{"id":{"repo_id":"ohiolink","oai_identifier":"oai:etd.ohiolink.edu:wright1389634127"},"canonical_url":"https://search.ndltd.org/etd/ohiolink/oai:etd.ohiolink.edu:wright1389634127","repository":{"repo_id":"ohiolink","name":"OhioLINK","base_url":"https://etd.ohiolink.edu/acprod/odb_etd/ws/oai/oai"},"display":{"title":"Utility of a Volume-Regulated Drive System for Direct Mechanical Ventricular Actuation","abstract":"Direct Mechanical Ventricular Actuation (DMVA) is a non-blood contacting cardiac assist device that augments ventricular function. The purpose of this study was to determine if a volume-regulated &#x201c;hand pump&#x201d; drive system and a pressure-regulated &#x201c;switch tank&#x201d; drive system provide equivalent levels of cardiac support. Canine (n=2) and swine (n=4) were instrumented for hemodynamic monitoring and intravascular echocardiography. DMVA support was assessed during both severe heart failure and fibrillation. Pump function was evaluated using hemodynamic measures to calculate stroke work. Myocardial function was assessed using echocardiographic speckle tracking to quantify strain rate. Results were compared between groups using paired t-tests. There were no significant differences in either pump function or myocardial strain rates between the hand pump versus switch tank during support of either the failing or fibrillating heart. These results suggest functional equivalency between the two drive system mechanisms that supports development of an automated volume-regulated system, with its corresponding benefits in reduced size, portability, and potential user-friendly control.","abstract_html":"Direct Mechanical Ventricular Actuation (DMVA) is a non-blood contacting cardiac assist device that augments ventricular function. The purpose of this study was to determine if a volume-regulated &amp;#x201c;hand pump&amp;#x201d; drive system and a pressure-regulated &amp;#x201c;switch tank&amp;#x201d; drive system provide equivalent levels of cardiac support. Canine (n=2) and swine (n=4) were instrumented for hemodynamic monitoring and intravascular echocardiography. DMVA support was assessed during both severe heart failure and fibrillation. Pump function was evaluated using hemodynamic measures to calculate stroke work. Myocardial function was assessed using echocardiographic speckle tracking to quantify strain rate. Results were compared between groups using paired t-tests. There were no significant differences in either pump function or myocardial strain rates between the hand pump versus switch tank during support of either the failing or fibrillating heart. These results suggest functional equivalency between the two drive system mechanisms that supports development of an automated volume-regulated system, with its corresponding benefits in reduced size, portability, and potential user-friendly control.","abstract_has_math":false,"creators":["Schmitt, Benjamin A."],"institution":"Wright State University","degree_name":"Master of Science in Engineering (MSEgr)","degree_level":"masters","degree_discipline":"Biomedical Engineering","degree_department":null,"school":null,"contributors":["Reynolds, David","Anstadt, Mark"],"advisors":[],"committee_chairs":[],"committee_members":[],"year":2013,"date_issued":"2013","date_published":"2013","updated_at":"2026-11-26T04:36:04Z","subjects":["Biomedical Engineering","Medical Imaging","Direct Mechanical Ventricular Actuation","Mechanical Circulatory Support","Echocardiography","Speckle Tracking","Strain Rate","Myocardial Stretch"],"languages":["English"],"rights":["unrestricted","This thesis or dissertation is protected by copyright: some rights reserved. It is licensed for use under a Creative Commons license. Specific terms and permissions are available from this document's record in the OhioLINK ETD Center."],"rights_urls":[],"identifier_entries":[],"person_identifier_entries":[]},"links":{"outbound_url":"http://rave.ohiolink.edu/etdc/view?acc_num=wright1389634127","outbound_label":"Repository record","outbound_source":"dc:identifier","record_url":"http://rave.ohiolink.edu/etdc/view?acc_num=wright1389634127","record_label":"Repository record","record_kind":"record","record_source":"dc:identifier","access_status":null,"embargo_until":null,"access_note":null},"oai_pmh_record":{"url":"https://union.ndltd.org/oai-pmh/?verb=GetRecord&metadataPrefix=oai_dc&identifier=oai%3Aunion.ndltd.org%3Arepo%3Aohiolink%3Aoai%3Aetd.ohiolink.edu%3Awright1389634127","prefix":"oai_dc"},"source_record":{"url":"https://etd.ohiolink.edu/acprod/odb_etd/ws/oai/oai?verb=GetRecord&metadataPrefix=oai_etdms&identifier=oai%3Aetd.ohiolink.edu%3Awright1389634127","prefix":"oai_etdms"},"metadata_groups":[{"id":"people","label":"People","entries":[{"key":"dc:contributor","label":"Contributor","values":["Reynolds, David","Anstadt, Mark"]},{"key":"dc:creator","label":"Author","values":["Schmitt, Benjamin A."]