{"id":{"repo_id":"ohiolink","oai_identifier":"oai:etd.ohiolink.edu:case1720022424347684"},"canonical_url":"https://search.ndltd.org/etd/ohiolink/oai:etd.ohiolink.edu:case1720022424347684","repository":{"repo_id":"ohiolink","name":"OhioLINK","base_url":"https://etd.ohiolink.edu/acprod/odb_etd/ws/oai/oai"},"display":{"title":"Real-Time Kinematic Modeling of Magnetically-Actuated MRI-Guided Robotic Intravascular Cardiac Catheter","abstract":"This thesis presents modeling and control of the magnetically-actuated MRI-guided robotic intravascular cardiac catheter. The robotic catheters are embedded with current carrying coil actuators. The actuator contains one axial coil, and two orthogonal side coils. The catheter is steered by the magnetic moments generated on a set of current carrying actuation coils mounted on the catheter body by the static magnetic feld of an MRI scanner. The kinematics of the robotic catheter is modeled using the Cosserat rod theory as a series of fexible and rigid segments. In the proposed model, the equilibrium confguration of the catheter under external loads, including actuation moments, and the contact force at the tip of the catheter is calculated. A Jacobian-based iterative inverse kinematics and the open-loop control are presented and experimentally validated using hardware setup. This work also presents the model under tip contact position constraint of the MRI-actuated robotic catheter. In this model, the initial value problem parameter derivatives and tip contact force are calculated analytically. The analytical formulation of the relationship between the contact force and the actuation inputs under the tip contact position constraint, described by the contact force Jacobian, is presented.The proposed kinematic model of the MRI-actuated robotic catheter shows very good performance in performing the complex trajectories with high reproducibility during the experiments. These methods deliver consistent performance showing good results in terms of reproducibility and accuracy. This success sets a strong foundation for the potential implementation of a closed-loop control system with real-time MRI guidance information. The simulation based benchmark tests of the proposed analytical method for the calculation of the contact force Jacobian of the MRI-actuated robotic catheter using the Cosserat rod model and the analytical derivations of the IVP parameter derivatives of the MRI-actuated robotic catheter under tip contact force show that the analytically and numerically calculated contact force Jacobians are comparable. The average computational speed of the analytical method for solving the contact force Jacobian is significantly faster than the numerical method.","abstract_html":"This thesis presents modeling and control of the magnetically-actuated MRI-guided robotic intravascular cardiac catheter. The robotic catheters are embedded with current carrying coil actuators. The actuator contains one axial coil, and two orthogonal side coils. The catheter is steered by the magnetic moments generated on a set of current carrying actuation coils mounted on the catheter body by the static magnetic feld of an MRI scanner. The kinematics of the robotic catheter is modeled using the Cosserat rod theory as a series of fexible and rigid segments. In the proposed model, the equilibrium confguration of the catheter under external loads, including actuation moments, and the contact force at the tip of the catheter is calculated. A Jacobian-based iterative inverse kinematics and the open-loop control are presented and experimentally validated using hardware setup. This work also presents the model under tip contact position constraint of the MRI-actuated robotic catheter. In this model, the initial value problem parameter derivatives and tip contact force are calculated analytically. The analytical formulation of the relationship between the contact force and the actuation inputs under the tip contact position constraint, described by the contact force Jacobian, is presented.The proposed kinematic model of the MRI-actuated robotic catheter shows very good performance in performing the complex trajectories with high reproducibility during the experiments. These methods deliver consistent performance showing good results in terms of reproducibility and accuracy. This success sets a strong foundation for the potential implementation of a closed-loop control system with real-time MRI guidance information. The simulation based benchmark tests of the proposed analytical method for the calculation of the contact force Jacobian of the MRI-actuated robotic catheter using the Cosserat rod model and the analytical derivations of the IVP parameter derivatives of the MRI-actuated robotic catheter under tip contact force show that the analytically and numerically calculated contact