{"id":{"repo_id":"ohiolink","oai_identifier":"oai:etd.ohiolink.edu:ucin1428064830"},"canonical_url":"https://search.ndltd.org/etd/ohiolink/oai:etd.ohiolink.edu:ucin1428064830","repository":{"repo_id":"ohiolink","name":"OhioLINK","base_url":"https://etd.ohiolink.edu/acprod/odb_etd/ws/oai/oai"},"display":{"title":"The Frequency of Physiologic Monitor Alarms in a Children’s Hospital","abstract":"Background and Objective: In most hospitals, alarms from physiologic monitors are frequent, and the vast majority of alarms are not actionable. Concerns about alarm fatigue prompted The Joint Commission to urge hospitals to take immediate action to improve alarm management in 2013. Understanding the types of alarms that occur most frequently will guide hospitals striving to reduce alarm frequencies. The objective of this study was to evaluate the burden of alarm frequency and to describe the patterns of physiologic monitor alarms at a single, free-standing children’s hospital including the rates of alarms and the most common causes of alarms.Methods: This was a cross-sectional study of physiologic monitor alarms at a freestanding children’s hospital. Data were collected from all inpatient units over 7 consecutive days in August 2014 for each physiologic monitor alarm using a central alarm collection system. Alarms were classified by alarm cause (e.g. low blood oxygen level), and sensor (e.g. pulse oximetry probe). Procedural, radiology, rehabilitation, and psychiatry units were excluded. Each unit census was determined by four daily measurements during the study period from the electronic health record.Results: There were a total of 220813 alarms over the study. The most common cause of alarms during the week was technical alarms, or alarms generated for issues with the monitoring equipment (e.g. probe off or low battery), followed by clinical alarms for low blood oxygen level and high heart rate. Pulse oximetry probes were the most common sensor to cause alarms. Alarm rates varied significantly by unit and by cause.Conclusion: Alarm rates at a children’s hospital were quite high. Technical alarms, a type of alarm unrelated to a change in patient physiologic status, were the most common cause of alarm. Strategies targeted at reducing technical alarms from physiologic monitors may help reduce overall alarm frequencies.","abstract_html":"Background and Objective: In most hospitals, alarms from physiologic monitors are frequent, and the vast majority of alarms are not actionable. Concerns about alarm fatigue prompted The Joint Commission to urge hospitals to take immediate action to improve alarm management in 2013. Understanding the types of alarms that occur most frequently will guide hospitals striving to reduce alarm frequencies. The objective of this study was to evaluate the burden of alarm frequency and to describe the patterns of physiologic monitor alarms at a single, free-standing children’s hospital including the rates of alarms and the most common causes of alarms.Methods: This was a cross-sectional study of physiologic monitor alarms at a freestanding children’s hospital. Data were collected from all inpatient units over 7 consecutive days in August 2014 for each physiologic monitor alarm using a central alarm collection system. Alarms were classified by alarm cause (e.g. low blood oxygen level), and sensor (e.g. pulse oximetry probe). Procedural, radiology, rehabilitation, and psychiatry units were excluded. Each unit census was determined by four daily measurements during the study period from the electronic health record.Results: There were a total of 220813 alarms over the study. The most common cause of alarms during the week was technical alarms, or alarms generated for issues with the monitoring equipment (e.g. probe off or low battery), followed by clinical alarms for low blood oxygen level and high heart rate. Pulse oximetry probes were the most common sensor to cause alarms. Alarm rates varied significantly by unit and by cause.Conclusion: Alarm rates at a children’s hospital were quite high. Technical alarms, a type of alarm unrelated to a change in patient physiologic status, were the most common cause of alarm. Strategies targeted at reducing technical alarms from physiologic