{"id":{"repo_id":"ubc","oai_identifier":"oai:circle.library.ubc.ca:2429/4375"},"canonical_url":"https://search.ndltd.org/etd/ubc/oai:circle.library.ubc.ca:2429/4375","repository":{"repo_id":"ubc","name":"University of British Columbia","base_url":"http://circle.library.ubc.ca/oai/request"},"display":{"title":"Dexterity enhancement in microsurgery using a microgripper and motion-scaling system","abstract":"The design and control of a six-degree-of-freedom (6-DOF) force-reflecting motion-scaling teleoperation system was presented in [1]. In this thesis, a remotely controlled microgripper is developed as an end-effector for this system. The device features small size and weight, and large stroke and force compared to other designs. A stylus-shaped teleoperation master that measures the force at the fingers of the operator provides an intuitive means for operating the microgripper. This design also enables the microgripper to be used as a hand-held instrument. Force sensing enables the accurate measurement and control of tool-tissue forces, as well as the emulation of different mechanical devices. Issues concerning the design, control, and application to microsurgical tasks are addressed here. 6-DOF force/torque sensing has also been added to the teleoperation system, enabling the use of hand and environment forces to improve teleoperation transparency, and enabling the measurement of forces during microsurgery. Several methods for teleoperation control have been implemented, and their potential use in microsurgery is discussed. In addition, experiments have been conducted to quantify the effects of scaled motion and scaled force feedback on teleoperation performance in tasks involving sub-millimetre motions and contact forces from 3 to 15 grams. Significant improvements in accuracy of task execution as well as operator confidence and fatigue were observed when scaled motion and scaled force feedback were provided.","abstract_html":"The design and control of a six-degree-of-freedom (6-DOF) force-reflecting motion-scaling teleoperation system was presented in [1]. In this thesis, a remotely controlled microgripper is developed as an end-effector for this system. The device features small size and weight, and large stroke and force compared to other designs. A stylus-shaped teleoperation master that measures the force at the fingers of the operator provides an intuitive means for operating the microgripper. This design also enables the microgripper to be used as a hand-held instrument. Force sensing enables the accurate measurement and control of tool-tissue forces, as well as the emulation of different mechanical devices. Issues concerning the design, control, and application to microsurgical tasks are addressed here. 6-DOF force/torque sensing has also been added to the teleoperation system, enabling the use of hand and environment forces to improve teleoperation transparency, and enabling the measurement of forces during microsurgery. Several methods for teleoperation control have been implemented, and their potential use in microsurgery is discussed. In addition, experiments have been conducted to quantify the effects of scaled motion and scaled force feedback on teleoperation performance in tasks involving sub-millimetre motions and contact forces from 3 to 15 grams. Significant improvements in accuracy of task execution as well as operator confidence and fatigue were observed when scaled motion and scaled force feedback were provided.","abstract_has_math":false,"creators":["Ku, Shyan"],"institution":"University of British Columbia","degree_name":"Master of Applied Science - MASc","degree_level":"master's","degree_discipline":"Electrical and Computer Engineering","degree_department":null,"school":null,"contributors":[],"advisors":[],"committee_chairs":[],"committee_members":[],"year":1996,"date_issued":"1996","date_published":"1996","updated_at":"2026-11-19T14:24:53Z","subjects":[],"languages":["eng"],"rights":["For non-commercial purposes only, such as research, private study and education. Additional conditions apply, see Terms of Use https://open.library.ubc.ca/terms_of_use."],"rights_urls":[],"identifier_entries":[],"person_identifier_entries":[]},"links":{"outbound_url":"http://hdl.handle.net/2429/4375","outbound_label":"Handle","outbound_source":"dc:identifier","record_url":"http://hdl.handle.net/2429/4375","record_label":"Repository record","record_kind":"handle","record_source":"dc:identifier","download_url":"http://circle.library.ubc.ca/bitstream/2429/4375/1/ubc_1996-0234.pdf","download_label":"Download PDF","download_source":"dc:identifier","access_status":null,"embargo_until":null,"access_note":null},"source_record":{"url":"http://circle.library.ubc.ca/oai/request?verb=GetRecord&metadataPrefix=etdms&identifier=oai%3Acircle.library.ubc.ca%3A2429%2F4375","prefix":"etdms"},"metadata_groups":[{"id":"people","label":"People","entries":[{"key":"dc:creator","label":"Author","values":["Ku, Shyan"]}]},{"id":"academic_context","label":"Academic Context","entries":[{"key":"dc:date","label":"Dc Date","values":["1996"]},{"key":"dc:publisher","label":"Institution","values":["University of British Columbia"]},{"key":"dc:type","label":"Dc Type","values":["Text"]},{"key":"thesis:degree_discipline","label":"Discipline","values":["Electrical and Computer Engineering"]},{"key":"thesis:degree_level","label":"Degree Level","values":["master's"]},{"key":"thesis:degree_name","label":"Degree Name","values":["Master of Applied Science - MASc"]},{"key":"thesis:institution_name","label":"Thesis Institution Name","values":["University of British Columbia"]}]},{"id":"language_rights","label":"Language and Rights","entries":[{"key":"dc:language","label":"Dc Language","values":["eng"]},{"key":"dc:rights","label":"Dc Rights","values":["For non-commercial purposes only, such as research, private study and education. Additional conditions apply, see Terms of Use https://open.library.ubc.ca/terms_of_use."]