{"id":{"repo_id":"ohiolink","oai_identifier":"oai:etd.ohiolink.edu:case1704293346229248"},"canonical_url":"https://search.ndltd.org/etd/ohiolink/oai:etd.ohiolink.edu:case1704293346229248","repository":{"repo_id":"ohiolink","name":"OhioLINK","base_url":"https://etd.ohiolink.edu/acprod/odb_etd/ws/oai/oai"},"display":{"title":"The Effects of Socket Fit and Residual-Limb Load on 3-Dof, Simultaneous and Proportional, Myo-Prosthesis Hand Control and Chronically Implanted, Intramuscular EMG Patterns","abstract":"Myo-electric prosthesis hands (myo-prostheses) intend to restore independence in individuals with upper-limb loss, however, 44% of consumers discontinue device use due to unreliable function [1]. Literature suggests that residual-limb changes may affect the myo-control algorithm’s ability to decode user-intended movement from skin-surface recorded electromyography (EMG). Chronically implanted, intramuscular EMG (ciEMG) improves signal integrity, but the extent to which ciEMG inputs mitigate myo-control degradation is unknown [2]–[4]. This study investigates how 3 Degree-of-Freedom (DoF), Simultaneous and Proportional (SP) myo-control performance and ciEMG hand-posture patterns are affected by loose socket fit and added residual-limb load. Load did degrade myo-control performance during a posture matching test, but only with loose socket wear. Examining postural ciEMG patterns revealed that Extensor Digitorum (ED) activity was greater during loose socket wear, as well. Correcting socket fit may improve postural ciEMG pattern consistency during different residual-limb loads, thereby maintaining reliable myo-prosthesis function.","abstract_html":"Myo-electric prosthesis hands (myo-prostheses) intend to restore independence in individuals with upper-limb loss, however, 44% of consumers discontinue device use due to unreliable function [1]. Literature suggests that residual-limb changes may affect the myo-control algorithm’s ability to decode user-intended movement from skin-surface recorded electromyography (EMG). Chronically implanted, intramuscular EMG (ciEMG) improves signal integrity, but the extent to which ciEMG inputs mitigate myo-control degradation is unknown [2]–[4]. This study investigates how 3 Degree-of-Freedom (DoF), Simultaneous and Proportional (SP) myo-control performance and ciEMG hand-posture patterns are affected by loose socket fit and added residual-limb load. Load did degrade myo-control performance during a posture matching test, but only with loose socket wear. Examining postural ciEMG patterns revealed that Extensor Digitorum (ED) activity was greater during loose socket wear, as well. Correcting socket fit may improve postural ciEMG pattern consistency during different residual-limb loads, thereby maintaining reliable myo-prosthesis function.","abstract_has_math":false,"creators":["Dsouza, Karina Theresa"],"institution":"Case Western Reserve University School of Graduate Studies","degree_name":"Master of Sciences (Engineering)","degree_level":"masters","degree_discipline":"Biomedical Engineering","degree_department":null,"school":null,"contributors":[],"advisors":["Tyler, Dustin"],"committee_chairs":["Williams, Matthew","Taylor, Brian"],"committee_members":[],"year":2024,"date_issued":"2024-01-30","date_published":"2024-01-30","updated_at":"2026-11-27T04:59:06Z","subjects":["Biomedical Engineering","neuroprosthetics","implanted intramuscular electrodes","myoelectric control","prosthesis limbs","prosthesis socket fit"],"languages":["English"],"rights":["unrestricted","This thesis or dissertation is protected by copyright: some rights reserved. It is licensed for use under a Creative Commons license. Specific terms and permissions are available from this document's record in the OhioLINK ETD Center."],"rights_urls":[],"identifier_entries":[],"person_identifier_entries":[]},"links":{"outbound_url":"http://rave.ohiolink.edu/etdc/view?acc_num=case1704293346229248","outbound_label":"Repository record","outbound_source":"dc:identifier","record_url":"http://rave.ohiolink.edu/etdc/view?acc_num=case1704293346229248","record_label":"Repository record","record_kind":"record","record_source":"dc:identifier","access_status":null,"embargo_until":null,"access_note":null},"oai_pmh_record":{"url":"https://union.ndltd.org/oai-pmh/?verb=GetRecord&metadataPrefix=oai_dc&identifier=oai%3Aunion.ndltd.org%3Arepo%3Aohiolink%3Aoai%3Aetd.ohiolink.edu%3Acase1704293346229248","prefix":"oai_dc"},"source_record":{"url":"https://etd.ohiolink.edu/acprod/odb_etd/ws/oai/oai?verb=GetRecord&metadataPrefix=oai_etdms&identifier=oai%3Aetd.ohiolink.edu%3Acase1704293346229248","prefix":"oai_etdms"},"metadata_groups":[{"id":"people","label":"People","entries":[{"key":"dc:contributor.advisor","label":"Advisor","values":["Tyler, Dustin"]},{"key":"dc:contributor.committeechair","label":"Committee Chair","values":["Williams, Matthew","Taylor, Brian"]},{"key":"dc:creator","label":"Author","values":["Dsouza, Karina Theresa"]}]},{"id":"academic_context","label":"Academic Context","entries":[{"key":"dc:date","label":"Dc Date","values":["2024-01-30"]},{"key":"dc:publisher","label":"Institution","values":["Case Western Reserve University School of Graduate Studies / OhioLINK"]},{"key":"dc:type","label":"Dc Type","values":["Electronic Thesis or Dissertation"]},{"key":"thesis:degree_discipline","label":"Discipline","values":["Biomedical