{"id":{"repo_id":"ohiolink","oai_identifier":"oai:etd.ohiolink.edu:akron1205764260"},"canonical_url":"https://search.ndltd.org/etd/ohiolink/oai:etd.ohiolink.edu:akron1205764260","repository":{"repo_id":"ohiolink","name":"OhioLINK","base_url":"https://etd.ohiolink.edu/acprod/odb_etd/ws/oai/oai"},"display":{"title":"A Real-Time Technique for the Correction of Invasive Blood Pressure Measurements using Counter Pressure","abstract":"In critical care, a fluid-filled catheter transducer system is commonly used to continuously monitor blood pressure. The fluid-filled catheter transducer system, as a first approximation, behaves as a second order dynamic system. The dynamic characteristics of the system are affected by variations in assembly technique and time dependent changes in the system, often resulting in distortion and inaccurate measurements. A previous simulation study employed a counter pressure source in tandem with the transducer. The counter pressure generated minimized the fluid flow in the pressure monitoring system. When the flow in the system was zero the counter pressure generated closely approximated the true blood pressure. The current study developed a real-time technique to generate accurate and dependable counter pressure. To validate this technique, one experimental model (second order dynamic system) and two simulation models (second order dynamic system and fourth order dynamic system) of catheter transducer system were developed and tested under varying system conditions. The real-time technique developed, successfully reproduced the true blood pressure waveforms, regardless of variations in the system characteristics and changes in the system over time.","abstract_html":"In critical care, a fluid-filled catheter transducer system is commonly used to continuously monitor blood pressure. The fluid-filled catheter transducer system, as a first approximation, behaves as a second order dynamic system. The dynamic characteristics of the system are affected by variations in assembly technique and time dependent changes in the system, often resulting in distortion and inaccurate measurements. A previous simulation study employed a counter pressure source in tandem with the transducer. The counter pressure generated minimized the fluid flow in the pressure monitoring system. When the flow in the system was zero the counter pressure generated closely approximated the true blood pressure. The current study developed a real-time technique to generate accurate and dependable counter pressure. To validate this technique, one experimental model (second order dynamic system) and two simulation models (second order dynamic system and fourth order dynamic system) of catheter transducer system were developed and tested under varying system conditions. The real-time technique developed, successfully reproduced the true blood pressure waveforms, regardless of variations in the system characteristics and changes in the system over time.","abstract_has_math":false,"creators":["Patel, Darshan Shyam"],"institution":"University of Akron","degree_name":"Master of Science in Engineering","degree_level":"masters","degree_discipline":"Biomedical Engineering","degree_department":null,"school":null,"contributors":[],"advisors":["Taylor, Bruce"],"committee_chairs":[],"committee_members":[],"year":2008,"date_issued":"2008-05-12","date_published":"2008-05-12","updated_at":"2026-11-27T04:13:13Z","subjects":["Biomedical Research","COUNTER PRESSURE","catheter transducer system","BLOOD PRESSURE","transducer system","catheter transducer","PRESSURE WAVE","counter pressure source"],"languages":["English"],"rights":["unrestricted","This thesis or dissertation is protected by copyright: all rights reserved. It may not be copied or redistributed beyond the terms of applicable copyright laws."],"rights_urls":[],"identifier_entries":[],"person_identifier_entries":[]},"links":{"outbound_url":"http://rave.ohiolink.edu/etdc/view?acc_num=akron1205764260","outbound_label":"Repository record","outbound_source":"dc:identifier","record_url":"http://rave.ohiolink.edu/etdc/view?acc_num=akron1205764260","record_label":"Repository record","record_kind":"record","record_source":"dc:identifier","access_status":null,"embargo_until":null,"access_note":null},"oai_pmh_record":{"url":"https://union.ndltd.org/oai-pmh/?verb=GetRecord&metadataPrefix=oai_dc&identifier=oai%3Aunion.ndltd.org%3Arepo%3Aohiolink%3Aoai%3Aetd.ohiolink.edu%3Aakron1205764260","prefix":"oai_dc"},"source_record":{"url":"https://etd.ohiolink.edu/acprod/odb_etd/ws/oai/oai?verb=GetRecord&metadataPrefix=oai_etdms&identifier=oai%3Aetd.ohiolink.edu%3Aakron1205764260","prefix":"oai_etdms"},"metadata_groups":[{"id":"people","label":"People","entries":[{"key":"dc:contributor.advisor","label":"Advisor","values":["Taylor, Bruce"]},{"key":"dc:creator","label":"Author","values":["Patel, Darshan Shyam"]}]},{"id":"academic_context","label":"Academic Context","entries":[{"key":"dc:date","label":"Dc Date","values":["2008-05-12"]},{"key":"dc:publisher","label":"Institution","values":["University of Akron / OhioLINK"]},{"key":"dc:type","label":"Dc Type","values":["Electronic Thesis or