{"id":{"repo_id":"ohiolink","oai_identifier":"oai:etd.ohiolink.edu:case1462899086"},"canonical_url":"https://search.ndltd.org/etd/ohiolink/oai:etd.ohiolink.edu:case1462899086","repository":{"repo_id":"ohiolink","name":"OhioLINK","base_url":"https://etd.ohiolink.edu/acprod/odb_etd/ws/oai/oai"},"display":{"title":"In vivo Biocompatibilty and Time-Dependent Changes in Mechanical Properties of Woven Collagen Meshes: Comparison to Xenograft and Synthetic Mid-Urethral Sling Materials","abstract":"Woven collagen meshes crosslinked using either genipin or 1-ethyl-3-(3-dimethylaminopropyl)carbodiimide (EDC) and N-hydroxysuccinimide (NHS), as well as synthetic and xenograft materials used for the management of stress urinary incontinence, were implanted in a subcutaneous rat model to evaluate biocompatibility and changes in mechanical properties over time. The performance of materials were compared to determine if woven collagen meshes could be used for management of stress urinary incontinence.Both woven collagen meshes elicited a more facile chronic inflammatory response than synthetic or xenograft materials. However, genipin crosslinked collagen meshes initiated patent tissue ingrowth while EDC/NHS crosslinked collagen meshes did not. Because of this, the mechanical properties of genipin crosslinked collagen meshes most closely matched those of native tissues over time. Overall, the host response and mechanical properties of genipin crosslinked collagen meshes in this subcutaneous animal model show promise for their potential as a new material for management of stress urinary incontinence.","abstract_html":"Woven collagen meshes crosslinked using either genipin or 1-ethyl-3-(3-dimethylaminopropyl)carbodiimide (EDC) and N-hydroxysuccinimide (NHS), as well as synthetic and xenograft materials used for the management of stress urinary incontinence, were implanted in a subcutaneous rat model to evaluate biocompatibility and changes in mechanical properties over time. The performance of materials were compared to determine if woven collagen meshes could be used for management of stress urinary incontinence.Both woven collagen meshes elicited a more facile chronic inflammatory response than synthetic or xenograft materials. However, genipin crosslinked collagen meshes initiated patent tissue ingrowth while EDC/NHS crosslinked collagen meshes did not. Because of this, the mechanical properties of genipin crosslinked collagen meshes most closely matched those of native tissues over time. Overall, the host response and mechanical properties of genipin crosslinked collagen meshes in this subcutaneous animal model show promise for their potential as a new material for management of stress urinary incontinence.","abstract_has_math":false,"creators":["Chapin, Katherine Joan"],"institution":"Case Western Reserve University School of Graduate Studies","degree_name":"Master of Sciences (Engineering)","degree_level":"masters","degree_discipline":"Biomedical Engineering","degree_department":null,"school":null,"contributors":["Akkus, Ozan"],"advisors":[],"committee_chairs":[],"committee_members":[],"year":2016,"date_issued":"2016-05-30","date_published":"2016-05-30","updated_at":"2026-11-26T05:29:22Z","subjects":["Biomedical Engineering","Biomedical Research","stress urinary incontinence, biomaterial, collagen, biocompatibility, mid-urethral sling"],"languages":["English"],"rights":["unrestricted","This thesis or dissertation is protected by copyright: all rights reserved. It may not be copied or redistributed beyond the terms of applicable copyright laws."],"rights_urls":[],"identifier_entries":[],"person_identifier_entries":[]},"links":{"outbound_url":"http://rave.ohiolink.edu/etdc/view?acc_num=case1462899086","outbound_label":"Repository record","outbound_source":"dc:identifier","record_url":"http://rave.ohiolink.edu/etdc/view?acc_num=case1462899086","record_label":"Repository record","record_kind":"record","record_source":"dc:identifier","access_status":null,"embargo_until":null,"access_note":null},"oai_pmh_record":{"url":"https://union.ndltd.org/oai-pmh/?verb=GetRecord&metadataPrefix=oai_dc&identifier=oai%3Aunion.ndltd.org%3Arepo%3Aohiolink%3Aoai%3Aetd.ohiolink.edu%3Acase1462899086","prefix":"oai_dc"},"source_record":{"url":"https://etd.ohiolink.edu/acprod/odb_etd/ws/oai/oai?verb=GetRecord&metadataPrefix=oai_etdms&identifier=oai%3Aetd.ohiolink.edu%3Acase1462899086","prefix":"oai_etdms"},"metadata_groups":[{"id":"people","label":"People","entries":[{"key":"dc:contributor","label":"Contributor","values":["Akkus, Ozan"]},{"key":"dc:creator","label":"Author","values":["Chapin, Katherine Joan"]}]},{"id":"academic_context","label":"Academic Context","entries":[{"key":"dc:date","label":"Dc Date","values":["2016-05-30"]},{"key":"dc:publisher","label":"Institution","values":["Case Western Reserve University School of Graduate Studies / OhioLINK"]},{"key":"dc:type","label":"Dc Type","values":["Electronic Thesis or Dissertation"]},{"key":"thesis:degree_discipline","label":"Discipline","values":["Biomedical