{"id":{"repo_id":"ohiolink","oai_identifier":"oai:etd.ohiolink.edu:ucin1100536748"},"canonical_url":"https://search.ndltd.org/etd/ohiolink/oai:etd.ohiolink.edu:ucin1100536748","repository":{"repo_id":"ohiolink","name":"OhioLINK","base_url":"https://etd.ohiolink.edu/acprod/odb_etd/ws/oai/oai"},"display":{"title":"EXPERIMENTAL CHARACTERIZATION AND ACTIVE CONTROL SIMULATION OF THE ACOUSTIC NOISE RESPONSE OF A HIGH-FIELD, HIGH RATE MRI SCANNER","abstract":"In this project, the time-frequency characteristics of the harmful acoustic noise response produced by a 4-Tesla MRI scanner operated with echo planer imaging (EPI) sequence are experimentally quantified, and the effects of structural acoustics and gradient pulse excitations on the emitted audible sound field are determined, such that feasible noise control measures can be developed. By employing a high-speed dynamic data acquisition system, the acoustic noise emissions at the patient ear and mouth locations were measured and simultaneously recorded with the electrical current signals of applied gradient pulses. Two forms of gradient waveforms (impulsive and operating types) were studied. The impulsive test results yielded the magnetic-structural acoustic transfer functions that can be applied to predict the sound pressure response spectrum of any gradient pulse excitation. Also, the measured transfer functions, when analyzed in conjunction with the operating data, gave further insights into the fundamental nature of the MRI acoustic noise generation and transmission processes. The overall sound pressure levels emitted by this MRI scanner are found to be in the range of 115-125 decibels (dB). The most intense acoustic noise energy concentrations lie within the audible frequency range of 1-4 kHz. There are mainly three types of sound pressure response observed in the EPI sequence, namely harmonic, non-harmonic, and broadband spectrum. The effect of each gradient axis on the acoustic noise emitted differs from each other with the frequency-encoding Y-gradient being the most dominant one and produces generally odd number harmonics and non-harmonics. Also, the operating sound spectrum predicted from the magnetic-structural acoustic transfer functions compares very well with the direct measurements. In addition, the primary frequency content of the audible sound matches with the range where most normal speech communications occur. The strong harmonic content of the sound field and the quantifiable relationships between the acoustic noise and gradient pulse signals make this problem an ideal candidate for active noise control (ANC) application. The ANC simulation performed show a potential of reducing the peak resonance response by about 20 dB. The findings are currently being analyzed further for the purpose of developing an actual working ANC prototype.","abstract_html":"In this project, the time-frequency characteristics of the harmful acoustic noise response produced by a 4-Tesla MRI scanner operated with echo planer imaging (EPI) sequence are experimentally quantified, and the effects of structural acoustics and gradient pulse excitations on the emitted audible sound field are determined, such that feasible noise control measures can be developed. By employing a high-speed dynamic data acquisition system, the acoustic noise emissions at the patient ear and mouth locations were measured and simultaneously recorded with the electrical current signals of applied gradient pulses. Two forms of gradient waveforms (impulsive and operating types) were studied. The impulsive test results yielded the magnetic-structural acoustic transfer functions that can be applied to predict the sound pressure response spectrum of any gradient pulse excitation. Also, the measured transfer functions, when analyzed in conjunction with the operating data, gave further insights into the fundamental nature of the MRI acoustic noise generation and transmission processes. The overall sound pressure levels emitted by this MRI scanner are found to be in the range of 115-125 decibels (dB). The most intense acoustic noise energy concentrations lie within the audible frequency range of 1-4 kHz. There are mainly three types of sound pressure response observed in the EPI sequence, namely harmonic, non-harmonic, and broadband spectrum. The effect of each gradient axis on the acoustic noise emitted differs from each other with the frequency-encoding Y-gradient being the most dominant one and produces generally odd number harmonics and non-harmonics. Also, the operating sound spectrum predicted from the magnetic-structural acoustic transfer functions compares very well with the direct measurements. In addition, the primary frequency content of the audible sound matches with the range where most normal speech communications occur. The strong harmonic content of the sound field and the quantifiable relationships between