{"id":{"repo_id":"ohiolink","oai_identifier":"oai:etd.ohiolink.edu:wright1187369612"},"canonical_url":"https://search.ndltd.org/etd/ohiolink/oai:etd.ohiolink.edu:wright1187369612","repository":{"repo_id":"ohiolink","name":"OhioLINK","base_url":"https://etd.ohiolink.edu/acprod/odb_etd/ws/oai/oai"},"display":{"title":"Comparative Investigation of Dosimetric Tools in IMRT","abstract":"Verification of the isodose distribution delivered to the planned tumor volume is required for quality assurance in the clinic. Verification in this context means ensuring that the actual dose delivered to a volume matches the dose that a computerized planning system intended that volume to receive. The computerized planning is performed by a Computerized Medical Systems (CMS) Intensity Modulated Radiation Therapy (IMRT) plan. Two methods of measuring the delivered dose are film dosimetry and diode dosimetry. This project will compare the relative isodose distribution profiles of film dosimetry and MapCheck (Sun Nuclear Corporation, a diode-based dosimetry system) to the relative isodose distribution profile of the CMS IMRT plan. Two IMRT plans were used that were intended to deliver 7500centigray to the prostate. These plans were then performed while measuring the isodose distributions using the above two techniques. The two techniques were compared on the basis of time efficiency and accuracy.","abstract_html":"Verification of the isodose distribution delivered to the planned tumor volume is required for quality assurance in the clinic. Verification in this context means ensuring that the actual dose delivered to a volume matches the dose that a computerized planning system intended that volume to receive. The computerized planning is performed by a Computerized Medical Systems (CMS) Intensity Modulated Radiation Therapy (IMRT) plan. Two methods of measuring the delivered dose are film dosimetry and diode dosimetry. This project will compare the relative isodose distribution profiles of film dosimetry and MapCheck (Sun Nuclear Corporation, a diode-based dosimetry system) to the relative isodose distribution profile of the CMS IMRT plan. Two IMRT plans were used that were intended to deliver 7500centigray to the prostate. These plans were then performed while measuring the isodose distributions using the above two techniques. The two techniques were compared on the basis of time efficiency and accuracy.","abstract_has_math":false,"creators":["Garcia, Aaron Nicholas"],"institution":"Wright State University","degree_name":"Master of Science (MS)","degree_level":"masters","degree_discipline":"Physics","degree_department":null,"school":null,"contributors":["Foy, Brent"],"advisors":[],"committee_chairs":[],"committee_members":[],"year":2007,"date_issued":"2007-08-20","date_published":"2007-08-20","updated_at":"2026-11-26T04:31:43Z","subjects":["Physics, Radiation","IMRT"],"languages":["English"],"rights":["unrestricted","This thesis or dissertation is protected by copyright: all rights reserved. It may not be copied or redistributed beyond the terms of applicable copyright laws."],"rights_urls":[],"identifier_entries":[],"person_identifier_entries":[]},"links":{"outbound_url":"http://rave.ohiolink.edu/etdc/view?acc_num=wright1187369612","outbound_label":"Repository record","outbound_source":"dc:identifier","record_url":"http://rave.ohiolink.edu/etdc/view?acc_num=wright1187369612","record_label":"Repository record","record_kind":"record","record_source":"dc:identifier","access_status":null,"embargo_until":null,"access_note":null},"oai_pmh_record":{"url":"https://union.ndltd.org/oai-pmh/?verb=GetRecord&metadataPrefix=oai_dc&identifier=oai%3Aunion.ndltd.org%3Arepo%3Aohiolink%3Aoai%3Aetd.ohiolink.edu%3Awright1187369612","prefix":"oai_dc"},"source_record":{"url":"https://etd.ohiolink.edu/acprod/odb_etd/ws/oai/oai?verb=GetRecord&metadataPrefix=oai_etdms&identifier=oai%3Aetd.ohiolink.edu%3Awright1187369612","prefix":"oai_etdms"},"metadata_groups":[{"id":"people","label":"People","entries":[{"key":"dc:contributor","label":"Contributor","values":["Foy, Brent"]},{"key":"dc:creator","label":"Author","values":["Garcia, Aaron Nicholas"]}]},{"id":"academic_context","label":"Academic Context","entries":[{"key":"dc:date","label":"Dc Date","values":["2007-08-20"]},{"key":"dc:publisher","label":"Institution","values":["Wright State University / OhioLINK"]},{"key":"dc:type","label":"Dc Type","values":["Electronic Thesis or Dissertation"]},{"key":"thesis:degree_discipline","label":"Discipline","values":["Physics"]},{"key":"thesis:degree_level","label":"Degree Level","values":["masters"]},{"key":"thesis:degree_name","label":"Degree Name","values":["Master of Science (MS)"]},{"key":"thesis:institution_name","label":"Thesis Institution Name","values":["Wright State University"]}]},{"id":"subjects_keywords","label":"Subjects and Keywords","entries":[{"key":"dc:subject","label":"Dc Subject","values":["Physics, Radiation","IMRT"]}]},{"id":"language_rights","label":"Language and Rights","entries":[{"key":"dc:language","label":"Dc Language","values":["English"]},{"key":"dc:rights","label":"Dc Rights","values":["unrestricted","This thesis or dissertation is protected by copyright: all rights reserved. It may not be copied or redistributed beyond the terms of applicable copyright laws."]}]},{"id":"identifiers","label":"Identifiers","entries":[{"key":"dc:identifier","label":"Identifier","values":["http://rave.ohiolink.edu/etdc/view?acc_num=wright1187369612"]}]},{"id":"additional","label":"Additional Metadata","entries":[{"key":"dc:description","label":"Description","values":["Verification of the isodose distribution delivered to the planned tumor volume is required for quality assurance in the clinic. Verification in this context means ensuring that the actual dose delivered to a volume matches the dose that a computerized planning system intended that volume to receive. The computerized planning is performed by a Computerized Medical Systems (CMS) Intensity Modulated Radiation Therapy (IMRT) plan. Two methods of measuring the delivered dose are film dosimetry and diode dosimetry. This project will compare the relative isodose distribution profiles of film dosimetry and MapCheck (Sun Nuclear Corporation, a diode-based dosimetry system) to the relative isodose distribution profile of the CMS IMRT plan. Two IMRT plans were used that were intended to deliver 7500centigray to the prostate. These plans were then performed while measuring the isodose distributions using the above two techniques. The two techniques were compared on the basis of time efficiency and accuracy."]},{"key":"dc:format","label":"Dc Format","values":["application/pdf","p.75","2.45 MB"]},{"key":"dc:title","label":"Title","values":["Comparative Investigation of Dosimetric Tools in IMRT"]}]}],"canonical_facts":{"dc:contributor":["Foy, Brent"],"dc:creator":["Garcia, Aaron Nicholas"],"dc:date":["2007-08-20"],"dc:description":["Verification of the isodose distribution delivered to the planned tumor volume is required for quality assurance in the clinic. Verification in this context means ensuring that the actual dose delivered to a volume matches the dose that a computerized planning system intended that volume to receive. The computerized planning is performed by a Computerized Medical Systems (CMS) Intensity Modulated Radiation Therapy (IMRT) plan. Two methods of measuring the delivered dose are film dosimetry and diode dosimetry. This project will compare the relative isodose distribution profiles of film dosimetry and MapCheck (Sun Nuclear Corporation, a diode-based dosimetry system) to the relative isodose distribution profile of the CMS IMRT plan. Two IMRT plans were used that were intended to deliver 7500centigray to the prostate. These plans were then performed while measuring the isodose distributions using the above two techniques. The two techniques were compared on the basis of time efficiency and accuracy."],"dc:format":["application/pdf","p.75","2.45 MB"],"dc:identifier":["http://rave.ohiolink.edu/etdc/view?acc_num=wright1187369612"],"dc:language":["English"],"dc:publisher":["Wright State University / OhioLINK"],"dc:rights":["unrestricted","This thesis or dissertation is protected by copyright: all rights reserved. It may not be copied or redistributed beyond the terms of applicable copyright laws."],"dc:subject":["Physics, Radiation","IMRT"],"dc:title":["Comparative Investigation of Dosimetric Tools in IMRT"],"dc:type":["Electronic Thesis or Dissertation"],"thesis:degree_discipline":["Physics"],"thesis:degree_level":["masters"],"thesis:degree_name":["Master of Science (MS)"],"thesis:institution_name":["Wright State University"]},"updated_at":"2026-11-26T04:31:43Z"}