Engineering"]},{"key":"thesis:degree_level","label":"Degree Level","values":["masters"]},{"key":"thesis:degree_name","label":"Degree Name","values":["Master of Sciences (Engineering)"]},{"key":"thesis:institution_name","label":"Thesis Institution Name","values":["Case Western Reserve University School of Graduate Studies"]}]},{"id":"subjects_keywords","label":"Subjects and Keywords","entries":[{"key":"dc:subject","label":"Dc Subject","values":["Biomedical Engineering","Biomedical Research","stress urinary incontinence, biomaterial, collagen, biocompatibility, mid-urethral sling"]}]},{"id":"language_rights","label":"Language and Rights","entries":[{"key":"dc:language","label":"Dc Language","values":["English"]},{"key":"dc:rights","label":"Dc Rights","values":["unrestricted","This thesis or dissertation is protected by copyright: all rights reserved. It may not be copied or redistributed beyond the terms of applicable copyright laws."]}]},{"id":"identifiers","label":"Identifiers","entries":[{"key":"dc:identifier","label":"Identifier","values":["http://rave.ohiolink.edu/etdc/view?acc_num=case1462899086"]}]},{"id":"additional","label":"Additional Metadata","entries":[{"key":"dc:description","label":"Description","values":["Woven collagen meshes crosslinked using either genipin or 1-ethyl-3-(3-dimethylaminopropyl)carbodiimide (EDC) and N-hydroxysuccinimide (NHS), as well as synthetic and xenograft materials used for the management of stress urinary incontinence, were implanted in a subcutaneous rat model to evaluate biocompatibility and changes in mechanical properties over time. The performance of materials were compared to determine if woven collagen meshes could be used for management of stress urinary incontinence.Both woven collagen meshes elicited a more facile chronic inflammatory response than synthetic or xenograft materials. However, genipin crosslinked collagen meshes initiated patent tissue ingrowth while EDC/NHS crosslinked collagen meshes did not. Because of this, the mechanical properties of genipin crosslinked collagen meshes most closely matched those of native tissues over time. Overall, the host response and mechanical properties of genipin crosslinked collagen meshes in this subcutaneous animal model show promise for their potential as a new material for management of stress urinary incontinence."]},{"key":"dc:format","label":"Dc Format","values":["application/pdf","p.125","4.06 MB"]},{"key":"dc:title","label":"Title","values":["In vivo Biocompatibilty and Time-Dependent Changes in Mechanical Properties of Woven Collagen Meshes: Comparison to Xenograft and Synthetic Mid-Urethral Sling Materials"]}]}],"canonical_facts":{"dc:contributor":["Akkus, Ozan"],"dc:creator":["Chapin, Katherine Joan"],"dc:date":["2016-05-30"],"dc:description":["Woven collagen meshes crosslinked using either genipin or 1-ethyl-3-(3-dimethylaminopropyl)carbodiimide (EDC) and N-hydroxysuccinimide (NHS), as well as synthetic and xenograft materials used for the management of stress urinary incontinence, were implanted in a subcutaneous rat model to evaluate biocompatibility and changes in mechanical properties over time. The performance of materials were compared to determine if woven collagen meshes could be used for management of stress urinary incontinence.Both woven collagen meshes elicited a more facile chronic inflammatory response than synthetic or xenograft materials. However, genipin crosslinked collagen meshes initiated patent tissue ingrowth while EDC/NHS crosslinked collagen meshes did not. Because of this, the mechanical properties of genipin crosslinked collagen meshes most closely matched those of native tissues over time. Overall, the host response and mechanical properties of genipin crosslinked collagen meshes in this subcutaneous animal model show promise for their potential as a new material for management of stress urinary incontinence."],"dc:format":["application/pdf","p.125","4.06 MB"],"dc:identifier":["http://rave.ohiolink.edu/etdc/view?acc_num=case1462899086"],"dc:language":["English"],"dc:publisher":["Case Western Reserve University School of Graduate Studies / OhioLINK"],"dc:rights":["unrestricted","This thesis or dissertation is protected by copyright: all rights reserved. It may not be copied or redistributed beyond the terms of applicable copyright laws."],"dc:subject":["Biomedical Engineering","Biomedical Research","stress urinary incontinence, biomaterial, collagen, biocompatibility, mid-urethral sling"],"dc:title":["In vivo Biocompatibilty and Time-Dependent Changes in Mechanical Properties of Woven Collagen Meshes: Comparison to Xenograft and Synthetic Mid-Urethral Sling Materials"],"dc:type":["Electronic Thesis or Dissertation"],"thesis:degree_discipline":["Biomedical Engineering"],"thesis:degree_level":["masters"],"thesis:degree_name":["Master of Sciences (Engineering)"],"thesis:institution_name":["Case Western Reserve University School of Graduate Studies"]},"updated_at":"2026-11-26T05:29:22Z"}