}]},{"id":"academic_context","label":"Academic Context","entries":[{"key":"dc:date","label":"Dc Date","values":["2013"]},{"key":"dc:publisher","label":"Institution","values":["Wright State University / OhioLINK"]},{"key":"dc:type","label":"Dc Type","values":["Electronic Thesis or Dissertation"]},{"key":"thesis:degree_discipline","label":"Discipline","values":["Biomedical Engineering"]},{"key":"thesis:degree_level","label":"Degree Level","values":["masters"]},{"key":"thesis:degree_name","label":"Degree Name","values":["Master of Science in Engineering (MSEgr)"]},{"key":"thesis:institution_name","label":"Thesis Institution Name","values":["Wright State University"]}]},{"id":"subjects_keywords","label":"Subjects and Keywords","entries":[{"key":"dc:subject","label":"Dc Subject","values":["Biomedical Engineering","Medical Imaging","Direct Mechanical Ventricular Actuation","Mechanical Circulatory Support","Echocardiography","Speckle Tracking","Strain Rate","Myocardial Stretch"]}]},{"id":"language_rights","label":"Language and Rights","entries":[{"key":"dc:language","label":"Dc Language","values":["English"]},{"key":"dc:rights","label":"Dc Rights","values":["unrestricted","This thesis or dissertation is protected by copyright: some rights reserved. It is licensed for use under a Creative Commons license. Specific terms and permissions are available from this document's record in the OhioLINK ETD Center."]}]},{"id":"identifiers","label":"Identifiers","entries":[{"key":"dc:identifier","label":"Identifier","values":["http://rave.ohiolink.edu/etdc/view?acc_num=wright1389634127"]}]},{"id":"additional","label":"Additional Metadata","entries":[{"key":"dc:description","label":"Description","values":["Direct Mechanical Ventricular Actuation (DMVA) is a non-blood contacting cardiac assist device that augments ventricular function. The purpose of this study was to determine if a volume-regulated &#x201c;hand pump&#x201d; drive system and a pressure-regulated &#x201c;switch tank&#x201d; drive system provide equivalent levels of cardiac support. Canine (n=2) and swine (n=4) were instrumented for hemodynamic monitoring and intravascular echocardiography. DMVA support was assessed during both severe heart failure and fibrillation. Pump function was evaluated using hemodynamic measures to calculate stroke work. Myocardial function was assessed using echocardiographic speckle tracking to quantify strain rate. Results were compared between groups using paired t-tests. There were no significant differences in either pump function or myocardial strain rates between the hand pump versus switch tank during support of either the failing or fibrillating heart. These results suggest functional equivalency between the two drive system mechanisms that supports development of an automated volume-regulated system, with its corresponding benefits in reduced size, portability, and potential user-friendly control."]},{"key":"dc:format","label":"Dc Format","values":["application/pdf","p.46","1.01 MB"]},{"key":"dc:title","label":"Title","values":["Utility of a Volume-Regulated Drive System for Direct Mechanical Ventricular Actuation"]}]}],"canonical_facts":{"dc:contributor":["Reynolds, David","Anstadt, Mark"],"dc:creator":["Schmitt, Benjamin A."],"dc:date":["2013"],"dc:description":["Direct Mechanical Ventricular Actuation (DMVA) is a non-blood contacting cardiac assist device that augments ventricular function. The purpose of this study was to determine if a volume-regulated &#x201c;hand pump&#x201d; drive system and a pressure-regulated &#x201c;switch tank&#x201d; drive system provide equivalent levels of cardiac support. Canine (n=2) and swine (n=4) were instrumented for hemodynamic monitoring and intravascular echocardiography. DMVA support was assessed during both severe heart failure and fibrillation. Pump function was evaluated using hemodynamic measures to calculate stroke work. Myocardial function was assessed using echocardiographic speckle tracking to quantify strain rate. Results were compared between groups using paired t-tests. There were no significant differences in either pump function or myocardial strain rates between the hand pump versus switch tank during support of either the failing or fibrillating heart. These results suggest functional equivalency between the two drive system mechanisms that supports development of an automated volume-regulated system, with its corresponding benefits in reduced size, portability, and potential user-friendly control."],"dc:format":["application/pdf","p.46","1.01 MB"],"dc:identifier":["http://rave.ohiolink.edu/etdc/view?acc_num=wright1389634127"],"dc:language":["English"],"dc:publisher":["Wright State University / OhioLINK"],"dc:rights":["unrestricted","This thesis or dissertation is protected by copyright: some rights reserved. It is licensed for use under a Creative Commons license. Specific terms and permissions are available from this document's record in the OhioLINK ETD Center."],"dc:subject":["Biomedical Engineering","Medical Imaging","Direct Mechanical Ventricular Actuation","Mechanical Circulatory Support","Echocardiography","Speckle Tracking","Strain Rate","Myocardial Stretch"],"dc:title":["Utility of a Volume-Regulated Drive System for Direct Mechanical Ventricular Actuation"],"dc:type":["Electronic Thesis or Dissertation"],"thesis:degree_discipline":["Biomedical Engineering"],"thesis:degree_level":["masters"],"thesis:degree_name":["Master of Science in Engineering (MSEgr)"],"thesis:institution_name":["Wright State University"]},"updated_at":"2026-11-26T04:36:04Z"}