{"id":{"repo_id":"nps","oai_identifier":"oai:calhoun.nps.edu:10945/11047"},"canonical_url":"https://search.ndltd.org/etd/nps/oai:calhoun.nps.edu:10945/11047","repository":{"repo_id":"nps","name":"Naval Postgraduate School","base_url":"https://calhoun.nps.edu/server/oai/request"},"display":{"title":"Bending strength test and evaluation of a transtibial prosthetic socket fabricated by selective laser sintering","abstract":"Transtibial prosthetic sockets made using selective laser sintering (SLS) of Duraform (™) PA have no documented failure data. In order to produce prostheses with safe weight and usage limits, a non-standard bending test was needed to determine the maximimi safe loading of the socket and to recommend design improvements. The bending test was designed to replicate forces experienced by a wearer stepping down and forward as from a slight elevation. This test, along with a finite element analysis, provided information about the force limitations and weakest point of the socket. Suggestions for improving the design incorporated this information and laid the groundwork for further improvements.","abstract_html":"Transtibial prosthetic sockets made using selective laser sintering (SLS) of Duraform (™) PA have no documented failure data. In order to produce prostheses with safe weight and usage limits, a non-standard bending test was needed to determine the maximimi safe loading of the socket and to recommend design improvements. The bending test was designed to replicate forces experienced by a wearer stepping down and forward as from a slight elevation. This test, along with a finite element analysis, provided information about the force limitations and weakest point of the socket. Suggestions for improving the design incorporated this information and laid the groundwork for further improvements.","abstract_has_math":false,"creators":["Brooks, Amanda Joyce"],"institution":"University of Texas at Austin","degree_name":"Master of Science in Engineering","degree_level":"Masters","degree_discipline":"Engineering","degree_department":null,"school":null,"contributors":[],"advisors":["Crawford, Richard"],"committee_chairs":[],"committee_members":[],"year":2004,"date_issued":"2004-05","date_published":"2004-05","updated_at":"2026-11-19T06:02:42Z","subjects":[],"languages":[],"rights":["This publication is a work of the U.S. Government as defined in Title 17, United States Code, Section 101. Copyright protection is not available for this work in the United States.","Distribution Statement A. Approved for public release: Distribution is unlimited."],"rights_urls":[],"identifier_entries":[],"person_identifier_entries":[]},"links":{"outbound_url":"https://hdl.handle.net/10945/11047","outbound_label":"Handle","outbound_source":"dc:identifier.uri","record_url":"https://hdl.handle.net/10945/11047","record_label":"Repository record","record_kind":"handle","record_source":"dc:identifier.uri","download_url":null,"download_label":null,"download_source":null,"access_status":null,"embargo_until":null,"access_note":null},"source_record":{"url":"https://calhoun.nps.edu/server/oai/request?verb=GetRecord&metadataPrefix=dim&identifier=oai%3Acalhoun.nps.edu%3A10945%2F11047","prefix":"dim"},"metadata_groups":[{"id":"people","label":"People","entries":[{"key":"dc:contributor.advisor","label":"Advisor","values":["Crawford, Richard"]},{"key":"dc:contributor.other","label":"Dc Contributor Other","values":["CIVINS (Civilian Institutions)"]},{"key":"dc:creator","label":"Author","values":["Brooks, Amanda Joyce"]}]},{"id":"academic_context","label":"Academic Context","entries":[{"key":"dc:date.accessioned","label":"Dc Date Accessioned","values":["2012-08-22T15:34:52Z"]},{"key":"dc:date.available","label":"Dc Date Available","values":["2012-08-22T15:34:52Z"]},{"key":"dc:date.issued","label":"Date","values":["2004-05"]},{"key":"dc:publisher","label":"Institution","values":["University of Texas at Austin"]},{"key":"dc:relation.ispartofseries","label":"Dc Relation Ispartofseries","values":["CIVINS (Civilian Institutions) Theses and Dissertations"]},{"key":"dc:relation.orgunit","label":"Dc Relation Orgunit","values":["c6eb8929-c883-4d31-84f5-166ce3ae8c52"]},{"key":"dc:relation.seriesentity","label":"Dc Relation Seriesentity","values":["0855bc76-cf0d-4c5f-84ba-3b1c54b01904"]},{"key":"dc:type","label":"Dc