{"id":{"repo_id":"soton","oai_identifier":"oai:eprints.soton.ac.uk:67337"},"canonical_url":"https://search.ndltd.org/etd/soton/oai:eprints.soton.ac.uk:67337","repository":{"repo_id":"soton","name":"University of Southampton","base_url":"https://eprints.soton.ac.uk/cgi/oai2"},"display":{"title":"Abrasion-corrosion of cast CoCrMo in simulated hip joint environments","abstract":"Metal-on-metal (MoM) hip joint replacements have been increasingly used for younger and more active patients in recent years due to their improved wear performance compared to conventional metal-on-polymer bearings. MoM bearings operate at body temperature within a corrosive joint environment and therefore are inevitably being subjected to wear and corrosion as well as the combined action of tribo-corrosion. Issues such as metal sensitivity/metallosis associated with high levels of metal ion release triggered by the wear and corrosion products remain critical concerns. During the past few decades, significant research has been conducted into understanding the wear/lubrication mechanisms within the MoM hip joints in order to improve their performance and thereby prolonging their life. However, not much attention has been given to the combined effect of wear and corrosion of such devices in the hip joint environment, in addition, the role of third body particles and the effects of proteins have not been well understood.<br/>In this work, a systemic approach is presented for the first time for the mapping of abrasion and tribo-corrosion performance of a cast CoCrMo (F75) in simulated hip joint environments. The effects of third body particles have been studied in the MoM context using 4 ?m SiC, 1 ?m and 300 nm Al2O3, as well as sub-micron BaSO4. Modified tribo-testers (micro-abrasion, nanoindenter/scratching) incorporating a novel electrochemical cell have been used to monitor the abrasion-corrosion behaviour of the alloy in situ. The effects of solution chemistry, abrasives size / concentration and presence of proteins on the wear / corrosion level, wear-corrosion mechanisms, and the depassivation/repassivation kinetics of the CoCrMo have been explored. A variety of surface and sub-surface characterization techniques have been employed to identify the microstructual wear mechanism interactions. Results show that the change of protein concentration (0, 25% and 50% bovine serum) and pH (pH 7.4 and pH 4.0) of the test solutions can significantly influence the protein adsorption behaviour, which subsequently influence the wear rates (synergy), wear mechanisms as well as the wear-induced corrosion currents of the CoCrMo. For abrasion-corrosion tests, reducing abrasive size from 4 ?m to 300 nm and/or abrasive volume concentration from 0.238 vol% to 0.006 vol% results in different abrasion-corrosion wear mechanisms (rolling or grooving abrasion) and the average wear-induced corrosion currents show a linear correlation with wear rates for 4 ?m and 1 ?m abrasives. For low volume concentration (< 0.03 vol%) slurries containing bovine serum, organo-metallic conglomerates have been found within the wear scars. These conglomerates help separate the surfaces, impose less damage to the surface passive film and polish the wear scars through a chemical mechanical polishing mechanism. In addition, tribo-corrosion tests at micro-/nano- scales reveal the effects of single abrasive particle on the surface/sub-surface microstructual change. This investigation has revealed the nanoscale wear mechanisms that generate nanoscale wear debris, the mechanical mixing of the surface nanostructure with adsorbed denatured protein and also the slip/dislocation systems that are present near and on abraded surfaces that are likely to disrupt the surface passive films. The findings give a better understanding of the evolution of the sub-surface nanocrystalline structures and tribo-layers formation seen for the retrieved implants. This near surface nanostructure layer and phase transformation might offer better wear resistance through these inherent self-protecting mechanisms (i.e. increased hardness); conversely, it may become the precursors to debris ejection and enhanced ion-release into the CoCrMo joints.<br/>This work established an experimental technique that gives greater understanding of the tribocorrosion behaviour of cast CoCrMo in simulated hip joint environments. In particular, the roles of third body abrasive particles and proteins have been addressed, which are relevant to clinical applications. The material multi-scale wear mechanisms as well as the evolution of the surface / subsurface microstructures and tribo-layers have been elucidated, which provide new insights into the in vivo wear mechanisms of CoCrMo. The findings of this study may provide some important indications for improved MoM joint materials, design, manufacture and evaluation.","abstract_html":"Metal-on-metal (MoM) hip joint replacements have been increasingly used for younger and more active patients in recent years due to their improved wear performance compared to conventional metal-on-polymer bearings. MoM bearings operate at body temperature within a corrosive joint environment and therefore are inevitably being subjected to wear and corrosion as well as the combined action of tribo-corrosion. Issues such as metal sensitivity/metallosis associated with high levels of metal ion release triggered by the wear and corrosion products remain critical concerns. During the past few decades, significant research has been conducted into understanding the wear/lubrication mechanisms within the MoM hip joints in order to improve their performance and thereby prolonging their life. However, not much attention has been given to the combined effect of wear and corrosion of such devices in the hip joint environment, in addition, the role of third body particles and the effects of proteins have not been well understood.