{"id":{"repo_id":"ohiolink","oai_identifier":"oai:etd.ohiolink.edu:ucin1407410171"},"canonical_url":"https://search.ndltd.org/etd/ohiolink/oai:etd.ohiolink.edu:ucin1407410171","repository":{"repo_id":"ohiolink","name":"OhioLINK","base_url":"https://etd.ohiolink.edu/acprod/odb_etd/ws/oai/oai"},"display":{"title":"Validation of LeadCare (c) II Portable Blood Lead Analyzer","abstract":"This study compares a LeadCare © II blood lead analyzer with a LeadCare © I blood lead analyzer and determines the correlation of the analytical results. Subjects recruited for a blood lead level monitoring study during a building demolition/renovation project participated in this study. A total of 12 eligible study participants were recruited. Two background samples (background 1 and background 2) were collected via venipuncture. Background 1 sample results from the LeadCare © I analyzer ranged from < 1.4 to 4.1 µg/dL. The background 1 sample mean was 1.75 µg/dL with a SD ± 0.80. Results for the background 2 sample ranged from < 1.4 to 2.7 with a mean of 1.53 µg/dL and a SD ± 0.39. The lower reportable range of the LeadCare © II is 3.3 µg/dL and 1.4 µg/dL for the LeadCare © I. Most of the sample results reported by the LeadCare © I were less than 3.3 µg/dL making a comparison with the LeadCare © II impractical.","abstract_html":"This study compares a LeadCare © II blood lead analyzer with a LeadCare © I blood lead analyzer and determines the correlation of the analytical results. Subjects recruited for a blood lead level monitoring study during a building demolition/renovation project participated in this study. A total of 12 eligible study participants were recruited. Two background samples (background 1 and background 2) were collected via venipuncture. Background 1 sample results from the LeadCare © I analyzer ranged from &lt; 1.4 to 4.1 µg/dL. The background 1 sample mean was 1.75 µg/dL with a SD ± 0.80. Results for the background 2 sample ranged from &lt; 1.4 to 2.7 with a mean of 1.53 µg/dL and a SD ± 0.39. The lower reportable range of the LeadCare © II is 3.3 µg/dL and 1.4 µg/dL for the LeadCare © I. Most of the sample results reported by the LeadCare © I were less than 3.3 µg/dL making a comparison with the LeadCare © II impractical.","abstract_has_math":false,"creators":["Kuyat, Shelly"],"institution":"University of Cincinnati","degree_name":"MS","degree_level":"masters","degree_discipline":"Medicine: Industrial Hygiene (Environmental Health)","degree_department":null,"school":null,"contributors":[],"advisors":[],"committee_chairs":["Talaska, Glenn"],"committee_members":[],"year":2014,"date_issued":"2014-10-13","date_published":"2014-10-13","updated_at":"2026-11-27T04:39:55Z","subjects":["Occupational Safety","LeadCare","Portable blood lead analyzer"],"languages":["English"],"rights":["unrestricted","This thesis or dissertation is protected by copyright: all rights reserved. It may not be copied or redistributed beyond the terms of applicable copyright laws."],"rights_urls":[],"identifier_entries":[],"person_identifier_entries":[]},"links":{"outbound_url":"http://rave.ohiolink.edu/etdc/view?acc_num=ucin1407410171","outbound_label":"Repository record","outbound_source":"dc:identifier","record_url":"http://rave.ohiolink.edu/etdc/view?acc_num=ucin1407410171","record_label":"Repository record","record_kind":"record","record_source":"dc:identifier","access_status":null,"embargo_until":null,"access_note":null},"oai_pmh_record":{"url":"https://union.ndltd.org/oai-pmh/?verb=GetRecord&metadataPrefix=oai_dc&identifier=oai%3Aunion.ndltd.org%3Arepo%3Aohiolink%3Aoai%3Aetd.ohiolink.edu%3Aucin1407410171","prefix":"oai_dc"},"source_record":{"url":"https://etd.ohiolink.edu/acprod/odb_etd/ws/oai/oai?verb=GetRecord&metadataPrefix=oai_etdms&identifier=oai%3Aetd.ohiolink.edu%3Aucin1407410171","prefix":"oai_etdms"},"metadata_groups":[{"id":"people","label":"People","entries":[{"key":"dc:contributor.committeechair","label":"Committee Chair","values":["Talaska, Glenn"]},{"key":"dc:creator","label":"Author","values":["Kuyat, Shelly"]}]},{"id":"academic_context","label":"Academic Context","entries":[{"key":"dc:date","label":"Dc Date","values":["2014-10-13"]},{"key":"dc:publisher","label":"Institution","values":["University of Cincinnati / OhioLINK"]},{"key":"dc:type","label":"Dc