force Jacobians are comparable. The average computational speed of the analytical method for solving the contact force Jacobian is significantly faster than the numerical method.","abstract_has_math":false,"creators":["Itsarachaiyot, Yuttana"],"institution":"Case Western Reserve University School of Graduate Studies","degree_name":"Doctor of Philosophy","degree_level":"doctoral","degree_discipline":"EECS - Electrical Engineering","degree_department":null,"school":null,"contributors":[],"advisors":["Cavusoglu, Murat Cenk"],"committee_chairs":[],"committee_members":[],"year":2024,"date_issued":"2024-08-22","date_published":"2024-08-22","updated_at":"2026-11-27T04:59:23Z","subjects":["Robotics","Electrical Engineering","robotic catheter","kinematic modeling","magnetically-actuated robotic catheter"],"languages":["English"],"rights":["unrestricted","This thesis or dissertation is protected by copyright: all rights reserved. It may not be copied or redistributed beyond the terms of applicable copyright laws."],"rights_urls":[],"identifier_entries":[],"person_identifier_entries":[]},"links":{"outbound_url":"http://rave.ohiolink.edu/etdc/view?acc_num=case1720022424347684","outbound_label":"Repository record","outbound_source":"dc:identifier","record_url":"http://rave.ohiolink.edu/etdc/view?acc_num=case1720022424347684","record_label":"Repository record","record_kind":"record","record_source":"dc:identifier","access_status":null,"embargo_until":null,"access_note":null},"oai_pmh_record":{"url":"https://union.ndltd.org/oai-pmh/?verb=GetRecord&metadataPrefix=oai_dc&identifier=oai%3Aunion.ndltd.org%3Arepo%3Aohiolink%3Aoai%3Aetd.ohiolink.edu%3Acase1720022424347684","prefix":"oai_dc"},"source_record":{"url":"https://etd.ohiolink.edu/acprod/odb_etd/ws/oai/oai?verb=GetRecord&metadataPrefix=oai_etdms&identifier=oai%3Aetd.ohiolink.edu%3Acase1720022424347684","prefix":"oai_etdms"},"metadata_groups":[{"id":"people","label":"People","entries":[{"key":"dc:contributor.advisor","label":"Advisor","values":["Cavusoglu, Murat Cenk"]},{"key":"dc:creator","label":"Author","values":["Itsarachaiyot, Yuttana"]}]},{"id":"academic_context","label":"Academic Context","entries":[{"key":"dc:date","label":"Dc Date","values":["2024-08-22"]},{"key":"dc:publisher","label":"Institution","values":["Case Western Reserve University School of Graduate Studies / OhioLINK"]},{"key":"dc:type","label":"Dc Type","values":["Electronic Thesis or Dissertation"]},{"key":"thesis:degree_discipline","label":"Discipline","values":["EECS - Electrical Engineering"]},{"key":"thesis:degree_level","label":"Degree Level","values":["doctoral"]},{"key":"thesis:degree_name","label":"Degree Name","values":["Doctor of Philosophy"]},{"key":"thesis:institution_name","label":"Thesis Institution Name","values":["Case Western Reserve University School of Graduate Studies"]}]},{"id":"subjects_keywords","label":"Subjects and Keywords","entries":[{"key":"dc:subject","label":"Dc Subject","values":["Robotics","Electrical Engineering","robotic catheter","kinematic modeling","magnetically-actuated robotic catheter"]}]},{"id":"language_rights","label":"Language and Rights","entries":[{"key":"dc:language","label":"Dc Language","values":["English"]},{"key":"dc:rights","label":"Dc Rights","values":["unrestricted","This thesis or dissertation is protected by copyright: all rights reserved. It may not be copied or redistributed beyond the terms of applicable copyright laws."]}]},{"id":"identifiers","label":"Identifiers","entries":[{"key":"dc:identifier","label":"Identifier","values":["http://rave.ohiolink.edu/etdc/view?acc_num=case1720022424347684"]}]},{"id":"additional","label":"Additional Metadata","entries":[{"key":"dc:description","label":"Description","values":["This thesis presents modeling and control of the magnetically-actuated MRI-guided robotic intravascular cardiac catheter. The robotic catheters are embedded with current carrying coil actuators. The actuator contains one axial coil, and two orthogonal side coils. The catheter is steered by the magnetic moments generated on a set of current carrying actuation coils mounted on the catheter body by the static magnetic feld of an MRI scanner. The kinematics of the robotic catheter is modeled using the Cosserat rod theory as a series of fexible and rigid segments. In the proposed model, the equilibrium confguration of the catheter under external loads, including actuation moments, and the contact force at the tip of the catheter is calculated. A Jacobian-based iterative inverse kinematics and the open-loop control are presented and experimentally validated using hardware setup. This work also presents the model under tip contact position constraint of the MRI-actuated robotic catheter. In this model, the initial value problem parameter derivatives and tip contact force are calculated analytically. The analytical