&lt;br/&gt;In this work, a systemic approach is presented for the first time for the mapping of abrasion and tribo-corrosion performance of a cast CoCrMo (F75) in simulated hip joint environments. The effects of third body particles have been studied in the MoM context using 4 ?m SiC, 1 ?m and 300 nm Al2O3, as well as sub-micron BaSO4. Modified tribo-testers (micro-abrasion, nanoindenter/scratching) incorporating a novel electrochemical cell have been used to monitor the abrasion-corrosion behaviour of the alloy in situ. The effects of solution chemistry, abrasives size / concentration and presence of proteins on the wear / corrosion level, wear-corrosion mechanisms, and the depassivation/repassivation kinetics of the CoCrMo have been explored. A variety of surface and sub-surface characterization techniques have been employed to identify the microstructual wear mechanism interactions. Results show that the change of protein concentration (0, 25% and 50% bovine serum) and pH (pH 7.4 and pH 4.0) of the test solutions can significantly influence the protein adsorption behaviour, which subsequently influence the wear rates (synergy), wear mechanisms as well as the wear-induced corrosion currents of the CoCrMo. For abrasion-corrosion tests, reducing abrasive size from 4 ?m to 300 nm and/or abrasive volume concentration from 0.238 vol% to 0.006 vol% results in different abrasion-corrosion wear mechanisms (rolling or grooving abrasion) and the average wear-induced corrosion currents show a linear correlation with wear rates for 4 ?m and 1 ?m abrasives. For low volume concentration (&lt; 0.03 vol%) slurries containing bovine serum, organo-metallic conglomerates have been found within the wear scars. These conglomerates help separate the surfaces, impose less damage to the surface passive film and polish the wear scars through a chemical mechanical polishing mechanism. In addition, tribo-corrosion tests at micro-/nano- scales reveal the effects of single abrasive particle on the surface/sub-surface microstructual change. This investigation has revealed the nanoscale wear mechanisms that generate nanoscale wear debris, the mechanical mixing of the surface nanostructure with adsorbed denatured protein and also the slip/dislocation systems that are present near and on abraded surfaces that are likely to disrupt the surface passive films. The findings give a better understanding of the evolution of the sub-surface nanocrystalline structures and tribo-layers formation seen for the retrieved implants. This near surface nanostructure layer and phase transformation might offer better wear resistance through these inherent self-protecting mechanisms (i.e. increased hardness); conversely, it may become the precursors to debris ejection and enhanced ion-release into the CoCrMo joints.&lt;br/&gt;This work established an experimental technique that gives greater understanding of the tribocorrosion behaviour of cast CoCrMo in simulated hip joint environments. In particular, the roles of third body abrasive particles and proteins have been addressed, which are relevant to clinical applications. The material multi-scale wear mechanisms as well as the evolution of the surface / subsurface microstructures and tribo-layers have been elucidated, which provide new insights into the in vivo wear mechanisms of CoCrMo. The findings of this study may provide some important indications for improved MoM joint materials, design, manufacture and evaluation.","abstract_has_math":false,"creators":["Sun, Dan"],"institution":"University of Southampton","degree_name":"Ph.D.","degree_level":"doctoral","degree_discipline":null,"degree_department":null,"school":null,"contributors":[],"advisors":["Wood, R.J.K.","Wharton, J.A."],"committee_chairs":[],"committee_members":[],"year":2009,"date_issued":"2009-05","date_published":"2009-05","updated_at":"2026-11-19T09:29:52Z","subjects":[],"languages":["eng"],"rights":[],"rights_urls":[],"identifier_entries":[],"person_identifier_entries":[{"key":"person:identifier.orcid","label":"ORCID","values":["0000000306819239","000000023439017X"],"render_values":[{"text":"0000-0003-0681-9239","href":"https://orcid.org/0000-0003-0681-9239","code":true},{"text":"0000-0002-3439-017X","href":"https://orcid.org/0000-0002-3439-017X","code":true}]}]},"links":{"outbound_url":null,"outbound_label":null,"outbound_source":null,"record_url":null,"record_label":null,"record_kind":null,"record_source":null,"download_url":"https://eprints.soton.ac.uk/id/eprint/67337/1/Binder1.pdf","download_label":"Download PDF","download_source":"dc:identifier.uri","access_status":null,"embargo_until":null,"access_note":null},"source_record":{"url":"https://eprints.soton.ac.uk/cgi/oai2?verb=GetRecord&metadataPrefix=uketd_dc&identifier=oai%3Aeprints.soton.ac.uk%3A67337","prefix":"uketd_dc"},"metadata_groups":[{"id":"people","label":"People","entries":[{"key":"dc:contributor.advisor","label":"Advisor","values":["Wood, R.J.K.","Wharton, J.A."]},{"key":"dc:creator","label":"Author","values":["Sun, Dan"]},{"key":"person:identifier.orcid","label":"ORCID","values":["0000000306819239","000000023439017X"]}]},{"id":"academic_context","label":"Academic Context","entries":[{"key":"dc:date","label":"Dc Date","values":["2009-05"]},{"key":"dc:date.issued","label":"Date","values":["2009-05"]},{"key":"dc:publisher.department","label":"Dc Publisher Department","values":["Bioengineering Sciences (pre 2011 reorg)","Engineering Mats & Surface Engineerg Gp (pre 2018 reorg)","School of Engineering Sciences"]},{"key":"dc:publisher.institution","label":"Dc Publisher Institution","values":["University of Southampton"]},{"key":"dc:relation.isreferencedby","label":"Dc Relation Isreferencedby","values":["https://eprints.soton.ac.uk/id/eprint/67337/"]},{"key":"dc:relation.references","label":"Dc Relation References","values":["1. Wu, X., Tao, N., Hong, Y., Liu, G., Xu, B., Lu, J., Lu, K., Strain-induced grain refinement of cobalt during surface mechanical attrition treatment. Acta Materialia, 2005. 53(3): p. 681-691. 2. Williams, S., Tipper, J.L., Ingham, E., Stone, M.H., Fisher, J., In vitro analysis of the wear, wear debris and biological activity of surface-engineered coatings for use in metal-on-metal total hip replacements. Proc Inst Mech Eng, 2003. 217(H): p. 155-163. 3. http://ortho.muhealth.org/hipandknee/images/HipReplace.jpg. 4. http://www.hipsforyou.com/strykerceramichip.php. 5. Dowson, D., New joints for the millennium: wear control in total replacement hip joints. Proc I Mech E (H), 2001. 215(H4): p. 335-358. 6. Buford, A., Goswami, T., Review of wear mechanisms in hip implants: Paper I - General. Materials and Design, 2004. 25(5): p. 385-393. 7. Dowson, D., Hardaker, C., Flett, M., Issac, G.H., A hip joint simulator study of the performance of metal-on-metal joints. Part I: The role of materials. The Journal of Arthroplasty, 2004. 19(8): p. 118-123. 8. Jin, Z.M., Stone, M.H., Ingham, E., Fisher, J, Biotribology. Current Orthopaedics, 2006. 20: p. 32-40. 9. Roberts, P., Grigoris, P., Bosch, H., Talwaker, N, Resurfacing arthroplasty of the hip. Current Orthopaedics, 2005. 19: p. 263-279. 10. Hughs, D., Hip resurfacing versus hip replacement - What, Why, Who and When? http://personal.strath.ac.uk/david.hughes/Papers/Hip_replacement_versus_hip_ resurfacing.pdf. 11. Adams, D., Quigley, S., Hip resurfacing: Past, present and future. Journal of Orthopaedic Nursing, 2005. 9: p. 87-94. 12. Wang, A., Yue, S., Bobyn, J.D., Chan, F.W., and Medley, J.B., Surface characterization of metal-on-metal hip implants tested in a hip simulator. Wear, 1999. 225-229(pt.2): p. 708-715. 13. Catelas, I., Bobyn, J.D., Medley, J.B., Krygier, J.J., Zukor, D.J., Huk, O.L., Size, shape, and composition of wear particles from metal-metal hip simulator testing: Effects of alloy and number of loading cycles. Journal of Biomedical Materials Research - Part A, 2003. 