monitors may help reduce overall alarm frequencies.","abstract_has_math":false,"creators":["Schondelmeyer, Amanda C., M.D."],"institution":"University of Cincinnati","degree_name":"MS","degree_level":"masters","degree_discipline":"Medicine: Clinical and Translational Research","degree_department":null,"school":null,"contributors":[],"advisors":[],"committee_chairs":["Haynes, Erin"],"committee_members":[],"year":2015,"date_issued":"2015-09-01","date_published":"2015-09-01","updated_at":"2026-11-27T04:40:14Z","subjects":["Surgery","Phsyiologic monitor alarm","Alarm fatigue","Pulse oximetry"],"languages":["English"],"rights":["unrestricted","This thesis or dissertation is protected by copyright: all rights reserved. It may not be copied or redistributed beyond the terms of applicable copyright laws."],"rights_urls":[],"identifier_entries":[],"person_identifier_entries":[]},"links":{"outbound_url":"http://rave.ohiolink.edu/etdc/view?acc_num=ucin1428064830","outbound_label":"Repository record","outbound_source":"dc:identifier","record_url":"http://rave.ohiolink.edu/etdc/view?acc_num=ucin1428064830","record_label":"Repository record","record_kind":"record","record_source":"dc:identifier","access_status":null,"embargo_until":null,"access_note":null},"oai_pmh_record":{"url":"https://union.ndltd.org/oai-pmh/?verb=GetRecord&metadataPrefix=oai_dc&identifier=oai%3Aunion.ndltd.org%3Arepo%3Aohiolink%3Aoai%3Aetd.ohiolink.edu%3Aucin1428064830","prefix":"oai_dc"},"source_record":{"url":"https://etd.ohiolink.edu/acprod/odb_etd/ws/oai/oai?verb=GetRecord&metadataPrefix=oai_etdms&identifier=oai%3Aetd.ohiolink.edu%3Aucin1428064830","prefix":"oai_etdms"},"metadata_groups":[{"id":"people","label":"People","entries":[{"key":"dc:contributor.committeechair","label":"Committee Chair","values":["Haynes, Erin"]},{"key":"dc:creator","label":"Author","values":["Schondelmeyer, Amanda C., M.D."]}]},{"id":"academic_context","label":"Academic Context","entries":[{"key":"dc:date","label":"Dc Date","values":["2015-09-01"]},{"key":"dc:publisher","label":"Institution","values":["University of Cincinnati / OhioLINK"]},{"key":"dc:type","label":"Dc Type","values":["Electronic Thesis or Dissertation"]},{"key":"thesis:degree_discipline","label":"Discipline","values":["Medicine: Clinical and Translational Research"]},{"key":"thesis:degree_level","label":"Degree Level","values":["masters"]},{"key":"thesis:degree_name","label":"Degree Name","values":["MS"]},{"key":"thesis:institution_name","label":"Thesis Institution Name","values":["University of Cincinnati"]}]},{"id":"subjects_keywords","label":"Subjects and Keywords","entries":[{"key":"dc:subject","label":"Dc Subject","values":["Surgery","Phsyiologic monitor alarm","Alarm fatigue","Pulse oximetry"]}]},{"id":"language_rights","label":"Language and Rights","entries":[{"key":"dc:language","label":"Dc Language","values":["English"]},{"key":"dc:rights","label":"Dc Rights","values":["unrestricted","This thesis or dissertation is protected by copyright: all rights reserved. It may not be copied or redistributed beyond the terms of applicable copyright laws."]}]},{"id":"identifiers","label":"Identifiers","entries":[{"key":"dc:identifier","label":"Identifier","values":["http://rave.ohiolink.edu/etdc/view?acc_num=ucin1428064830"]}]},{"id":"additional","label":"Additional Metadata","entries":[{"key":"dc:description","label":"Description","values":["Background and Objective: In most hospitals, alarms from physiologic monitors are frequent, and the vast majority of alarms are not actionable. Concerns about alarm fatigue prompted The Joint Commission to urge hospitals to take immediate action to improve alarm management in 2013. Understanding the types of alarms that occur most frequently will guide hospitals striving to reduce alarm frequencies. The objective of this study was to evaluate the burden of alarm frequency and to describe the patterns of physiologic monitor alarms at a single, free-standing children’s hospital including the rates of alarms and the most common causes of alarms.Methods: This was a cross-sectional study of physiologic monitor alarms at a freestanding children’s hospital. Data were collected from all inpatient units over 7 consecutive days in August 2014 for each