the acoustic noise and gradient pulse signals make this problem an ideal candidate for active noise control (ANC) application. The ANC simulation performed show a potential of reducing the peak resonance response by about 20 dB. The findings are currently being analyzed further for the purpose of developing an actual working ANC prototype.","abstract_has_math":false,"creators":["MORE, SHASHIKANT R."],"institution":"University of Cincinnati","degree_name":"MS","degree_level":"masters","degree_discipline":"Engineering : Mechanical Engineering","degree_department":null,"school":null,"contributors":[],"advisors":["Lim, Dr. Teik"],"committee_chairs":[],"committee_members":[],"year":2004,"date_issued":"2004","date_published":"2004","updated_at":"2026-11-27T04:36:58Z","subjects":["Engineering, Mechanical","MRI","acoustic","noise","sound","audible","time-frequency","active control","ANC"],"languages":["English"],"rights":["unrestricted","This thesis or dissertation is protected by copyright: all rights reserved. It may not be copied or redistributed beyond the terms of applicable copyright laws."],"rights_urls":[],"identifier_entries":[],"person_identifier_entries":[]},"links":{"outbound_url":"http://rave.ohiolink.edu/etdc/view?acc_num=ucin1100536748","outbound_label":"Repository record","outbound_source":"dc:identifier","record_url":"http://rave.ohiolink.edu/etdc/view?acc_num=ucin1100536748","record_label":"Repository record","record_kind":"record","record_source":"dc:identifier","access_status":null,"embargo_until":null,"access_note":null},"oai_pmh_record":{"url":"https://union.ndltd.org/oai-pmh/?verb=GetRecord&metadataPrefix=oai_dc&identifier=oai%3Aunion.ndltd.org%3Arepo%3Aohiolink%3Aoai%3Aetd.ohiolink.edu%3Aucin1100536748","prefix":"oai_dc"},"source_record":{"url":"https://etd.ohiolink.edu/acprod/odb_etd/ws/oai/oai?verb=GetRecord&metadataPrefix=oai_etdms&identifier=oai%3Aetd.ohiolink.edu%3Aucin1100536748","prefix":"oai_etdms"},"metadata_groups":[{"id":"people","label":"People","entries":[{"key":"dc:contributor.advisor","label":"Advisor","values":["Lim, Dr. Teik"]},{"key":"dc:creator","label":"Author","values":["MORE, SHASHIKANT R."]}]},{"id":"academic_context","label":"Academic Context","entries":[{"key":"dc:date","label":"Dc Date","values":["2004"]},{"key":"dc:publisher","label":"Institution","values":["University of Cincinnati / OhioLINK"]},{"key":"dc:type","label":"Dc Type","values":["Electronic Thesis or Dissertation"]},{"key":"thesis:degree_discipline","label":"Discipline","values":["Engineering : Mechanical Engineering"]},{"key":"thesis:degree_level","label":"Degree Level","values":["masters"]},{"key":"thesis:degree_name","label":"Degree Name","values":["MS"]},{"key":"thesis:institution_name","label":"Thesis Institution Name","values":["University of Cincinnati"]}]},{"id":"subjects_keywords","label":"Subjects and Keywords","entries":[{"key":"dc:subject","label":"Dc Subject","values":["Engineering, Mechanical","MRI","acoustic","noise","sound","audible","time-frequency","active control","ANC"]}]},{"id":"language_rights","label":"Language and Rights","entries":[{"key":"dc:language","label":"Dc Language","values":["English"]},{"key":"dc:rights","label":"Dc Rights","values":["unrestricted","This thesis or dissertation is protected by copyright: all rights reserved. It may not be copied or redistributed beyond the terms of applicable copyright laws."]}]},{"id":"identifiers","label":"Identifiers","entries":[{"key":"dc:identifier","label":"Identifier","values":["http://rave.ohiolink.edu/etdc/view?acc_num=ucin1100536748"]}]},{"id":"additional","label":"Additional Metadata","entries":[{"key":"dc:description","label":"Description","values":["In this project, the time-frequency characteristics of the harmful acoustic noise response produced by a 4-Tesla MRI scanner operated with echo planer imaging (EPI) sequence are experimentally quantified, and the effects of structural acoustics and gradient pulse excitations on the emitted audible sound field are determined, such that feasible noise control measures can be developed. By employing a high-speed dynamic data acquisition system, the acoustic noise emissions at the patient ear and mouth locations were measured and simultaneously recorded with the electrical current signals of applied gradient pulses. Two forms of gradient waveforms (impulsive and operating types) were studied. The impulsive test results yielded the magnetic-structural acoustic transfer functions that can be applied to predict the sound pressure response spectrum of any gradient pulse excitation. Also, the measured transfer functions, when analyzed in conjunction with the operating data, gave further insights into the fundamental nature of the MRI acoustic noise generation and transmission processes. The overall sound pressure levels emitted by this MRI scanner are found to be in the range of 115-125 decibels (dB). The