Type","values":["Electronic Thesis or Dissertation"]},{"key":"thesis:degree_discipline","label":"Discipline","values":["Medicine: Industrial Hygiene (Environmental Health)"]},{"key":"thesis:degree_level","label":"Degree Level","values":["masters"]},{"key":"thesis:degree_name","label":"Degree Name","values":["MS"]},{"key":"thesis:institution_name","label":"Thesis Institution Name","values":["University of Cincinnati"]}]},{"id":"subjects_keywords","label":"Subjects and Keywords","entries":[{"key":"dc:subject","label":"Dc Subject","values":["Occupational Safety","LeadCare","Portable blood lead analyzer"]}]},{"id":"language_rights","label":"Language and Rights","entries":[{"key":"dc:language","label":"Dc Language","values":["English"]},{"key":"dc:rights","label":"Dc Rights","values":["unrestricted","This thesis or dissertation is protected by copyright: all rights reserved. It may not be copied or redistributed beyond the terms of applicable copyright laws."]}]},{"id":"identifiers","label":"Identifiers","entries":[{"key":"dc:identifier","label":"Identifier","values":["http://rave.ohiolink.edu/etdc/view?acc_num=ucin1407410171"]}]},{"id":"additional","label":"Additional Metadata","entries":[{"key":"dc:description","label":"Description","values":["This study compares a LeadCare © II blood lead analyzer with a LeadCare © I blood lead analyzer and determines the correlation of the analytical results. Subjects recruited for a blood lead level monitoring study during a building demolition/renovation project participated in this study. A total of 12 eligible study participants were recruited. Two background samples (background 1 and background 2) were collected via venipuncture. Background 1 sample results from the LeadCare © I analyzer ranged from < 1.4 to 4.1 µg/dL. The background 1 sample mean was 1.75 µg/dL with a SD ± 0.80. Results for the background 2 sample ranged from < 1.4 to 2.7 with a mean of 1.53 µg/dL and a SD ± 0.39. The lower reportable range of the LeadCare © II is 3.3 µg/dL and 1.4 µg/dL for the LeadCare © I. Most of the sample results reported by the LeadCare © I were less than 3.3 µg/dL making a comparison with the LeadCare © II impractical."]},{"key":"dc:format","label":"Dc Format","values":["application/pdf","p.32","2.37 MB"]},{"key":"dc:title","label":"Title","values":["Validation of LeadCare (c) II Portable Blood Lead Analyzer"]}]}],"canonical_facts":{"dc:contributor.committeechair":["Talaska, Glenn"],"dc:creator":["Kuyat, Shelly"],"dc:date":["2014-10-13"],"dc:description":["This study compares a LeadCare © II blood lead analyzer with a LeadCare © I blood lead analyzer and determines the correlation of the analytical results. Subjects recruited for a blood lead level monitoring study during a building demolition/renovation project participated in this study. A total of 12 eligible study participants were recruited. Two background samples (background 1 and background 2) were collected via venipuncture. Background 1 sample results from the LeadCare © I analyzer ranged from < 1.4 to 4.1 µg/dL. The background 1 sample mean was 1.75 µg/dL with a SD ± 0.80. Results for the background 2 sample ranged from < 1.4 to 2.7 with a mean of 1.53 µg/dL and a SD ± 0.39. The lower reportable range of the LeadCare © II is 3.3 µg/dL and 1.4 µg/dL for the LeadCare © I. Most of the sample results reported by the LeadCare © I were less than 3.3 µg/dL making a comparison with the LeadCare © II impractical."],"dc:format":["application/pdf","p.32","2.37 MB"],"dc:identifier":["http://rave.ohiolink.edu/etdc/view?acc_num=ucin1407410171"],"dc:language":["English"],"dc:publisher":["University of Cincinnati / OhioLINK"],"dc:rights":["unrestricted","This thesis or dissertation is protected by copyright: all rights reserved. It may not be copied or redistributed beyond the terms of applicable copyright laws."],"dc:subject":["Occupational Safety","LeadCare","Portable blood lead analyzer"],"dc:title":["Validation of LeadCare (c) II Portable Blood Lead Analyzer"],"dc:type":["Electronic Thesis or Dissertation"],"thesis:degree_discipline":["Medicine: Industrial Hygiene (Environmental Health)"],"thesis:degree_level":["masters"],"thesis:degree_name":["MS"],"thesis:institution_name":["University of Cincinnati"]},"updated_at":"2026-11-27T04:39:55Z"}