67(1): p. 312-327. 14. Goldsmith, A.A.J., Dowson, D., Isaac, G.H., and Lanscaster, J.G., A Comparative Joint Simulator Study of the Wear of Metal-On-Metal and Alternative Material Combinations in Hip Replacements. Proceedings of the I MECH E Part H Journal of Engineering in Medicine, 2000. 214: p. 39-47. 15. Walker, P.S., Erkman, M. J., Metal-on-metal lubrication in artificial human joints. Wear, 1972. 21(2): p. 377-392. 16. Liu, F., Jin, Z. M., Hirt, F., Rieker, C., Roberts, P., Grigoris, P., Effect of wear of bearing surfaces on elastohydrodynamic lubrication of metal-on-metal hip implants. Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine, 2005. 219(5): p. 319-328. 17. Dowson, D., Jin, Z.M., Metal-on-metal hip joint tribology. Proc. IMechE, Part H: J. Engineering in Medicine, 2005. 220: p. 107-118. 18. Medley, J.B., Bobyn, J.D., Krygler, J.J., Chan, F.W., Tanzer, M., Roter, G.E. Elastohydrodynamic lubrication and wear of metal-on-metal hip implant. in World Tribology Forum in Arthroplasty. 2001. Bern: Hans Huber. 19. Varano, R., Bobyn, D., Medley, J.B., Yue, S., The effect of microstructure on the wear of cobalt-based alloys used in metal-on-metal hip implants Proc I Mech E (H), 2006. 220: p. 145-159. 20. Chan, F.W., Bobyn, J.D., Medley, J.B., Krygier, J.J., Yue, S., Tanzer, M., Engineering issues and wear performance of metal on metal hip implants. Clinical orthopaedics and related research 1996( 333): p. 96-107. 21. Chan, F.W., Bobyn, J.D., Medley, J.B., Krygier, J.J., Tanzer, M., Wear and lubrication of metal-on-metal hip implants. Clin Orthop, 1999. 369: p. 10-24. 22. Walker, P.S. and B.L. Gold, Tribology (friction, lubrication and wear) of allmetal artificial hip joints. Wear, 1971. 17(4): p. 285-299. 23. Hernandez-Rodriguez, M.A.L., Mercado-Solis, R. D., Perez-Unzueta, A. J., Martinez-Delgado, D. I., Cantu-Sifuentes, M., Wear of cast metal-metal pairs for total replacement hip prostheses. Wear, 2005. 259(7-12): p. 958-963. 24. Hallab, N.J. and J.J. Jacobs, Orthopedic implant fretting corrosion. Corrosion Reviews, 2003. 21(2-3): p. 183-213. 25. Hallab, N.J., Messina, C., Skipor, A., Jacobs, J.J., Difference in the fretting corrosion of metal-metal and ceramic-metal modular junctions of total hip replacements. Journal of Orthopaedic Research, 2004. 22: p. 250-259. 26. Gilbert, J.L., Buckley, C.A., and Jacobs, J.J., In-Vivo Corrosion of Modular Hip Prosthesis Components in Mixed and Similar Metal Combinations: The Effect of Crevice, Stress, Motion and Alloy Coupling. J Biomed Mater Res, 1993. 27(12): p. 1533.1544. 27. Ratzer-Schiebe, H.J., The relationship between repassivation behavior and pitting corrosion for ti and Ti-6Al-4V. Corrosion 1978. 34: p. 437-442. 28. Dumbleton, J.H., Tribology of natural and artificial joints. 1981, Amsterdam: Elsevier. 29. Tipper, J.L., Ingham, E., Jin, Z.M., Fisher, J., The science of metal-on-metal articulation. Current Orthopaedics, 2005. 19: p. 280-287. 30. Callister, W.D., Materials science and engineering: an introduction. 5th ed. 1999: John Wiley & Sons Inc. p.579. 31. Venugopalan, R., Gaydon, J. , A review of corrosion behaviour of surgical implant alloys. Technical Review Note, Perkin Elmer Instruments, Princeton Applied Research, 1999-2001: p. 1-10. 32. Dearnley, P.A., A review of metallic, ceramic and surface treated materials used for bearing surfaces in human joint replacements. Proc Inst Mech Eng, 1999. 3(H): p. 107-135. 33. Mckee, G.K., Watson-Farrar, J., Replacement of arthritic hips by the Mckee- Farrar prosthsis. J Bone & Joint Surgery, 1966. 48(B): p. 245-259. 34. Marti, A., Cobalt-base alloys used in bone surgery. Injury, Int J. Care Injured, 2000. 31: p. S-D 18-21. 35. Park, J.B., Kon Kim, Y., Metallic Biomaterials, in The Biomedical Engineering Handbook, 2nd Ed., J.D. Bronzino, Editor. 2000, CRC Press LLC: Boca Raton. 36. Yan, Y., Neville, A., Dowson, D., Williams, S., Tribocorrosion in implantsassessing high carbon and low carbon Co-Cr-Mo alloys by in situ electrochemical measurements. Tribology International, 2006. 39(12): p. 1509- 1517. 37. Metcalf, J.E.P., Cawley, J., Band, T.J., Cobalt Chromium Molybdenum metalon- metal resurfacing orthopaedic hip devices. Business briefing: medical device manufacturing & technology, 2004: p. 1-7. 38. Salinas-Rodriguez, A., The Role of the FCC-HCP phase transformation during the plastic deformation of Co-Cr-Mo-C alloys for biomedical applications, in ASTM Special Technical Publication 1365, R.L.K. J.A. Disegi, R. Pilliar, Editor. 1999, American Society for Testing and Materials: West Conshokochen, PA. p. 108-121. 39. Lopez, H.F., Saldivar, A. Huang, P. Development and properties of ƒÃ-martensite in CoCrMo alloys for biomaterials applications. in the Proceedings of 2nd International Alloy Conference. 1999. Davos, Switzerland: Kluwer Academic/Plenum Publishers, New York. 40. Buscher, R., Fischer, A., The pathways of dynamic recrystallization in all-metal hip joints. Wear 2005. 259: p. 887-897. 41. Weber, B.G., Experience with the Metasul total hip bearing system. Clin Orthop, 1996. 3295: p. S78-88. 42. Harlan, C.A., M.D., History of Hip Resurfacing, http://www.jrioh. com/jri_hip_surface.php. Access date: 01.Dec 2007. 43. Wagner, M., Wagner, H., Prelinimary results of uncemented metal-on-metal stemmed and resrfacing hip replacement arthroplasty. clin Orthop, 1996. 329S: p. S89-98. 44. Jin, Z.M., Tipper, J.L., Stone, M.H., Ingham, E., Fisher, J, Hip joints, aritificial, in Encyclopedia of Medical Devices and Instrumentation, J.G. Webster, Editor. 2006, John Wiley & Sons, Inc. 45. Cawley, J., Metcalf, J.E.P., Jones, A.H., Band, T.J., Skupien, D.S., A tribological study of cobalt chromium molybdenum alloys used in metal-onmetal resurfacing hip arthroplasty. Wear, 2003. 255(pt.2): p. 999-1006. 46. Wimmer, M.A., Loos, J., Nassutt, R., Heitkemper, M., Fischer, A., The acting wear mechanisms on metal-on-metal hip joint bearings: in vitro results. Wear, 2001. 250: p. 129-139. 47. http://www.wheelessonline.com/ortho/gait. Access date 01-01-2009. 48. Paul, J.P., Force actions transmitted by joints in the human body. Proceedings of the Royal Society of London. Series B, Biological Sciences, 1976. 192(1107): p. 163-172. 49. Litsky, A.S., Spector, M., Biomaterials, in Orthopaedic basic science, S.R. Simon, Editor. 1994, American academy of orthopaedic surgeons. p. 470-473. 50. Bergmann, G., Graichen, F., Rohlmann, A., Hip joint loading during walking and running, measured in two patients. J. Biomech, 1993. 26: p. 969. 51. Bergmann, G., Graichen, F., Rohlmann, A., Is staircase walking a risk for the fixation of hip implants? Journal of Biomechanics, 1995. 28(5): p. 535-53. 52. Jin, Z.M., Dowson, D., Fisher, J., Analysis of fluid film lubrication in artificial hip joint replacements with surfaces of high elastic modulus. Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine, 1997. 