}]},{"id":"identifiers","label":"Identifiers","entries":[{"key":"dc:identifier","label":"Identifier","values":["http://hdl.handle.net/2429/4375","http://circle.library.ubc.ca/bitstream/2429/4375/1/ubc_1996-0234.pdf"]}]},{"id":"additional","label":"Additional Metadata","entries":[{"key":"dc:description","label":"Description","values":["The design and control of a six-degree-of-freedom (6-DOF) force-reflecting motion-scaling teleoperation system was presented in [1]. In this thesis, a remotely controlled microgripper is developed as an end-effector for this system. The device features small size and weight, and large stroke and force compared to other designs. A stylus-shaped teleoperation master that measures the force at the fingers of the operator provides an intuitive means for operating the microgripper. This design also enables the microgripper to be used as a hand-held instrument. Force sensing enables the accurate measurement and control of tool-tissue forces, as well as the emulation of different mechanical devices. Issues concerning the design, control, and application to microsurgical tasks are addressed here. 6-DOF force/torque sensing has also been added to the teleoperation system, enabling the use of hand and environment forces to improve teleoperation transparency, and enabling the measurement of forces during microsurgery. Several methods for teleoperation control have been implemented, and their potential use in microsurgery is discussed. In addition, experiments have been conducted to quantify the effects of scaled motion and scaled force feedback on teleoperation performance in tasks involving sub-millimetre motions and contact forces from 3 to 15 grams. Significant improvements in accuracy of task execution as well as operator confidence and fatigue were observed when scaled motion and scaled force feedback were provided."]},{"key":"dc:format","label":"Dc Format","values":["10086742","application/pdf"]},{"key":"dc:title","label":"Title","values":["Dexterity enhancement in microsurgery using a microgripper and motion-scaling system"]}]}],"canonical_facts":{"dc:creator":["Ku, Shyan"],"dc:date":["1996"],"dc:description":["The design and control of a six-degree-of-freedom (6-DOF) force-reflecting motion-scaling teleoperation system was presented in [1]. In this thesis, a remotely controlled microgripper is developed as an end-effector for this system. The device features small size and weight, and large stroke and force compared to other designs. A stylus-shaped teleoperation master that measures the force at the fingers of the operator provides an intuitive means for operating the microgripper. This design also enables the microgripper to be used as a hand-held instrument. Force sensing enables the accurate measurement and control of tool-tissue forces, as well as the emulation of different mechanical devices. Issues concerning the design, control, and application to microsurgical tasks are addressed here. 6-DOF force/torque sensing has also been added to the teleoperation system, enabling the use of hand and environment forces to improve teleoperation transparency, and enabling the measurement of forces during microsurgery. Several methods for teleoperation control have been implemented, and their potential use in microsurgery is discussed. In addition, experiments have been conducted to quantify the effects of scaled motion and scaled force feedback on teleoperation performance in tasks involving sub-millimetre motions and contact forces from 3 to 15 grams. Significant improvements in accuracy of task execution as well as operator confidence and fatigue were observed when scaled motion and scaled force feedback were provided."],"dc:format":["10086742","application/pdf"],"dc:identifier":["http://hdl.handle.net/2429/4375","http://circle.library.ubc.ca/bitstream/2429/4375/1/ubc_1996-0234.pdf"],"dc:language":["eng"],"dc:publisher":["University of British Columbia"],"dc:rights":["For non-commercial purposes only, such as research, private study and education. Additional conditions apply, see Terms of Use https://open.library.ubc.ca/terms_of_use."],"dc:title":["Dexterity enhancement in microsurgery using a microgripper and motion-scaling system"],"dc:type":["Text"],"thesis:degree_discipline":["Electrical and Computer Engineering"],"thesis:degree_level":["master's"],"thesis:degree_name":["Master of Applied Science - MASc"],"thesis:institution_name":["University of British Columbia"]},"updated_at":"2026-11-19T14:24:53Z"}