Type","values":["Thesis"]},{"key":"thesis:degree_discipline","label":"Discipline","values":["Engineering"]},{"key":"thesis:degree_level","label":"Degree Level","values":["Masters"]},{"key":"thesis:degree_name","label":"Degree Name","values":["Master of Science in Engineering"]},{"key":"thesis:institution_name","label":"Thesis Institution Name","values":["University of Texas at Austin"]}]},{"id":"language_rights","label":"Language and Rights","entries":[{"key":"dc:rights","label":"Dc Rights","values":["This publication is a work of the U.S. Government as defined in Title 17, United States Code, Section 101. Copyright protection is not available for this work in the United States.","Distribution Statement A. Approved for public release: Distribution is unlimited."]}]},{"id":"identifiers","label":"Identifiers","entries":[{"key":"dc:identifier.uri","label":"Identifier URI","values":["https://hdl.handle.net/10945/11047"]}]},{"id":"additional","label":"Additional Metadata","entries":[{"key":"dc:description.abstract","label":"Abstract","values":["Transtibial prosthetic sockets made using selective laser sintering (SLS) of Duraform (™) PA have no documented failure data. In order to produce prostheses with safe weight and usage limits, a non-standard bending test was needed to determine the maximimi safe loading of the socket and to recommend design improvements. The bending test was designed to replicate forces experienced by a wearer stepping down and forward as from a slight elevation. This test, along with a finite element analysis, provided information about the force limitations and weakest point of the socket. Suggestions for improving the design incorporated this information and laid the groundwork for further improvements."]},{"key":"dc:description.uri","label":"Dc Description URI","values":["http://archive.org/details/bendingstrengtht1094511047"]},{"key":"dc:format.extent","label":"Dc Format Extent","values":["55 p."]},{"key":"dc:title","label":"Title","values":["Bending strength test and evaluation of a transtibial prosthetic socket fabricated by selective laser sintering"]},{"key":"dspace:entity.type","label":"Dspace Entity Type","values":["Publication"]}]}],"canonical_facts":{"dc:contributor.advisor":["Crawford, Richard"],"dc:contributor.other":["CIVINS (Civilian Institutions)"],"dc:creator":["Brooks, Amanda Joyce"],"dc:date.accessioned":["2012-08-22T15:34:52Z"],"dc:date.available":["2012-08-22T15:34:52Z"],"dc:date.issued":["2004-05"],"dc:description.abstract":["Transtibial prosthetic sockets made using selective laser sintering (SLS) of Duraform (™) PA have no documented failure data. In order to produce prostheses with safe weight and usage limits, a non-standard bending test was needed to determine the maximimi safe loading of the socket and to recommend design improvements. The bending test was designed to replicate forces experienced by a wearer stepping down and forward as from a slight elevation. This test, along with a finite element analysis, provided information about the force limitations and weakest point of the socket. Suggestions for improving the design incorporated this information and laid the groundwork for further improvements."],"dc:description.uri":["http://archive.org/details/bendingstrengtht1094511047"],"dc:format.extent":["55 p."],"dc:identifier.uri":["https://hdl.handle.net/10945/11047"],"dc:publisher":["University of Texas at Austin"],"dc:relation.ispartofseries":["CIVINS (Civilian Institutions) Theses and Dissertations"],"dc:relation.orgunit":["c6eb8929-c883-4d31-84f5-166ce3ae8c52"],"dc:relation.seriesentity":["0855bc76-cf0d-4c5f-84ba-3b1c54b01904"],"dc:rights":["This publication is a work of the U.S. Government as defined in Title 17, United States Code, Section 101. Copyright protection is not available for this work in the United States.","Distribution Statement A. Approved for public release: Distribution is unlimited."],"dc:title":["Bending strength test and evaluation of a transtibial prosthetic socket fabricated by selective laser sintering"],"dc:type":["Thesis"],"dspace:entity.type":["Publication"],"thesis:degree_discipline":["Engineering"],"thesis:degree_level":["Masters"],"thesis:degree_name":["Master of Science in Engineering"],"thesis:institution_name":["University of Texas at Austin"]},"updated_at":"2026-11-19T06:02:42Z"}