211(3): p. 247-256. 53. Semlitsch, M., Willert H.G., Clinical wear behavior of ultra-high molecular weight polyethylene cups paired with metal and ceramic ball heads in comparison to metal-on-metal pairings of hip joint replacements. Proc Inst Mech Eng, 1997. 211(H): p. 73-88. 54. Doorn, P.F., Campbell, P.A., Worrall, J., Benya, P.D., McKellop, H.A., Amstutz, H.C., Metal wear particle characterization from metal on metal total hip replacements: transimission electron microscopy study of periprosthetic tissues and isolated particles. J. Biomed. Mater. Res, 1998. 42: p. 103-111. 55. Hanawa, T., Metal ion release from metal implants. Materials Science and Engineering C, 2004. 24(6-8 SPEC ISS): p. 745-752. 56. Peters, T.J., Serum Albumin. Advances in Protein Chemistry, 1985. 37: p. 161- 245. 57. Van Oss, C.J., Good, R.J., Orientation of the water molecules of hydration of human serum albumin Jounral of Protein Chemistry, 1988. 7(2): p. 179-183. 58. Burnstock, G., Purine-mediated signalling in pain and visceral perception. Trends Pharmacol. Sci., 2001. 22 p. 182-188. 59. Wood, D.J., The characterization of particulate debris obtained from failed orthopedic implants. 1993, San Jose State University, College of Material Engineering. 60. Steen, K.H., Reeh, P.W., Sustained graded pain and hyperalgesia from harmless experimental tissue acidosis in human skin. Neuroscience letter, 1993. 154: p. 113-116. 61. Ratner, B.D., Hoffman, A. S., Schoen, F.J., Lemons, J.E., Biomaterials Science, an Introduction to Materials in Medicine, 2nd. 2004: Elsevier Academic Press. 62. Moulton, S.E., Barisci , J.N., Bath, A., Stella, R. , Wallace, G.G. , Investigation of protein adsorption and electrochemical behavior at a gold electrode. J Colloid Interface Sci. , 2003. 15(261(2)): p. 312-319. 63. Hiromoto, S., K. Noda, and T. Hanawa, Electrochemical properties of an interface between titanium and fibroblasts L929. Electrochimica Acta, 2002. 48(4): p. 387-396. 64. Gettens, R.T.T., Gettens, R. T. T., Gilbert, J. L. Fibrinogen Adsorption onto 316L Stainless Steel: Voltage Effects. in Bioengineering Conference, 2006. Proceedings of the IEEE 32nd Annual Northeast. 2006. 65. de Ficquelmont-Loizos, M.M., Takenouti, H., Kante, W., Long-time and shorttime investigation of the electrode interface through electrochemical impedance measurements. Application to adsorption of human serum albumin onto glassy carbon rotating disc electrode. Journal of Electroanalytical Chemistry, 1997. 428(1-2): p. 129-140. 66. Kleijn, J.M., Barten, Desiree, Stuart, Martien A. Cohen, Adsorption of charged macromolecules at a gold electrode. Langmuir, 2004. 20(22): p. 9703-9713. 67. Serro, A.P., Gispert, M. P., Martins, M. C. L., Brogueira, P., Colaco, R., Saramago, B., Adsorption of albumin on prosthetic materials: Implication for tribological behavior. J Bio Mat Res - Part A, 2006. 78(3): p. 581-589. 68. Sharma., C.P., Clark, G.C.F., Williams, D.F., The adsorption of proteins on metal surfaces. Engineering in medicine, 1981. 10(1): p. 11-16. 69. Bernabeu, P., Tamisier, L., de Cesare, A., Caprani, A., Study of the adsorption of albumin on a platinum rotating disk electrode using impedance measurements. Electrochimica Acta, 1988. 33(8): p. 1129-1136. 70. Choi, K.H., Friedt, J. M., Laureyn, W., Frederix, F., Campitelli, A., Borghs, G., Investigation of protein adsorption with simultaneous measurements of atomic force microscope and quartz crystal microbalance. Journal of Vacuum Science & Technology B (Microelectronics and Nanometer Structures), 2003. 21(4): p. 1433-6. 71. Ying, P., Viana, A. S., Abrantes, L. M., Jin, G., Adsorption of human serum albumin onto gold: A combined electrochemical and ellipsometric study. Journal of Colloid and Interface Science, 2004. 279(1): p. 95-99. 72. Malmsten, M., Formation of adsorbed protein layers. Journal of Colloid and Interface Science, 1998. 207: p. 186-199. 73. Merritt, K., Brown, S. A., Effect of proteins and pH on fretting corrosion and metal ion release. Journal of Biomedical Materials Research, 1988. 22(2): p. 111-120. 74. Sven, P., Determination of transferrin isoproteins in human cerebrospinal fluid using isoelectric focusing and zone immunoelectrophoresis assay. Electrophoresis, 1987. 8(11): p. 515-517. 75. Salim, M., O'Sullivan, B., McArthur, S. L.,Wright, P.C., Characterization of fibrinogen adsorption onto glass microcapillary surfaces by ELISA. Lab Chip, 2007. 7: p. 64 - 70. 76. Geborek, P., Wollheim, F.A., Synovial fluid, in Mechanics of Human Joints, Physiology, Pathophysiology and Treatment, Verna Wright and E.L. Radin, Editors. 1993, Marcel Dekker: New York. p. 117-136. 77. Jones, K.L., O'Melia, C. R., Protein and humic acid adsorption onto hydrophilic membrane surfaces: Effects of pH and ionic strength. Journal of Membrane Science, 2000. 165(1): p. 31-46. 78. Cabilio, N.R., Omanovic, S., Roscoe, S.G., Electrochemical studies of the effect of temperature and pH on the adsorption of alpha-lactalbumin at Pt. Langmuir, 2000. 16(22): p. 8480-8488. 79. Hughes Wassell, D.T., Embery, G., Adsorption of bovine serum albumin on to titanium powder. Biomaterials, 1996. 17(9): p. 859-864. 80. Malmsten, M., Biopolymers at Interfaces. 1998, New York: Marcel Dekker. 81. Okazaki, Y., Tateishi, T., Ito, Y., Corrosion resistance of implant alloys in pseudo physiological solution and role of alloying elements in passive films. Materials Transactions, JIM, 1997. 38(1): p. 78-84. 82. Lacour, F., Torresi, R., Gabrielli, C., Caprani, A., Comparison of the quarzcrystal microbalance and the double-layer capacitance methods for measuring the kinetics of the adsorption of bovine serum albumin onto a gold electrode. Journal of the Electrochemical Society, 1992. 139(6): p. 1619-1622. 83. Omanovic, S., Roscoe, S. G., Electrochemical studies of the adsorption behavior of bovine serum albumin on stainless steel. Langmuir, 1999. 15(23): p. 8315-8321. 84. Moulton, S.E., Barisci, J. N., Bath, A., Stella, R., Wallace, G. G., Studies of double layer capacitance and electron transfer at a gold electrode exposed to protein solutions. Electrochimica Acta, 2004. 49(24): p. 4223-4230. 85. Gassa, L., De Mele, Monica Fernandez Lorenzo. Influence of the composition of the biological media on the electrochemical response of the metallic components of biomaterials. 2005. Lisbon, Portugal. 86. Tang, Q., Xu, C.H., Shi S. Q., Zhou, L M., Formation and characterization of protein patterns on the surfaces with different properties. Synthetic Metals, 2004. 147(1-3): p. 247-52. 87. Tuemmler, H.P., Thull, R., Model of the metal/tissue connection of implants made of titanium or tantalum. Biological and Biomechanical Performance of Biomaterials. 1986, Paris, Fr: Elsevier, Amsterdam, Neth. 403-408. 88. Gilbert, J.L., Bai, Zhijun, Bearinger, Jane, Megremis, Spiro. The dynamics of oxide films on metallic biomaterials. 2003. Anaheim, CA., United States: ASM International. 