{"id":{"repo_id":"ohiolink","oai_identifier":"oai:etd.ohiolink.edu:mco1438985667"},"canonical_url":"https://search.ndltd.org/etd/ohiolink/oai:etd.ohiolink.edu:mco1438985667","repository":{"repo_id":"ohiolink","name":"OhioLINK","base_url":"https://etd.ohiolink.edu/acprod/odb_etd/ws/oai/oai"},"display":{"title":"Evaluation of Phantoms Used in Image Quality Performance Testing of Dental Cone Beam Computed Tomography Systems","abstract":"Cone beam computed tomography (CBCT) units dedicated for dental and maxillofacial imaging have gained widespread use in recent years. So far there are no standardized testing methods to evaluate the image quality or the dose on these units. Although the vendors commonly provide an image quality phantom with the machine, the procedures, image quality parameters evaluated, and performance criteria are limited and specific to the vendors. The goal of this study is to evaluate the available phantoms as a testing tool for image quality assessment of dental CBCT units. In addition, the optimal diameter and features of a QA phantom for dental CBCT testing was assessed.Two commercially available phantoms were evaluated to assess the adequacy of each for use in the standardized testing procedures. These included ACR CT phantom (Gammex 464) and CATPHAN (Phantom Laboratory). Additionally, a prototype dental CT phantom (CIRS Inc.) was evaluated. Scans were made on three different machines; i-CAT Classic (Imaging Sciences International), ILUMA (IMTEC/3M) and GALILEOS Comfort (Sirona). A CT scan of the three phantoms were also performed with a conventional CT scanner (Toshiba Aquilion 16) to verify the pixel values. The image quality parameters that were evaluated included: image noise, image uniformity, pixel value accuracy, pixel value linearity, contrast scale, and high contrast spatial resolution. There is not one phantom evaluated that provided superior results for all image quality tests. The prototype phantom was adequate for all tests with the exception of the high contrast resolution test whereas the ACR phantom and the CATPHAN due to larger diameter relative to the prototype demonstrated higher noise and non-uniformity, pixel value inaccuracy, and lower contrast scale because of more beam hardening artifact occurring with the large diameter. The optimal diameter for a phantom specialized in dental CBCT testing was found to be between 16 and 17 cm in diameter.","abstract_html":"Cone beam computed tomography (CBCT) units dedicated for dental and maxillofacial imaging have gained widespread use in recent years. So far there are no standardized testing methods to evaluate the image quality or the dose on these units. Although the vendors commonly provide an image quality phantom with the machine, the procedures, image quality parameters evaluated, and performance criteria are limited and specific to the vendors. The goal of this study is to evaluate the available phantoms as a testing tool for image quality assessment of dental CBCT units. In addition, the optimal diameter and features of a QA phantom for dental CBCT testing was assessed.Two commercially available phantoms were evaluated to assess the adequacy of each for use in the standardized testing procedures. These included ACR CT phantom (Gammex 464) and CATPHAN (Phantom Laboratory). Additionally, a prototype dental CT phantom (CIRS Inc.) was evaluated. Scans were made on three different machines; i-CAT Classic (Imaging Sciences International), ILUMA (IMTEC/3M) and GALILEOS Comfort (Sirona). A CT scan of the three phantoms were also performed with a conventional CT scanner (Toshiba Aquilion 16) to verify the pixel values. The image quality parameters that were evaluated included: image noise, image uniformity, pixel value accuracy, pixel value linearity, contrast scale, and high contrast spatial resolution. There is not one phantom evaluated that provided superior results for all image quality tests. The prototype phantom was adequate for all tests with the exception of the high contrast resolution test whereas the ACR phantom and the CATPHAN due to larger diameter relative to the prototype demonstrated higher noise and non-uniformity, pixel value inaccuracy, and lower contrast scale because of more beam hardening artifact occurring with the large diameter. The optimal diameter for a phantom specialized in dental CBCT testing was found to be between 16 and 17 cm in