{"id":{"repo_id":"ohiolink","oai_identifier":"oai:etd.ohiolink.edu:mco1222876677"},"canonical_url":"https://search.ndltd.org/etd/ohiolink/oai:etd.ohiolink.edu:mco1222876677","repository":{"repo_id":"ohiolink","name":"OhioLINK","base_url":"https://etd.ohiolink.edu/acprod/odb_etd/ws/oai/oai"},"display":{"title":"Development and Implementation of a GafChromic EBT In-Vivo Personal Dosimetry System","abstract":"RadioChromic EBT film was used to create an in-vivo dosimetry system. Generalproperties of chromic film were investigated. Statistical analysis of previous diode doseswas used to create dose ranges encompassing 95% of clinical applications. Calibrationcurves were generated for 6 MV, 10 MV, 18 MV, 6 MeV, 9 MeV, and 12 MeV energiesand fitted to polynomial equations of the third power. Calibration equations were used tocreate clinical dose calculation forms. Statistical analysis was performed to measure dosereproducibility and error. Skin doses and Dmax doses were measured and comparedagainst accepted values. Dmax measurements were shown to have an average percent errorof 4.47%. Skin measurements had average percent errors of 5.67% for 6 MV, 3.40% for18 MV, 2.18% for 6 MeV, and 5.86% for 12 MeV. Film handling, storage, and scanninginstructions were created to make the dosimetry system user friendly with reproducibleoutcomes.","abstract_html":"RadioChromic EBT film was used to create an in-vivo dosimetry system. Generalproperties of chromic film were investigated. Statistical analysis of previous diode doseswas used to create dose ranges encompassing 95% of clinical applications. Calibrationcurves were generated for 6 MV, 10 MV, 18 MV, 6 MeV, 9 MeV, and 12 MeV energiesand fitted to polynomial equations of the third power. Calibration equations were used tocreate clinical dose calculation forms. Statistical analysis was performed to measure dosereproducibility and error. Skin doses and Dmax doses were measured and comparedagainst accepted values. Dmax measurements were shown to have an average percent errorof 4.47%. Skin measurements had average percent errors of 5.67% for 6 MV, 3.40% for18 MV, 2.18% for 6 MeV, and 5.86% for 12 MeV. Film handling, storage, and scanninginstructions were created to make the dosimetry system user friendly with reproducibleoutcomes.","abstract_has_math":false,"creators":["Bugno, Jacob R."],"institution":"University of Toledo Health Science Campus","degree_name":"Master of Science in Biomedical Sciences (MSBS)","degree_level":"masters","degree_discipline":"College of Graduate Studies","degree_department":null,"school":null,"contributors":[],"advisors":[],"committee_chairs":["Parsai, E. Ishmael"],"committee_members":[],"year":2008,"date_issued":"2008-11-12","date_published":"2008-11-12","updated_at":"2026-11-27T04:17:19Z","subjects":["Physics","Radiology","radiochromic","dosimetry","medical physics","film"],"languages":["English"],"rights":["unrestricted","This thesis or dissertation is protected by copyright: some rights reserved. It is licensed for use under a Creative Commons license. Specific terms and permissions are available from this document's record in the OhioLINK ETD Center."],"rights_urls":[],"identifier_entries":[],"person_identifier_entries":[]},"links":{"outbound_url":"http://rave.ohiolink.edu/etdc/view?acc_num=mco1222876677","outbound_label":"Repository record","outbound_source":"dc:identifier","record_url":"http://rave.ohiolink.edu/etdc/view?acc_num=mco1222876677","record_label":"Repository record","record_kind":"record","record_source":"dc:identifier","access_status":null,"embargo_until":null,"access_note":null},"oai_pmh_record":{"url":"https://union.ndltd.org/oai-pmh/?verb=GetRecord&metadataPrefix=oai_dc&identifier=oai%3Aunion.ndltd.org%3Arepo%3Aohiolink%3Aoai%3Aetd.ohiolink.edu%3Amco1222876677","prefix":"oai_dc"},"source_record":{"url":"https://etd.ohiolink.edu/acprod/odb_etd/ws/oai/oai?verb=GetRecord&metadataPrefix=oai_etdms&identifier=oai%3Aetd.ohiolink.edu%3Amco1222876677","prefix":"oai_etdms"},"metadata_groups":[{"id":"people","label":"People","entries":[{"key":"dc:contributor.committeechair","label":"Committee Chair","values":["Parsai, E. Ishmael"]},{"key":"dc:creator","label":"Author","values":["Bugno, Jacob R."]