89. Lacour, F., Frayret, J. P., Caprani, A., Effect of the electrode roughness on the kinetics of albumin adsorption at a rotating disk electrode investigated by double-layer capacitance measurements. Journal of the Electrochemical Society, 1992. 139(1): p. 153-159. 90. Contu, F., Elsener, B., Bohni, H., Serum effect on the electrochemical behaviour of titanium, Ti6Al4V and Ti6Al7Nb alloys in sulp"]},{"key":"dc:type","label":"Dc Type","values":["Thesis"]},{"key":"dc:type.qualificationlevel","label":"Dc Type Qualificationlevel","values":["doctoral"]},{"key":"dc:type.qualificationname","label":"Dc Type Qualificationname","values":["Ph.D."]}]},{"id":"language_rights","label":"Language and Rights","entries":[{"key":"dc:language.dcterms_iso639_2","label":"Language (ISO 639-2)","values":["eng"]}]},{"id":"identifiers","label":"Identifiers","entries":[{"key":"dc:identifier.uri","label":"Identifier URI","values":["https://eprints.soton.ac.uk/id/eprint/67337/1/Binder1.pdf"]}]},{"id":"additional","label":"Additional Metadata","entries":[{"key":"dc:description.abstract","label":"Abstract","values":["Metal-on-metal (MoM) hip joint replacements have been increasingly used for younger and more active patients in recent years due to their improved wear performance compared to conventional metal-on-polymer bearings. MoM bearings operate at body temperature within a corrosive joint environment and therefore are inevitably being subjected to wear and corrosion as well as the combined action of tribo-corrosion. Issues such as metal sensitivity/metallosis associated with high levels of metal ion release triggered by the wear and corrosion products remain critical concerns. During the past few decades, significant research has been conducted into understanding the wear/lubrication mechanisms within the MoM hip joints in order to improve their performance and thereby prolonging their life. However, not much attention has been given to the combined effect of wear and corrosion of such devices in the hip joint environment, in addition, the role of third body particles and the effects of proteins have not been well understood.<br/>In this work, a systemic approach is presented for the first time for the mapping of abrasion and tribo-corrosion performance of a cast CoCrMo (F75) in simulated hip joint environments. The effects of third body particles have been studied in the MoM context using 4 ?m SiC, 1 ?m and 300 nm Al2O3, as well as sub-micron BaSO4. Modified tribo-testers (micro-abrasion, nanoindenter/scratching) incorporating a novel electrochemical cell have been used to monitor the abrasion-corrosion behaviour of the alloy in situ. The effects of solution chemistry, abrasives size / concentration and presence of proteins on the wear / corrosion level, wear-corrosion mechanisms, and the depassivation/repassivation kinetics of the CoCrMo have been explored. A variety of surface and sub-surface characterization techniques have been employed to identify the microstructual wear mechanism interactions. Results show that the change of protein concentration (0, 25% and 50% bovine serum) and pH (pH 7.4 and pH 4.0) of the test solutions can significantly influence the protein adsorption behaviour, which subsequently influence the wear rates (synergy), wear mechanisms as well as the wear-induced corrosion currents of the CoCrMo. For abrasion-corrosion tests, reducing abrasive size from 4 ?m to 300 nm and/or abrasive volume concentration from 0.238 vol% to 0.006 vol% results in different abrasion-corrosion wear mechanisms (rolling or grooving abrasion) and the average wear-induced corrosion currents show a linear correlation with wear rates for 4 ?m and 1 ?m abrasives. For low volume concentration (< 0.03 vol%) slurries containing bovine serum, organo-metallic conglomerates have been found within the wear scars. These conglomerates help separate the surfaces, impose less damage to the surface passive film and polish the wear scars through a chemical mechanical polishing mechanism. In addition, tribo-corrosion tests at micro-/nano- scales reveal the effects of single abrasive particle on the surface/sub-surface microstructual change. This investigation has revealed the nanoscale wear mechanisms that generate nanoscale wear debris, the mechanical mixing of the surface nanostructure with adsorbed denatured protein and also the slip/dislocation systems that are present near and on abraded surfaces that are likely to disrupt the surface passive films. The findings give a better understanding of the evolution of the sub-surface nanocrystalline structures and tribo-layers formation seen for the retrieved implants. This near surface nanostructure layer and phase transformation might offer better wear resistance through these inherent self-protecting mechanisms (i.e. increased hardness); conversely, it may become the precursors to debris ejection and enhanced ion-release into the CoCrMo joints.<br/>This work established an experimental technique that gives greater understanding of the tribocorrosion behaviour of cast CoCrMo in simulated hip joint environments. In particular, the roles of third body abrasive particles and proteins have been addressed, which are relevant to clinical applications. The material multi-scale wear mechanisms as well as the evolution of the surface / subsurface microstructures and tribo-layers have been elucidated, which provide new insights into the in vivo wear mechanisms of CoCrMo. The findings of this study may provide some important indications for improved MoM joint materials, design, manufacture and evaluation."]},{"key":"dc:format","label":"Dc Format","values":["text"]},{"key":"dcterms:accessrights","label":"Dcterms Accessrights","values":["public"]},{"key":"dc:title","label":"Title","values":["Abrasion-corrosion of cast CoCrMo in simulated hip joint environments"]}]}],"canonical_facts":{"dc:contributor.advisor":["Wood, R.J.K.","Wharton, J.A."],"dc:creator":["Sun, Dan"],"dc:date":["2009-05"],"dc:date.issued":["2009-05"],"dc:description.abstract":["Metal-on-metal (MoM) hip joint replacements have been increasingly used for younger and more active patients in recent years due to their improved wear performance compared to conventional metal-on-polymer bearings. MoM bearings operate at body temperature within a corrosive joint environment and therefore are inevitably being subjected to wear and corrosion as well as the combined action of tribo-corrosion. Issues such as metal sensitivity/metallosis associated with high levels of metal ion release triggered by the wear and corrosion products remain critical concerns. During the past few decades, significant research has been conducted into understanding the wear/lubrication mechanisms within the MoM hip joints in order to improve their performance and thereby prolonging their life. However, not much attention has been given to the combined effect of wear and corrosion of such devices in the hip joint environment, in addition, the role of third body particles and the effects of proteins have not been well understood.