diameter.","abstract_has_math":false,"creators":["Alahmad, Haitham N."],"institution":"University of Toledo Health Science Campus","degree_name":"Master of Science in Biomedical Sciences (MSBS)","degree_level":"masters","degree_discipline":"Biomedical Sciences (Diagnostic Radiology)","degree_department":null,"school":null,"contributors":[],"advisors":[],"committee_chairs":["Parsai, E. Ishmael"],"committee_members":[],"year":2015,"date_issued":"2015","date_published":"2015","updated_at":"2026-11-27T04:40:34Z","subjects":["Medical Imaging","Radiation","Dental CBCT, Cone beam CT, Image Quality Testing,Computed Tomography"],"languages":["English"],"rights":["unrestricted","This thesis or dissertation is protected by copyright: all rights reserved. It may not be copied or redistributed beyond the terms of applicable copyright laws."],"rights_urls":[],"identifier_entries":[],"person_identifier_entries":[]},"links":{"outbound_url":"http://rave.ohiolink.edu/etdc/view?acc_num=mco1438985667","outbound_label":"Repository record","outbound_source":"dc:identifier","record_url":"http://rave.ohiolink.edu/etdc/view?acc_num=mco1438985667","record_label":"Repository record","record_kind":"record","record_source":"dc:identifier","access_status":null,"embargo_until":null,"access_note":null},"oai_pmh_record":{"url":"https://union.ndltd.org/oai-pmh/?verb=GetRecord&metadataPrefix=oai_dc&identifier=oai%3Aunion.ndltd.org%3Arepo%3Aohiolink%3Aoai%3Aetd.ohiolink.edu%3Amco1438985667","prefix":"oai_dc"},"source_record":{"url":"https://etd.ohiolink.edu/acprod/odb_etd/ws/oai/oai?verb=GetRecord&metadataPrefix=oai_etdms&identifier=oai%3Aetd.ohiolink.edu%3Amco1438985667","prefix":"oai_etdms"},"metadata_groups":[{"id":"people","label":"People","entries":[{"key":"dc:contributor.committeechair","label":"Committee Chair","values":["Parsai, E. Ishmael"]},{"key":"dc:creator","label":"Author","values":["Alahmad, Haitham N."]}]},{"id":"academic_context","label":"Academic Context","entries":[{"key":"dc:date","label":"Dc Date","values":["2015"]},{"key":"dc:publisher","label":"Institution","values":["University of Toledo Health Science Campus / OhioLINK"]},{"key":"dc:type","label":"Dc Type","values":["Electronic Thesis or Dissertation"]},{"key":"thesis:degree_discipline","label":"Discipline","values":["Biomedical Sciences (Diagnostic Radiology)"]},{"key":"thesis:degree_level","label":"Degree Level","values":["masters"]},{"key":"thesis:degree_name","label":"Degree Name","values":["Master of Science in Biomedical Sciences (MSBS)"]},{"key":"thesis:institution_name","label":"Thesis Institution Name","values":["University of Toledo Health Science Campus"]}]},{"id":"subjects_keywords","label":"Subjects and Keywords","entries":[{"key":"dc:subject","label":"Dc Subject","values":["Medical Imaging","Radiation","Dental CBCT, Cone beam CT, Image Quality Testing,Computed Tomography"]}]},{"id":"language_rights","label":"Language and Rights","entries":[{"key":"dc:language","label":"Dc Language","values":["English"]},{"key":"dc:rights","label":"Dc Rights","values":["unrestricted","This thesis or dissertation is protected by copyright: all rights reserved. It may not be copied or redistributed beyond the terms of applicable copyright laws."]}]},{"id":"identifiers","label":"Identifiers","entries":[{"key":"dc:identifier","label":"Identifier","values":["http://rave.ohiolink.edu/etdc/view?acc_num=mco1438985667"]}]},{"id":"additional","label":"Additional Metadata","entries":[{"key":"dc:description","label":"Description","values":["Cone beam computed tomography (CBCT) units dedicated for dental and maxillofacial imaging have gained widespread use in recent years. So far there are no standardized testing methods to evaluate the image quality or the dose on these units. Although the vendors commonly provide an image quality phantom with the machine, the procedures, image quality parameters evaluated, and performance criteria are limited and specific to the vendors. The goal of this study is to evaluate the available phantoms as a testing tool for image quality assessment of dental CBCT units. In addition, the optimal diameter and features of a QA phantom for dental CBCT testing was assessed.Two commercially available phantoms were evaluated to assess the adequacy of each for use in the standardized testing procedures. These included ACR CT phantom (Gammex 464) and CATPHAN (Phantom Laboratory). Additionally, a prototype dental CT phantom (CIRS Inc.) was