formulation of the relationship between the contact force and the actuation inputs under the tip contact position constraint, described by the contact force Jacobian, is presented.The proposed kinematic model of the MRI-actuated robotic catheter shows very good performance in performing the complex trajectories with high reproducibility during the experiments. These methods deliver consistent performance showing good results in terms of reproducibility and accuracy. This success sets a strong foundation for the potential implementation of a closed-loop control system with real-time MRI guidance information. The simulation based benchmark tests of the proposed analytical method for the calculation of the contact force Jacobian of the MRI-actuated robotic catheter using the Cosserat rod model and the analytical derivations of the IVP parameter derivatives of the MRI-actuated robotic catheter under tip contact force show that the analytically and numerically calculated contact force Jacobians are comparable. The average computational speed of the analytical method for solving the contact force Jacobian is significantly faster than the numerical method."]},{"key":"dc:format","label":"Dc Format","values":["application/pdf","p.76","8.32 MB"]},{"key":"dc:title","label":"Title","values":["Real-Time Kinematic Modeling of Magnetically-Actuated MRI-Guided Robotic Intravascular Cardiac Catheter"]}]}],"canonical_facts":{"dc:contributor.advisor":["Cavusoglu, Murat Cenk"],"dc:creator":["Itsarachaiyot, Yuttana"],"dc:date":["2024-08-22"],"dc:description":["This thesis presents modeling and control of the magnetically-actuated MRI-guided robotic intravascular cardiac catheter. The robotic catheters are embedded with current carrying coil actuators. The actuator contains one axial coil, and two orthogonal side coils. The catheter is steered by the magnetic moments generated on a set of current carrying actuation coils mounted on the catheter body by the static magnetic feld of an MRI scanner. The kinematics of the robotic catheter is modeled using the Cosserat rod theory as a series of fexible and rigid segments. In the proposed model, the equilibrium confguration of the catheter under external loads, including actuation moments, and the contact force at the tip of the catheter is calculated. A Jacobian-based iterative inverse kinematics and the open-loop control are presented and experimentally validated using hardware setup. This work also presents the model under tip contact position constraint of the MRI-actuated robotic catheter. In this model, the initial value problem parameter derivatives and tip contact force are calculated analytically. The analytical formulation of the relationship between the contact force and the actuation inputs under the tip contact position constraint, described by the contact force Jacobian, is presented.The proposed kinematic model of the MRI-actuated robotic catheter shows very good performance in performing the complex trajectories with high reproducibility during the experiments. These methods deliver consistent performance showing good results in terms of reproducibility and accuracy. This success sets a strong foundation for the potential implementation of a closed-loop control system with real-time MRI guidance information. The simulation based benchmark tests of the proposed analytical method for the calculation of the contact force Jacobian of the MRI-actuated robotic catheter using the Cosserat rod model and the analytical derivations of the IVP parameter derivatives of the MRI-actuated robotic catheter under tip contact force show that the analytically and numerically calculated contact force Jacobians are comparable. The average computational speed of the analytical method for solving the contact force Jacobian is significantly faster than the numerical method."],"dc:format":["application/pdf","p.76","8.32 MB"],"dc:identifier":["http://rave.ohiolink.edu/etdc/view?acc_num=case1720022424347684"],"dc:language":["English"],"dc:publisher":["Case Western Reserve University School of Graduate Studies / OhioLINK"],"dc:rights":["unrestricted","This thesis or dissertation is protected by copyright: all rights reserved. It may not be copied or redistributed beyond the terms of applicable copyright laws."],"dc:subject":["Robotics","Electrical Engineering","robotic catheter","kinematic modeling","magnetically-actuated robotic catheter"],"dc:title":["Real-Time Kinematic Modeling of Magnetically-Actuated MRI-Guided Robotic Intravascular Cardiac Catheter"],"dc:type":["Electronic Thesis or Dissertation"],"thesis:degree_discipline":["EECS - Electrical Engineering"],"thesis:degree_level":["doctoral"],"thesis:degree_name":["Doctor of Philosophy"],"thesis:institution_name":["Case Western Reserve University School of Graduate Studies"]},"updated_at":"2026-11-27T04:59:23Z"}