<br/>In this work, a systemic approach is presented for the first time for the mapping of abrasion and tribo-corrosion performance of a cast CoCrMo (F75) in simulated hip joint environments. The effects of third body particles have been studied in the MoM context using 4 ?m SiC, 1 ?m and 300 nm Al2O3, as well as sub-micron BaSO4. Modified tribo-testers (micro-abrasion, nanoindenter/scratching) incorporating a novel electrochemical cell have been used to monitor the abrasion-corrosion behaviour of the alloy in situ. The effects of solution chemistry, abrasives size / concentration and presence of proteins on the wear / corrosion level, wear-corrosion mechanisms, and the depassivation/repassivation kinetics of the CoCrMo have been explored. A variety of surface and sub-surface characterization techniques have been employed to identify the microstructual wear mechanism interactions. Results show that the change of protein concentration (0, 25% and 50% bovine serum) and pH (pH 7.4 and pH 4.0) of the test solutions can significantly influence the protein adsorption behaviour, which subsequently influence the wear rates (synergy), wear mechanisms as well as the wear-induced corrosion currents of the CoCrMo. For abrasion-corrosion tests, reducing abrasive size from 4 ?m to 300 nm and/or abrasive volume concentration from 0.238 vol% to 0.006 vol% results in different abrasion-corrosion wear mechanisms (rolling or grooving abrasion) and the average wear-induced corrosion currents show a linear correlation with wear rates for 4 ?m and 1 ?m abrasives. For low volume concentration (< 0.03 vol%) slurries containing bovine serum, organo-metallic conglomerates have been found within the wear scars. These conglomerates help separate the surfaces, impose less damage to the surface passive film and polish the wear scars through a chemical mechanical polishing mechanism. In addition, tribo-corrosion tests at micro-/nano- scales reveal the effects of single abrasive particle on the surface/sub-surface microstructual change. This investigation has revealed the nanoscale wear mechanisms that generate nanoscale wear debris, the mechanical mixing of the surface nanostructure with adsorbed denatured protein and also the slip/dislocation systems that are present near and on abraded surfaces that are likely to disrupt the surface passive films. The findings give a better understanding of the evolution of the sub-surface nanocrystalline structures and tribo-layers formation seen for the retrieved implants. This near surface nanostructure layer and phase transformation might offer better wear resistance through these inherent self-protecting mechanisms (i.e. increased hardness); conversely, it may become the precursors to debris ejection and enhanced ion-release into the CoCrMo joints.<br/>This work established an experimental technique that gives greater understanding of the tribocorrosion behaviour of cast CoCrMo in simulated hip joint environments. In particular, the roles of third body abrasive particles and proteins have been addressed, which are relevant to clinical applications. The material multi-scale wear mechanisms as well as the evolution of the surface / subsurface microstructures and tribo-layers have been elucidated, which provide new insights into the in vivo wear mechanisms of CoCrMo. The findings of this study may provide some important indications for improved MoM joint materials, design, manufacture and evaluation."],"dc:format":["text"],"dc:identifier.uri":["https://eprints.soton.ac.uk/id/eprint/67337/1/Binder1.pdf"],"dc:language.dcterms_iso639_2":["eng"],"dc:publisher.department":["Bioengineering Sciences (pre 2011 reorg)","Engineering Mats & Surface Engineerg Gp (pre 2018 reorg)","School of Engineering Sciences"],"dc:publisher.institution":["University of Southampton"],"dc:relation.isreferencedby":["https://eprints.soton.ac.uk/id/eprint/67337/"],"dc:relation.references":["1. Wu, X., Tao, N., Hong, Y., Liu, G., Xu, B., Lu, J., Lu, K., Strain-induced grain refinement of cobalt during surface mechanical attrition treatment. Acta Materialia, 2005. 53(3): p. 681-691. 2. Williams, S., Tipper, J.L., Ingham, E., Stone, M.H., Fisher, J., In vitro analysis of the wear, wear debris and biological activity of surface-engineered coatings for use in metal-on-metal total hip replacements. Proc Inst Mech Eng, 2003. 217(H): p. 155-163. 3. http://ortho.muhealth.org/hipandknee/images/HipReplace.jpg. 4. http://www.hipsforyou.com/strykerceramichip.php. 5. Dowson, D., New joints for the millennium: wear control in total replacement hip joints. Proc I Mech E (H), 2001. 215(H4): p. 335-358. 6. Buford, A., Goswami, T., Review of wear mechanisms in hip implants: Paper I - General. Materials and Design, 2004. 25(5): p. 385-393. 7. Dowson, D., Hardaker, C., Flett, M., Issac, G.H., A hip joint simulator study of the performance of metal-on-metal joints. Part I: The role of materials. The Journal of Arthroplasty, 2004. 19(8): p. 118-123. 8. Jin, Z.M., Stone, M.H., Ingham, E., Fisher, J, Biotribology. Current Orthopaedics, 2006. 20: p. 32-40. 9. Roberts, P., Grigoris, P., Bosch, H., Talwaker, N, Resurfacing arthroplasty of the hip. Current Orthopaedics, 2005. 19: p. 263-279. 10. Hughs, D., Hip resurfacing versus hip replacement - What, Why, Who and When? http://personal.strath.ac.uk/david.hughes/Papers/Hip_replacement_versus_hip_ resurfacing.pdf. 11. Adams, D., Quigley, S., Hip resurfacing: Past, present and future. Journal of Orthopaedic Nursing, 2005. 9: p. 87-94. 12. Wang, A., Yue, S., Bobyn, J.D., Chan, F.W., and Medley, J.B., Surface characterization of metal-on-metal hip implants tested in a hip simulator. Wear, 1999. 225-229(pt.2): p. 708-715. 13. Catelas, I., Bobyn, J.D., Medley, J.B., Krygier, J.J., Zukor, D.J., Huk, O.L., Size, shape, and composition of wear particles from metal-metal hip simulator testing: Effects of alloy and number of loading cycles. Journal of Biomedical Materials Research - Part A, 2003. 67(1): p. 312-327. 14. Goldsmith, A.A.J., Dowson, D., Isaac, G.H., and Lanscaster, J.G., A Comparative Joint Simulator Study of the Wear of Metal-On-Metal and Alternative Material Combinations in Hip Replacements. Proceedings of the I MECH E Part H Journal of Engineering in Medicine, 2000. 214: p. 39-47. 15. Walker, P.S., Erkman, M. J., Metal-on-metal lubrication in artificial human joints. Wear, 1972. 21(2): p. 377-392. 16. Liu, F., Jin, Z. M., Hirt, F., Rieker, C., Roberts, P., Grigoris, P., Effect of wear of bearing surfaces on elastohydrodynamic lubrication of metal-on-metal hip implants. Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine, 2005. 219(5): p. 319-328. 