Engineering"]},{"key":"thesis:degree_level","label":"Degree Level","values":["masters"]},{"key":"thesis:degree_name","label":"Degree Name","values":["Master of Sciences (Engineering)"]},{"key":"thesis:institution_name","label":"Thesis Institution Name","values":["Case Western Reserve University School of Graduate Studies"]}]},{"id":"subjects_keywords","label":"Subjects and Keywords","entries":[{"key":"dc:subject","label":"Dc Subject","values":["Biomedical Engineering","neuroprosthetics","implanted intramuscular electrodes","myoelectric control","prosthesis limbs","prosthesis socket fit"]}]},{"id":"language_rights","label":"Language and Rights","entries":[{"key":"dc:language","label":"Dc Language","values":["English"]},{"key":"dc:rights","label":"Dc Rights","values":["unrestricted","This thesis or dissertation is protected by copyright: some rights reserved. It is licensed for use under a Creative Commons license. Specific terms and permissions are available from this document's record in the OhioLINK ETD Center."]}]},{"id":"identifiers","label":"Identifiers","entries":[{"key":"dc:identifier","label":"Identifier","values":["http://rave.ohiolink.edu/etdc/view?acc_num=case1704293346229248"]}]},{"id":"additional","label":"Additional Metadata","entries":[{"key":"dc:description","label":"Description","values":["Myo-electric prosthesis hands (myo-prostheses) intend to restore independence in individuals with upper-limb loss, however, 44% of consumers discontinue device use due to unreliable function [1]. Literature suggests that residual-limb changes may affect the myo-control algorithm’s ability to decode user-intended movement from skin-surface recorded electromyography (EMG). Chronically implanted, intramuscular EMG (ciEMG) improves signal integrity, but the extent to which ciEMG inputs mitigate myo-control degradation is unknown [2]–[4]. This study investigates how 3 Degree-of-Freedom (DoF), Simultaneous and Proportional (SP) myo-control performance and ciEMG hand-posture patterns are affected by loose socket fit and added residual-limb load. Load did degrade myo-control performance during a posture matching test, but only with loose socket wear. Examining postural ciEMG patterns revealed that Extensor Digitorum (ED) activity was greater during loose socket wear, as well. Correcting socket fit may improve postural ciEMG pattern consistency during different residual-limb loads, thereby maintaining reliable myo-prosthesis function."]},{"key":"dc:format","label":"Dc Format","values":["application/pdf","p.44","1.65 MB"]},{"key":"dc:title","label":"Title","values":["The Effects of Socket Fit and Residual-Limb Load on 3-Dof, Simultaneous and Proportional, Myo-Prosthesis Hand Control and Chronically Implanted, Intramuscular EMG Patterns"]}]}],"canonical_facts":{"dc:contributor.advisor":["Tyler, Dustin"],"dc:contributor.committeechair":["Williams, Matthew","Taylor, Brian"],"dc:creator":["Dsouza, Karina Theresa"],"dc:date":["2024-01-30"],"dc:description":["Myo-electric prosthesis hands (myo-prostheses) intend to restore independence in individuals with upper-limb loss, however, 44% of consumers discontinue device use due to unreliable function [1]. Literature suggests that residual-limb changes may affect the myo-control algorithm’s ability to decode user-intended movement from skin-surface recorded electromyography (EMG). Chronically implanted, intramuscular EMG (ciEMG) improves signal integrity, but the extent to which ciEMG inputs mitigate myo-control degradation is unknown [2]–[4]. This study investigates how 3 Degree-of-Freedom (DoF), Simultaneous and Proportional (SP) myo-control performance and ciEMG hand-posture patterns are affected by loose socket fit and added residual-limb load. Load did degrade myo-control performance during a posture matching test, but only with loose socket wear. Examining postural ciEMG patterns revealed that Extensor Digitorum (ED) activity was greater during loose socket wear, as well. Correcting socket fit may improve postural ciEMG pattern consistency during different residual-limb loads, thereby maintaining reliable myo-prosthesis function."],"dc:format":["application/pdf","p.44","1.65 MB"],"dc:identifier":["http://rave.ohiolink.edu/etdc/view?acc_num=case1704293346229248"],"dc:language":["English"],"dc:publisher":["Case Western Reserve University School of Graduate Studies / OhioLINK"],"dc:rights":["unrestricted","This thesis or dissertation is protected by copyright: some rights reserved. It is licensed for use under a Creative Commons license. Specific terms and permissions are available from this document's record in the OhioLINK ETD Center."],"dc:subject":["Biomedical Engineering","neuroprosthetics","implanted intramuscular electrodes","myoelectric control","prosthesis limbs","prosthesis socket fit"],"dc:title":["The Effects of Socket Fit and Residual-Limb Load on 3-Dof, Simultaneous and Proportional, Myo-Prosthesis Hand Control and Chronically Implanted, Intramuscular EMG Patterns"],"dc:type":["Electronic Thesis or Dissertation"],"thesis:degree_discipline":["Biomedical Engineering"],"thesis:degree_level":["masters"],"thesis:degree_name":["Master of Sciences (Engineering)"],"thesis:institution_name":["Case Western Reserve University School of Graduate Studies"]},"updated_at":"2026-11-27T04:59:06Z"}