most intense acoustic noise energy concentrations lie within the audible frequency range of 1-4 kHz. There are mainly three types of sound pressure response observed in the EPI sequence, namely harmonic, non-harmonic, and broadband spectrum. The effect of each gradient axis on the acoustic noise emitted differs from each other with the frequency-encoding Y-gradient being the most dominant one and produces generally odd number harmonics and non-harmonics. Also, the operating sound spectrum predicted from the magnetic-structural acoustic transfer functions compares very well with the direct measurements. In addition, the primary frequency content of the audible sound matches with the range where most normal speech communications occur. The strong harmonic content of the sound field and the quantifiable relationships between the acoustic noise and gradient pulse signals make this problem an ideal candidate for active noise control (ANC) application. The ANC simulation performed show a potential of reducing the peak resonance response by about 20 dB. The findings are currently being analyzed further for the purpose of developing an actual working ANC prototype."]},{"key":"dc:format","label":"Dc Format","values":["application/pdf","p.127","4.59 MB"]},{"key":"dc:title","label":"Title","values":["EXPERIMENTAL CHARACTERIZATION AND ACTIVE CONTROL SIMULATION OF THE ACOUSTIC NOISE RESPONSE OF A HIGH-FIELD, HIGH RATE MRI SCANNER"]}]}],"canonical_facts":{"dc:contributor.advisor":["Lim, Dr. Teik"],"dc:creator":["MORE, SHASHIKANT R."],"dc:date":["2004"],"dc:description":["In this project, the time-frequency characteristics of the harmful acoustic noise response produced by a 4-Tesla MRI scanner operated with echo planer imaging (EPI) sequence are experimentally quantified, and the effects of structural acoustics and gradient pulse excitations on the emitted audible sound field are determined, such that feasible noise control measures can be developed. By employing a high-speed dynamic data acquisition system, the acoustic noise emissions at the patient ear and mouth locations were measured and simultaneously recorded with the electrical current signals of applied gradient pulses. Two forms of gradient waveforms (impulsive and operating types) were studied. The impulsive test results yielded the magnetic-structural acoustic transfer functions that can be applied to predict the sound pressure response spectrum of any gradient pulse excitation. Also, the measured transfer functions, when analyzed in conjunction with the operating data, gave further insights into the fundamental nature of the MRI acoustic noise generation and transmission processes. The overall sound pressure levels emitted by this MRI scanner are found to be in the range of 115-125 decibels (dB). The most intense acoustic noise energy concentrations lie within the audible frequency range of 1-4 kHz. There are mainly three types of sound pressure response observed in the EPI sequence, namely harmonic, non-harmonic, and broadband spectrum. The effect of each gradient axis on the acoustic noise emitted differs from each other with the frequency-encoding Y-gradient being the most dominant one and produces generally odd number harmonics and non-harmonics. Also, the operating sound spectrum predicted from the magnetic-structural acoustic transfer functions compares very well with the direct measurements. In addition, the primary frequency content of the audible sound matches with the range where most normal speech communications occur. The strong harmonic content of the sound field and the quantifiable relationships between the acoustic noise and gradient pulse signals make this problem an ideal candidate for active noise control (ANC) application. The ANC simulation performed show a potential of reducing the peak resonance response by about 20 dB. The findings are currently being analyzed further for the purpose of developing an actual working ANC prototype."],"dc:format":["application/pdf","p.127","4.59 MB"],"dc:identifier":["http://rave.ohiolink.edu/etdc/view?acc_num=ucin1100536748"],"dc:language":["English"],"dc:publisher":["University of Cincinnati / OhioLINK"],"dc:rights":["unrestricted","This thesis or dissertation is protected by copyright: all rights reserved. It may not be copied or redistributed beyond the terms of applicable copyright laws."],"dc:subject":["Engineering, Mechanical","MRI","acoustic","noise","sound","audible","time-frequency","active control","ANC"],"dc:title":["EXPERIMENTAL CHARACTERIZATION AND ACTIVE CONTROL SIMULATION OF THE ACOUSTIC NOISE RESPONSE OF A HIGH-FIELD, HIGH RATE MRI SCANNER"],"dc:type":["Electronic Thesis or Dissertation"],"thesis:degree_discipline":["Engineering : Mechanical Engineering"],"thesis:degree_level":["masters"],"thesis:degree_name":["MS"],"thesis:institution_name":["University of Cincinnati"]},"updated_at":"2026-11-27T04:36:58Z"}