17. Dowson, D., Jin, Z.M., Metal-on-metal hip joint tribology. Proc. IMechE, Part H: J. Engineering in Medicine, 2005. 220: p. 107-118. 18. Medley, J.B., Bobyn, J.D., Krygler, J.J., Chan, F.W., Tanzer, M., Roter, G.E. Elastohydrodynamic lubrication and wear of metal-on-metal hip implant. in World Tribology Forum in Arthroplasty. 2001. Bern: Hans Huber. 19. Varano, R., Bobyn, D., Medley, J.B., Yue, S., The effect of microstructure on the wear of cobalt-based alloys used in metal-on-metal hip implants Proc I Mech E (H), 2006. 220: p. 145-159. 20. Chan, F.W., Bobyn, J.D., Medley, J.B., Krygier, J.J., Yue, S., Tanzer, M., Engineering issues and wear performance of metal on metal hip implants. Clinical orthopaedics and related research 1996( 333): p. 96-107. 21. Chan, F.W., Bobyn, J.D., Medley, J.B., Krygier, J.J., Tanzer, M., Wear and lubrication of metal-on-metal hip implants. Clin Orthop, 1999. 369: p. 10-24. 22. Walker, P.S. and B.L. Gold, Tribology (friction, lubrication and wear) of allmetal artificial hip joints. Wear, 1971. 17(4): p. 285-299. 23. Hernandez-Rodriguez, M.A.L., Mercado-Solis, R. D., Perez-Unzueta, A. J., Martinez-Delgado, D. I., Cantu-Sifuentes, M., Wear of cast metal-metal pairs for total replacement hip prostheses. Wear, 2005. 259(7-12): p. 958-963. 24. Hallab, N.J. and J.J. Jacobs, Orthopedic implant fretting corrosion. Corrosion Reviews, 2003. 21(2-3): p. 183-213. 25. Hallab, N.J., Messina, C., Skipor, A., Jacobs, J.J., Difference in the fretting corrosion of metal-metal and ceramic-metal modular junctions of total hip replacements. Journal of Orthopaedic Research, 2004. 22: p. 250-259. 26. Gilbert, J.L., Buckley, C.A., and Jacobs, J.J., In-Vivo Corrosion of Modular Hip Prosthesis Components in Mixed and Similar Metal Combinations: The Effect of Crevice, Stress, Motion and Alloy Coupling. J Biomed Mater Res, 1993. 27(12): p. 1533.1544. 27. Ratzer-Schiebe, H.J., The relationship between repassivation behavior and pitting corrosion for ti and Ti-6Al-4V. Corrosion 1978. 34: p. 437-442. 28. Dumbleton, J.H., Tribology of natural and artificial joints. 1981, Amsterdam: Elsevier. 29. Tipper, J.L., Ingham, E., Jin, Z.M., Fisher, J., The science of metal-on-metal articulation. Current Orthopaedics, 2005. 19: p. 280-287. 30. Callister, W.D., Materials science and engineering: an introduction. 5th ed. 1999: John Wiley & Sons Inc. p.579. 31. Venugopalan, R., Gaydon, J. , A review of corrosion behaviour of surgical implant alloys. Technical Review Note, Perkin Elmer Instruments, Princeton Applied Research, 1999-2001: p. 1-10. 32. Dearnley, P.A., A review of metallic, ceramic and surface treated materials used for bearing surfaces in human joint replacements. Proc Inst Mech Eng, 1999. 3(H): p. 107-135. 33. Mckee, G.K., Watson-Farrar, J., Replacement of arthritic hips by the Mckee- Farrar prosthsis. J Bone & Joint Surgery, 1966. 48(B): p. 245-259. 34. Marti, A., Cobalt-base alloys used in bone surgery. Injury, Int J. Care Injured, 2000. 31: p. S-D 18-21. 35. Park, J.B., Kon Kim, Y., Metallic Biomaterials, in The Biomedical Engineering Handbook, 2nd Ed., J.D. Bronzino, Editor. 2000, CRC Press LLC: Boca Raton. 36. Yan, Y., Neville, A., Dowson, D., Williams, S., Tribocorrosion in implantsassessing high carbon and low carbon Co-Cr-Mo alloys by in situ electrochemical measurements. Tribology International, 2006. 39(12): p. 1509- 1517. 37. Metcalf, J.E.P., Cawley, J., Band, T.J., Cobalt Chromium Molybdenum metalon- metal resurfacing orthopaedic hip devices. Business briefing: medical device manufacturing & technology, 2004: p. 1-7. 38. Salinas-Rodriguez, A., The Role of the FCC-HCP phase transformation during the plastic deformation of Co-Cr-Mo-C alloys for biomedical applications, in ASTM Special Technical Publication 1365, R.L.K. J.A. Disegi, R. Pilliar, Editor. 1999, American Society for Testing and Materials: West Conshokochen, PA. p. 108-121. 39. Lopez, H.F., Saldivar, A. Huang, P. Development and properties of ƒÃ-martensite in CoCrMo alloys for biomaterials applications. in the Proceedings of 2nd International Alloy Conference. 1999. Davos, Switzerland: Kluwer Academic/Plenum Publishers, New York. 40. Buscher, R., Fischer, A., The pathways of dynamic recrystallization in all-metal hip joints. Wear 2005. 259: p. 887-897. 41. Weber, B.G., Experience with the Metasul total hip bearing system. Clin Orthop, 1996. 3295: p. S78-88. 42. Harlan, C.A., M.D., History of Hip Resurfacing, http://www.jrioh. com/jri_hip_surface.php. Access date: 01.Dec 2007. 43. Wagner, M., Wagner, H., Prelinimary results of uncemented metal-on-metal stemmed and resrfacing hip replacement arthroplasty. clin Orthop, 1996. 329S: p. S89-98. 44. Jin, Z.M., Tipper, J.L., Stone, M.H., Ingham, E., Fisher, J, Hip joints, aritificial, in Encyclopedia of Medical Devices and Instrumentation, J.G. Webster, Editor. 2006, John Wiley & Sons, Inc. 45. Cawley, J., Metcalf, J.E.P., Jones, A.H., Band, T.J., Skupien, D.S., A tribological study of cobalt chromium molybdenum alloys used in metal-onmetal resurfacing hip arthroplasty. Wear, 2003. 255(pt.2): p. 999-1006. 46. Wimmer, M.A., Loos, J., Nassutt, R., Heitkemper, M., Fischer, A., The acting wear mechanisms on metal-on-metal hip joint bearings: in vitro results. Wear, 2001. 250: p. 129-139. 47. http://www.wheelessonline.com/ortho/gait. Access date 01-01-2009. 48. Paul, J.P., Force actions transmitted by joints in the human body. Proceedings of the Royal Society of London. Series B, Biological Sciences, 1976. 192(1107): p. 163-172. 49. Litsky, A.S., Spector, M., Biomaterials, in Orthopaedic basic science, S.R. Simon, Editor. 1994, American academy of orthopaedic surgeons. p. 470-473. 50. Bergmann, G., Graichen, F., Rohlmann, A., Hip joint loading during walking and running, measured in two patients. J. Biomech, 1993. 26: p. 969. 51. Bergmann, G., Graichen, F., Rohlmann, A., Is staircase walking a risk for the fixation of hip implants? Journal of Biomechanics, 1995. 28(5): p. 535-53. 52. Jin, Z.M., Dowson, D., Fisher, J., Analysis of fluid film lubrication in artificial hip joint replacements with surfaces of high elastic modulus. Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine, 1997. 211(3): p. 247-256. 53. Semlitsch, M., Willert H.G., Clinical wear behavior of ultra-high molecular weight polyethylene cups paired with metal and ceramic ball heads in comparison to metal-on-metal pairings of hip joint replacements. Proc Inst Mech Eng, 1997. 211(H): p. 73-88. 54. Doorn, P.F., Campbell, P.A., Worrall, J., Benya, P.D., McKellop, H.A., Amstutz, H.C., Metal wear particle characterization from metal on metal total hip replacements: transimission electron microscopy study of periprosthetic tissues and isolated particles. J. Biomed. Mater. Res, 1998. 42: p. 103-111. 55. Hanawa, T., Metal ion release from metal implants. Materials Science and Engineering C, 2004. 24(6-8 SPEC ISS): p. 745-752. 