evaluated. Scans were made on three different machines; i-CAT Classic (Imaging Sciences International), ILUMA (IMTEC/3M) and GALILEOS Comfort (Sirona). A CT scan of the three phantoms were also performed with a conventional CT scanner (Toshiba Aquilion 16) to verify the pixel values. The image quality parameters that were evaluated included: image noise, image uniformity, pixel value accuracy, pixel value linearity, contrast scale, and high contrast spatial resolution. There is not one phantom evaluated that provided superior results for all image quality tests. The prototype phantom was adequate for all tests with the exception of the high contrast resolution test whereas the ACR phantom and the CATPHAN due to larger diameter relative to the prototype demonstrated higher noise and non-uniformity, pixel value inaccuracy, and lower contrast scale because of more beam hardening artifact occurring with the large diameter. The optimal diameter for a phantom specialized in dental CBCT testing was found to be between 16 and 17 cm in diameter."]},{"key":"dc:format","label":"Dc Format","values":["application/pdf","p.71","1.59 MB"]},{"key":"dc:title","label":"Title","values":["Evaluation of Phantoms Used in Image Quality Performance Testing of Dental Cone Beam Computed Tomography Systems"]}]}],"canonical_facts":{"dc:contributor.committeechair":["Parsai, E. Ishmael"],"dc:creator":["Alahmad, Haitham N."],"dc:date":["2015"],"dc:description":["Cone beam computed tomography (CBCT) units dedicated for dental and maxillofacial imaging have gained widespread use in recent years. So far there are no standardized testing methods to evaluate the image quality or the dose on these units. Although the vendors commonly provide an image quality phantom with the machine, the procedures, image quality parameters evaluated, and performance criteria are limited and specific to the vendors. The goal of this study is to evaluate the available phantoms as a testing tool for image quality assessment of dental CBCT units. In addition, the optimal diameter and features of a QA phantom for dental CBCT testing was assessed.Two commercially available phantoms were evaluated to assess the adequacy of each for use in the standardized testing procedures. These included ACR CT phantom (Gammex 464) and CATPHAN (Phantom Laboratory). Additionally, a prototype dental CT phantom (CIRS Inc.) was evaluated. Scans were made on three different machines; i-CAT Classic (Imaging Sciences International), ILUMA (IMTEC/3M) and GALILEOS Comfort (Sirona). A CT scan of the three phantoms were also performed with a conventional CT scanner (Toshiba Aquilion 16) to verify the pixel values. The image quality parameters that were evaluated included: image noise, image uniformity, pixel value accuracy, pixel value linearity, contrast scale, and high contrast spatial resolution. There is not one phantom evaluated that provided superior results for all image quality tests. The prototype phantom was adequate for all tests with the exception of the high contrast resolution test whereas the ACR phantom and the CATPHAN due to larger diameter relative to the prototype demonstrated higher noise and non-uniformity, pixel value inaccuracy, and lower contrast scale because of more beam hardening artifact occurring with the large diameter. The optimal diameter for a phantom specialized in dental CBCT testing was found to be between 16 and 17 cm in diameter."],"dc:format":["application/pdf","p.71","1.59 MB"],"dc:identifier":["http://rave.ohiolink.edu/etdc/view?acc_num=mco1438985667"],"dc:language":["English"],"dc:publisher":["University of Toledo Health Science Campus / OhioLINK"],"dc:rights":["unrestricted","This thesis or dissertation is protected by copyright: all rights reserved. It may not be copied or redistributed beyond the terms of applicable copyright laws."],"dc:subject":["Medical Imaging","Radiation","Dental CBCT, Cone beam CT, Image Quality Testing,Computed Tomography"],"dc:title":["Evaluation of Phantoms Used in Image Quality Performance Testing of Dental Cone Beam Computed Tomography Systems"],"dc:type":["Electronic Thesis or Dissertation"],"thesis:degree_discipline":["Biomedical Sciences (Diagnostic Radiology)"],"thesis:degree_level":["masters"],"thesis:degree_name":["Master of Science in Biomedical Sciences (MSBS)"],"thesis:institution_name":["University of Toledo Health Science Campus"]},"updated_at":"2026-11-27T04:40:34Z"}