Dissertation"]},{"key":"thesis:degree_discipline","label":"Discipline","values":["Biomedical Engineering"]},{"key":"thesis:degree_level","label":"Degree Level","values":["masters"]},{"key":"thesis:degree_name","label":"Degree Name","values":["Master of Science in Engineering"]},{"key":"thesis:institution_name","label":"Thesis Institution Name","values":["University of Akron"]}]},{"id":"subjects_keywords","label":"Subjects and Keywords","entries":[{"key":"dc:subject","label":"Dc Subject","values":["Biomedical Research","COUNTER PRESSURE","catheter transducer system","BLOOD PRESSURE","transducer system","catheter transducer","PRESSURE WAVE","counter pressure source"]}]},{"id":"language_rights","label":"Language and Rights","entries":[{"key":"dc:language","label":"Dc Language","values":["English"]},{"key":"dc:rights","label":"Dc Rights","values":["unrestricted","This thesis or dissertation is protected by copyright: all rights reserved. It may not be copied or redistributed beyond the terms of applicable copyright laws."]}]},{"id":"identifiers","label":"Identifiers","entries":[{"key":"dc:identifier","label":"Identifier","values":["http://rave.ohiolink.edu/etdc/view?acc_num=akron1205764260"]}]},{"id":"additional","label":"Additional Metadata","entries":[{"key":"dc:description","label":"Description","values":["In critical care, a fluid-filled catheter transducer system is commonly used to continuously monitor blood pressure. The fluid-filled catheter transducer system, as a first approximation, behaves as a second order dynamic system. The dynamic characteristics of the system are affected by variations in assembly technique and time dependent changes in the system, often resulting in distortion and inaccurate measurements. A previous simulation study employed a counter pressure source in tandem with the transducer. The counter pressure generated minimized the fluid flow in the pressure monitoring system. When the flow in the system was zero the counter pressure generated closely approximated the true blood pressure. The current study developed a real-time technique to generate accurate and dependable counter pressure. To validate this technique, one experimental model (second order dynamic system) and two simulation models (second order dynamic system and fourth order dynamic system) of catheter transducer system were developed and tested under varying system conditions. The real-time technique developed, successfully reproduced the true blood pressure waveforms, regardless of variations in the system characteristics and changes in the system over time."]},{"key":"dc:format","label":"Dc Format","values":["application/pdf","758.6 KB"]},{"key":"dc:title","label":"Title","values":["A Real-Time Technique for the Correction of Invasive Blood Pressure Measurements using Counter Pressure"]}]}],"canonical_facts":{"dc:contributor.advisor":["Taylor, Bruce"],"dc:creator":["Patel, Darshan Shyam"],"dc:date":["2008-05-12"],"dc:description":["In critical care, a fluid-filled catheter transducer system is commonly used to continuously monitor blood pressure. The fluid-filled catheter transducer system, as a first approximation, behaves as a second order dynamic system. The dynamic characteristics of the system are affected by variations in assembly technique and time dependent changes in the system, often resulting in distortion and inaccurate measurements. A previous simulation study employed a counter pressure source in tandem with the transducer. The counter pressure generated minimized the fluid flow in the pressure monitoring system. When the flow in the system was zero the counter pressure generated closely approximated the true blood pressure. The current study developed a real-time technique to generate accurate and dependable counter pressure. To validate this technique, one experimental model (second order dynamic system) and two simulation models (second order dynamic system and fourth order dynamic system) of catheter transducer system were developed and tested under varying system conditions. The real-time technique developed, successfully reproduced the true blood pressure waveforms, regardless of variations in the system characteristics and changes in the system over time."],"dc:format":["application/pdf","758.6 KB"],"dc:identifier":["http://rave.ohiolink.edu/etdc/view?acc_num=akron1205764260"],"dc:language":["English"],"dc:publisher":["University of Akron / OhioLINK"],"dc:rights":["unrestricted","This thesis or dissertation is protected by copyright: all rights reserved. It may not be copied or redistributed beyond the terms of applicable copyright laws."],"dc:subject":["Biomedical Research","COUNTER PRESSURE","catheter transducer system","BLOOD PRESSURE","transducer system","catheter transducer","PRESSURE WAVE","counter pressure source"],"dc:title":["A Real-Time Technique for the Correction of Invasive Blood Pressure Measurements using Counter Pressure"],"dc:type":["Electronic Thesis or Dissertation"],"thesis:degree_discipline":["Biomedical Engineering"],"thesis:degree_level":["masters"],"thesis:degree_name":["Master of Science in Engineering"],"thesis:institution_name":["University of Akron"]},"updated_at":"2026-11-27T04:13:13Z"}