56. Peters, T.J., Serum Albumin. Advances in Protein Chemistry, 1985. 37: p. 161- 245. 57. Van Oss, C.J., Good, R.J., Orientation of the water molecules of hydration of human serum albumin Jounral of Protein Chemistry, 1988. 7(2): p. 179-183. 58. Burnstock, G., Purine-mediated signalling in pain and visceral perception. Trends Pharmacol. Sci., 2001. 22 p. 182-188. 59. Wood, D.J., The characterization of particulate debris obtained from failed orthopedic implants. 1993, San Jose State University, College of Material Engineering. 60. Steen, K.H., Reeh, P.W., Sustained graded pain and hyperalgesia from harmless experimental tissue acidosis in human skin. Neuroscience letter, 1993. 154: p. 113-116. 61. Ratner, B.D., Hoffman, A. S., Schoen, F.J., Lemons, J.E., Biomaterials Science, an Introduction to Materials in Medicine, 2nd. 2004: Elsevier Academic Press. 62. Moulton, S.E., Barisci , J.N., Bath, A., Stella, R. , Wallace, G.G. , Investigation of protein adsorption and electrochemical behavior at a gold electrode. J Colloid Interface Sci. , 2003. 15(261(2)): p. 312-319. 63. Hiromoto, S., K. Noda, and T. Hanawa, Electrochemical properties of an interface between titanium and fibroblasts L929. Electrochimica Acta, 2002. 48(4): p. 387-396. 64. Gettens, R.T.T., Gettens, R. T. T., Gilbert, J. L. Fibrinogen Adsorption onto 316L Stainless Steel: Voltage Effects. in Bioengineering Conference, 2006. Proceedings of the IEEE 32nd Annual Northeast. 2006. 65. de Ficquelmont-Loizos, M.M., Takenouti, H., Kante, W., Long-time and shorttime investigation of the electrode interface through electrochemical impedance measurements. Application to adsorption of human serum albumin onto glassy carbon rotating disc electrode. Journal of Electroanalytical Chemistry, 1997. 428(1-2): p. 129-140. 66. Kleijn, J.M., Barten, Desiree, Stuart, Martien A. Cohen, Adsorption of charged macromolecules at a gold electrode. Langmuir, 2004. 20(22): p. 9703-9713. 67. Serro, A.P., Gispert, M. P., Martins, M. C. L., Brogueira, P., Colaco, R., Saramago, B., Adsorption of albumin on prosthetic materials: Implication for tribological behavior. J Bio Mat Res - Part A, 2006. 78(3): p. 581-589. 68. Sharma., C.P., Clark, G.C.F., Williams, D.F., The adsorption of proteins on metal surfaces. Engineering in medicine, 1981. 10(1): p. 11-16. 69. Bernabeu, P., Tamisier, L., de Cesare, A., Caprani, A., Study of the adsorption of albumin on a platinum rotating disk electrode using impedance measurements. Electrochimica Acta, 1988. 33(8): p. 1129-1136. 70. Choi, K.H., Friedt, J. M., Laureyn, W., Frederix, F., Campitelli, A., Borghs, G., Investigation of protein adsorption with simultaneous measurements of atomic force microscope and quartz crystal microbalance. Journal of Vacuum Science & Technology B (Microelectronics and Nanometer Structures), 2003. 21(4): p. 1433-6. 71. Ying, P., Viana, A. S., Abrantes, L. M., Jin, G., Adsorption of human serum albumin onto gold: A combined electrochemical and ellipsometric study. Journal of Colloid and Interface Science, 2004. 279(1): p. 95-99. 72. Malmsten, M., Formation of adsorbed protein layers. Journal of Colloid and Interface Science, 1998. 207: p. 186-199. 73. Merritt, K., Brown, S. A., Effect of proteins and pH on fretting corrosion and metal ion release. Journal of Biomedical Materials Research, 1988. 22(2): p. 111-120. 74. Sven, P., Determination of transferrin isoproteins in human cerebrospinal fluid using isoelectric focusing and zone immunoelectrophoresis assay. Electrophoresis, 1987. 8(11): p. 515-517. 75. Salim, M., O'Sullivan, B., McArthur, S. L.,Wright, P.C., Characterization of fibrinogen adsorption onto glass microcapillary surfaces by ELISA. Lab Chip, 2007. 7: p. 64 - 70. 76. Geborek, P., Wollheim, F.A., Synovial fluid, in Mechanics of Human Joints, Physiology, Pathophysiology and Treatment, Verna Wright and E.L. Radin, Editors. 1993, Marcel Dekker: New York. p. 117-136. 77. Jones, K.L., O'Melia, C. R., Protein and humic acid adsorption onto hydrophilic membrane surfaces: Effects of pH and ionic strength. Journal of Membrane Science, 2000. 165(1): p. 31-46. 78. Cabilio, N.R., Omanovic, S., Roscoe, S.G., Electrochemical studies of the effect of temperature and pH on the adsorption of alpha-lactalbumin at Pt. Langmuir, 2000. 16(22): p. 8480-8488. 79. Hughes Wassell, D.T., Embery, G., Adsorption of bovine serum albumin on to titanium powder. Biomaterials, 1996. 17(9): p. 859-864. 80. Malmsten, M., Biopolymers at Interfaces. 1998, New York: Marcel Dekker. 81. Okazaki, Y., Tateishi, T., Ito, Y., Corrosion resistance of implant alloys in pseudo physiological solution and role of alloying elements in passive films. Materials Transactions, JIM, 1997. 38(1): p. 78-84. 82. Lacour, F., Torresi, R., Gabrielli, C., Caprani, A., Comparison of the quarzcrystal microbalance and the double-layer capacitance methods for measuring the kinetics of the adsorption of bovine serum albumin onto a gold electrode. Journal of the Electrochemical Society, 1992. 139(6): p. 1619-1622. 83. Omanovic, S., Roscoe, S. G., Electrochemical studies of the adsorption behavior of bovine serum albumin on stainless steel. Langmuir, 1999. 15(23): p. 8315-8321. 84. Moulton, S.E., Barisci, J. N., Bath, A., Stella, R., Wallace, G. G., Studies of double layer capacitance and electron transfer at a gold electrode exposed to protein solutions. Electrochimica Acta, 2004. 49(24): p. 4223-4230. 85. Gassa, L., De Mele, Monica Fernandez Lorenzo. Influence of the composition of the biological media on the electrochemical response of the metallic components of biomaterials. 2005. Lisbon, Portugal. 86. Tang, Q., Xu, C.H., Shi S. Q., Zhou, L M., Formation and characterization of protein patterns on the surfaces with different properties. Synthetic Metals, 2004. 147(1-3): p. 247-52. 87. Tuemmler, H.P., Thull, R., Model of the metal/tissue connection of implants made of titanium or tantalum. Biological and Biomechanical Performance of Biomaterials. 1986, Paris, Fr: Elsevier, Amsterdam, Neth. 403-408. 88. Gilbert, J.L., Bai, Zhijun, Bearinger, Jane, Megremis, Spiro. The dynamics of oxide films on metallic biomaterials. 2003. Anaheim, CA., United States: ASM International. 89. Lacour, F., Frayret, J. P., Caprani, A., Effect of the electrode roughness on the kinetics of albumin adsorption at a rotating disk electrode investigated by double-layer capacitance measurements. Journal of the Electrochemical Society, 1992. 139(1): p. 153-159. 90. Contu, F., Elsener, B., Bohni, H., Serum effect on the electrochemical behaviour of titanium, Ti6Al4V and Ti6Al7Nb alloys in sulp"],"dcterms:accessrights":["public"],"dc:title":["Abrasion-corrosion of cast CoCrMo in simulated hip joint environments"],"dc:type":["Thesis"],"dc:type.qualificationlevel":["doctoral"],"dc:type.qualificationname":["Ph.D."],"person:identifier.orcid":["0000000306819239","000000023439017X"]},"updated_at":"2026-11-19T09:29:52Z"}