}]},{"id":"academic_context","label":"Academic Context","entries":[{"key":"dc:date","label":"Dc Date","values":["2008-11-12"]},{"key":"dc:publisher","label":"Institution","values":["University of Toledo Health Science Campus / OhioLINK"]},{"key":"dc:type","label":"Dc Type","values":["Electronic Thesis or Dissertation"]},{"key":"thesis:degree_discipline","label":"Discipline","values":["College of Graduate Studies"]},{"key":"thesis:degree_level","label":"Degree Level","values":["masters"]},{"key":"thesis:degree_name","label":"Degree Name","values":["Master of Science in Biomedical Sciences (MSBS)"]},{"key":"thesis:institution_name","label":"Thesis Institution Name","values":["University of Toledo Health Science Campus"]}]},{"id":"subjects_keywords","label":"Subjects and Keywords","entries":[{"key":"dc:subject","label":"Dc Subject","values":["Physics","Radiology","radiochromic","dosimetry","medical physics","film"]}]},{"id":"language_rights","label":"Language and Rights","entries":[{"key":"dc:language","label":"Dc Language","values":["English"]},{"key":"dc:rights","label":"Dc Rights","values":["unrestricted","This thesis or dissertation is protected by copyright: some rights reserved. It is licensed for use under a Creative Commons license. Specific terms and permissions are available from this document's record in the OhioLINK ETD Center."]}]},{"id":"identifiers","label":"Identifiers","entries":[{"key":"dc:identifier","label":"Identifier","values":["http://rave.ohiolink.edu/etdc/view?acc_num=mco1222876677"]}]},{"id":"additional","label":"Additional Metadata","entries":[{"key":"dc:description","label":"Description","values":["RadioChromic EBT film was used to create an in-vivo dosimetry system. Generalproperties of chromic film were investigated. Statistical analysis of previous diode doseswas used to create dose ranges encompassing 95% of clinical applications. Calibrationcurves were generated for 6 MV, 10 MV, 18 MV, 6 MeV, 9 MeV, and 12 MeV energiesand fitted to polynomial equations of the third power. Calibration equations were used tocreate clinical dose calculation forms. Statistical analysis was performed to measure dosereproducibility and error. Skin doses and Dmax doses were measured and comparedagainst accepted values. Dmax measurements were shown to have an average percent errorof 4.47%. Skin measurements had average percent errors of 5.67% for 6 MV, 3.40% for18 MV, 2.18% for 6 MeV, and 5.86% for 12 MeV. Film handling, storage, and scanninginstructions were created to make the dosimetry system user friendly with reproducibleoutcomes."]},{"key":"dc:format","label":"Dc Format","values":["application/pdf","p.60","908.6 KB"]},{"key":"dc:title","label":"Title","values":["Development and Implementation of a GafChromic EBT In-Vivo Personal Dosimetry System"]}]}],"canonical_facts":{"dc:contributor.committeechair":["Parsai, E. Ishmael"],"dc:creator":["Bugno, Jacob R."],"dc:date":["2008-11-12"],"dc:description":["RadioChromic EBT film was used to create an in-vivo dosimetry system. Generalproperties of chromic film were investigated. Statistical analysis of previous diode doseswas used to create dose ranges encompassing 95% of clinical applications. Calibrationcurves were generated for 6 MV, 10 MV, 18 MV, 6 MeV, 9 MeV, and 12 MeV energiesand fitted to polynomial equations of the third power. Calibration equations were used tocreate clinical dose calculation forms. Statistical analysis was performed to measure dosereproducibility and error. Skin doses and Dmax doses were measured and comparedagainst accepted values. Dmax measurements were shown to have an average percent errorof 4.47%. Skin measurements had average percent errors of 5.67% for 6 MV, 3.40% for18 MV, 2.18% for 6 MeV, and 5.86% for 12 MeV. Film handling, storage, and scanninginstructions were created to make the dosimetry system user friendly with reproducibleoutcomes."],"dc:format":["application/pdf","p.60","908.6 KB"],"dc:identifier":["http://rave.ohiolink.edu/etdc/view?acc_num=mco1222876677"],"dc:language":["English"],"dc:publisher":["University of Toledo Health Science Campus / OhioLINK"],"dc:rights":["unrestricted","This thesis or dissertation is protected by copyright: some rights reserved. It is licensed for use under a Creative Commons license. Specific terms and permissions are available from this document's record in the OhioLINK ETD Center."],"dc:subject":["Physics","Radiology","radiochromic","dosimetry","medical physics","film"],"dc:title":["Development and Implementation of a GafChromic EBT In-Vivo Personal Dosimetry System"],"dc:type":["Electronic Thesis or Dissertation"],"thesis:degree_discipline":["College of Graduate Studies"],"thesis:degree_level":["masters"],"thesis:degree_name":["Master of Science in Biomedical Sciences (MSBS)"],"thesis:institution_name":["University of Toledo Health Science Campus"]},"updated_at":"2026-11-27T04:17:19Z"}