{"id":{"repo_id":"rit","oai_identifier":"oai:repository.rit.edu:theses-9878"},"canonical_url":"https://search.ndltd.org/etd/rit/oai:repository.rit.edu:theses-9878","repository":{"repo_id":"rit","name":"Rochester Institute of Technology","base_url":"https://repository.rit.edu/do/oai/"},"display":{"title":"Non-­‐Invasive Fault Detection in Axial Flow Blood Pump Used As Ventricular Assistive Device","abstract":"A Ventricular Assistive Device (VAD) is a mechanical pump used to assist the functioning of a weak heart. A catastrophic obstruction in the VAD system could cost the patient their life. This thesis presents a set of fault detection techniques using the commercially available Jarvik 2000 Flowmaker® VAD in a closed loop circuit that incorporates the ability to alter common causes of VAD congestion. The first goals of this research is to determine the nominal pressure-low performance of the Jarvik pump, which is a graphical presentation of the static head and guides us about the major and minor losses in the system. Second goal is to characterize the health of the VAD system using frequency analysis of the acoustic signature. Principal Component Analysis, a data compression technique used to discover patterns in data of high dimension, is implemented on the frequency analysis and is followed by a health classification based on Bayes theorem. The classification results indicate that this technique is accurate to a high degree in detecting different levels of obstruction in the VAD system.","abstract_html":"A Ventricular Assistive Device (VAD) is a mechanical pump used to assist the functioning of a weak heart. A catastrophic obstruction in the VAD system could cost the patient their life. This thesis presents a set of fault detection techniques using the commercially available Jarvik 2000 Flowmaker® VAD in a closed loop circuit that incorporates the ability to alter common causes of VAD congestion. The first goals of this research is to determine the nominal pressure-low performance of the Jarvik pump, which is a graphical presentation of the static head and guides us about the major and minor losses in the system. Second goal is to characterize the health of the VAD system using frequency analysis of the acoustic signature. Principal Component Analysis, a data compression technique used to discover patterns in data of high dimension, is implemented on the frequency analysis and is followed by a health classification based on Bayes theorem. The classification results indicate that this technique is accurate to a high degree in detecting different levels of obstruction in the VAD system.","abstract_has_math":false,"creators":["Rana, Rohit"],"institution":null,"degree_name":"Mechanical Engineering (MS)","degree_level":"Thesis","degree_discipline":"Mechanical Engineering (KGCOE)","degree_department":null,"school":null,"contributors":["Jason Kolodziej"],"advisors":[],"committee_chairs":[],"committee_members":[],"year":null,"date_issued":"","date_published":null,"updated_at":"2026-11-19T09:02:08Z","subjects":["Fault detection","Ivad","Non invasive","Ventricular assistive device"],"languages":[],"rights":[],"rights_urls":[],"identifier_entries":[],"person_identifier_entries":[]},"links":{"outbound_url":"https://repository.rit.edu/theses/8711","outbound_label":"Repository record","outbound_source":"dc:identifier","record_url":"https://repository.rit.edu/theses/8711","record_label":"Repository record","record_kind":"record","record_source":"dc:identifier","download_url":null,"download_label":null,"download_source":null,"access_status":null,"embargo_until":null,"access_note":null},"source_record":{"url":"https://repository.rit.edu/do/oai/?verb=GetRecord&metadataPrefix=oai_etdms&identifier=oai%3Arepository.rit.edu%3Atheses-9878","prefix":"oai_etdms"},"metadata_groups":[{"id":"people","label":"People","entries":[{"key":"dc:contributor","label":"Contributor","values":["Jason Kolodziej"]},{"key":"dc:creator","label":"Author","values":["Rana, Rohit"]}]},{"id":"academic_context","label":"Academic Context","entries":[{"key":"dc:date.available","label":"Dc Date Available","values":["1970-01-01T08:00:00Z"]},{"key":"dc:date.created","label":"Dc Date Created","values":["2014-07-01T07:00:00Z"]},{"key":"thesis:degree_discipline","label":"Discipline","values":["Mechanical Engineering (KGCOE)"]},{"key":"thesis:degree_level","label":"Degree Level","values":["Thesis"]},{"key":"thesis:degree_name","label":"Degree Name","values":["Mechanical Engineering (MS)"]}]},{"id":"subjects_keywords","label":"Subjects and Keywords","entries":[{"key":"dc:subject","label":"Dc Subject","values":["Fault detection","Ivad","Non invasive","Ventricular assistive device"]}]},{"id":"identifiers","label":"Identifiers","entries":[{"key":"dc:identifier","label":"Identifier","values":["https://repository.rit.edu/theses/8711"]}]},{"id":"additional","label":"Additional Metadata","entries":[{"key":"dc:description.abstract","label":"Abstract","values":["A Ventricular Assistive Device (VAD) is a mechanical pump used to assist the functioning of a weak heart. A catastrophic obstruction in the VAD system could cost the patient their life. This thesis presents a set of fault detection techniques using the commercially available Jarvik 2000 Flowmaker® VAD in a closed loop circuit that incorporates the ability to alter common causes of VAD congestion. The first goals of this research is to determine the nominal pressure-low performance of the Jarvik pump, which is a graphical presentation of the static head and guides us about the major and minor losses in the system. Second goal is to characterize the health of the VAD system using frequency analysis of the acoustic signature. Principal Component Analysis, a data compression technique used to discover patterns in data of high dimension, is implemented on the frequency analysis and is followed by a health classification based on Bayes theorem. The classification results indicate that this technique is accurate to a high degree in detecting different levels of obstruction in the VAD system."]},{"key":"dc:title","label":"Title","values":["Non-­‐Invasive Fault Detection in Axial Flow Blood Pump Used As Ventricular Assistive Device"]}]}],"canonical_facts":{"dc:contributor":["Jason Kolodziej"],"dc:creator":["Rana, Rohit"],"dc:date.available":["1970-01-01T08:00:00Z"],"dc:date.created":["2014-07-01T07:00:00Z"],"dc:description.abstract":["A Ventricular Assistive Device (VAD) is a mechanical pump used to assist the functioning of a weak heart. A catastrophic obstruction in the VAD system could cost the patient their life. This thesis presents a set of fault detection techniques using the commercially available Jarvik 2000 Flowmaker® VAD in a closed loop circuit that incorporates the ability to alter common causes of VAD congestion. The first goals of this research is to determine the nominal pressure-low performance of the Jarvik pump, which is a graphical presentation of the static head and guides us about the major and minor losses in the system. Second goal is to characterize the health of the VAD system using frequency analysis of the acoustic signature. Principal Component Analysis, a data compression technique used to discover patterns in data of high dimension, is implemented on the frequency analysis and is followed by a health classification based on Bayes theorem. The classification results indicate that this technique is accurate to a high degree in detecting different levels of obstruction in the VAD system."],"dc:identifier":["https://repository.rit.edu/theses/8711"],"dc:subject":["Fault detection","Ivad","Non invasive","Ventricular assistive device"],"dc:title":["Non-­‐Invasive Fault Detection in Axial Flow Blood Pump Used As Ventricular Assistive Device"],"thesis:degree_discipline":["Mechanical Engineering (KGCOE)"],"thesis:degree_level":["Thesis"],"thesis:degree_name":["Mechanical Engineering (MS)"]},"updated_at":"2026-11-19T09:02:08Z"}