physiologic monitor alarm using a central alarm collection system. Alarms were classified by alarm cause (e.g. low blood oxygen level), and sensor (e.g. pulse oximetry probe). Procedural, radiology, rehabilitation, and psychiatry units were excluded. Each unit census was determined by four daily measurements during the study period from the electronic health record.Results: There were a total of 220813 alarms over the study. The most common cause of alarms during the week was technical alarms, or alarms generated for issues with the monitoring equipment (e.g. probe off or low battery), followed by clinical alarms for low blood oxygen level and high heart rate. Pulse oximetry probes were the most common sensor to cause alarms. Alarm rates varied significantly by unit and by cause.Conclusion: Alarm rates at a children’s hospital were quite high. Technical alarms, a type of alarm unrelated to a change in patient physiologic status, were the most common cause of alarm. Strategies targeted at reducing technical alarms from physiologic monitors may help reduce overall alarm frequencies."]},{"key":"dc:format","label":"Dc Format","values":["application/pdf","p.18","1.85 MB"]},{"key":"dc:title","label":"Title","values":["The Frequency of Physiologic Monitor Alarms in a Children’s Hospital"]}]}],"canonical_facts":{"dc:contributor.committeechair":["Haynes, Erin"],"dc:creator":["Schondelmeyer, Amanda C., M.D."],"dc:date":["2015-09-01"],"dc:description":["Background and Objective: In most hospitals, alarms from physiologic monitors are frequent, and the vast majority of alarms are not actionable. Concerns about alarm fatigue prompted The Joint Commission to urge hospitals to take immediate action to improve alarm management in 2013. Understanding the types of alarms that occur most frequently will guide hospitals striving to reduce alarm frequencies. The objective of this study was to evaluate the burden of alarm frequency and to describe the patterns of physiologic monitor alarms at a single, free-standing children’s hospital including the rates of alarms and the most common causes of alarms.Methods: This was a cross-sectional study of physiologic monitor alarms at a freestanding children’s hospital. Data were collected from all inpatient units over 7 consecutive days in August 2014 for each physiologic monitor alarm using a central alarm collection system. Alarms were classified by alarm cause (e.g. low blood oxygen level), and sensor (e.g. pulse oximetry probe). Procedural, radiology, rehabilitation, and psychiatry units were excluded. Each unit census was determined by four daily measurements during the study period from the electronic health record.Results: There were a total of 220813 alarms over the study. The most common cause of alarms during the week was technical alarms, or alarms generated for issues with the monitoring equipment (e.g. probe off or low battery), followed by clinical alarms for low blood oxygen level and high heart rate. Pulse oximetry probes were the most common sensor to cause alarms. Alarm rates varied significantly by unit and by cause.Conclusion: Alarm rates at a children’s hospital were quite high. Technical alarms, a type of alarm unrelated to a change in patient physiologic status, were the most common cause of alarm. Strategies targeted at reducing technical alarms from physiologic monitors may help reduce overall alarm frequencies."],"dc:format":["application/pdf","p.18","1.85 MB"],"dc:identifier":["http://rave.ohiolink.edu/etdc/view?acc_num=ucin1428064830"],"dc:language":["English"],"dc:publisher":["University of Cincinnati / OhioLINK"],"dc:rights":["unrestricted","This thesis or dissertation is protected by copyright: all rights reserved. It may not be copied or redistributed beyond the terms of applicable copyright laws."],"dc:subject":["Surgery","Phsyiologic monitor alarm","Alarm fatigue","Pulse oximetry"],"dc:title":["The Frequency of Physiologic Monitor Alarms in a Children’s Hospital"],"dc:type":["Electronic Thesis or Dissertation"],"thesis:degree_discipline":["Medicine: Clinical and Translational Research"],"thesis:degree_level":["masters"],"thesis:degree_name":["MS"],"thesis:institution_name":["University of Cincinnati"]},"updated_at":"2026-11-27T04:40:14Z"}