium : a knee simulator study"]}]}],"canonical_facts":{"dc:contributor":["Clarence W. Mayott","Michael Pappas","Georghios Makris"],"dc:creator":["Canonaco, Alessandro F."],"dc:date.created":["1995-01-31T08:00:00Z"],"dc:description.abstract":["Two tests were conducted to examine the wear characteristics of tibial bearings used in total knee replacement systems. Each test consisted of six A/P Glide Tibial Bearings each having a conical control arm. The plastic portion of the conical bearings were all made of UHMWPe. Each of these bearing systems was mounted onto a Co-Cr alloy tibial platform and Co-Cr alloy LCS (low contact stress) femoral component . These test samples were mounted onto the New Jersey Mark III Knee Simulator System. The simulator was configured to produce flexion and axial rotation to simulates normal gait. Each test ran at 2 Hz with saline being sprayed between articulating surfaces. Simulation I tested six UHMWPe bearings with an off-center load applied to the bearing by the femoral component 25 from the articulating surface segment tangent. Simulation II tested three Hylamer and three UHMWPe bearings without an off-center load. Hylamer 's volumetric loss and wear rate were found to be higher then UHMWPe. Hylamer had a maximum volumetric loss of 12.86 mm^3 and a maximum wear rate of 6.19 mm^3/million cycles while UHMWPe had a maximum volumetric loss of 3.57 mm^3 and a maximum wear rate of 1.67 mm^3/million cycles. Hylamer increase in crystallinity slightly increases its yield strength and ultimate tensile strength. However, by increasing the crystallinity, stiffness is also increased. This increase in stiffness increases the contact stress which in turn increases the wear. Although a slight increase in strength is gained when using Hylamer , wear resistance, an important characteristic for total knee replacement systems, is reduced."],"dc:identifier":["https://digitalcommons.njit.edu/theses/1173"],"dc:subject":["Artificial knee.","Implants","Artificial--Materials.","Polyethylene.","Mechanical wear--Testing.","Cobalt alloys.","Chromium alloys.","Biomedical Engineering and Bioengineering"],"dc:title":["Wear of polyethylene and hylamer on cobalt-chromium : a knee simulator study"],"dc:type":["Thesis"],"thesis:degree_discipline":["Biomedical Engineering Committee"],"thesis:degree_name":["Master of Science in Biomedical Engineering - (M.S.)"]},"updated_at":"2026-11-19T08:40:35Z"}